Radiation Dose and Image Quality in Pediatric CT: Effect of Technical Factors and Phantom Size and Shape
Abstract
PURPOSE: To evaluate effects of varying tube current and voltage on radiation dose, image noise, and image contrast with different phantom sizes and shapes. MATERIALS AND METHODS: Four round lucite phantoms with 8-32-cm diameters were scanned with multi-detector row computed tomography (CT) and 80-120 kVp. Radiation dose was based on CT dose index, image noise, and iodine contrast and measured with constant and variable tube currents that were age appropriate for each tube voltage. Radiation dose and image noise and contrast were compared in round and oval 24-cm phantoms. For various combinations of technical factors and phantom sizes and shapes, percentage differences were calculated for radiation dose and image noise and contrast. Associations between tube voltage and radiation dose, image noise, and image contrast in round and oval phantoms were determined by fitting second-degree polynomials to data. Differences in radiation dose and image noise and contrast, which were attributable to differences in tube voltage, were tested with paired t tests. RESULTS: With 165-mAs tube current, radiation doses with 140- and 80-kVp tube voltages were 103% ([41.9 mGy - 20.6 mGy]/20.6 mGy) and 58% ([10.2 mGy - 4.2 mGy]/10.1 mGy) higher in the 8-cm phantom than in the 32-cm phantom. When tube current was adapted for phantom size, radiation dose at 80 kVp in the 8-cm phantom was reduced by 82% ([10.1 mGy - 1.8 mGy]/10.1 mGy). In the 8-cm phantom, tube voltage was decreased from 120 to 80 kVp and tube current remained at 165 mAs, resulting in a 68% noise increase ([3.1 HU - 1.8 HU]/1.8 HU). With variable tube current, 80-kVp tube voltage in the 8-cm phantom led to a 138% noise increase ([7.3 HU - 3.1 HU]/3.1 HU). With reduced tube voltage, image contrast increased. In the 8-cm phantom, with a constant 165-mAs tube current and a decrease in tube voltage from 120 to 80 kVp, there was a 35% ([333 HU - 217 HU]/333 HU) increase in contrast. No difference was noted in radiation dose or noise between round and oval phantoms (P = .604 and P = .06, respectively), but a small statistically significant difference (1%) in contrast attenuation was demonstrated (P = .025). CONCLUSION: Reduced tube voltage for pediatric contrast material-enhanced CT reduces radiation dose and maintains image contrast. Image noise increases, but the effect is minimal in smaller phantoms. An additional reduction in tube current further reduces radiation dose.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Data Reduction and Error Analysis for the Physical Sciences
Bevington, P. R., Robinson, D. K., Blair, J. M., Mallinckrodt, A. J., & McKay, S. (1993) · Computers in Physics · doi:10.1063/1.4823194 · Cited by 14,824
International Electrotechnical Commission
Buck, J. (2010) · Handbook of Transnational Economic Governance Regimes · doi:10.1163/ej.9789004163300.i-1081.494 · Cited by 2,742
International Commission on Radiological Protection
(1969) · Radiology · doi:10.1148/93.6.1298 · Cited by 2,154
Photon mass attenuation and energy-absorption coefficients
Hubbell, J. H. (1982) · The International Journal of Applied Radiation and Isotopes · doi:10.1016/0020-708x(82)90248-4 · Cited by 1,459
Strategies for CT Radiation Dose Optimization
Kalra, M. K., Maher, M. M., Toth, T. L., Hamberg, L. M., Blake, M. A., Shepard, J.-A., & Saini, S. (2004) · Radiology · doi:10.1148/radiol.2303021726 · Cited by 1,080
CT scanning: patterns of use and dose
Mettler, F. A., Wiest, P. W., Locken, J. A., & Kelsey, C. A. (2000) · Journal of Radiological Protection · doi:10.1088/0952-4746/20/4/301 · Cited by 664
AAPM/RSNA Physics Tutorial for Residents: Topics in CT: Radiation Dose in CT
McNitt-Gray, M. F. (2002) · RadioGraphics · doi:10.1148/rg.226025128 · Cited by 612
Estimating cancer risks from pediatric CT: going from the qualitative to the quantitative
Brenner, D. J. (2002) · Pediatric Radiology · doi:10.1007/s00247-002-0671-1 · Cited by 554
Donnelly, L. F., Emery, K. H., Brody, A. S., Laor, T., Gylys-Morin, V. M., Anton, C. G., Thomas, S. R., & Frush, D. P. (2001) · American Journal of Roentgenology · doi:10.2214/ajr.176.2.1760303 · Cited by 480
Helical CT of the Body: Are Settings Adjusted for Pediatric Patients?
Paterson, A., Frush, D. P., & Donnelly, L. F. (2001) · American Journal of Roentgenology · doi:10.2214/ajr.176.2.1760297 · Cited by 422
Dose Reduction in Pediatric CT: A Rational Approach
Boone, J. M., Geraghty, E. M., Seibert, J. A., & Wootton-Gorges, S. L. (2003) · Radiology · doi:10.1148/radiol.2282020471 · Cited by 253
Radiation Exposures to Patients from CT: Reality, Public Perception, and Policy
Nickoloff, E. L., & Alderson, P. O. (2001) · American Journal of Roentgenology · doi:10.2214/ajr.177.2.1770285 · Cited by 220
Siegel, M. J. (2003) · Radiology · doi:10.1148/radiol.2293020999 · Cited by 180
Self-citationThe ALARA Concept in Pediatric CT: Myth or Reality?
