Accuracy of Devices Used for Self-Monitoring of Blood Glucose
Open accessThe DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.
Abstract
We measured the inaccuracy of 17 home blood glucose monitors (two visually read, eight colorimetric and seven amperometric). Using strips from a single batch, blood glucose measurements were performed by three medical laboratory technologists on at least 50 capillary blood specimens from patients attending two diabetes clinics. Additional capillary blood was deproteinized and assayed with hexokinase to give a whole blood glucose result. A dedicated glucose analyser was also studied to cross-validate the methodology. At a mean glucose concentration of about 9 mmol/L, monitor readings differed from the reference results by −5.1 to + 19.5% with three systems failing to meet the American Diabetes Association guideline for total error of less than 15%. This problem would be alleviated by manufacturers adopting a common policy on calibration and on reporting as a plasma or whole blood value.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Self-Monitoring of Blood Glucose
American Diabetes Association (1994) · Diabetes Care · doi:10.2337/diacare.17.1.81 · Cited by 255
Cited by
Monitoring blood glucose control in diabetes mellitus: a systematic review.
Coster, Gulliford, Seed, Powrie, & Swaminathan (2000) · Health Technology Assessment · doi:10.3310/hta4120 · Cited by 245
Approved IFCC recommendation on reporting results for blood glucose (abbreviated)
D'Orazio, P. (2005) · Clinical Chemistry · doi:10.1373/clinchem.2005.051979 · Cited by 208
Skeie, S., Thue, G., Nerhus, K., & Sandberg, S. (2002) · Clinical Chemistry · doi:10.1093/clinchem/48.7.994 · Cited by 131
IFCC recommendation on reporting results for blood glucose
(2001) · Clinica Chimica Acta · doi:10.1016/s0009-8981(01)00431-4 · Cited by 110
Evolution of analytical performance in portable glucose meters in the last decade
Böhme, P., Floriot, M., Sirveaux, M.-A., Durain, D., Ziegler, O., Drouin, P., & Guerci, B. (2003) · Diabetes Care · doi:10.2337/diacare.26.4.1170 · Cited by 51
Metadata
- Authors
- Johnson, R. N., & Baker, J. R.
- Venue
- Annals of Clinical Biochemistry: International Journal of Laboratory Medicine
- Publisher
- SAGE Publications
- Published
- 1998-01
- Volume
- 35
- Issue
- 1
- Pages
- 68-74
- ISSN
- 0004-5632 (ISSN-L)
- 1758-1001
- OpenAlex
- W2334444847
- Cited by
- 44
- References
- 2
- FWCI
- 2.56
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Accuracy of Devices Used for Self-Monitoring of Blood Glucose. https://app.cochat.ai/papers/10.1177/000456329803500108
- The paper
- Johnson, R. N., & Baker, J. R. (1998). Accuracy of Devices Used for Self-Monitoring of Blood Glucose. Annals of Clinical Biochemistry: International Journal of Laboratory Medicine. https://doi.org/10.1177/000456329803500108
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.