Renal Volume Measurements: Accuracy and Repeatability of US Compared with That of MR Imaging
Abstract
PURPOSE: To determine the accuracy and repeatability of ultrasonography (US) with the ellipsoid formula in calculating the renal volume. MATERIALS AND METHODS: The renal volumes in 20 volunteers aged 19-51 years were determined by using US with the ellipsoid formula and magnetic resonance (MR) imaging with the voxel-count method by two independent observers for each modality. The observers performed all measurements twice, with an interval between the first and second examinations. The voxel-count method was the reference standard. Repeatability was evaluated by calculating the SD of the difference (method of Bland and Altman). RESULTS: Renal volume was underestimated with US by 45 mL (25%) on average. A comparable underestimation was found when the ellipsoid formula was applied to MR images. This indicates that the inaccuracy of US renal volume measurements (a) occurred because the kidney does not resemble an ellipsoid and (b) was not primarily related to the imaging modality. Intra- and interobserver variations in US volume measurements were poor; the SD of the difference was 21-32 mL. For comparison, the SD of the difference in reference-standard measurements was 5-10 mL. CONCLUSION: Use of US with the ellipsoid formula is not appropriate for accurate and reproducible calculation of renal volume.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Cited by
Applying the right statistics: analyses of measurement studies
Bland, J. M., & Altman, D. G. (2003) · Ultrasound in Obstetrics & Gynecology · doi:10.1002/uog.122 · Cited by 1,311
Irazabal, M. V., Rangel, L. J., Bergstralh, E. J., Osborn, S. L., Harmon, A. J., Sundsbak, J. L., Bae, K. T., Chapman, A. B., Grantham, J. J., Mrug, M., et al. (2015) · Journal of the American Society of Nephrology · doi:10.1681/asn.2013101138 · Cited by 625
Chapman, A. B., Devuyst, O., Eckardt, K.-U., Gansevoort, R. T., Harris, T., Horie, S., Kasiske, B. L., Odland, D., Pei, Y., Perrone, R. D., et al. (2015) · Kidney International · doi:10.1038/ki.2015.59 · Cited by 611
Normal Values for Renal Length and Volume as Measured by Magnetic Resonance Imaging
Cheong, B., Muthupillai, R., Rubin, M. F., & Flamm, S. D. (2007) · Clinical Journal of the American Society of Nephrology · doi:10.2215/cjn.00930306 · Cited by 265
Metadata
- Authors
- Bakker, J., Olree, M., Kaatee, R., de Lange, E. E., Moons, K. G. M., Beutler, J. J., & Beek, F. J. A.
- Venue
- Radiology
- Published
- 1999-06
- Volume
- 211
- Issue
- 3
- Pages
- 623-628
- ISSN
- 0033-8419 (ISSN-L)
- 1527-1315
- OpenAlex
- W2150877210
- Cited by
- 248
- References
- 1
- FWCI
- 1.32
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Renal Volume Measurements: Accuracy and Repeatability of US Compared with That of MR Imaging. https://app.cochat.ai/papers/10.1148/radiology.211.3.r99jn19623
- The paper
- Bakker, J., Olree, M., Kaatee, R., de Lange, E. E., Moons, K. G. M., Beutler, J. J., & Beek, F. J. A. (1999). Renal Volume Measurements: Accuracy and Repeatability of US Compared with That of MR Imaging. Radiology. https://doi.org/10.1148/radiology.211.3.r99jn19623
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.