Background: LC-MS/MS is rapidly becoming the technology of choice for measuring steroid hormones. We have developed a rapid LC-MS/MS assay for the routine analysis of serum cortisol. We have used this assay to investigate the effects of gender and exogenous steroid interference on the immunoassay measurement of serum cortisol. Methods: Zinc sulphate (40 µL) was added to 20 µL of sample. This was vortexed for 10 s followed by the addition of 100 µL of internal standard in methanol. Following mixing and centrifugation, 10 µL of sample was injected into an Acquity LC system coupled to a Quattro Premier tandem mass spectrometer. Serum samples ( n = 149) were analysed by LC-MS/MS and two commercial immunoassays. Results were then compared for all samples and for gender differences. A further set of serum samples ( n = 171) was analysed by the LC-MS/MS assay and a GC-MS assay. Results: Cortisol had a retention time of 0.98 min and the assay had an injection-to-injection time of 2.6 min per sample. Mean recovery was 99% and mean CV was 8%. The immunoassays gave comparisons of: Roche = 1.23 × LC-MS/MS −1.12 nmol/L and Abbott = 0.94 × LC-MS/MS + 11.97. The comparison with GC-MS showed LC-MS/MS = 1.11 × GC-MS – 22.90. Discussion: We have developed an LC-MS/MS assay for serum cortisol analysis that is suitable for routine clinical use and has been in use in our laboratory for 12 months. The availability of this assay will give more reliable results in patients receiving exogenous steroid therapy.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1177/0004563212473448

Cites

3 of 3 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 3 references

Cited by

See all 46

Metadata

Authors
Owen, L. J., Adaway, J. E., Davies, S., Neale, S., El-Farhan, N., Ducroq, D., Evans, C., Rees, D. A., MacKenzie, F., & Keevil, B. G.
Venue
Annals of Clinical Biochemistry: International Journal of Laboratory Medicine
Published
2013-06-12
Volume
50
Issue
4
Pages
345-352
ISSN
  • 0004-5632 (ISSN-L)
  • 1758-1001
OpenAlex
W2120794284
Cited by
46
References
3
FWCI
2.65
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Development of a rapid assay for the analysis of serum cortisol and its implementation into a routine service laboratory. https://app.cochat.ai/papers/10.1177/0004563212473448
The paper
Owen, L. J., Adaway, J. E., Davies, S., Neale, S., El-Farhan, N., Ducroq, D., Evans, C., Rees, D. A., MacKenzie, F., & Keevil, B. G. (2013). Development of a rapid assay for the analysis of serum cortisol and its implementation into a routine service laboratory. Annals of Clinical Biochemistry: International Journal of Laboratory Medicine. https://doi.org/10.1177/0004563212473448

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.