CoChat

Feasibility of Factory Calibration for Subcutaneous Glucose Sensors in Subjects With Diabetes

Open access
Hoss, Udo, Budiman, Erwin S., Liu, Hanqing, Christiansen, Mark P. · 2014 · Journal of Diabetes Science and Technology
Cited by 53

The DOI matches its Crossref record; 3 of the 3 references it cites were found in a registry or an index.

Abstract

Background: Continuous glucose monitoring using subcutaneously inserted sensors currently requires blood glucose tests for sensor calibration. Alternatively, sensors precalibrated during the manufacturing process may eliminate the need for fingerstick calibrations. In this study we evaluated the feasibility of sensor factory calibration in subjects with diabetes. Methods: A total of 33 subjects with diabetes were asked to wear 4 sensors in parallel, 2 on the arm and 2 on the abdomen. Sensors from a lot with low in vitro sensitivity coefficient of variation were used in the study. Based on frequent capillary blood glucose measurements, the average glucose sensitivity of each sensor was determined over a 5-day wear time. The in vivo sensitivities were analyzed for inter- and intrasubject variation. Mean absolute relative difference (MARD) calculation and consensus error grid analysis (EGA) were performed using a single calibration factor for all sensors, to simulate factory calibration and compared against conventional finger-stick calibration. Results: The sensitivity coefficient of variation between sensors increased from 2.9% in vitro to 6.0% in vivo. No difference in sensor response between subjects ( P = .069) as well as between insertion sites (arm and abdomen) was detected ( P = .104). Applying one calibration factor to all sensors in the study resulted in an MARD of 13.4%, and 83.5% of the values fell in consensus EGA zone A. Multiple fingerstick calibration resulted in an MARD of 12.7% and 84.1% in zone A. Conclusions: Feasibility of factory calibration was demonstrated in subjects with diabetes using sensors based on “wired enzyme” technology, resulting in accuracy metrics similar to sensors calibrated with capillary blood glucose.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1177/1932296813511747

Cites

3 of 3 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 3 references

Cited by

See all 53

Metadata

Authors
Hoss, U., Budiman, E. S., Liu, H., & Christiansen, M. P.
Venue
Journal of Diabetes Science and Technology
Published
2014-01
Volume
8
Issue
1
Pages
89-94
ISSN
1932-2968 (ISSN-L)
OpenAlex
W2151247283
Cited by
53
References
3
FWCI
5.03
Type
journal article
Language
en
Open access
green

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Feasibility of Factory Calibration for Subcutaneous Glucose Sensors in Subjects With Diabetes. https://app.cochat.ai/papers/10.1177/1932296813511747
The paper
Hoss, U., Budiman, E. S., Liu, H., & Christiansen, M. P. (2014). Feasibility of Factory Calibration for Subcutaneous Glucose Sensors in Subjects With Diabetes. Journal of Diabetes Science and Technology. https://doi.org/10.1177/1932296813511747

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.