Evaluation of Linearity and Interference Effect on SMBG and POCT Devices, Showing Drastic High Values, Low Values, or Error Messages
Open accessThe DOI matches its Crossref record; 7 of the 7 references it cites were found in a registry or an index.
Abstract
Background: When a patient with diabetes measures blood glucose levels using a simple blood glucose meter, an error message or abnormal value may be displayed because of interfering substances, potentially leading to unnecessary medical treatment. Methods: Here, we tested 10 self-monitoring of blood glucose (SMBG) and point-of-care testing (POCT) devices to investigate under what conditions they fail to measure blood glucose levels or display abnormal values instead. We also evaluated the influence and linearity of several well-known interfering substances and hematocrit (Ht). Each adjusted sample was measured by each device twice and the values compared with the reference values. Results: Linearity was favorable in all but one model. The impact of the interfering substances and Ht has been confirmed in many SMBG and POCT devices. In particular, some models showed significant pseudo-high or -low levels for samples that contained ascorbic acid, which is used widely in infusion solutions. Three devices showed error messages to avoid presenting a false value. Conclusions: Since pseudo-high levels may result in the excessive administration of insulin and lead to adverse events, this report serves as a critical reminder of the importance of device choice for specific indications. Three new devices have safety functions that prevent the showing of false values.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Many meters incorporate changes that eliminate or greatly ameliorate most interferences, but interferences persist (137,138).”
From Guidelines and Recommendations for Laboratory Analysis in the Diagnosis and Management of Diabetes Mellitus (2023) · cites this paper as [137]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Miller, K. M., Beck, R. W., Bergenstal, R. M., Goland, R. S., Haller, M. J., McGill, J. B., Rodriguez, H., Simmons, J. H., Hirsch, I. B., & for the T1D Exchange Clinic Network (2013) · Diabetes Care · doi:10.2337/dc12-1770 · Cited by 505
Ziegler, R., Heidtmann, B., Hilgard, D., Hofer, S., Rosenbauer, J., Holl, R., & DPV-Wiss-Initiative, F. T. (2011) · Pediatric Diabetes · doi:10.1111/j.1399-5448.2010.00650.x · Cited by 334
Grant, R. W., & Kirkman, M. S. (2015) · Diabetes Care · doi:10.2337/dc14-2142 · Cited by 190
Harashima, S., Fukushima, T., Sasaki, M., Nishi, Y., Fujimoto, S., Ogura, M., Yamane, S., Tanaka, D., Harada, N., Hamasaki, A., et al. (2013) · Diabetes/Metabolism Research and Reviews · doi:10.1002/dmrr.2363 · Cited by 26
Sekido, K., Sekido, T., Kaneko, A., Hosokawa, M., Sato, A., Sato, Y., Yamazaki, M., & Komatsu, M. (2017) · Endocrine Journal · doi:10.1507/endocrj.ej17-0193 · Cited by 23
Kim, S. K., Hahm, J. R., Kim, H., Kim, S., Jung, T. S., Jung, J. H., Lee, H. R., & Kim, D. R. (2013) · Yonsei Medical Journal · doi:10.3349/ymj.2013.54.5.1289 · Cited by 16
Interference by Pralidoxime (PAM) salts in clinical laboratory tests
Nagase, S., Kohguchi, K., Tohyama, K., Watanabe, M., & Iwatani, Y. (2012) · Clinica Chimica Acta · doi:10.1016/j.cca.2012.11.017 · Cited by 4
Research works printed but not linked: 15
- 1. American Diabetes Association. Standards of medical care in diabetes—2015. Diabetes Care. 2015;38:S1-S94. [DOI] [PubMed] [Google Scholar]
- 7. Kawashima H. Effect of hematocrit on devices for self-monitoring of blood glucose: comparison between devices compliant with the new and old ISO15197. J Med Pharm Sci. 2016;73:211-220. [Google Scholar]
- 8. Usami M, Tateishi A, Yamauchi M, et al. Evaluation of the user-friendliness and performance of new devices for self-monitoring of blood glucose with colored liquid crystal display. J Clin Lab Instr Reag. 2013;36:207-216. [Google Scholar]
- 9. Nagatomo R, Kaneko M, Saito Y, et al. Study of the factors contributing to errors and differences between models of devices for self-monitoring of blood glucose. Practice. 2013;30:103-112. [Google Scholar]
- 10. Tsuchida K, Ishida T, Ishikawa T, et al. Adequate knowledge and selection of blood glucose meters: information that helps alleviate stress in diabetic patients. Jap J Med Tech. 2017;66:110-116. [Google Scholar]
