Blood Pressure Estimation Using Photoplethysmography Only: Comparison between Different Machine Learning Approaches
Open accessAbstract
Introduction. Blood pressure (BP) has been a potential risk factor for cardiovascular diseases. BP measurement is one of the most useful parameters for early diagnosis, prevention, and treatment of cardiovascular diseases. At present, BP measurement mainly relies on cuff-based techniques that cause inconvenience and discomfort to users. Although some of the present prototype cuffless BP measurement techniques are able to reach overall acceptable accuracies, they require an electrocardiogram (ECG) and a photoplethysmograph (PPG) that make them unsuitable for true wearable applications. Therefore, developing a single PPG-based cuffless BP estimation algorithm with enough accuracy would be clinically and practically useful. Methods. The University of Queensland vital sign dataset (online database) was accessed to extract raw PPG signals and its corresponding reference BPs (systolic BP and diastolic BP). The online database consisted of PPG waveforms of 32 cases from whom 8133 (good quality) signal segments (5 s for each) were extracted, preprocessed, and normalised in both width and amplitude. Three most significant pulse features (pulse area, pulse rising time, and width 25%) with their corresponding reference BPs were used to train and test three machine learning algorithms (regression tree, multiple linear regression (MLR), and support vector machine (SVM)). A 10-fold cross-validation was applied to obtain overall BP estimation accuracy, separately for the three machine learning algorithms. Their estimation accuracies were further analysed separately for three clinical BP categories (normotensive, hypertensive, and hypotensive). Finally, they were compared with the ISO standard for noninvasive BP device validation (average difference no greater than 5 mmHg and SD no greater than 8 mmHg). Results. In terms of overall estimation accuracy, the regression tree achieved the best overall accuracy for SBP (mean and SD of difference: −0.1 ± 6.5 mmHg) and DBP (mean and SD of difference: −0.6 ± 5.2 mmHg). MLR and SVM achieved the overall mean difference less than 5 mmHg for both SBP and DBP, but their SD of difference was >8 mmHg. Regarding the estimation accuracy in each BP categories, only the regression tree achieved acceptable ISO standard for SBP (−1.1 ± 5.7 mmHg) and DBP (−0.03 ± 5.6 mmHg) in the normotensive category. MLR and SVM did not achieve acceptable accuracies in any BP categories. Conclusion. This study developed and compared three machine learning algorithms to estimate BPs using PPG only and revealed that the regression tree algorithm was the best approach with overall acceptable accuracy to ISO standard for BP device validation. Furthermore, this study demonstrated that the regression tree algorithm achieved acceptable measurement accuracy only in the normotensive category, suggesting that future algorithm development for BP estimation should be more specific for different BP categories.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Cuffless Blood Pressure Estimation Algorithms for Continuous Health-Care Monitoring
Kachuee, M., Kiani, M. M., Mohammadzade, H., & Shabany, M. (2017) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2016.2580904 · Cited by 621
Seitsonen, E. R. J., Korhonen, I. K. J., Van Gils, M. J., Huiku, M., Lötjönen, J. M. P., Korttila, K. T., & Yli‐Hankala, A. M. (2005) · Acta Anaesthesiologica Scandinavica · doi:10.1111/j.1399-6576.2005.00654.x · Cited by 120 · cited 1×
““pulse area” feature of the PPG segment reflects the vascular tone changes [22].”
Background
Cited by
El-Hajj, C., & Kyriacou, P. A. (2020) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2020.101870 · Cited by 294
Photoplethysmography signal processing and synthesis
Mejía-Mejía, E., Allen, J., Budidha, K., El-Hajj, C., Kyriacou, P. A., & Charlton, P. H. (2022) · Photoplethysmography · doi:10.1016/b978-0-12-823374-0.00015-3 · Cited by 138
Current Status and Prospects of Health-Related Sensing Technology in Wearable Devices
Cho, J. (2019) · Journal of Healthcare Engineering · doi:10.1155/2019/3924508 · Cited by 33
Metadata
- Authors
- Khalid, S. G., Zhang, J., Chen, F., & Zheng, D.
- Venue
- Journal of Healthcare Engineering
- Publisher
- Wiley
- Published
- 2018-10-23
- Volume
- 2018
- Pages
- 1-13
- ISSN
- 2040-2295 (ISSN-L)
- 2040-2309
- OpenAlex
- W2894890204
- Cited by
- 171
- References
- 2
- FWCI
- 6.80
- Type
- journal article
- Language
- en
- Open access
- hybrid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Blood Pressure Estimation Using Photoplethysmography Only: Comparison between Different Machine Learning Approaches. https://app.cochat.ai/papers/10.1155/2018/1548647
- The paper
- Khalid, S. G., Zhang, J., Chen, F., & Zheng, D. (2018). Blood Pressure Estimation Using Photoplethysmography Only: Comparison between Different Machine Learning Approaches. Journal of Healthcare Engineering. https://doi.org/10.1155/2018/1548647
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.