Autonomous Spinal Robotic System for Transforaminal Lumbar Epidural Injections: A Proof of Concept of Study
Open accessAbstract
Study Design: Phantom study Objective: The aim of our study is to demonstrate in a proof-of-concept model whether the use of a marker less autonomous robotic controlled injection delivery system will increase accuracy in the lumbar spine. Methods: Ideal transforaminal epidural injection trajectories (bilateral L2/3, L3/4, L4/5, L5/S1 and S1) were planned out on a virtual pre-operative planning software by 1 experienced provider. Twenty transforaminal epidural injections were administered in a lumbar spine phantom model, 10 using a freehand procedure, and 10 using a marker less autonomous spinal robotic system. Procedural accuracy, defined as the difference between pre-operative planning and actual post-operative needle tip distance (mm) and angular orientation (degrees), were assessed between the freehand and robotic procedures. Results: Procedural accuracy for robotically placed transforaminal epidural injections was significantly higher with the difference in pre- and post-operative needle tip distance being 20.1 (±5.0) mm in the freehand procedure and 11.4 (±3.9) mm in the robotically placed procedure ( P < .001). Needle tip precision for the freehand technique was 15.6 mm (26.3 – 10.7) compared to 10.1 mm (16.3 – 6.1) for the robotic technique. Differences in needle angular orientation deviation were 5.6 (±3.3) degrees in the robotically placed procedure and 12.0 (±4.8) degrees in the freehand procedure ( P = .003). Conclusion: The robotic system allowed for comparable placement of transforaminal epidural injections as a freehand technique by an experienced provider, with additional benefits of improved accuracy and precision.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Registration pose error, TRE, and success rate are general measures to report 2D/3D registration performance, commonly used in the literature.
“Our results presented using registration pose error, TRE, and success rate are general measures to report the 2D/3D registration performance, which are commonly used in the literature [51], [18], [24].”
From A Fully Differentiable Framework for 2D/3D Registration and the Projective Spatial Transformers (2024) · cites this paper as [24]
The Robotic Technique section describes a 2D/3D registration pipeline, but does not identify measures for evaluating its performance. Statistical Analysis reports needle-tip distance and angular-orientation errors for the injection procedure, not registration pose error, TRE, or registration success rate. The searched passages do not address whether those three measures are commonly used in the literature; page numbers are unavailable in our copy. (LLM, conf=0.75)
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
No references could be resolved yet.
Research works printed but not linked: 25
- [1]Hession WG Stanczak JD Davis KW Choi JJ. Epidural steroid injections. Semin Roentgenol. 2004; 39(1): 7- 23. doi: 10.1016/j.ro.2003.10.010. 14976834 · DOI
- [2]Mathis JM. Epidural steroid injections. Neuroimaging Clin N Am. 2010; 20(2): 193- 202. doi: 10.1016/j.nic.2010.02.006. 20439015 · DOI
- [3]Bicket MC Gupta A Brown CH 4th Cohen SP. Epidural injections for spinal pain: A systematic review and meta-analysis evaluating the “control” injections in randomized controlled trials. Anesthesiology. 2013; 119(4): 907- 931. doi: 10.1097/ALN.0b013e31829c2ddd. 24195874 · DOI
- [4]Mandell JC Czuczman GJ Gaviola GC Ghazikhanian V Cho CH. The lumbar neural foramen and transforaminal epidural steroid injections: An anatomic review with key safety considerations in planning the percutaneous approach. AJR Am J Roentgenol. 2017; 209(1): W26- W35. doi: 10.2214/AJR.16.17471. 28504548 · DOI
- [5]Bui J Bogduk N. A systematic review of the effectiveness of CT-guided, lumbar transforaminal injection of steroids. Pain Med. 2013; 14(12): 1860- 1865. doi: 10.1111/pme.12243. 24330228 · DOI
- [6]White AH Derby R Wynne G. Epidural injections for the diagnosis and treatment of low-back pain. Spine. 1980; 5(1): 78- 86. doi: 10.1097/00007632-198001000-00014. 6444766 · DOI
- [7]Vad VB Bhat AL Lutz GE Cammisa F. Transforaminal epidural steroid injections in lumbosacral radiculopathy: A prospective randomized study. Spine. 2002; 27(1): 11- 15. doi: 10.1097/00007632-200201010-00005. 11805628 · DOI
- [8]Benny BV Patel MY. Predicting epidural steroid injections with laboratory markers and imaging techniques. Spine J. 2014; 14(10): 2500- 2508. doi: 10.1016/j.spinee.2014.04.003. 24743064 · DOI
- [9]Lee MH Yang KS Kim YH JungDo H Lim SJ Moon DE. Accuracy of live fluoroscopy to detect intravascular injection during lumbar transforaminal epidural injections. Korean J Pain. 2010; 23(1): 18- 23. doi: 10.3344/kjp.2010.23.1.18. 20552068 PMC2884208 · DOI
- [10]Smuck M Fuller BJ Chiodo A, et al. Accuracy of intermittent fluoroscopy to detect intravascular injection during transforaminal epidural injections. Spine. 2008; 33(7): E205- E210. doi: 10.1097/BRS.0b013e31816960fe. 18379390 · DOI
- [11]Faul F Erdfelder E Buchner A Lang A-G. Statistical power analyses using G*Power 3.1: Tests for correlation and regression analyses. Behav Res Methods. 2009; 41(4): 1149- 1160. doi: 10.3758/BRM.41.4.1149. 19897823 · DOI · Paper page
- [12]Bellingham GA Peng PWH. A low-cost ultrasound phantom of the lumbosacral spine. Reg Anesth Pain Med. 2010; 35(3): 290- 293. doi: 10.1097/AAP.0b013e3181c75a76. 20921841 · DOI
- [13]Park JW Cheon MW Lee MH. Phantom study of a new laser-etched needle for improving visibility during ultrasonography-guided lumbar medial branch access with novices. Ann Rehabil Med. 2016; 40(4): 575- 582. doi: 10.5535/arm.2016.40.4.575. 27606263 PMC5012968 · DOI
- [14]Fedorov A Beichel R Kalpathy-Cramer J, et al. 3D Slicer as an image computing platform for the quantitative imaging network. Magn Reson Imaging. 2012; 30(9): 1323- 1341. doi: 10.1016/j.mri.2012.05.001. 22770690 PMC3466397 · DOI
- [15]Hartley R Zisserman A. Multiple View Geometry in Computer Vision. Cambridge: Cambridge University Press; 2004.
