Pose Estimation of Periacetabular Osteotomy Fragments With Intraoperative X-Ray Navigation
Open accessAbstract
Objective: State-of-the-art navigation systems for pelvic osteotomies use optical systems with external fiducials. In this paper, we propose the use of X-ray navigation for pose estimation of periacetabular fragments without fiducials. Methods: A two-dimensional/three-dimensional (2-D/3-D) registration pipeline was developed to recover fragment pose. This pipeline was tested through an extensive simulation study and six cadaveric surgeries. Using osteotomy boundaries in the fluoroscopic images, the preoperative plan was refined to more accurately match the intraoperative shape. Results: In simulation, average fragment pose errors were 1.3 °/1.7 mm when the planned fragment matched the intraoperative fragment, 2.2 °/2.1 mm when the plan was not updated to match the true shape, and 1.9 °/2.0 mm when the fragment shape was intraoperatively estimated. In cadaver experiments, the average pose errors were 2.2 °/2.2 mm, 3.8 °/2.5 mm, and 3.5 °/2.2 mm when registering with the actual fragment shape, a preoperative plan, and an intraoperatively refined plan, respectively. Average errors of the lateral center edge angle were less than 20 for all fragment shapes in simulation and cadaver experiments. Conclusion: The proposed pipeline is capable of accurately reporting femoral head coverage within a range clinically identified for long-term joint survivability. Significance: Human interpretation of fragment pose is challenging and usually restricted to rotation about a single anatomical axis. The proposed pipeline provides an intraoperative estimate of rigid pose with respect to all anatomical axes, is compatible with minimally invasive incisions, and has no dependence on external fiducials.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Each contextualization study that considered multiple objects did not explicitly isolate parameters for each object, but used a single NN with output layer indexed by object.
“However, we found that each of the contextualization studies which considered multiple objects (Grupp et al., 2020c; Doerr et al., 2020; Wang et al., 2020; Esfandiari et al., 2021) did not explicitly isolate the parameters corresponding to each object, but instead used a single NN with data in the output layer indexed according to the appropriate object.”
From The Impact of Machine Learning on 2D/3D Registration for Image-Guided Interventions: A Systematic Review and Perspective (2021) · cites this paper as Grupp et al., 2020c
The Materials and Methods section (Intraoperative Registration Strategy; page unavailable) describes successive optimizations of individual objects, followed by simultaneous optimization of their poses. It also specifies different optimization parameters for the pelvis than for the femur and fragment. This contradicts the attribution that the study did not isolate object-specific parameters and instead used a single neural network with object-indexed outputs. (LLM, conf=0.90)
Checked by a reasoning model — may be wrong.
Experimental — accuracy not yet measuredIterative optimization with differentiable rendering from a fixed initialization achieved an SMSR of only 7.6%.
“In contrast, iterative optimization with differentiable rendering from a fixed initialization [20] also does not require patient-specific training, but achieved an SMSR of only 7.6%.”
From Rapid patient-specific neural networks for intraoperative X-ray to volume registration (2025) · cites this paper as [20]
The Intraoperative Registration Strategy describes iterative optimization using digitally reconstructed radiographs, supporting that part of the description. It specifies derivative-free optimization and landmark-based initialization, however, not differentiable rendering from a fixed initialization. The Results report other accuracy measures; a search for “7.6” found no SMSR of 7.6%. (LLM, conf=0.87)
Experimental — accuracy not yet measuredxReg did not achieve sufficient accuracy, producing SMSRs of just 0.3% and 3.3% for the fixed and landmark initializations, respectively.
“xReg did not achieve sufficient accuracy, producing SMSRs of just 0.3% and 3.3% for the fixed [20] and landmark initializations [5], respectively (Fig. 4a).”
From Rapid patient-specific neural networks for intraoperative X-ray to volume registration (2025) · cites this paper as [20]
The Intraoperative Registration Strategy section describes a single-landmark initialization, so this source supports the identification of the landmark method cited as [5]. The Results and Discussion report pose and LCE errors, but do not report a 3.3% SMSR or conclude that xReg using landmark initialization had insufficient accuracy. The sentence's performance figure is therefore not supported by this source. (LLM, conf=0.83)
Experimental — accuracy not yet measuredTraditional renderers are non-differentiable, so pose estimation is commonly driven by gradient-free optimization.
