Hypoglycemia Prevention via Pump Attenuation and Red-Yellow-Green “Traffic” Lights Using Continuous Glucose Monitoring and Insulin Pump Data
Open accessThe DOI matches its Crossref record; 5 of the 5 references it cites were found in a registry or an index.
Abstract
Background: Hypoglycemia has been identified as a primary barrier to optimal management of diabetes. This observation, in conjunction with the introduction of continuous glucose monitoring (CGM) devices, has set the stage for achieving tight glycemic control with systems that adjust the insulin pump settings based on measured glucose concentrations. Because system safety and system reliability are key considerations, there is a need for algorithms that reduce the risk of hypoglycemia in closed-loop, open-loop, and advisory-mode systems. More specifically, the algorithm presented here is formulated as a component of the independent safety system module proposed in the modular control-to-range architecture. Methods: We developed two algorithms for attenuating insulin pump injections, which we refer to as Brakes and Power Brakes: Brakes is a pump attenuation function computed using CGM information only, while Power Brakes is an attenuation function in which a metabolic state observer with insulin input is used in addition to CGM information to inform the level of pump attenuation. These algorithms modulate the insulin pump delivery so that the insulin injection rate is dramatically reduced when the risk of hypoglycemia is high. Additionally, we combined these algorithms with an alert system that indicates a level of hypoglycemic risk to the user. Results: We demonstrated the effectiveness of Brakes and Power Brakes in reducing the incidence of hypoglycemia in two simulated scenarios: An elevated basal rate scenario and a scenario in which a bolus is delivered for a meal that is skipped. For these scenarios, the incidence of hypoglycemia using Power Brakes was reduced by 88 and 94%, respectively, where we defined hypoglycemia based on the American Diabetes Association guidelines for defining and reporting as 70 mg/dl. In the elevated basal rate scenario, no rebounds above 180 mg/dl (the desired upper limit of the control-to-range protocol) following hypoglycemia were shown to occur. We demonstrated the way the hypoglycemia alert system can trigger the intake of carbohydrates to reduce the incidence of hypoglycemia by 98%. Conclusions: This article offers, for the first time, a method for smoothly reducing insulin delivery rate to prevent hypoglycemia in individuals with type 1 diabetes mellitus based on a mathematically formal assessment of hypoglycemic risk. In silico, we demonstrate the way this method can prevent hypoglycemia while avoiding hyperglycemia rebounds that exceed 180 mg/dl. In conjunction with the pump attenuation algorithms, this article also proposes a system for alerting an individual of their hypoglycemic risk that contains three “levels” of alerts in the form of a traffic light. This alert system can be used in an advisory mode setting to alert the user to take action when hypoglycemia is imminent (“red” light) or in a closed-loop setting where initiation of rescue action begins when the red light alert is triggered.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Quantitative estimation of insulin sensitivity.
Bergman, R. N., Ider, Y. Z., Bowden, C. R., & Cobelli, C. (1979) · American Journal of Physiology-Endocrinology and Metabolism · doi:10.1152/ajpendo.1979.236.6.e667 · Cited by 1,874
Cryer, P. E., Davis, S. N., & Shamoon, H. (2003) · Diabetes Care · doi:10.2337/diacare.26.6.1902 · Cited by 1,160
Defining and Reporting Hypoglycemia in Diabetes
American Diabetes Association Workgroup on Hypoglycemia (2005) · Diabetes Care · doi:10.2337/diacare.28.5.1245 · Cited by 1,146
Hypoglycaemia: the limiting factor in the glycaemic management of type I and type II diabetes.
Cryer, P. (2002) · Diabetologia · doi:10.1007/s00125-002-0822-9 · Cited by 794
In Silico Preclinical Trials: A Proof of Concept in Closed-Loop Control of Type 1 Diabetes
Kovatchev, B. P., Breton, M., Dalla Man, C., & Cobelli, C. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300106 · Cited by 760
Self-citation
Cited by
Artificial Pancreas: Past, Present, Future
Cobelli, C., Renard, E., & Kovatchev, B. (2011) · Diabetes · doi:10.2337/db11-0654 · Cited by 600
Self-citationCappon, G., Vettoretti, M., Sparacino, G., & Facchinetti, A. (2019) · Diabetes & Metabolism Journal · doi:10.4093/dmj.2019.0121 · Cited by 477
Nocturnal Glucose Control with an Artificial Pancreas at a Diabetes Camp
Phillip, M., Battelino, T., Atlas, E., Kordonouri, O., Bratina, N., Miller, S., Biester, T., Avbelj Stefanija, M., Muller, I., Nimri, R., et al. (2013) · New England Journal of Medicine · doi:10.1056/nejmoa1206881 · Cited by 435
Fully Integrated Artificial Pancreas in Type 1 Diabetes
Breton, M., Farret, A., Bruttomesso, D., Anderson, S., Magni, L., Patek, S., Dalla Man, C., Place, J., Demartini, S., Del Favero, S., et al. (2012) · Diabetes · doi:10.2337/db11-1445 · Cited by 315
“, body weight, insulin-to-carbohydrate ratio, and basal insulin delivery) (24).”
BackgroundMethodSelf-citationAssessing Sensor Accuracy for Non-Adjunct Use of Continuous Glucose Monitoring
Kovatchev, B. P., Patek, S. D., Ortiz, E. A., & Breton, M. D. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0272 · Cited by 234
Self-citation
Metadata
- Authors
- Hughes, C. S., Patek, S. D., Breton, M. D., & Kovatchev, B. P.
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2010-09
- Volume
- 4
- Issue
- 5
- Pages
- 1146-1155
- ISSN
- 1932-2968 (ISSN-L)
- OpenAlex
- W2128394871
- Cited by
- 89
- References
- 5
- FWCI
- 5.72
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Hypoglycemia Prevention via Pump Attenuation and Red-Yellow-Green “Traffic” Lights Using Continuous Glucose Monitoring and Insulin Pump Data. https://app.cochat.ai/papers/10.1177/193229681000400513
- The paper
- Hughes, C. S., Patek, S. D., Breton, M. D., & Kovatchev, B. P. (2010). Hypoglycemia Prevention via Pump Attenuation and Red-Yellow-Green “Traffic” Lights Using Continuous Glucose Monitoring and Insulin Pump Data. Journal of Diabetes Science and Technology. https://doi.org/10.1177/193229681000400513
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.