Incorrect order of draw could be mitigate the patient safety: a phlebotomy management case report

Open access
Lima-Oliveira, Gabriel, Lippi, Giuseppe, Salvagno, Gian Luca, Montagnana, Martina, Picheth, Geraldo, Guidi, Gian Cesare·2013·Biochemia Medica

Abstract

Procedures involving phlebotomy are critical for obtaining diagnostic blood specimens and represent a well known and recognized problem, probably among the most important issues in laboratory medicine. The aim of this report is to show spurious hyperkalemia and hypocalcemia due to inadequate phlebotomy procedure. The diagnostic blood specimens were collected from a male outpatient 45 years old, with no clinical complaints. The tubes drawing order were as follows: i) clot activator and gel separator (serum vacuum tube), ii) K,EDTA, iii) a needleless blood gas dedicated-syringe with 80 I.U. lithium heparin, directly connected to the vacuum tube holder system. The laboratory testing results from serum vacuum tube and dedicated syringe were 4.8 and 8.5 mmol/L for potassium, 2.36 and 1.48 mmol/L for total calcium, respectively. Moreover 0.15 mmol/L of free calcium was observed in dedicated syringe. A new blood collection was performed without K3EDTA tube. Different results were found for potassium (4.7 and 4.5 mmol/L) and total calcium (2.37 and 2.38 mmol/L) from serum vacuum tube and dedicated syringe, respectively. Also free calcium showed different concentration (1.21 mmol/L) in this new sample when compared with the first blood specimen. Based on this case we do not encourage the laboratory managers training the phlebotomists to insert the dedicated syringes in needle-holder system at the end of all vacuum tubes. To avoid double vein puncture the dedicated syringe for free calcium determination should be inserted immediately after serum tubes before EDTA vacuum tubes.

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.11613/bm.2013.026

Cites

1 of 1 references verified0 citation errors · 0 not checkable

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • Contamination still occurs more commonly than might be expected and can be difficult to identify.

    “there is firm evidence showing that contamination still occurs more commonly than might be expected and can be difficult to identify [79], [80], [81], [82].”

    From Joint EFLM-COLABIOCLI Recommendation for venous blood sampling (2018) · cites this paper as [80]

    The Discussion reports that EDTA contamination accounted for about 25% of the hyperkalaemic samples in a cited study, supporting the point that it can be common in that group. Our copy does not itself discuss how difficult contamination is to identify; that point appears only in the title of a work listed in the References. Neither passage establishes how common contamination is across samples generally. (LLM, conf=0.80)