“As traditional renderers are non-differentiable, pose estimation is commonly driven by gradient-free optimization [24].”
From Intraoperative 2D/3D Image Registration via Differentiable X-Ray Rendering (2024) · cites this paper as [24]
The Intraoperative Registration Strategy section (page unavailable) says the paper uses CMA-ES for pose optimization and explicitly calls it derivative-free; it also uses BOBYQA at a finer resolution. Neither that section nor Related Work (page unavailable) says traditional renderers are non-differentiable or identifies that as the reason for using gradient-free optimization. The paper supports the optimization practice, but not the claimed explanation. (LLM, conf=0.86)
Experimental — accuracy not yet measuredTraditional gradient-free optimization (xReg) successfully recovers the poses of some X-rays but frequently converges to suboptimal minima.
“Traditional gradient-free optimization (xReg [24]) is able to successfully recover the poses of some X-rays but frequently converges to suboptimal minima.”
From Intraoperative 2D/3D Image Registration via Differentiable X-Ray Rendering (2024) · cites this paper as [24]
The Intraoperative Registration Strategy describes derivative-free optimization, and the Results report successful pose estimates alongside some large errors. But the paper does not establish that the method frequently converges to suboptimal minima; it describes CMA-ES as robust to local minima and BOBYQA as less robust, without attributing the observed errors to such convergence. Page numbers are unavailable in our copy. (LLM, conf=0.84)
Experimental — accuracy not yet measured
Evaluation on synthetic data may include data sampled from a broad range of viewpoints or deformations that are approximately uniformly distributed, whereas real data is not uniformly sampled but clustered around certain viewpoints.
“For example, evaluation on synthetic data may include data sampled from a broad range of viewpoints or deformations that are approximately uniformly distributed. Real data, on the other hand, is not uniformly sampled from such a distribution, but rather, will be clustered around certain viewpoints (see (Grupp et al., 2020c) for a visualization of viewpoints used during a cadaveric surgery vs the”
From The Impact of Machine Learning on 2D/3D Registration for Image-Guided Interventions: A Systematic Review and Perspective (2021) · cites this paper as Grupp et al., 2020c
The Materials and Methods sections describe simulated views generated from a perturbed AP view and random C-arm rotations, and fluoroscopic views collected during cadaver surgeries (pages unavailable in our copy). They support the distinction between simulated and surgical viewpoints, but do not establish that the simulated viewpoints were approximately uniformly distributed or that the real viewpoints clustered around particular angles. The stated uniform sampling in the simulation passage concerns landmark-offset directions, not viewpoints. (LLM, conf=0.78)
Experimental — accuracy not yet measuredAlthough Grupp et al. (2020c) registered multiple objects, the regularization function employed learned landmark annotations from only a single object.
“Although Grupp et al. (2020c) registered multiple objects, the regularization function employed learned landmark annotations from a only a single object.”
From The Impact of Machine Learning on 2D/3D Registration for Image-Guided Interventions: A Systematic Review and Perspective (2021) · cites this paper as Grupp et al., 2020c
The Materials and Methods section describes multi-object registration, supporting the first part of the claim. It describes regularization of pose parameters, not the use of learned landmark annotations; the landmark used to initialize pelvis registration was manually annotated. Page numbers are unavailable in our copy. (LLM, conf=0.88)
Experimental — accuracy not yet measuredThe authors' group has previously released a dataset of CTs and >350 X-rays across multiple viewpoints of six cadaveric specimens prior to periacetabular osteotomy.
“To this end, our group has previously released a relatively large dataset of CTs and >350 X-rays across multiple viewpoints of six cadaveric specimens prior to undergoing periacetabular osteotomy (cf. Figure 4) (Grupp et al., 2020).”
