Time Savings Using a Standardized Glucose Reporting System and Ambulatory Glucose Profile
Open accessAbstract
BACKGROUND: Diabetes care is predominately done at home by the patient. When clinics do not have a reliable, easy process for obtaining this patient data, clinical decisions must be made with incomplete verbal recall reports. Unused or inaccessible glucose data represent a large information gap affecting clinical decision making. This study's purpose was to design an optimized glucose device download system with a standardized report and to evaluate its efficiency. METHODS: Observations and evaluations of glucose data retrieval occurred at two clinics; an additional clinic utilized the optimized process doing only post process timings. Patients/families and clinicians were surveyed about their experiences with the system and the standardized report (AGP). The study was approved by all the sites' IRBs. RESULTS: Optimized systems saved staff at least 3 min per patient. Standardized AGP reports and an optimized data system made the work flow of glucose data easier to complete. The AGP report was preferred by patients, families, and clinicians. CONCLUSIONS: An optimized system takes advantage of patient lobby downtime to download glucose devices and ensures that diabetes clinical decisions are made utilizing all available data. Staff and patients liked the software lobby system and found it a valuable time-saving tool.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The AGP recommendation was endorsed at the international consensus conference on CGM metrics and referenced in ADA 2019 Standards of Care and AACE consensus.
“This recommendation was later endorsed at the aforementioned international consensus conference on CGM metrics (18) and is referenced as an example in the American Diabetes Association (ADA) 2019 Standards of Care (16) and in an update to the American Association of Clinical Endocrinologists (AACE) consensus on use of CGM (69).”
From Clinical Targets for Continuous Glucose Monitoring Data Interpretation: Recommendations From the International Consensus on Time in Range (2019) · cites this paper as [69]
Cites
Abdulmalek, F., & Rajgopal, J. (2007) · International Journal of Production Economics · doi:10.1016/j.ijpe.2006.09.009 · Cited by 1,181 · see citing works
Continuous Glucose Monitoring: A Review of Successes, Challenges, and Opportunities
Rodbard, D. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0417 · Cited by 743 · see citing works
Ziegler, R., Heidtmann, B., Hilgard, D., Hofer, S., Rosenbauer, J., Holl, R., & DPV-Wiss-Initiative, F. T. (2011) · Pediatric Diabetes · doi:10.1111/j.1399-5448.2010.00650.x · Cited by 334 · see citing works
Bergenstal, R. M., Ahmann, A. J., Bailey, T., Beck, R. W., Bissen, J., Buckingham, B., Deeb, L., Dolin, R. H., Garg, S. K., Goland, R., et al. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0051 · Cited by 307 · see citing works
Self-citationAmbulatory Glucose Profile: Representation of Verified Self-Monitored Blood Glucose Data
Mazze, R. S., Lucido, D., Langer, O., Hartmann, K., & Rodbard, D. (1987) · Diabetes Care · doi:10.2337/diacare.10.1.111 · Cited by 144 · see citing works
Bergenstal, R. M. (2015) · Diabetes Care · doi:10.2337/dc15-0099 · Cited by 140 · see citing works
Self-citationMonitoring glycemic control: the importance of self-monitoring of blood glucose
Renard, E. (2005) · The American Journal of Medicine · doi:10.1016/j.amjmed.2005.07.052 · Cited by 99 · see citing works
Lean sigma--will it work for healthcare?
