Use of Subcutaneous Interstitial Fluid Glucose to Estimate Blood Glucose: Revisiting Delay and Sensor Offset
Open accessAbstract
Background: Estimates for delays in the interstitial fluid (ISF) glucose response to changes in blood glucose (BG) differ substantially among research groups. We review these findings along with arguments that continuous glucose monitoring (CGM) devices used to measure ISF delay contribute to the variability. We consider the impact of the ISF delay and review approaches to correct for it, including strategies pursued by the manufacturers of these devices. The focus on how the manufacturers have approached the problem is motivated by the observation that clinicians and researchers are often unaware of how the existing CGM devices process the ISF glucose signal. Methods: Numerous models and simulations were used to illustrate problems related to measurement and correction of ISF glucose delay. Results: We find that (1) there is no evidence that the true physiologic ISF glucose delay is longer than 5–10 min and that the values longer than this can be explained by delays in CGM filtering routines; (2) the primary impact of the true ISF delay is on sensor calibration algorithms, making it difficult to estimate calibration factors and offset (OS) currents; (3) inaccurate estimates of the sensor OS current result in overestimation of sensor glucose at low values, making it difficult to detect hypoglycemia; (4) many device companies introduce nonlinear components into their filters, which can be expected to confound attempts by investigators to reconstruct BG using linear deconvolution; and (5) algorithms advocated by academic groups are seldom compared to algorithms pursued by industry, making it difficult to ascertain their value. Conclusions: The absence of any direct comparisons between existing and new algorithms for correcting ISF delay and sensor OS current is, in part, due to the difficulty in extracting relevant details from industry patents and/or extracting unfiltered sensor signals from industry products. The model simulation environment, where all aspects of the signal can be derived, may be more appropriate for developing new filtering and calibration strategies. Nevertheless, clinicians, academic researchers, and the industry would benefit from collaborating when evaluating those strategies.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Under normal conditions (i.e., conditions in which glucose levels are not rapidly changing from activities such as exercising or eating), the physiological lag time between blood and interstitial fluid glucose ranges between 5 and 10 min.78,81”
From Biocompatible Materials for Continuous Glucose Monitoring Devices (2013) · cites this paper as [81]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Digital Signal Processing: Principles, Algorithms, and Applications
Proakis, J. G., & Manolakis, D. G. (1992) · Cited by 4,601
Digital control of dynamic systems
Franklin, G. F., Workman, M. L., & Powell, D. (1980) · International Linear Collider · Cited by 3,188
Digital control of dynamic systems
Anderson, B. D. O. (1981) · IEEE Transactions on Acoustics Speech and Signal Processing · doi:10.1109/tassp.1981.1163502 · Cited by 1,695
In Silico Preclinical Trials: A Proof of Concept in Closed-Loop Control of Type 1 Diabetes
Kovatchev, B. P., Breton, M., Dalla Man, C., & Cobelli, C. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300106 · Cited by 760
Boyne, M. S., Silver, D. M., Kaplan, J., & Saudek, C. D. (2003) · Diabetes · doi:10.2337/diabetes.52.11.2790 · Cited by 404
Rebrin, K., Steil, G. M., van Antwerp, W. P., & Mastrototaro, J. J. (1999) · American Journal of Physiology-Endocrinology and Metabolism · doi:10.1152/ajpendo.1999.277.3.e561 · Cited by 367
Self-citationAccuracy of the 5-Day FreeStyle Navigator Continuous Glucose Monitoring System
Weinstein, R. L., Schwartz, S. L., Brazg, R., Bugler, J., Peyser, T. A., & McGarraugh, G. V. (2007) · Diabetes Care · doi:10.2337/dc06-1602 · Cited by 228
Can Interstitial Glucose Assessment Replace Blood Glucose Measurements?
Rebrin, K., & Steil, G. M. (2000) · Diabetes Technology & Therapeutics · doi:10.1089/15209150050194332 · Cited by 227
Self-citationContinuous Glucose Monitoring: Real-Time Algorithms for Calibration, Filtering, and Alarms
Bequette, B. W. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400222 · Cited by 193
Monsod, T. P., Flanagan, D., Rife, F., Saenz, R., Caprio, S., Sherwin, R. S., & Tamborlane, W. V. (2002) · Diabetes Care · doi:10.2337/diacare.25.5.889 · Cited by 173
Research works printed but not linked: 2
- Velho G, 1989, Biomed Biochem Acta., 48, 957
- Youssef JE, Diabetes Technol Ther.
Cited by
Sorted by citations
The Performance and Usability of a Factory-Calibrated Flash Glucose Monitoring System
Bailey, T., Bode, B. W., Christiansen, M. P., Klaff, L. J., & Alva, S. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0378 · Cited by 685
Chang, H., Zheng, M., Yu, X., Than, A., Seeni, R. Z., Kang, R., Tian, J., Duong, P. K., Liu, L., Chen, P., et al. (2017) · Advanced Materials · doi:10.1002/adma.201702243 · Cited by 465
Biocompatible Materials for Continuous Glucose Monitoring Devices
Nichols, S. P., Koh, A., Storm, W. L., Shin, J. H., & Schoenfisch, M. H. (2013) · Chemical Reviews · doi:10.1021/cr300387j · Cited by 347
A Hydrogel Microneedle Patch for Point‐of‐Care Testing Based on Skin Interstitial Fluid
He, R., Niu, Y., Li, Z., Li, A., Yang, H., Xu, F., & Li, F. (2020) · Advanced Healthcare Materials · doi:10.1002/adhm.201901201 · Cited by 261
Soft Bioelectronics Based on Nanomaterials
Cho, K. W., Sunwoo, S., Hong, Y. J., Koo, J. H., Kim, J. H., Baik, S., Hyeon, T., & Kim, D. (2021) · Chemical Reviews · doi:10.1021/acs.chemrev.1c00531 · Cited by 254
Metadata
- Authors
- Rebrin, K., Sheppard, N. F., & Steil, G. M.
- Affiliations
- MicroCHIPS, Inc., Bedford, Massachusetts
- Bedford Research Foundation (US)
- Boston Children's Hospital (US)
- Corresponding author
- Steil, Garry M.
- ORCID
- Steil, Garry M.
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2010-09
- Volume
- 4
- Issue
- 5
- Pages
- 1087-1098
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC2956819
- OpenAlex
- W2028294547
- Cited by
- 152
- References
- 35
- Citations by year
- 2026: 11 · 2025: 9 · 2024: 6 · 2023: 5 · 2022: 11 · 2021: 15 · 2020: 8 · 2019: 14 · 2018: 9 · 2017: 8 · 2016: 9 · 2015: 8 · 2014: 13 · 2013: 13 · 2012: 5
- FWCI
- 6.27
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- Europe PMC (PubMed Central) (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Funders
- National Heart, Lung, and Blood Institute (R01HL88448, R01HL088448)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Physical Sciences › Chemistry › Analytical Chemistry › Electrochemical sensors and biosensors
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Hyperglycemia and glycemic control in critically ill and hospitalized patients
- Keywords
- interstitial glucose, blood glucose estimation, sensor offset, continuous glucose monitoring, sensor calibration, hypoglycemia detection, nonlinear filtering, deconvolution
- MeSH terms
- Algorithms, Animals, Blood Glucose, Calibration, Feasibility Studies, Glucose, Humans, Models, Biological, Monitoring, Physiologic, Signal Processing, Computer-Assisted, Time Factors, United States, United States Food and Drug Administration, Biosensing Techniques, Extracellular Fluid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review