From The Impact of Machine Learning on 2D/3D Registration for Image-Guided Interventions: A Systematic Review and Perspective (2021) · cites this paper as Grupp et al., 2020
The Materials and Methods (Simulation Study; page unavailable) reports CT scans from six cadaveric specimens and simulated fluoroscopic images from multiple views. The Cadaver Study describes collected intraoperative fluoroscopy, but our copy does not establish that the authors released a dataset containing more than 350 pre-osteotomy X-rays; it identifies only a supplementary video as available with the paper. (LLM, conf=0.85)
Experimental — accuracy not yet measuredThe CMAES benchmark method for pose estimation of bone anatomy and surgical devices has been tested to meet clinical requirements in various orthopedic applications.
“Due to its well-known robustness to local minima, the CMAES benchmark method for pose estimation of bone anatomy and surgical devices has been tested to meet clinical requirements in various orthopedic applications, including osteotomy [2], [13], femoroplasty [41], [4], core decompression of the hip [42], and transforaminal lumbar epidural injections [43], etc.”
From A Fully Differentiable Framework for 2D/3D Registration and the Projective Spatial Transformers (2024) · cites this paper as [2]
Only a few small labeled datasets exist for training deep networks for landmark localization, preventing generalization to new patients, procedures, and pathologies.
“While landmark localizers can be learned, only a few small labeled datasets exist for training deep networks [24, 41, 50], preventing the generalization of these methods to new patients, procedures, and pathologies.”
From Intraoperative 2D/3D Image Registration via Differentiable X-Ray Rendering (2024) · cites this paper as [24]
The Intraoperative Registration Strategy section describes manual landmark annotation, and the Discussion mentions neural-network landmark recognition as a possible alternative (page numbers unavailable). Neither passage addresses the availability or size of labeled training datasets, or says that dataset scarcity prevents generalization to new patients, procedures, or pathologies. Searches for labeled datasets, deep-network training, and generalization found no supporting passage. (LLM, conf=0.76)
Experimental — accuracy not yet measuredNeural networks trained on small datasets for 2D/3D registration fail on new patients or require impractical landmark supervision.
“neural networks trained on small datasets fail on new patients or require impractical landmark supervision”
From Intraoperative 2D/3D Image Registration via Differentiable X-Ray Rendering (2024) · cites this paper as [24]
The Discussion (page unavailable) mentions neural networks as a possible way to avoid manual landmark identification, but does not discuss their training datasets, performance on new patients, or landmark supervision during training. The Results > Cadaver Study passage measures manual annotation time for the paper’s registration method, not for a neural network. Searches for small datasets, unseen patients, and landmark supervision found no passage addressing the claim. (LLM, conf=0.76)
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Completely Derandomized Self-Adaptation in Evolution Strategies
Hansen, N., & Ostermeier, A. (2001) · Evolutionary Computation · doi:10.1162/106365601750190398 · Cited by 4,397 · see citing works
Chen, X., Graham, J., Hutchinson, C., & Muir, L. (2013) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2012.2226740 · Cited by 26 · see citing works
Flagged first, then by citations
Research works printed but not linked: 59
- [1]Murphy SB, Ganz R, and Müller M, “ The prognosis in untreated dysplasia of the hip. A study of radiographic factors that predict the outcome,” J. Bone Joint Surg. Amer, vol. 77, no. 7, pp. 985– 989, 1995. 7608241
- [2]Ganz R, “ A new periacetabular osteotomy for the treatment of hip dysplasias technique and preliminary results,” Clin. Orthopaedics Related Res, vol. 232, pp. 26– 36, 1988. 3383491
- [3]Armand M, “ Outcome of periacetabular osteotomy: joint contact pressure calculation using standing AP radiographs, 12 patients followed for average 2 years,” Acta Orthopaedics, vol. 76, no. 3, pp. 303– 313, 2005. PMC2745131 16156455
- [4]Armiger RS, “ Three-dimensional mechanical evaluation of joint contact pressure in 12 periacetabular osteotomy patients with 10-year follow-up,” Acta Orthopaedics, vol. 80, no. 2, pp. 155– 161, 2009. PMC2689368 19404795
- [5]Niknafs N, “ Biomechanical factors in planning of periacetabular osteotomy,” Front Bioeng. Biotechnol, vol. 1, p. 20, 2013. 25152876 PMC4126379
- [6]Hipp JA, “ Planning acetabular redirection osteotomies based on joint contact pressures,” Clin. Orthopaedics Related Res, vol. 364, pp. 134– 143, 1999. 10416402
- [7]Troelsen A, “ Surgical advances in periacetabular osteotomy for treatment of hip dysplasia in adults,” Acta Orthopaedics, vol. 80, no. suppl. 332, pp. 1– 33, 2009. 19757731
- [8]Wiberg G, “ Studies on dysplastic acetabulum and congenital subluxation of the hip joint with special reference to the complications of osteoarthritis,” Acta Chir. Scand, vol. 83, no. 58, pp. 53– 68, 1939.
