Particle Size-Selective Assessment of Protection of European Standard FFP Respirators and Surgical Masks against Particles-Tested with Human Subjects

Open access
Lee, Shu-An, Hwang, Dong-Chir, Li, He-Yi, Tsai, Chieh-Fu, Chen, Chun-Wan, Chen, Jen-Kun·2016·Journal of Healthcare Engineering

Abstract

This study was conducted to investigate the protection of disposable filtering half-facepiece respirators of different grades against particles between 0.093 and 1.61 μ m. A personal sampling system was used to particle size-selectively assess the protection of respirators. The results show that about 10.9% of FFP2 respirators and 28.2% of FFP3 respirators demonstrate assigned protection factors (APFs) below 10 and 20, which are the levels assigned for these respirators by the British Standard. On average, the protection factors of FFP respirators were 11.5 to 15.9 times greater than those of surgical masks. The minimum protection factors (PFs) were observed for particles between 0.263 and 0.384 μ m. No significant difference in PF results was found among FFP respirator categories and particle size. A strong association between fit factors and protection factors was found. The study indicates that FFP respirators may not achieve the expected protection level and the APFs may need to be revised for these classes of respirators.

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1155/2016/8572493

Cites

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • Respirators were found to be less protective when facing smaller particles.

    “Respirators were found to be less protective when facing smaller particles (Lee et al., 2005a(Lee et al., , 2008(Lee et al., , 2016(Lee et al., , 2017))”

    From Size-selective Assessment of Respirator Protection against Airborne Fungi and (1→3)-β-D-glucan in Farms (2018) · cites this paper as Lee et al., 2016

    The Conclusion reports the worst protection at an intermediate particle size, 0.263–0.384 μm, and says protection factors for FFP respirators were not size dependent across the tested range. The Discussion likewise says smaller particles can be deposited by diffusion, rather than showing that protection consistently decreases as particles get smaller. (LLM, conf=0.89)

    Checked by a reasoning model — may be wrong.

    Experimental — accuracy not yet measured
  • Particle size is known to affect the protection factors of respirators.

    “Particle size is known to affect the protection factors of respirators.”

    From Size-selective Assessment of Respirator Protection against Airborne Fungi and (1→3)-β-D-glucan in Farms (2018) · cites this paper as Lee et al., 2016

    The Discussion reports that earlier studies found protection factors to be significantly size-dependent for N95 respirators. But this paper’s Results (section 3.2) and Conclusion (section 5) found no significant size dependence for the FFP respirators it tested over 0.093–1.61 μm. The unqualified claim therefore overgeneralizes the evidence. (LLM, conf=0.88)

    Experimental — accuracy not yet measured
  • A moderate correlation between FFs and PFs was also found for N95 FFRs against airborne irons in the welding workplace (Han, 2002); EN-specified FFP respirators and surgical masks against NaCl aerosols (Lee et al., 2016); and full facepiece respirators with HEPA filters (Lee et al., 2017).

    “A moderate correlation between FFs and PFs was also found for the following: N95 FFRs against airborne irons in the welding workplace (Han, 2002); EN-specified FFP respirators and surgical masks against NaCl aerosols (Lee et al., 2016); full facepiece respirators with HEPA filters (Lee et al., 2017).”

    From Size-selective Assessment of Respirator Protection against Airborne Fungi and (1→3)-β-D-glucan in Farms (2018) · cites this paper as Lee et al., 2016

    Section 3.3 (page not available) reports significant correlations between fit factors and protection factors for FFP respirators (r = 0.378) and surgical masks (r = 0.482). The Methods (§§2.3 and 2.6; pages not available) confirm that NaCl aerosols were used. The paper describes the separate correlations as between weak and moderate, rather than simply moderate; it calls the combined-data correlation strong. (LLM, conf=0.86)

    Experimental — accuracy not yet measured

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

The citation lists haven’t been fetched yet

References 20 · Cited by 134

Research works printed but not linked: 17

  1. 10.1080/15428119791012874 · DOI
  2. 10.1080/1047322X.1995.10387652 · DOI
  3. 10.1016/j.jhin.2013.06.020 · DOI
  4. 10.1001/jama.2009.1466 · DOI · Paper page
  5. 10.1093/annhyg/mei058 · DOI
  6. 10.1080/15459620903120086 · DOI · Paper page
  7. 10.1093/annhyg/meh097 · DOI
  8. 10.1093/annhyg/meq085 · DOI
  9. 10.1080/15459620490250026 · DOI
  10. 10.1093/annhyg/mes074 · DOI
  11. 10.1080/15459620500391668 · DOI · Paper page
  12. 2004, Aerosol and Air Quality Research, 4, 56, 10.4209/aaqr.2004.07.0005 · DOI · Paper page
  13. 10.1080/15459624.2012.685838 · DOI
  14. 2015, Journal of Labor, Occupational Safety and Health, 23, 179
  15. 10.1080/15298669291360120 · DOI
  16. 10.1080/15459620701346925 · DOI · Paper page
  17. 10.1202/475.1 · DOI

Authors
Lee, S.-A., Hwang, D.-C., Li, H.-Y., Tsai, C.-F., Chen, C.-W., & Chen, J.-K.
Affiliations
  • Feng Chia University (TW)
  • Department of Environmental Engineering and Science, Feng Chia University, Taichung 40724, Taiwan
  • Institute of Labor, Occupational Safety And Health, Ministry of Labor (TW)
  • National Health Research Institutes (TW)
Corresponding author
Lee, Shu-An
Venue
Journal of Healthcare Engineering
Publisher
Wiley
Published
2016
Volume
2016
Pages
1-12
ISSN
  • 2040-2295 (ISSN-L)
  • 2040-2309
OpenAlex
W2301162259
Cited by
134
References
20
Citations by year
2025: 4 · 2024: 5 · 2023: 9 · 2022: 10 · 2021: 35 · 2020: 64 · 2019: 2 · 2018: 3 · 2017: 2
FWCI
2.09
Type
journal article
Language
en
Open access
Locations
  • Journal of Healthcare Engineering (publishedVersion, cc-by)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
Funders
Ministry of Labor, Taiwan (IOSH99-H311)
Topics
  • Health Sciences › Medicine › Pulmonary and Respiratory Medicine › Infection Control and Ventilation
  • Physical Sciences › Engineering › Electrical and Electronic Engineering › Aerosol Filtration and Electrostatic Precipitation
  • Health Sciences › Health Professions › Occupational Therapy › Occupational Health and Safety in Workplaces
Keywords
surgical masks, respiratory protection, personal protective equipment, fit factor
MeSH terms
Humans, Masks, Particle Size, Respiratory Protective Devices, Occupational Exposure, Research Subjects

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review