Wearable Heart Rate Monitor Technology Accuracy in Research: A Comparative Study Between PPG and ECG Technology

Weiler, Dustin T., Villajuan, Stefanie O., Edkins, Laura, Cleary, Sean, Saleem, Jason J.·2017·Proceedings of the Human Factors and Ergonomics Society Annual Meeting

Abstract

Wearable heart rate (HR) monitors typically implement photoplethysmography (PPG) technology and are used in research as an alternative to electrocardiogram (ECG). However, questions surrounding the accuracy of PPG technology exist. To provide an answer regarding the question of accuracy, we conducted a study to compare average HR readings of two different HR technologies (PPG vs. ECG) after an interval style cardio-based workout. A total of 30 trials were conducted and average HR readings from the two HR technologies were compared using an ANOVA. Results revealed no significant difference between the two technologies. However, when HR reached around 155-160 beats per minute, a difference of +/− 5 beats per minute was observed between the two technologies with PPG HR readings being less than ECG. As a result, future research could consider the wearable PPG HR technology as accurate, but with certain experimental design implications.

Integrity signals

Registry

Crossref · 10.1177/1541931213601804

Cites

1 of 1 references verified0 citation errors · 0 not checkable

Statistics

Recomputing p-values needs the paper’s full text.

Claims the paper makes

We couldn’t retrieve this paper’s full text, so its claims and self-check weren’t run. The registry and citation checks still apply.

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • Darker skin contains more melanin, which absorbs more green light than lighter skin.

    “likely because darker skin contains more melanin which absorbs more green light than lighter skin.”
  • Inaccurate PPG heart rate measurements occur up to 15% more frequently in dark skin than in light skin.

    “previous research demonstrated that inaccurate PPG HR measurements occur up to 15% more frequently in dark skin as compared to light skin”

Cites

Flagged first, then by citations

Cited by

Sorted by citations

Authors
Weiler, D. T., Villajuan, S. O., Edkins, L., Cleary, S., & Saleem, J. J.
Affiliations
University of Louisville (US)
Venue
Proceedings of the Human Factors and Ergonomics Society Annual Meeting
Published
2017-09
Volume
61
Issue
1
Pages
1292-1296
ISSN
  • 0163-5182 (ISSN-L)
  • 1071-1813
  • 2169-5067
OpenAlex
W2765157217
Cited by
87
References
1
Citations by year
2026: 3 · 2025: 11 · 2024: 10 · 2023: 13 · 2022: 18 · 2021: 13 · 2020: 5 · 2019: 12 · 2018: 2
FWCI
2.34
Type
journal article
Language
en
Open access
closed
Locations
Proceedings of the Human Factors and Ergonomics Society Annual Meeting (publishedVersion)
Topics
  • Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
  • Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Rate Variability and Autonomic Control
  • Health Sciences › Medicine › Complementary and alternative medicine › Cardiovascular and exercise physiology
Keywords
wearable heart rate monitor, photoplethysmography, electrocardiogram, heart rate accuracy, heart rate response to exercise, measurement error, wearable sensors

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review