A Comparative Effectiveness Analysis of Three Continuous Glucose Monitors: The Navigator, G4 Platinum, and Enlite
Open accessThe DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.
Abstract
Background: The effectiveness and safety of continuous glucose monitors (CGMs) is dependent on their accuracy and reliability. The objective of this study was to compare 3 CGMs in adult and pediatric subjects with type 1 diabetes under closed-loop blood-glucose (BG) control. Twenty-four subjects (12 adults) with type 1 diabetes each participated in one 48-hour closed-loop BG control experiment. Methods: Venous plasma glucose (PG) measurements obtained every 15 minutes (4657 values) were paired in time with corresponding CGM glucose (CGMG) measurements obtained from 3 CGMs (FreeStyle Navigator, Abbott Diabetes Care; G4 Platinum, Dexcom; Enlite, Medtronic) worn simultaneously by each subject. Results: The Navigator and G4 Platinum (G4) had the best overall accuracy, with an aggregate mean absolute relative difference (MARD) of all paired points of 12.3 ± 12.1% and 10.8 ± 9.9%, respectively. Both had lower MARDs of all paired points than Enlite (17.9 ± 15.8%, P < .005). Very large errors (MARD > 50%) were less common with the G4 (0.5%) than with the Enlite (4.3%, P = .0001) while the number of very large errors with the Navigator (1.4%) was intermediate between the G4 and Enlite ( P = .1 and P = .06, respectively). The average MARD for experiments in adolescent subjects were lower than in adult subjects for the Navigator and G4, while there was no difference for Enlite. All 3 devices had similar reliability. Conclusions: A comprehensive head-to-head-to-head comparison of 3 CGMs revealed marked differences in both accuracy and precision. The Navigator and G4 were found to outperform the Enlite in these areas.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Helton, K. L., Ratner, B. D., & Wisniewski, N. A. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500318 · Cited by 102
Cited by
Continuous Glucose Monitoring: A Review of Successes, Challenges, and Opportunities
Rodbard, D. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0417 · Cited by 737
The Performance and Usability of a Factory-Calibrated Flash Glucose Monitoring System
Bailey, T., Bode, B. W., Christiansen, M. P., Klaff, L. J., & Alva, S. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0378 · Cited by 685
Continuous glucose monitoring: a review of recent studies demonstrating improved glycemic outcomes
Rodbard, D. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0035 · Cited by 458
Diabetes: Advances in Diagnosis and Treatment
Nathan, D. M. (2015) · JAMA · doi:10.1001/jama.2015.9536 · Cited by 384
Self-citationAleppo, G., Ruedy, K. J., Riddlesworth, T. D., Kruger, D. F., Peters, A. L., Hirsch, I., Bergenstal, R. M., Toschi, E., Ahmann, A. J., Shah, V. N., et al. (2017) · Diabetes Care · doi:10.2337/dc16-2482 · Cited by 329
Metadata
- Authors
- Damiano, E. R., McKeon, K., El-Khatib, F. H., Zheng, H., Nathan, D. M., & Russell, S. J.
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2014-04-21
- Volume
- 8
- Issue
- 4
- Pages
- 699-708
- ISSN
- 1932-2968 (ISSN-L)
- OpenAlex
- W2136635912
- Cited by
- 139
- References
- 1
- FWCI
- 16.99
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.