Prosthesis Oversizing in Balloon-Expandable Transcatheter Aortic Valve Implantation Is Associated With Contained Rupture of the Aortic Root
Open accessAbstract
Background—: To retrospectively investigate the potential cause of contained rupture of the aortic root in balloon-expandable transcatheter aortic valve implantation (TAVI) by means of pre- and postinterventional multislice computed tomography. Methods and Results—: Seventy-two patients (mean age 82±7 years, mean aortic valve area 0.69±0.19 cm 2 ) underwent balloon-expandable TAVI using the EdwardsSAPIEN Transcatheter Heart Valve (23 mm, n=19; 26 mm, n=50; 29 mm, n=3). Aortic annulus dimensions were quantified by multislice computed tomography–based cross-sectional area assessment and average diameter calculation (CAAD) before and after TAVI. Post-TAVI multislice computed tomography data sets were available in 65 patients; contained aortic root rupture was diagnosed in 3 patients. Pre-TAVI CAAD was 23.1±1.8 mm; post-TAVI CAAD was 22.9±1.3 mm. Median relative change in CAAD pre- and post-TAVI was −0.5% (interquartile range, 3.6%). Relative increase of 5% to 10% was observed in 4 patients (1 with contained rupture), relative increase >10% in 2 patients, both with contained rupture. Mean relative oversizing, calculated as the relative difference in diameter between pre-TAVI CAAD and nominal diameter of the selected prosthesis, was 9.8%±7.8%. Relative oversizing was significantly higher in patients with contained rupture compared with patients without contained rupture (24.6%±5.4% versus 9.1%±6.6%; P <0.001). Relative oversizing ≥20% occurred in 6 patients (3 with contained rupture). Conclusions—: Contained rupture of the aortic root in balloon-expandable TAVI is associated with severe prosthesis oversizing. Multislice computed tomography–based assessment of aortic annulus dimension in conjunction with adapted sizing guidelines may reduce the incidence of severe oversizing.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Lang, R. M., Bierig, M., Devereux, R. B., Flachskampf, F. A., Foster, E., Pellikka, P. A., Picard, M. H., Roman, M. J., Seward, J., Shanewise, J. S., et al. (2005) · Journal of the American Society of Echocardiography · doi:10.1016/j.echo.2005.10.005 · Cited by 11,318
Transcatheter Aortic-Valve Implantation for Aortic Stenosis in Patients Who Cannot Undergo Surgery
Leon, M. B., Smith, C. R., Mack, M., Miller, D. C., Moses, J. W., Svensson, L. G., Tuzcu, E. M., Webb, J. G., Fontana, G. P., Makkar, R. R., et al. (2010) · New England Journal of Medicine · doi:10.1056/nejmoa1008232 · Cited by 7,205
Transcatheter versus Surgical Aortic-Valve Replacement in High-Risk Patients
Smith, C. R., Leon, M. B., Mack, M. J., Miller, D. C., Moses, J. W., Svensson, L. G., Tuzcu, E. M., Webb, J. G., Fontana, G. P., Makkar, R. R., et al. (2011) · New England Journal of Medicine · doi:10.1056/nejmoa1103510 · Cited by 6,345
Cited by
Barbanti, M., Yang, T.-H., Rodès Cabau, J., Tamburino, C., Wood, D. A., Jilaihawi, H., Blanke, P., Makkar, R. R., Latib, A., Colombo, A., et al. (2013) · Circulation · doi:10.1161/circulationaha.113.002947 · Cited by 601
Self-citationBinder, R. K., Webb, J. G., Willson, A. B., Urena, M., Hansson, N. C., Norgaard, B. L., Pibarot, P., Barbanti, M., Larose, E., Freeman, M., et al. (2013) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2013.04.036 · Cited by 335
Multi-modality imaging in aortic stenosis: an EACVI clinical consensus document
Dweck, M. R., Loganath, K., Bing, R., Treibel, T. A., McCann, G. P., Newby, D. E., Leipsic, J., Fraccaro, C., Paolisso, P., Cosyns, B., et al. (2023) · European Heart Journal - Cardiovascular Imaging · doi:10.1093/ehjci/jead153 · Cited by 79
Metadata
- Authors
- Blanke, P., Reinöhl, J., Schlensak, C., Siepe, M., Pache, G., Euringer, W., Geibel-Zehender, A., Bode, C., Langer, M., Beyersdorf, F., & Zehender, M.
- Venue
- Circulation: Cardiovascular Interventions
- Published
- 2012-08
- Volume
- 5
- Issue
- 4
- Pages
- 540-548
- ISSN
- 1941-7632 (ISSN-L)
- 1941-7640
- OpenAlex
- W2160308897
- Cited by
- 158
- References
- 4
- FWCI
- 11.51
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Prosthesis Oversizing in Balloon-Expandable Transcatheter Aortic Valve Implantation Is Associated With Contained Rupture of the Aortic Root. https://app.cochat.ai/papers/10.1161/circinterventions.111.967349
- The paper
- Blanke, P., Reinöhl, J., Schlensak, C., Siepe, M., Pache, G., Euringer, W., Geibel-Zehender, A., Bode, C., Langer, M., Beyersdorf, F., & Zehender, M. (2012). Prosthesis Oversizing in Balloon-Expandable Transcatheter Aortic Valve Implantation Is Associated With Contained Rupture of the Aortic Root. Circulation: Cardiovascular Interventions. https://doi.org/10.1161/circinterventions.111.967349
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.