Slovis, T. L. (2002) · Radiology · doi:10.1148/radiol.2231012100 · Cited by 180
Computer-Simulated Radiation Dose Reduction for Abdominal Multidetector CT of Pediatric Patients
Frush, D. P., Slack, C. C., Hollingsworth, C. L., Bisset, G. S., Donnelly, L. F., Hsieh, J., Lavin-Wensell, T., & Mayo, J. R. (2002) · American Journal of Roentgenology · doi:10.2214/ajr.179.5.1791107 · Cited by 174
Lucaya, J., Piqueras, J., García-Peña, P., Enríquez, G., García-Macías, M., & Sotil, J. (2000) · American Journal of Roentgenology · doi:10.2214/ajr.175.4.1750985 · Cited by 165
Haaga, J. R. (2001) · American Journal of Roentgenology · doi:10.2214/ajr.177.2.1770289 · Cited by 143
Radiation Exposure and Image Quality in Chest CT Examinations
Ravenel, J. G., Scalzetti, E. M., Huda, W., & Garrisi, W. (2001) · American Journal of Roentgenology · doi:10.2214/ajr.177.2.1770279 · Cited by 142
Brenner, D. J., Elliston, C. D., Hall, E. J., & Berdon, W. E. (2001) · Medical Physics · doi:10.1118/1.1415074 · Cited by 134
CODY, D. D., Moxley, D. M., Krugh, K. T., O'Daniel, J. C., Wagner, L. K., & Eftekhari, F. (2004) · American Journal of Roentgenology · doi:10.2214/ajr.182.4.1820849 · Cited by 117
Influence of phantom diameter, kVp and scan mode upon computed tomography dose index
Nickoloff, E. L., Dutta, A. K., & Lu, Z. F. (2003) · Medical Physics · doi:10.1118/1.1543149 · Cited by 106
Feigal, D. W. (2002) · International Journal of Trauma Nursing · doi:10.1067/mtn.2002.121511 · Cited by 87
Rogers, L. F. (2001) · American Journal of Roentgenology · doi:10.2214/ajr.176.2.1760287 · Cited by 80
Radiation dose reduction in CT of the pediatric pelvis.
Kamel, I. R., Hernandez, R. J., Martin, J. E., Schlesinger, A. E., Niklason, L. T., & Guire, K. E. (1994) · Radiology · doi:10.1148/radiology.190.3.8115611 · Cited by 68
Use of radiology in U.S. general short-term hospitals: 1980-1990.
Mettler, F. A., Briggs, J. E., Carchman, R. M., Altobelli, K., Hart, B. L., & Kelsey, C. A. (1993) · Radiology · doi:10.1148/radiology.189.2.8210363 · Cited by 52
Radiation dose reduction in paediatric cranial CT
Chan, C. Y., Wong, Y. P., Chau, L., Yu, S., & Lau, P.-C. (1999) · Pediatric Radiology · doi:10.1007/s002470050692 · Cited by 44
Thoracic CT Technique in Children
Brody, A. S. (2001) · Journal of Thoracic Imaging · doi:10.1097/00005382-200110000-00007 · Cited by 37
Cited by
Abbara, S., Blanke, P., Maroules, C. D., Cheezum, M., Choi, A. D., Han, B. K., Marwan, M., Naoum, C., Norgaard, B. L., Rubinshtein, R., et al. (2016) · Journal of Cardiovascular Computed Tomography · doi:10.1016/j.jcct.2016.10.002 · Cited by 1,095
Intravenous Contrast Medium Administration and Scan Timing at CT: Considerations and Approaches
Bae, K. T. (2010) · Radiology · doi:10.1148/radiol.10090908 · Cited by 1,067
Strategies for Reducing Radiation Dose in CT
McCollough, C. H., Primak, A. N., Braun, N., Kofler, J., Yu, L., & Christner, J. (2009) · Radiologic Clinics of North America · doi:10.1016/j.rcl.2008.10.006 · Cited by 842
Abbara, S., Arbab-Zadeh, A., Callister, T. Q., Desai, M. Y., Mamuya, W., Thomson, L., & Weigold, W. G. (2009) · Journal of Cardiovascular Computed Tomography · doi:10.1016/j.jcct.2009.03.004 · Cited by 580
The Measurement, Reporting, and Management of Radiation Dose in CT
McCollough, C., Cody, D., Edyvean, S., Geise, R., Gould, B., Keat, N., Huda, W., Judy, P., Kalender, W., McNitt-Gray, M., et al. (2008) · doi:10.37206/97 · Cited by 569
Metadata
- Authors
- Siegel, M. J., Schmidt, B., Bradley, D., Suess, C., & Hildebolt, C.
- Venue
- Radiology
- Published
- 2004-11
- Volume
- 233
- Issue
- 2
- Pages
- 515-522
- ISSN
- 0033-8419 (ISSN-L)
- 1527-1315
- OpenAlex
- W2147913236
- Cited by
- 294
- References
- 27
- FWCI
- 8.02
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.