Bahensky, J. A., Roe, J., & Bolton, R. (2005) · PubMed · Cited by 83 · see citing works
Consensus recommendations for the use of Ambulatory Glucose Profile in clinical practice
Matthaei, S., Antuna, R., Bosi, E., Evans, M. L., Geelhoed-Duijvestijn, N., & Joubert, M. (2014) · The British Journal of Diabetes · doi:10.15277/bjdvd.2014.046 · Cited by 38 · see citing works
Assessing the value of the Ambulatory Glucose Profile in clinical practice
Matthaei, S. (2014) · The British Journal of Diabetes · doi:10.15277/bjdvd.2014.045 · Cited by 27 · see citing works
Ambulatory glucose profile (AGP): utility in UK clinical practice
Evans, M. L., CRANSTON, I. C., & Bailey, C. J. (2017) · British Journal of Diabetes · doi:10.15277/bjd.2017.121 · Cited by 18 · see citing works
McElwee-Malloy, M. (2016) · PubMed · Cited by 1 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Battelino, T., Danne, T., Bergenstal, R. M., Amiel, S. A., Beck, R., Biester, T., Bosi, E., Buckingham, B. A., Cefalu, W. T., Close, K. L., et al. (2019) · Diabetes Care · doi:10.2337/dci19-0028 · Cited by 4,032 · see citing works
Self-citationLatres, E., Finan, D. A., Greenstein, J. L., Kowalski, A. J., & Kieffer, T. J. (2019) · Cell Metabolism · doi:10.1016/j.cmet.2019.02.007 · Cited by 118 · see citing works
Johnson, M. L., Martens, T. W., Criego, A. B., Carlson, A. L., Simonson, G. D., & Bergenstal, R. M. (2019) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2019.0034 · Cited by 104 · see citing works
“19 The ambulatory glucose profile (AGP) is a standardized, single-page glucose report designed to simplify analysis and interpretation of downloaded CGM data.”
Self-citationAleppo, G., & Webb, K. M. (2018) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818813581 · Cited by 54 · see citing works
“Standard deviation and coefficient of variation to determine glycemic variability and increase risk for hypoglycemia.(53,54) 5.”
Espinoza, J., Shah, P., & Raymond, J. K. (2020) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2019.0377 · Cited by 40 · see citing works
Leveraging advances in diabetes technologies in primary care: a narrative review
Bode, B. W., King, A., Russell-Jones, D. L., & Billings, L. K. (2021) · Annals of Medicine · doi:10.1080/07853890.2021.1931427 · Cited by 39 · see citing works
Inducible Pluripotent Stem Cells as a Potential Cure for Diabetes
Verhoeff, K., Henschke, S., Marfil-Garza, B., Dadheech, N., & Shapiro, A. M. (2021) · Cells · doi:10.3390/cells10020278 · Cited by 21 · see citing works
“â Improved dynamic glycemic understanding (real time glycemic targets, time spent in euglycemia, hypoglycemia, or hyperglycemia) [5,12,13] â Hyper/hypo glycemic alarms.”
BackgroundUtilizing continuous glucose monitoring in primary care practice: What the numbers mean
Edelman, S. V., Cavaiola, T. S., BOEDER, S. C., & Pettus, J. (2020) · Primary care diabetes · doi:10.1016/j.pcd.2020.10.013 · Cited by 25 · see citing works
The Ambulatory Glucose Profile: Opportunities for Enhancement
Rodbard, D. (2020) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2020.0524 · Cited by 20 · see citing works
Aleppo, G., Chmiel, R., Zurn, A., Bandoske, R., Creamer, P., Neubauer, N., Wong, J., Andrade, S. B., & Hauptman, A. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231196168 · Cited by 19 · see citing works
Role of Continuous Glucose Monitoring in Insulin-Requiring Patients with Diabetes
Welsh, J. B. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0100 · Cited by 14 · see citing works
“41 Data from Dexcom Clarity suite of mobile and web-based software for retrospectively sharing and analyzing CGM data confirm the benefits of these behaviors: in a recent analysis of 50,000 users of the Dexcom G5 Mobile app, patients who logged in four or more times to Clarity during August 2017 had…”
Standardized Hybrid Closed-Loop System Reporting
Shah, V. N., & Garg, S. K. (2020) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2020.0622 · Cited by 13 · see citing works
“33 Moreover, presenting nonstandardized data discourages providers and patients from downloading the devices, which may result in failure to optimize diabetes care.”
