Evaluation of a noninvasive continuous cardiac output monitoring system based on thoracic bioreactance
Abstract
Noninvasive cardiac output (CO) measurement can be useful in many clinical settings where invasive monitoring is not desired. Bioimpedance (intrabeat measurement of changes in transthoracic voltage amplitude in response to an injected high-frequency current) has been explored for this purpose but is limited in some clinical settings because of inherently low signal-to-noise ratio. Since changes in fluid content also induce changes in thoracic capacitive and inductive properties, we tested whether a noninvasive CO measurement could be obtained through measurement of the relative phase shift of an injected current (i.e., bioreactance). We constructed a prototype device that applies a 75-kHz current and determines the relative phase shift (dΦ/d t) of the recorded transthoracic voltage. CO was related to the product of peak dΦ/d t, heart rate, and ventricular ejection time. The preclinical study was done in nine open-chest pigs put on right heart bypass so that CO could be varied at known values. This was followed by a feasibility study in 27 postoperative patients who had a Swan-Ganz catheter (SGC). The measurements of noninvasive CO measurement and cardiopulmonary bypass pump correlated to each other ( r = 0.84) despite the large variation in CO and temperatures. Similarly, in patients, mean CO values were 5.18 and 5.17 l/min as measured by SGC and the noninvasive CO measurement system, respectively, and were highly correlated over the range of values studied ( r = 0.90). Preclinical and clinical data demonstrate the feasibility of using blood flow-related phase shifts of transthoracic electric signals to perform noninvasive continuous CO monitoring.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Sandham, J. D., Hull, R. D., Brant, R. F., Knox, L., Pineo, G. F., Doig, C. J., Laporta, D. P., Viner, S., Passerini, L., Devitt, H., et al. (2003) · New England Journal of Medicine · doi:10.1056/nejmoa021108 · Cited by 1,529
A new stroke volume equation for thoracic electrical bioimpedance: Theory and rationale
BERNSTEIN, D. P. (1986) · Critical Care Medicine · doi:10.1097/00003246-198610000-00017 · Cited by 456
Van De Water, J. M., Miller, T. W., Vogel, R. L., Mount, B. E., & Dalton, M. L. (2003) · Chest · doi:10.1378/chest.123.6.2028 · Cited by 219
Packer, M., Abraham, W. T., Mehra, M. R., Yancy, C. W., Lawless, C. E., Mitchell, J. E., Smart, F. W., Bijou, R., O’Connor, C. M., Massie, B. M., et al. (2006) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2005.12.071 · Cited by 215
Albert, N. M., Hail, M. D., Li, J., & Young, J. B. (2004) · American Journal of Critical Care · doi:10.4037/ajcc2004.13.6.469 · Cited by 134
Schmidt, C., Theilmeier, G., Aken, H. V., Korsmeier, P., Wirtz, S. P., Berendes, E., Hoffmeier, A., & Meissner, A. (2005) · British Journal of Anaesthesia · doi:10.1093/bja/aei224 · Cited by 124
The validity of cardiac output measurement by thoracic impedance: a meta-analysis.
Hd, F. (1992) · PubMed · Cited by 108
Moshkovitz, Y., Kaluski, E., Milo, O., Vered, Z., & Cotter, G. (2004) · Current Opinion in Cardiology · doi:10.1097/00001573-200405000-00008 · Cited by 107
A review of impedance cardiography
Jensen, L. A., Yakimets, J., & Teo, K. K. (1995) · Heart & Lung · doi:10.1016/s0147-9563(05)80036-6 · Cited by 86
Evaluation of a thoracic bioimpedance cardiac output monitor during cardiac catheterization
Barin, E. S., Haryadi, D. G., Schookin, S. I., Westenskow, D. R., Zubenko, V. G., Beliaev, K. R., & Morozov, A. A. (2000) · Critical Care Medicine · doi:10.1097/00003246-200003000-00016 · Cited by 49
Cited by
Toward ubiquitous blood pressure monitoring via pulse transit time: Theory and practice
Mukkamala, R., Hahn, J.-O., Inan, O. T., Mestha, L. K., Kim, C.-S., Toreyin, H., & Kyal, S. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2441951 · Cited by 935
“However, the disadvantage of reflectance-mode PPG is that the SNR is typically lower for any common location, since tissue is highly forward scattering such that the reflected energy back to the PD is low [72].”
BackgroundInfluentialHemodynamic parameters to guide fluid therapy
Marik, P. E., Monnet, X., & Teboul, J.-L. (2011) · Annals of Intensive Care · doi:10.1186/2110-5820-1-1 · Cited by 686
Bioimpedance at the Bedside: Current Applications, Limitations, and Opportunities
Mulasi, U., Kuchnia, A. J., Cole, A. J., & Earthman, C. P. (2015) · Nutrition in Clinical Practice · doi:10.1177/0884533614568155 · Cited by 337
Lukaski, H. C. (2013) · European Journal of Clinical Nutrition · doi:10.1038/ejcn.2012.149 · Cited by 303
Noninvasive Cardiac Output Monitors: A State-of the-Art Review
Marik, P. E. (2013) · Journal of Cardiothoracic and Vascular Anesthesia · doi:10.1053/j.jvca.2012.03.022 · Cited by 293
Metadata
- Authors
- Keren, H., Burkhoff, D., & Squara, P.
- Venue
- American Journal of Physiology-Heart and Circulatory Physiology
- Publisher
- American Physiological Society
- Published
- 2007-07
- Volume
- 293
- Issue
- 1
- Pages
- H583-H589
- ISSN
- 0363-6135 (ISSN-L)
- 1522-1539
- OpenAlex
- W2122575600
- Cited by
- 313
- References
- 11
- FWCI
- 7.34
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Evaluation of a noninvasive continuous cardiac output monitoring system based on thoracic bioreactance. https://app.cochat.ai/papers/10.1152/ajpheart.00195.2007
- The paper
- Keren, H., Burkhoff, D., & Squara, P. (2007). Evaluation of a noninvasive continuous cardiac output monitoring system based on thoracic bioreactance. American Journal of Physiology-Heart and Circulatory Physiology. https://doi.org/10.1152/ajpheart.00195.2007
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.