Dose Reduction in Pediatric CT: A Rational Approach
Abstract
PURPOSE: To determine size-dependent technique factors for pediatric computed tomography (CT) by using physically measured objective data. MATERIALS AND METHODS: Six phantom cylinders (10-32 cm in diameter) were scanned with a clinical multi-detector row CT scanner. CT noise was statistically characterized for CT technique factors from 80 to 140 kVp and from 10 to 300 mAs. Dose measurements were performed with each phantom. Dilute iodine and tissue contrast were determined with computer calculations validated with measured data. The dose, noise, and contrast data were computer fit, and pediatric CT technique factors (milliampere seconds) necessary to maintain the contrast-to-noise ratio (CNR) were computed. RESULTS: As compared with that in a reference cylindric adult abdomen of 28 cm in diameter, CNR was maintained at a constant level in pediatric patients of 25, 20, and 15 cm in diameter, respectively, when milliampere second values of 0.557, 0.196, and 0.054 of the adult milliampere second values were used. The corresponding doses were reduced to 0.642, 0.287, and 0.090 of the 28-cm-diameter adult dose, respectively. CT techniques for examination of pediatric heads measuring 15 and 13 cm, respectively, can involve the use of milliampere second values of 0.572 and 0.366 of those used for examination of a standard 17-cm-diameter adult head. CONCLUSION: CT technique charts for pediatric abdominal and head examinations were produced on the basis of physically measured data; use of these tables will enable pediatric radiation dose to be reduced while CNR is preserved.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Estimated Risks of Radiation-Induced Fatal Cancer from Pediatric CT
Brenner, D. J., Elliston, C. D., Hall, E. J., & Berdon, W. E. (2001) · American Journal of Roentgenology · doi:10.2214/ajr.176.2.1760289 · Cited by 3,222
An accurate method for computer‐generating tungsten anode x‐ray spectra from 30 to 140 kV
Boone, J. M., & Seibert, J. A. (1997) · Medical Physics · doi:10.1118/1.597953 · Cited by 706
Self-citationCT scanning: patterns of use and dose
Mettler, F. A., Wiest, P. W., Locken, J. A., & Kelsey, C. A. (2000) · Journal of Radiological Protection · doi:10.1088/0952-4746/20/4/301 · Cited by 664
Donnelly, L. F., Emery, K. H., Brody, A. S., Laor, T., Gylys-Morin, V. M., Anton, C. G., Thomas, S. R., & Frush, D. P. (2001) · American Journal of Roentgenology · doi:10.2214/ajr.176.2.1760303 · Cited by 480
A method for describing the doses delivered by transmission x-ray computed tomography
Shope, T. B., Gagne, R. M., & Johnson, G. C. (1981) · Medical Physics · doi:10.1118/1.594995 · Cited by 435
Helical CT of the Body: Are Settings Adjusted for Pediatric Patients?
Paterson, A., Frush, D. P., & Donnelly, L. F. (2001) · American Journal of Roentgenology · doi:10.2214/ajr.176.2.1760297 · Cited by 422
Lucaya, J., Piqueras, J., García-Peña, P., Enríquez, G., García-Macías, M., & Sotil, J. (2000) · American Journal of Roentgenology · doi:10.2214/ajr.175.4.1750985 · Cited by 165
Specification and Acceptance Testing of Computed Tomography Scanners
Lin, P. P., Beck, T., Borras, C., et al. (1993) · doi:10.37206/38 · Cited by 120
Radiation dose reduction in CT of the pediatric pelvis.
Kamel, I. R., Hernandez, R. J., Martin, J. E., Schlesinger, A. E., Niklason, L. T., & Guire, K. E. (1994) · Radiology · doi:10.1148/radiology.190.3.8115611 · Cited by 68
Suzuki, A., & Suzuki, M. N. (1978) · Medical Physics · doi:10.1118/1.594445 · Cited by 66
Cited by
Intravenous Contrast Medium Administration and Scan Timing at CT: Considerations and Approaches
Bae, K. T. (2010) · Radiology · doi:10.1148/radiol.10090908 · Cited by 1,067
Strategies for Reducing Radiation Dose in CT
McCollough, C. H., Primak, A. N., Braun, N., Kofler, J., Yu, L., & Christner, J. (2009) · Radiologic Clinics of North America · doi:10.1016/j.rcl.2008.10.006 · Cited by 842
CT Dose Reduction and Dose Management Tools: Overview of Available Options
McCollough, C. H., Bruesewitz, M. R., & Kofler, J. M. (2006) · RadioGraphics · doi:10.1148/rg.262055138 · Cited by 768
The Measurement, Reporting, and Management of Radiation Dose in CT
McCollough, C., Cody, D., Edyvean, S., Geise, R., Gould, B., Keat, N., Huda, W., Judy, P., Kalender, W., McNitt-Gray, M., et al. (2008) · doi:10.37206/97 · Cited by 569
National survey of doses from CT in the UK: 2003
Shrimpton, P. C., Hillier, M. C., Lewis, M. A., & Dunn, M. (2006) · The British Journal of Radiology · doi:10.1259/bjr/93277434 · Cited by 568
Metadata
- Authors
- Boone, J. M., Geraghty, E. M., Seibert, J. A., & Wootton-Gorges, S. L.
- Venue
- Radiology
- Published
- 2003-08
- Volume
- 228
- Issue
- 2
- Pages
- 352-360
- ISSN
- 0033-8419 (ISSN-L)
- 1527-1315
- OpenAlex
- W2136904992
- Cited by
- 253
- References
- 13
- FWCI
- 8.28
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Dose Reduction in Pediatric CT: A Rational Approach. https://app.cochat.ai/papers/10.1148/radiol.2282020471
- The paper
- Boone, J. M., Geraghty, E. M., Seibert, J. A., & Wootton-Gorges, S. L. (2003). Dose Reduction in Pediatric CT: A Rational Approach. Radiology. https://doi.org/10.1148/radiol.2282020471
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.