- 11. Fujisaki N, Otsuji M, Minobe M, et al. Imported hand cream containing various reducing substances, and its effect on the measured values obtained from self-monitoring of blood glucose. J Jap Diabet Soc. 2013;56:666-669. [Google Scholar]
- 14. Miura F, Yoshino M, Tomioka M, et al. Clinical evaluation of 5 types of simple glucometers: blood glucose levels, patients, and effects of the measurement environment. J Jap Diabet Soc. 2009;52:865-870. [Google Scholar]
- 15. Yamazaki. Separation of SMBG devices and POCT-compliant blood-glucose measurement devices. J Clin Lab Instr Reag. 2009;32:707-713. [Google Scholar]
- 16. POC Promotion Committee, Japan Society for Clinical Laboratory Automation (2013) POCT guidelines. 3rd ed. J Jap Soc Clin Lab Automat. 2013;38(suppl 1):8-115. [Google Scholar]
- 17. David B. POCT12-A3 Point-of-Care Blood Glucose Testing in Acute and Chronic Care Facilities; Approved Guideline. 3rd ed. Wayne, PA: Clinical and Laboratory Standards Institute; 2013. [Google Scholar]
- 18. Mitsui T, Hirose Y, Miyahara S, et al. Survey of the accuracy and usability of the latest blood glucose measurement devices for POCT. J New Remed Clin. 2013;62:2194-2201. [Google Scholar]
- 19. Yoshikawa Y, Kuribara S, Matsumoto S, Matsuda M, Hosokawa N, Kokubo T. Study of interfering substances and insufficient amounts of blood samples in the use of simple glucometers, and caution pertaining to their use in hospital settings. J Clin Lab Instr Reag. 2006;29:573-581. [Google Scholar]
- 20. Koide K, Nojima J, Kodama K, Atsumi Y. Basic consideration to make use of flash glucose monitoring systems for medical treatment guidance. Diabetes. 2017;66(suppl 1):A183. [Google Scholar]
- 24. Miura F, Yoshino M, Tomioka M, et al. Accuracy of five self-monitoring blood glucose devices. Diabetes Japan. 2009;52:865-870. [Google Scholar]
- 26. Sano S, Sugiura N, Yamaguchi K, et al. Comparative study of 10 different types of devices for self-monitoring of blood glucose: hematocrit and dissolved oxygen. J Clin Lab Instr Reag. 2005;28:349-356. [Google Scholar] Articles from Journal of Diabetes Science and Technology are provided here courtesy of Diabetes Technology Society
Cited by
7. Diabetes Technology: Standards of Medical Care in Diabetes—2022
American Diabetes Association Professional Practice Committee (2022) · Diabetes Care · doi:10.2337/dc22-s007 · Cited by 358
Sacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Lernmark, Å., Metzger, B. E., Nathan, D. M., & Kirkman, M. S. (2023) · Diabetes Care · doi:10.2337/dci23-0036 · Cited by 351
7. Diabetes Technology: Standards of Care in Diabetes—2025
American Diabetes Association Professional Practice Committee, ElSayed, N. A., McCoy, R. G., Aleppo, G., Balapattabi, K., Beverly, E. A., Briggs Early, K., Bruemmer, D., Echouffo-Tcheugui, J. B., Ekhlaspour, L., et al. (2025) · Diabetes Care · doi:10.2337/dc25-s007 · Cited by 348
7. Diabetes Technology: Standards of Care in Diabetes — 2023
ElSayed, N. A., Aleppo, G., Aroda, V. R., Bannuru, R. R., Brown, F. M., Bruemmer, D., Collins, B. S., Hilliard, M. E., Isaacs, D., Johnson, E. L., et al. (2023) · Diabetes Care · doi:10.2337/dc23-s007 · Cited by 301
7. Diabetes Technology: Standards of Care in Diabetes—2024
American Diabetes Association Professional Practice Committee, ElSayed, N. A., Aleppo, G., Bannuru, R. R., Bruemmer, D., Collins, B. S., Ekhlaspour, L., Hilliard, M. E., Johnson, E. L., Khunti, K., et al. (2024) · Diabetes Care · doi:10.2337/dc24-s007 · Cited by 281
Metadata
- Authors
- Sai, S., Urata, M., & Ogawa, I.
- Affiliations
- PL Hospital (JP)
- Corresponding author
- Sai, Shizuki
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2019-01-19
- Volume
- 13
- Issue
- 4
- Pages
- 734-743
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC6610602
- OpenAlex
- W2910826564
- Semantic Scholar
- 1a7b6ee76dc15f6f8d4eb7785fd71d9ca3aa9f7e
- Cited by
- 20
- References
- 7
- Citations by year
- 2026: 2 · 2025: 3 · 2024: 5 · 2023: 4 · 2022: 1 · 2021: 3 · 2020: 1 · 2019: 1
- FWCI
- 1.03
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Licence
- journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Public Health, Environmental and Occupational Health › Patient Safety and Medication Errors
- Keywords
- self-monitoring of blood glucose, point-of-care testing, blood glucose meters, measurement linearity, interfering substances, vitamin C, hematocrit, error messages, diabetes mellitus
- MeSH terms
- Blood Glucose, Hematocrit, Humans, Blood Glucose Self-Monitoring, Point-of-Care Systems
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.