An Image-based Approach to Understanding the Physics of MR Artifacts
The DOI matches its Crossref record; 16 of the 16 references it cites were found in a registry or an index.
Abstract
As clinical magnetic resonance (MR) imaging becomes more versatile and more complex, it is increasingly difficult to develop and maintain a thorough understanding of the physical principles that govern the changing technology. This is particularly true for practicing radiologists, whose primary obligation is to interpret clinical images and not necessarily to understand complex equations describing the underlying physics. Nevertheless, the physics of MR imaging plays an important role in clinical practice because it determines image quality, and suboptimal image quality may hinder accurate diagnosis. This article provides an image-based explanation of the physics underlying common MR imaging artifacts, offering simple solutions for remedying each type of artifact. Solutions that have emerged from recent technologic advances with which radiologists may not yet be familiar are described in detail. Types of artifacts discussed include those resulting from voluntary and involuntary patient motion, magnetic susceptibility, magnetic field inhomogeneities, gradient nonlinearity, standing waves, aliasing, chemical shift, and signal truncation. With an improved awareness and understanding of these artifacts, radiologists will be better able to modify MR imaging protocols so as to optimize clinical image quality, allowing greater confidence in diagnosis.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Generalized autocalibrating partially parallel acquisitions (GRAPPA)
Griswold, M. A., Jakob, P. M., Heidemann, R. M., Nittka, M., Jellus, V., Wang, J., Kiefer, B., & Haase, A. (2002) · Magnetic Resonance in Medicine · doi:10.1002/mrm.10171 · Cited by 5,303 · cited 1×
“Several data extrapolation algorithms are available for use in reconstructing image data obtained with a parallel imaging technique: GRAPPA (generalized autocalibrating partially parallel acquisition) fills missing lines in k-space before the Fourier transform, so that the entirety of k-space is filled before images a…”
MethodReduction of eddy‐current‐induced distortion in diffusion MRI using a twice‐refocused spin echo
Reese, T. G., Heid, O., Weisskoff, R. M., & Wedeen, V. J. (2003) · Magnetic Resonance in Medicine · doi:10.1002/mrm.10308 · Cited by 1,227 · cited 1×
“clinical diffusion-weighted imaging because of the high amplitude and long duration of diffusion-sensitizing gradients (22).”
BackgroundLee, M.-J., Kim, S., Lee, S.-A., Song, H.-T., Huh, Y.-M., Kim, D.-H., Han, S. H., & Suh, J.-S. (2007) · RadioGraphics · doi:10.1148/rg.273065087 · Cited by 513 · cited 1×
“sizes, using a short TE or short interecho interval, or increasing the sampling bandwidth (25).”
MethodBailes, D. R., Gilderdale, D. J., Bydder, G. M., Collins, A. G., & Firmin, D. N. (1985) · Journal of Computer Assisted Tomography · doi:10.1097/00004728-198507010-00039 · Cited by 313 · cited 1×
“With ROPE, the most positive phase encoding values are sampled during inspiration and the most negative phase encoding values are sampled during expiration, allowing a smooth arrangement of the k-space data throughout the respiratory cycle and minimization of ghosting (18).”
BackgroundMR image artifacts from periodic motion
Wood, M. L., & Henkelman, R. M. (1985) · Medical Physics · doi:10.1118/1.595782 · Cited by 302 · cited 1×
“Ghosting and Smearing Ghosting and smearing, common artifacts produced by voluntary or involuntary motion of the patient, have long plagued clinical MR imaging (10).”
BackgroundBailes, D. R., Gilderdale, D. J., Bydder, G. M., et al. (1985) · PubMed · Cited by 300
Parallel MR Imaging: A User’s Guide
Glockner, J. F., Hu, P., Stanley, D. W., Angelos, L., & King, K. (2005) · Radiographics · doi:10.1148/rg.255045202 · Cited by 269 · cited 1×
“The SNR is decreased by a factor proportional to 1/(g • √X), where g is a variable inversely related to the degree of coil element independence, and X is the parallel imaging factor (14).”
BackgroundWhy fat is bright in rare and fast spin‐echo imaging
Hardy, P., Henkelman, R. M., Bishop, J. E., Poon, E. C. S., & Plewes, D. B. (1992) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.1880020511 · Cited by 229 · cited 1×
“Moreover, J coupling and other factors in T2-weighted fast spin-echo imaging cause an appearance of high signal intensity in fatty structures (11).”
