CoChat

Comparison of Inappropriate Shocks and Other Health Outcomes Between Single‐ and Dual‐Chamber Implantable Cardioverter‐Defibrillators for Primary Prevention of Sudden Cardiac Death: Results From the Cardiovascular Research Network Longitudinal Study of Implantable Cardioverter‐Defibrillators

Open access
Peterson, Pamela N., Greenlee, Robert T., Go, Alan S., Magid, David J., Cassidy‐Bushrow, Andrea, Garcia‐Montilla, Romel, Glenn, Karen A., Gurwitz, Jerry H., Hammill, Stephen C., Hayes, John, Kadish, Alan, Reynolds, Kristi, Sharma, Param, Smith, David H., Varosy, Paul D., Vidaillet, Humberto, Zeng, Chan X., Normand, Sharon‐Lise T., Masoudi, Frederick A. · 2017 · Journal of the American Heart Association
Cited by 38

The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.

Abstract

Background: In US clinical practice, many patients who undergo placement of an implantable cardioverter‐defibrillator ( ICD ) for primary prevention of sudden cardiac death receive dual‐chamber devices. The superiority of dual‐chamber over single‐chamber devices in reducing the risk of inappropriate ICD shocks in clinical practice has not been established. The objective of this study was to compare risk of adverse outcomes, including inappropriate shocks, between single‐ and dual‐chamber ICD s for primary prevention. Methods and Results: We identified patients receiving a single‐ or dual‐chamber ICD for primary prevention who did not have an indication for pacing from 15 hospitals within 7 integrated health delivery systems in the Longitudinal Study of Implantable Cardioverter‐Defibrillators from 2006 to 2009. The primary outcome was time to first inappropriate shock. ICD shocks were adjudicated for appropriateness. Other outcomes included all‐cause hospitalization, heart failure hospitalization, and death. Patient, clinician, and hospital‐level factors were accounted for using propensity score weighting methods. Among 1042 patients without pacing indications, 54.0% (n=563) received a single‐chamber device and 46.0% (n=479) received a dual‐chamber device. In a propensity‐weighted analysis, device type was not significantly associated with inappropriate shock (hazard ratio, 0.91; 95% confidence interval, 0.59–1.38 [ P =0.65]), all‐cause hospitalization (hazard ratio, 1.03; 95% confidence interval, 0.87–1.21 [ P =0.76]), heart failure hospitalization (hazard ratio, 0.93; 95% confidence interval, 0.72–1.21 [ P =0.59]), or death (hazard ratio, 1.19; 95% confidence interval, 0.93–1.53 [ P =0.17]). Conclusions: Among patients who received an ICD for primary prevention without indications for pacing, dual‐chamber devices were not associated with lower risk of inappropriate shock or differences in hospitalization or death compared with single‐chamber devices. This study does not justify the use of dual‐chamber devices to minimize inappropriate shocks.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1161/jaha.117.006937

Cites

1 of 1 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 1 references

Cited by

See all 38

Metadata

Authors
Peterson, P. N., Greenlee, R. T., Go, A. S., Magid, D. J., Cassidy‐Bushrow, A., Garcia‐Montilla, R., Glenn, K. A., Gurwitz, J. H., Hammill, S. C., Hayes, J., Kadish, A., Reynolds, K., Sharma, P., Smith, D. H., Varosy, P. D., Vidaillet, H., Zeng, C. X., Normand, S. T., & Masoudi, F. A.
Venue
Journal of the American Heart Association
Published
2017-11
Volume
6
Issue
11
ISSN
2047-9980 (ISSN-L)
OpenAlex
W2767448569
Cited by
38
References
1
FWCI
2.90
Type
journal article
Language
en
Open access
gold

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Comparison of Inappropriate Shocks and Other Health Outcomes Between Single‐ and Dual‐Chamber Implantable Cardioverter‐Defibrillators for Primary Prevention of Sudden Cardiac Death: Results From the Cardiovascular Research Network Longitudinal Study of Implantable Cardioverter‐Defibrillators. https://app.cochat.ai/papers/10.1161/jaha.117.006937
The paper
Peterson, P. N., Greenlee, R. T., Go, A. S., Magid, D. J., Cassidy‐Bushrow, A., Garcia‐Montilla, R., Glenn, K. A., Gurwitz, J. H., Hammill, S. C., Hayes, J., Kadish, A., Reynolds, K., Sharma, P., Smith, D. H., Varosy, P. D., Vidaillet, H., Zeng, C. X., Normand, S. T., & Masoudi, F. A. (2017). Comparison of Inappropriate Shocks and Other Health Outcomes Between Single‐ and Dual‐Chamber Implantable Cardioverter‐Defibrillators for Primary Prevention of Sudden Cardiac Death: Results From the Cardiovascular Research Network Longitudinal Study of Implantable Cardioverter‐Defibrillators. Journal of the American Heart Association. https://doi.org/10.1161/jaha.117.006937

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.