Evaluating the Accuracy of Continuous Glucose-Monitoring Sensors

Open access
Kovatchev, Boris P., Gonder-Frederick, Linda A., Cox, Daniel J., Clarke, William L.·2004·Diabetes Care

Abstract

OBJECTIVE—The objective of this study was to introduce continuous glucose–error grid analysis (CG-EGA) as a method of evaluating the accuracy of continuous glucose-monitoring sensors in terms of both accurate blood glucose (BG) values and accurate direction and rate of BG fluctuations and to illustrate the application of CG-EGA with data from the TheraSense Freestyle Navigator. RESEARCH DESIGN AND METHODS—We approach the design of CG-EGA from the understanding that continuous glucose sensors (CGSs) allow the observation of BG fluctuations as a process in time. We account for specifics of process characterization (location, speed, and direction) and for biological limitations of the observed processes (time lags associated with interstitial sensors). CG-EGA includes two interacting components: 1) point–error grid analysis (P-EGA) evaluates the sensor’s accuracy in terms of correct presentation of BG values and 2) rate–error grid analysis (R-EGA) assesses the sensor’s ability to capture the direction and rate of BG fluctuations. RESULTS—CG-EGA revealed that the accuracy of the Navigator, measured as a percentage of accurate readings plus benign errors, was significantly different at hypoglycemia (73.5%), euglycemia (99%), and hyperglycemia (95.4%). Failure to detect hypoglycemia was the most common error. The point accuracy of the Navigator was relatively stable over a wide range of BG rates of change, and its rate accuracy decreased significantly at high BG levels. CONCLUSIONS—Traditional self-monitoring of BG device evaluation methods fail to capture the important temporal characteristics of the continuous glucose-monitoring process. CG-EGA addresses this problem, thus providing a comprehensive assessment of sensor accuracy that appears to be a useful adjunct to other CGS performance measures.

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.2337/diacare.27.8.1922

Cites

13 of 13 references verified0 citation errors · 0 not checkable

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • Accuracy of CGM should be assessed from three viewpoints: the analytical viewpoint, clinical accuracy, and accuracy of the trend.

    “Other authors have emphasized the need to assess accuracy from 3 viewpoints: the analytical viewpoint (i.e., evaluation of deviations between CGM values and blood glucose concentrations measured with a suitable comparator), the clinical accuracy, and the accuracy of the trend (i.e., the characterization of deviations between the glucose concentration rate of change and blood glucose concentrations”
  • The continuous glucose-error grid comprises a point-error grid analysis for blood glucose measurement accuracy and a rate-error grid analysis for prediction capability.

    “Because CGM systems also provide information on glucose fluctuations, the continuous glucose-error grid (CG-EGA) has been introduced comprising of (1) a point-error grid analysis (P-EGA) that evaluates the sensor accuracy in terms of accurate blood glucose measurements; and, (2) a rate-error grid analysis (R-EGA) that assesses the prediction capability of the sensor.77”

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Cites

Flagged first, then by citations

Cited by

Sorted by citations

Authors
Kovatchev, B. P., Gonder-Frederick, L. A., Cox, D. J., & Clarke, W. L.
Affiliations
University of Virginia Health System (US)
Venue
Diabetes Care
Published
2004-08-01
Volume
27
Issue
8
Pages
1922-1928
ISSN
  • 0149-5992 (ISSN-L)
  • 1935-5548
OpenAlex
W2116146322
Cited by
312
References
13
Citations by year
2026: 8 · 2025: 10 · 2024: 11 · 2023: 7 · 2022: 5 · 2021: 8 · 2020: 9 · 2019: 12 · 2018: 12 · 2017: 13 · 2016: 10 · 2015: 15 · 2014: 15 · 2013: 26 · 2012: 23
FWCI
8.89
Type
journal article
Language
en
Open access
Locations
  • Diabetes Care (publishedVersion)
  • PubMed (publishedVersion)
Funders
  • National Institutes of Health (5r01dk051562-23)
  • National Institute of Diabetes and Digestive and Kidney Diseases (R01DK 51562, RO1 DK 51562)
Topics
  • Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
  • Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Hyperglycemia and glycemic control in critically ill and hospitalized patients
  • Health Sciences › Medicine › Physiology › Clinical Laboratory Practices and Quality Control
Keywords
continuous glucose monitoring, sensor accuracy, error grid analysis, blood glucose fluctuation, hypoglycemia detection, time delay, self-monitoring of blood glucose
MeSH terms
Adult, Aged, Blood Glucose, Clinical Trials as Topic, Diabetes Mellitus, Type 1, Female, Humans, Male, Middle Aged, Reproducibility of Results, Monitoring, Ambulatory

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review