BackgroundMR Artifacts, Safety, and Quality Control
Zhuo, J., & Gullapalli, R. P. (2006) · RadioGraphics · doi:10.1148/rg.261055134 · Cited by 212
Metal Artifact Reduction Sequence: Early Clinical Applications
Olsen, R. V., Munk, P. L., Lee, M. J., Janzen, D. L., MacKay, A. L., Xiang, Q.-S., & Masri, B. (2000) · RadioGraphics · doi:10.1148/radiographics.20.3.g00ma10699 · Cited by 208 · cited 1×
“Artifact resulting from the presence of metallic objects, most commonly surgical hardware, not only distorts nearby structures but also may result in signal dropout, depending on the sequence used (24).”
BackgroundChemical Shift: The Artifact and Clinical Tool Revisited
Hood, M. N., Ho, V. B., Smirniotopoulos, J. G., & Szumowski, J. (1999) · RadioGraphics · doi:10.1148/radiographics.19.2.g99mr07357 · Cited by 180 · cited 1×
“The resultant chemical shift artifacts on clinical MR images manifest most prominently at fat-water interfaces such as those in the bladder and the periorbital regions (27).”
BackgroundEiber, M., Beer, A. J., Holzapfel, K., Tauber, R., Ganter, C., Weirich, G., Krause, B. J., Rummeny, E. J., & Gaa, J. (2009) · Investigative Radiology · doi:10.1097/rli.0b013e3181bbdc2f · Cited by 153
Gloviczki, M. L., Glockner, J. F., Gomez, S., Romero, J. C., Lerman, L. O., McKusick, M. A., & Textor, S. C. (2009) · Investigative Radiology · doi:10.1097/rli.0b013e3181b4c1e8 · Cited by 89
MR imaging of multiple sclerosis: comparison with clinical and CT examinations in 74 patients
Jj, S., Siddharthan, R., Tobias, J. K., Sheremata, W. A., Soila, K., & Viamonte, M. I. (1985) · American Journal of Roentgenology · doi:10.2214/ajr.145.5.957 · Cited by 65
Use of 3.0-T MR Imaging for Evaluation of the Abdomen
Ertürk, Ş. M., Alberich‐Bayarri, Á., Herrmann, K. A., Martí‐Bonmatí, L., & Ros, P. R. (2009) · Radiographics · doi:10.1148/rg.296095516 · Cited by 59 · cited 1×
“When the RF wavelength is shorter than the dimensions of the anatomic structures examined at clinical MR imaging, standing waves may result (23).”
BackgroundAttenberger, U. I., Runge, V. M., Stemmer, A., Williams, K. D., Naul, L. G., Michaely, H. J., Schoenberg, S. O., Reiser, M. F., & Wintersperger, B. J. (2009) · Investigative Radiology · doi:10.1097/rli.0b013e3181af3f0e · Cited by 47 · cited 2×
“The standard line-by-line rectilinear filling of k-space may be replaced by radial k-space filling with the use of a multishot radial acquisition technique (eg, syngoBLADE, Siemens Healthcare, Erlangen, Germany; PROPELLER, GE Healthcare, Milwaukee, Wis) (16).”
MethodSelf-citation
Cited by
Will Artificial Intelligence Replace Radiologists?
Langlotz, C. P. (2019) · Radiology: Artificial Intelligence · doi:10.1148/ryai.2019190058 · Cited by 244
Cardiovascular magnetic resonance physics for clinicians: part II
Biglands, J. D., Radjenovic, A., & Ridgway, J. P. (2012) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-14-66 · Cited by 241
“Explanations of these artefacts are given elsewhere [48,49].”
Alexander-Bloch, A., Clasen, L., Stockman, M., Ronan, L., Lalonde, F., Giedd, J., & Raznahan, A. (2016) · Human Brain Mapping · doi:10.1002/hbm.23180 · Cited by 178
“The physics of distortions caused by motion within the acquisition of a single image have been well documented [Morelli et al., 2011; Wood and Henkelman, 1985].”
BackgroundIscan, Z., Jin, T. B., Kendrick, A., Szeglin, B., Lu, H., Trivedi, M., Fava, M., McGrath, P. J., Weissman, M., Kurian, B. T., et al. (2015) · Human Brain Mapping · doi:10.1002/hbm.22856 · Cited by 164
“These include smearing/blurring, ringing/rippling/striping, ghosting, and radiofrequency leak, which are caused by subject motion, scanner setup, signal interference, and soforth [Bellon et al., 1986; Morelli et al., 2011].”
Background
Metadata
- Authors
- Morelli, J. N., Runge, V. M., Ai, F., Attenberger, U., Vu, L., Schmeets, S. H., Nitz, W. R., & Kirsch, J. E.
- Venue
- RadioGraphics
- Published
- 2011-05
- Volume
- 31
- Issue
- 3
- Pages
- 849-866
- ISSN
- 0271-5333 (ISSN-L)
- 1527-1323
- OpenAlex
- W2122630024
- Cited by
- 171
- References
- 16
- FWCI
- 8.27
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.