- [16]Gao C Farvardin A Grupp RB, et al. Fiducial-free 2D/3D registration for robot-assisted femoroplasty. IEEE Trans Med Robot bionics. 2020; 2(3): 437- 446. doi: 10.1109/tmrb.2020.3012460. 33763632 PMC7982989 · DOI · Paper page
- [17]Grupp RB Armand M Taylor RH. Patch-based image similarity for intraoperative 2D/3D pelvis registration during periacetabular osteotomy. Lect Notes Comput Sci. 2018; 11041 LNCS: 153- 163. doi: 10.1007/978-3-030-01201-4_17. · DOI · Paper page
- [18]Hansen N Ostermeier A. Completely derandomized self-adaptation in evolution strategies. Evol Comput. 2001; 9(2): 159- 195. doi: 10.1162/106365601750190398. 11382355 · DOI · Paper page
- [19]Grupp RB Hegeman RA Murphy RJ, et al. Pose estimation of periacetabular osteotomy fragments with intraoperative X-Ray navigation. IEEE Trans Biomed Eng. 2020; 67(2): 441- 452. doi: 10.1109/TBME.2019.2915165. 31059424 PMC7297497 · DOI · Paper page
- [20]Beyer LP Michalik K Niessen C, et al. Evaluation of a robotic assistance-system for percutaneous computed tomography-guided (CT-guided) facet joint injection: A phantom study. Med Sci Mon Int Med J Exp Clin Res. 2016; 22: 3334- 3339. doi: 10.12659/MSM.900686. PMC5042119 27648509 · DOI
- [21]Li G Patel NA Melzer A Sharma K Iordachita I Cleary K. MRI-guided lumbar spinal injections with body-mounted robotic system: cadaver studies. Minim Invasive Ther Allied Technol. 2020; 31: 297- 305. doi: 10.1080/13645706.2020.1799017. 32729771 PMC7855543 · DOI
- [22]Shoham M Lieberman IH Benzel EC, et al. Robotic assisted spinal surgery - from concept to clinical practice. Comput Aided Surg. 2007; 12: 105- 115. doi: 10.1080/10929080701243981. 17487660 · DOI
- [23]D'Souza M Gendreau J Feng A Kim LH Ho AL Veeravagu A. Robotic-assisted spine surgery: History, efficacy, cost, and future trends. Rob Surg Res Rev. 2019; 6: 9- 23. doi: 10.2147/rsrr.s190720. PMC6844237 31807602 · DOI
- [24]Condon A. Robotics in Spine Surgery: 17 Notes for Surgeons, ASCs & Administrators: Becker’s Spine Review; 2020. https://www.beckersspine.com/robotics/item/50394-robotics-in-spine-surgery-17-notes-for-surgeons-ascs-administrators.html.
- [25]Hoshide R Feldman E Taylor W. Cadaveric analysis of the Kambin’s triangle. Cureus. 2016; 8(2): e475. doi: 10.7759/cureus.475. 27004152 PMC4780690 · DOI
Cited by
Sorted by citations
A Fully Differentiable Framework for 2D/3D Registration and the Projective Spatial Transformers
Gao, C., Feng, A., Liu, X., Taylor, R. H., Armand, M., & Unberath, M. (2024) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2023.3299588 · Cited by 36 · see citing works
“Several researchers formulated such learning-based 2D/3D registration with reinforcement learning paradigms [22], [23], [24], but they still perform pose regression at each iteration.”
BackgroundResultSelf-citation
- Authors
- Margalit, A., Phalen, H., Gao, C., Ma, J., Suresh, K. V., Jain, P., Farvardin, A., Taylor, R. H., Armand, M., Chattre, A., & Jain, A.
- Affiliations
- Johns Hopkins University (US)
- Corresponding author
- Jain, Amit
- ORCID
- Venue
- Global Spine Journal
- Publisher
- SAGE Publications
- Published
- 2022-04-25
- Volume
- 14
- Issue
- 1
- Pages
- 138-145
- ISSN
- 2192-5682 (ISSN-L)
- 2192-5690
- PMC
- PMC10676186
- OpenAlex
- W4224470041
- Cited by
- 15
- References
- 0
- Citations by year
- 2026: 4 · 2025: 4 · 2024: 4 · 2023: 3
- FWCI
- 2.63
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- Global Spine Journal (publishedVersion, cc-by-nc-nd)
- PubMed Central (publishedVersion, cc-by-nc-nd)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Funders
- NIH (R01EB023939)
- T32 postdoctoral training (T32AR067708-06)
- National Science Foundation Graduate Research (DGE-1746891)
- Topics
- Health Sciences › Medicine › Surgery › Spinal Fractures and Fixation Techniques
- Health Sciences › Medicine › Surgery › Anesthesia and Pain Management
- Health Sciences › Medicine › Pathology and Forensic Medicine › Spine and Intervertebral Disc Pathology
- Keywords
- autonomous robots, needle placement accuracy, surgical planning
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review