Performance of a Factory-Calibrated, Real-Time Continuous Glucose Monitoring System in Pediatric Participants With Type 1 Diabetes
Open accessThe DOI matches its Crossref record; 5 of the 5 references it cites were found in a registry or an index.
Abstract
Background: The perceived value and consistent use of continuous glucose monitoring (CGM) systems depends in part on their accuracy. We assessed the performance of a sixth-generation CGM system (Dexcom G6) in children and adolescents. Methods: Forty-nine participants (6-17 years of age, mean ± SD of 13.5 ± 3.3 years), all with type 1 diabetes, enrolled and data were available from 37. Each participant wore 1 sensor for up to 10 days and was asked to undergo an in-clinic visit lasting 6-12 hours for frequent blood glucose (BG) sample testing on one of the sensor wear days. Estimated glucose values (EGVs) from the G6 system were compared with venous BG values measured with a laboratory reference instrument (YSI). Results: The overall mean absolute relative difference (MARD) for 1387 EGV-YSI pairs was 7.7%, and the overall percentage of EGVs within 20% or 20 mg/dL of the YSI reference value (for YSI > or ⩽100 mg/dL, respectively, the “%20/20”) was 96.2%. The %20/20 was 92.1% on Day 1 and 91.0% on Day 10 of sensor wear. For EGVs <70 mg/dL, 92.6% of the YSI values were within 20 mg/dL and for EGVs >250 mg/dL, 100% of the YSI values were within 20%. Differences between EGVs and YSI values in over 99.9% of the pairs posed no or only slight clinical risk as evaluated by surveillance error grid analysis. Conclusions: The accuracy of the G6 CGM system in pediatrics may encourage consistent use of the system and contribute to improved glycemic outcomes in this population.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Continuous glucose monitoring and intensive treatment of type 1 diabetes
(2008) · New England Journal of Medicine · doi:10.1056/nejmoa0805017 · Cited by 1,559
Agiostratidou, G., Anhalt, H., Ball, D., Blonde, L., Gourgari, E., Harriman, K. N., Kowalski, A. J., Madden, P., McAuliffe-Fogarty, A. H., McElwee-Malloy, M., et al. (2017) · Diabetes Care · doi:10.2337/dc17-1624 · Cited by 397
Mauras, N., Beck, R., Xing, D., Ruedy, K., Buckingham, B., Tansey, M., White, N. H., Weinzimer, S. A., Tamborlane, W., Kollman, C., et al. (2012) · Diabetes Care · doi:10.2337/dc11-1746 · Cited by 204
Tsalikian, E., Fox, L., Weinzimer, S., Buckingham, B., White, N. H., Beck, R., Kollman, C., Xing, D., Ruedy, K., & Diabetes Research in Children Network (DirecNet) Study Group (2012) · Pediatric Diabetes · doi:10.1111/j.1399-5448.2011.00837.x · Cited by 83
Cited by
One year clinical experience of the first commercial hybrid closed-loop system
Lal, R. A., Basina, M., Maahs, D. M., Hood, K., Buckingham, B., & Wilson, D. M. (2019) · Diabetes Care · doi:10.2337/dc19-0855 · Cited by 215
“Dexcom released the G6 system, which is the first CGM not requiring calibrations, resistant to acetaminophen interference, andapproved for nonadjunctive use (17,18).”
BackgroundAccuracy and Safety of Dexcom G7 Continuous Glucose Monitoring in Adults with Diabetes
Garg, S. K., Kipnes, M., Castorino, K., Bailey, T. S., Akturk, H. K., Welsh, J. B., Christiansen, M. P., Balo, A. K., Brown, S. A., Reid, J. L., et al. (2022) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2022.0011 · Cited by 179
Self-citationFreckmann, G., Eichenlaub, M., Waldenmaier, D., Pleus, S., Wehrstedt, S., Haug, C., Witthauer, L., Jendle, J., Hinzmann, R., Thomas, A., et al. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231190941 · Cited by 86
Karakus, K. E., Akturk, H. K., Alonso, G. T., Snell-Bergeon, J. K., & Shah, V. N. (2023) · Diabetes Care · doi:10.2337/dc23-0495 · Cited by 52
Baker, M., Musselman, M. E., Rogers, R., & Hellman, R. (2022) · American Journal of Health-System Pharmacy · doi:10.1093/ajhp/zxab456 · Cited by 27
Metadata
- Authors
- Welsh, J. B., Zhang, X., Puhr, S. A., Johnson, T. K., Walker, T. C., Balo, A. K., & Price, D.
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2018-09-10
- Volume
- 13
- Issue
- 2
- Pages
- 254-258
- ISSN
- 1932-2968 (ISSN-L)
- OpenAlex
- W2889768647
- Cited by
- 33
- References
- 5
- FWCI
- 2.05
- Type
- journal article
- Language
- en
- Open access
- hybrid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Performance of a Factory-Calibrated, Real-Time Continuous Glucose Monitoring System in Pediatric Participants With Type 1 Diabetes. https://app.cochat.ai/papers/10.1177/1932296818798816
- The paper
- Welsh, J. B., Zhang, X., Puhr, S. A., Johnson, T. K., Walker, T. C., Balo, A. K., & Price, D. (2019). Performance of a Factory-Calibrated, Real-Time Continuous Glucose Monitoring System in Pediatric Participants With Type 1 Diabetes. Journal of Diabetes Science and Technology. https://doi.org/10.1177/1932296818798816
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.