Integration of Diabetes Technology in Clinical Practice
Peters, A. L. (2019) · Endocrinology and Metabolism Clinics of North America · doi:10.1016/j.ecl.2019.10.002 · Cited by 8 · see citing works
Deeb, A., Muammar, T., Alsaffar, H., Sedaghat, S., Hassani, N. A., Odeh, R., Alkhayyat, H., Sinani, A. A., Attia, N. A., Adhami, S., et al. (2021) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2021.108671 · Cited by 7 · see citing works
Yap, P. X. C., Koh, L., Gallagher, P., & Lum, Z. K. (2021) · Journal of Diabetology · doi:10.4103/jod.jod_49_21 · Cited by 4 · see citing works
Kontinuierliche Glukosemessung
Gehr, B., & Biester, T. (2018) · Der Diabetologe · doi:10.1007/s11428-018-0383-2 · Cited by 2 · see citing works
Mackett, K., Gerstein, H. C., & Santesso, N. A. (2023) · Canadian Journal of Diabetes · doi:10.1016/j.jcjd.2023.01.001 · Cited by 2 · see citing works
Maines, E., Pertile, R., Cauvin, V., Soffiati, M., & Franceschi, R. (2024) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2024.111720 · Cited by 2 · see citing works
The Role of a “Technology Navigator” in an Academic Diabetes Clinic: A Feasibility Evaluation
Edmiston, P., Klippel, S. L., Baran, J. D., Khakpour, D., Hirsch, I. B., & Trikudanathan, S. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0406 · Cited by 1 · see citing works
New Trends: Time in Range and the Use of Continuous Glucose Monitoring Devices on Glycemic Control
Souza, K. L. V. D., Rassi, M. M. B., de Sá, D. A. R., Bussuan, R. M., Rodrigues, L. C. D. S., & Arbex, A. K. (2021) · European Journal of Medical and Health Sciences · doi:10.24018/ejmed.2021.3.1.491
“complications, it will help to prevent new complications or worsening of existing ones [140]-[144].”
BackgroundMethodAanstoot, H., Last, S., Riegman, N., Mul, D., Hoogendam, A., Akkerman, M., & Veeze, H. J. (2023) · Nederlands Tijdschrift voor Diabetologie · doi:10.1007/s12467-023-1216-5
Technological innovations in the treatment of diabetes over the last decade
Sonia, T. A., & Sharma, C. P. (2025) · Elsevier eBooks · doi:10.1016/b978-0-443-30272-5.00011-1
- Authors
- Mullen, D. M., Bergenstal, R., Criego, A., Arnold, K. C., Goland, R., & Richter, S.
- Affiliations
- International Diabetes Center, St. Louis Park, MN, USA
- Park Nicollet Clinic (US)
- The Diabetes Center, Ocean Springs, MS, USA
- Columbia University Irving Medical Center (US)
- Professional Data Analysts, Inc, Minneapolis, MN, USA
- Corresponding author
- Mullen, Deborah M.
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2017-11-24
- Volume
- 12
- Issue
- 3
- Pages
- 614-621
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC6154242
- OpenAlex
- W2769928215
- Cited by
- 22
- References
- 12
- Citations by year
- 2025: 1 · 2024: 1 · 2023: 4 · 2021: 5 · 2020: 4 · 2019: 4 · 2018: 3
- FWCI
- 2.38
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- Europe PMC (PubMed Central) (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Funders
- Leona M. and Harry B. Helmsley Charitable Trust
- Dexcom
- Abbott Diabetes Care
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Health Professions › Health Information Management › Electronic Health Records Systems
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
- Keywords
- ambulatory glucose profile, diabetes management, clinical workflow efficiency, home blood glucose monitoring
- MeSH terms
- Blood Glucose, Child, Female, Humans, Male, Pilot Projects, Software, Blood Glucose Self-Monitoring, Workflow
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review