    Experimental — accuracy not yet measured

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Cites

Flagged first, then by citations

Research works printed but not linked: 25

  1. [1]Simundic AM Lippi G Preanalytical phase - a continuous challenge for laboratory professionals Biochem Med 2012 22 145 9 http://dx.doi.org/10.11613/BM.2012.017 PMC4062337 22838180 · DOI
  2. [2]Bilic-Zulle L Simundic AM Supak-Smolcic V Nikolac N Honovic L Self reported routines and procedures for the extra-analytical phase of laboratory practice in Croatia -cross-sectional survey study Biochem Med 2010 20 64 74 http://dx.doi.org/10.11613/BM.2010.008 · DOI
  3. [3]Lippi G Salvagno GL Montagnana M Guidi GC The skilled phlebotomist Arch Pathol Lab Med 2006 130 1260 1 16948506
  4. [4]Lippi G Salvagno GL Montagnana M Lima-Oliveira G Guidi GC Favaloro EJ Quality Standards for Sample Collection in Coagulation Testing Semin Thromb Hemost 2012 38 565 75 http://dx.doi.org/10.1055/s-0032-1315961 22669757 · DOI
  5. [5]Lima-Oliveira G Guidi GC Salvagno GL Montagnana M Rego FGM Lippi G Is phlebotomy part of the dark side in the clinical laboratory struggle for quality? Lab Med 2012 43 17 21 http://dx.doi.org/10.1309/LMZ7YARD6ZSDIID · DOI
  6. [6]Lippi G Simundic AM Total quality in laboratory diagnostics. It’s time to think outside the box Biochem Med 2010 20 5 8 http://dx.doi.org/10.11613/BM.2010.001 · DOI
  7. [7]Lima-Oliveira G Lippi G Salvagno GL Montagnana M Mangueira C Sumita N New ways to deal with known preanalytical issues: use of transilluminator instead of tourniquet for easing vein access and eliminating stasis on clinical biochemistry Biochem Med 2011 21 152 9 http://dx.doi.org/10.11613/BM.2011.024 22135855 · DOI · Paper page
  8. [8]Lima-Oliveira G Lippi G Salvagno GL Montagnana M Scartezini M Guidi GC Transillumination: a new tool to eliminate the impact of venous stasis during the procedure for the collection of diagnostic blood specimens for routine haematological testing Int J Lab Hematol 2011 33 457 62 21414180
  9. [11]Kost GJ The significance of ionized calcium in cardiac and critical care. Availability and critical limits at US medical centers and chidren’s hospitals Arch Pathol Lab Med 1993 117 890 6 8368900
  10. [12]Kost GJ Using critical limits to improve patient outcome MLO Med Lab Obs 1993 25 22 7 10171391
  11. [15]Calam RR Cooper MH Recommended “order of draw” for collecting blood specimens into additive-containing tubes Clin Chem 1982 28 1399 7074955
  12. [16]Lima-Oliveira G Lippi G Salvagno GL Montagnana M Picheth G Guidi GC Impact of the phlebotomy training based on CLSI/NCCLS H03-A6 - procedures for the collection of diagnostic blood specimens by venipuncture Biochem Med 2012 22 342 51 http://dx.doi.org/10.11613/BM.2012.036 PMC3900046 23092065 · DOI · Paper page
  13. [17]Sulaiman RA Cornes MP Whitehead SJ Othonos N Ford C Gama R Effect of order of draw of blood samples during phlebotomy on routine biochemistry results J Clin Pathol 2011 64 1019 20 http://dx.doi.org/10.1136/jclinpath-2011-200206 21778299 · DOI · Paper page
  14. [18]Cornes MP Sulaiman RA Whitehead SJ Othonos N Ford C Gama R Incorrect order of draw of blood samples does not causes potassium EDTA sample contamination Brit J Biomed Sci 2012 69 1 2 23057163
  15. [19]Lima-Oliveira G Lippi G Salvagno GL Montagnana M Picheth G Guidi GC Different manufacturers of syringes: A new source of variability in blood gas, acid-base balance and related laboratory test? Clin Biochem 2012 45 683 7 http://dx.doi.org/10.1016/j.clinbiochem.2012.03.007 22440459 · DOI
  16. [20]Lima-Oliveira G Lippi G Salvagno GL Montagnana M Poli G Solero GP K3EDTA vacuum tubes validation for routine hematological testing ISRN Hematology 2012 2012 875357 22888448 PMC3409519
  17. [21]Lima-Oliveira G Lippi G Salvagno GL Montagnana M Picheth G Guidi GC Pre analytical management: serum vacuum tubes validation for routine clinical chemistry Biochem Med 2012 22 180 6 http://dx.doi.org/10.11613/BM.2012.021. PMC4062344 22838184 · DOI
  18. [22]Lima-Oliveira G Lippi G Salvagno GL Montagnana M Picheth G Guidi GC Sodium citrate vacuum tubes validation: preventing preanalytical variability in routine coagulation testing Blood Coagul Fibrinolysis 2013 24 252 5 http://dx.doi.org/10.1097/MBC.0b013e32835b72ea 23160241 · DOI
  19. [23]Lima-Oliveira G Lippi G Salvagno GL Montagnana M Poli G Solero GP The brand of K2EDTA vacuum tubes as a new source of preanalytical variability in routine hematology testing Brit J Biomed Sci 2013 70 6 9 23617091
  20. [24]Cornes MP Ford C Gama R Spurious hyperkalaemia due to EDTA contamination: common and not always easy to identify Ann Clin Biochem 2008 45 601 3 http://dx.doi.org/10.1258/acb.2008.007241 18768536 · DOI · Paper page
  21. [25]White G Serum ethylenediaminetetraacetic acid concentrations in routine samples submitted for biochemical analysis Ann Clin Biochem 2010 47 485 6 http://dx.doi.org/10.1258/acb.2010.009306. 20595407 · DOI
  22. [26]Naguib MT Evans N Combined false hyperkalemia and hypocalcemia due to specimen contamination during routine phlebotomy South Med J 2002 95 1218 20 12425515
  23. [28]Ricos C Alvarez V Cava F Garcia-Lario JV Hernandez A Jimenez CV Current databases on biological variation: pros, cons and progress Scand J Clin Lab Invest 1999 59 491 500 http://dx.doi.org/10.1080/00365519950185229 10667686 · DOI · Paper page
  24. [29]Cembrowski GS Tran DV Higgins TN The use of serial patient blood gas, electrolyte and glucose results to derive biologic variation: a new tool to assess the acceptability of intensive care unit testing Clin Chem Lab Med 2010 48 1447 54 http://dx.doi.org/10.1515/cclm.2010.286 20716012 · DOI
  25. [30]Ricos C Cava F Garcia-Lario JV Hernandez A Iglesias N Jimenez CV The reference change value: a proposal to interpret laboratory reports in serial testing based on biological variation Scand J Clin Lab Invest 2004 64 175 84 http://dx.doi.org/10.1080/00365510410004885 15222627 · DOI

Cited by

Sorted by citations

Authors
Lima-Oliveira, G., Lippi, G., Salvagno, G. L., Montagnana, M., Picheth, G., & Guidi, G. C.
Affiliations
  • University of Verona (IT)
  • University of Parma (IT)
  • Universidade Federal do Paraná (BR)
Corresponding author
Lima-Oliveira, Gabriel
Venue
Biochemia Medica
Published
2013
Volume
23
Issue
2
Pages
218-223
ISSN
  • 1330-0962 (ISSN-L)
  • 1846-7482
OpenAlex
W2155073043
Cited by
33
References
1
Citations by year
2024: 4 · 2023: 1 · 2022: 1 · 2021: 1 · 2020: 4 · 2019: 5 · 2018: 1 · 2017: 3 · 2016: 3 · 2015: 1 · 2014: 6 · 2013: 3
FWCI
3.02
Type
journal article
Language
en
Open access
Licence
cc-by
Locations
  • Biochemia Medica (publishedVersion, cc-by)
  • Europe PMC (PubMed Central) (publishedVersion, other-oa)
  • PubMed Central (publishedVersion, cc-by-nc)
  • PubMed (publishedVersion)
  • Hrčak Portal of scientific journals of Croatia (University Computing Centre) (submittedVersion, other-oa)
  • Padua Research Archive (University of Padova) (submittedVersion, cc-by)
Topics
  • Health Sciences › Medicine › Physiology › Clinical Laboratory Practices and Quality Control
  • Health Sciences › Health Professions › Emergency Medical Services › Patient Safety and Medication Errors
  • Health Sciences › Medicine › Pulmonary and Respiratory Medicine › Potassium and Related Disorders
Keywords
phlebotomy, specimen contamination, free calcium, blood collection tubes, lithium heparin, patient safety
MeSH terms
Humans, Male, Middle Aged, Specimen Handling, Phlebotomy, Patient Safety

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review