A Multicenter Evaluation of the Performance and Usability of a Novel Glucose Monitoring System in Chinese Adults With Diabetes
Open accessThe DOI matches its Crossref record; 7 of the 7 references it cites were found in a registry or an index.
Abstract
Objective: Flash glucose monitoring is a new glucose sensing technique that measures interstitial glucose levels for up to 14 days and does not require any calibration. The aim of this study is to evaluate the performance of the new system in Chinese patients with diabetes. Methods: A multicenter, prospective, masked study was performed in a total of 45 subjects with diabetes. Subjects wore 2 sensors at the same time, for up to 14 days. The accuracy was evaluated against capillary blood glucose (BG) and venous Yellow Springs Instrument (YSI; Yellow Springs, OH) measurements. During all 14 days, subjects were asked to perform at least 8 capillary BG tests per day. Each subject attended 3 days of 8-hour clinic sessions to measure YSI and sensor readings every 15 minutes. Results: Forty subjects had evaluable glucose readings, with 6687 of 6696 (99.9%) sensor and capillary BG pairs within consensus error grid zones A and B, including 5824 (87.0%) in zone A. The 6969 sensor and venous YSI pairs resulted in 6965 (99.9%) pairs within zones A and B, including 5755 (82.6%) in zone A. The sensor pairs with BG and YSI result in mean absolute relative difference (MARD) of 10.0% and 10.7%, respectively. Overall between-sensor coefficient of variation (CV) was 8.0%, and the mean lag time was 3.1 (95% confidence interval 2.54 to 4.29) minutes. Conclusions: The system works well for people with diabetes in China, and it is easy to wear and use.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Evaluating clinical accuracy of systems for self-monitoring of blood glucose
Clarke, W. L., Cox, D., Gonder-Frederick, L. A., Carter, W., & Pohl, S. L. (1987) · Diabetes Care · doi:10.2337/diacare.10.5.622 · Cited by 1,435
The Performance and Usability of a Factory-Calibrated Flash Glucose Monitoring System
Bailey, T., Bode, B. W., Christiansen, M. P., Klaff, L. J., & Alva, S. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0378 · Cited by 685
Parkes, J. L., Slatin, S. L., Pardo, S., & Ginsberg, B. H. (2000) · Diabetes Care · doi:10.2337/diacare.23.8.1143 · Cited by 620
Weinstock, R. S., Xing, D., Maahs, D. M., Michels, A., Rickels, M. R., Peters, A. L., Bergenstal, R. M., Harris, B., DuBose, S. N., Miller, K. M., et al. (2013) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2013-1589 · Cited by 338
Evaluating the Accuracy of Continuous Glucose-Monitoring Sensors
Kovatchev, B. P., Gonder-Frederick, L. A., Cox, D. J., & Clarke, W. L. (2004) · Diabetes Care · doi:10.2337/diacare.27.8.1922 · Cited by 312
Vigersky, R. A., Fonda, S. J., Chellappa, M., Walker, M. S., & Ehrhardt, N. M. (2012) · Diabetes Care · doi:10.2337/dc11-1438 · Cited by 301
Ehrhardt, N. M., Chellappa, M., Walker, M. S., Fonda, S. J., & Vigersky, R. A. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500320 · Cited by 199
Cited by
Reddy, M., Jugnee, N., El Laboudi, A., Spanudakis, E., Anantharaja, S., & Oliver, N. (2018) · Diabetic Medicine · doi:10.1111/dme.13561 · Cited by 245
7. Diabetes Technology: Standards of Medical Care in Diabetes—2019
American Diabetes Association (2019) · Diabetes Care · doi:10.2337/dc19-s007 · Cited by 179
Freckmann, G., Eichenlaub, M., Waldenmaier, D., Pleus, S., Wehrstedt, S., Haug, C., Witthauer, L., Jendle, J., Hinzmann, R., Thomas, A., et al. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231190941 · Cited by 86
Anjana, R. M., Kesavadev, J., Neeta, D., Tiwaskar, M., Pradeepa, R., Jebarani, S., Thangamani, S., Sastry, N. G., Brijendra Kumar, S., Ramu, M., et al. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0063 · Cited by 42
Evaluation of flash glucose monitoring after long-term use: a pediatric survey
Vergier, J., Samper, M., Dalla-Vale, F., Ventura, V., Baucher, F., Joubert, F., Pons, C., De Oliveira, M., Salet, R., Faure-Galon, N., et al. (2019) · Primary Care Diabetes · doi:10.1016/j.pcd.2018.08.004 · Cited by 32
Metadata
- Authors
- Ji, L., Guo, X., Guo, L., Ren, Q., Yu, N., & Zhang, J.
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2016-09-25
- Volume
- 11
- Issue
- 2
- Pages
- 290-295
- ISSN
- 1932-2968 (ISSN-L)
- OpenAlex
- W2509560015
- Cited by
- 61
- References
- 7
- FWCI
- 6.64
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: A Multicenter Evaluation of the Performance and Usability of a Novel Glucose Monitoring System in Chinese Adults With Diabetes. https://app.cochat.ai/papers/10.1177/1932296816662884
- The paper
- Ji, L., Guo, X., Guo, L., Ren, Q., Yu, N., & Zhang, J. (2017). A Multicenter Evaluation of the Performance and Usability of a Novel Glucose Monitoring System in Chinese Adults With Diabetes. Journal of Diabetes Science and Technology. https://doi.org/10.1177/1932296816662884
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.