Cardiovascular Collapse Caused by Electrocardiographically Silent 60-Hz Intracardiac Leakage Current: Implications for Electrical Safety

Open access
Swerdlow, Charles D., Olson, Walter H., O’Connor, Mark E., Gallik, Donna M., Malkin, Robert A., Laks, Michael · 1999 · Circulation
Cited by 60

The DOI matches its Crossref record.

Abstract

Background —The national standard for safe 60-Hz intracardiac leakage current under a single-fault condition is 50 μA. This standard is intended to protect patients from alternating current (AC) at levels below the threshold for sensation, but the minimum unsafe level for AC in closed-chest humans is not known. To determine this value, we studied 40 patients at testing of implantable cardioverter-defibrillators using a programmable source of 60-Hz AC. Methods and Results —We applied AC for 5-second test periods in increasing strengths until ventricular fibrillation (VF) was induced or 1 mA was reached. Two current paths were tested: bipolar, between tip and ring electrodes of a right ventricular pacing catheter, and unipolar, from tip to a remote electrode. We observed a characteristic sequence of 3 responses as AC was increased: (1) intermittent ventricular capture with QRS morphology identical to pacing through the electrodes (minimum value, 20 μA); (2) continuous capture at cycle length 282±88 ms (minimum value, 32 μA); and (3) VF persisting after AC termination (minimum value, 49 μA). Continuous capture caused loss of pulsatile arterial pressure and cardiovascular collapse (mean arterial pressure, 32±8 mm Hg) for the duration of AC with no ECG evidence of AC stimulation. Thus, the clinical picture was that of hypotensive ventricular tachycardia (VT). The continuous-capture threshold was ≤50 μA in 9 patients (22%) for bipolar AC and in 5 (12%) for unipolar AC. All patients showed continuous capture over a wide range for both bipolar AC (68±18 to 216±238 μA) and unipolar AC (84±27 to 278±226 μA). Conclusions —Leakage current causes cardiovascular collapse at levels below the VF threshold. Stimulation by silent AC that is neither felt nor visible on the ECG presents as hypotensive VT. In patients with intracardiac electrodes, leakage current less than or equal to the present standard of 50 μA may cause VT or VF. The safety standard for leakage current lasting ≥5 seconds should be ≤20 μA. This standard should be based on the continuous-capture threshold.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1161/01.cir.99.19.2559

Cites

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

See all 60

Metadata

Authors
Swerdlow, C. D., Olson, W. H., O’Connor, M. E., Gallik, D. M., Malkin, R. A., & Laks, M.
Venue
Circulation
Published
1999-05-18
Volume
99
Issue
19
Pages
2559-2564
ISSN
  • 0009-7322 (ISSN-L)
  • 1524-4539
OpenAlex
W2017687241
Cited by
60
References
0
FWCI
2.26
Type
journal article
Language
en
Open access
bronze

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Cardiovascular Collapse Caused by Electrocardiographically Silent 60-Hz Intracardiac Leakage Current: Implications for Electrical Safety. https://app.cochat.ai/papers/10.1161/01.cir.99.19.2559
The paper
Swerdlow, C. D., Olson, W. H., O’Connor, M. E., Gallik, D. M., Malkin, R. A., & Laks, M. (1999). Cardiovascular Collapse Caused by Electrocardiographically Silent 60-Hz Intracardiac Leakage Current: Implications for Electrical Safety. Circulation. https://doi.org/10.1161/01.cir.99.19.2559

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.