- [9]Hartig-Andreasen C, “ What factors predict failure 4 to 12 years after periacetabular osteotomy?” Clin. Orthopaedics Related Res, vol. 470, no. 11, pp. 2978– 2987, 2012. PMC3462869 22576934
- [10]Troelsen A, “ Assessment of hip dysplasia and osteoarthritis: Variability of different methods,” Acta Radiol, vol. 51, no. 2, pp. 187– 193, 2010. 20144145
- [11]Langlotz F, “ Computer assistance for pelvic osteotomies,” Clin. Orthopaedics Related Res, vol. 354, pp. 92– 102, 1998. 9755768
- [12]Murphy RJ, “ Development of a biomechanical guidance system for periacetabular osteotomy,” Int. J. Comput. Assisted Radiol. Surgery, vol. 10, no. 4, pp. 497– 508, 2015. PMC4476510 25234421
- [13]Mayman DJ, “ The kingston periacetabular osteotomy utilizing computer enhancement: A new technique,” Comput. Aided Surg, vol. 7, no. 3, pp. 179– 186, 2002. 12362378
- [14]Liu L, “ Computer assisted planning and navigation of periacetabular osteotomy with range of motion optimization,” in Proc. Med. Image Comput. Comput. Assisted Intervention, 2014, pp. 643– 650. 25485434
- [15]Akiyama H, “ Computed tomography-based navigation for curved periacetabular osteotomy,” J. Orthopaedics Sci, vol. 15, no. 6, pp. 829– 833, 2010. 21116903
- [16]Takao M, “ Comparison of rotational acetabular osteotomy performed with navigation by surgeons with different levels of experience of osteotomies,” Int. J. Comput. Assisted Radiol. Surgery, vol. 12, no. 5, pp. 841– 853, 2017. 27783201
- [17]Troelsen A, Elmengaard B, and Søballe K, “ A new minimally invasive transsartorial approach for periacetabular osteotomy,” J. Bone Joint Surg. Amer, vol. 90, no. 3, pp. 493– 498, 2008. 18310698
- [18]Troelsen A, Elmengaard B, and Søballe K, “ Comparison of the minimally invasive and ilioinguinal approaches for periacetabular osteotomy 263 single-surgeon procedures in well-defined study groups,” Acta Orthopaedics, vol. 79, no. 6, pp. 777– 784, 2008. 19085494
- [19]Belei P, “ Fluoroscopic navigation system for hip surface replacement,” Comput. Aided Surg, vol. 12, no. 3, pp. 160– 167, 2007. 17538788
- [21]Dennis DA, “ In vivo determination of normal and anterior cruciate ligament-deficient knee kinematics,” J. Biomechan, vol. 38, no. 2, pp. 241– 253, 2005. 15598450
- [22]Gong RH, Stewart J, and Abolmaesumi P, “ Multiple-object 2-D–3-D registration for noninvasive pose identification of fracture fragments,” IEEE Trans. Biomed. Eng, vol. 58, no. 6, pp. 1592– 1601, Jun. 2011. 21233043
- [23]Guéziec A, “ Anatomy-based registration of CT-scan and intraoperative X-ray images for guiding a surgical robot,” IEEE Trans. Med. Imag, vol. 17, no. 5, pp. 715– 728, Oct. 1998. 9874295
- [24]Joskowicz L, “ FRACAS: A system for computer-aided image-guided long bone fracture surgery,” Comput. Aided Surg, vol. 3, no. 6, pp. 271– 288, 1998. 10379977
- [25]Mahfouz MR, “ A robust method for registration of three-dimensional knee implant models to two-dimensional fluoroscopy images,” IEEE Trans. Med. Imag, vol. 22, no. 12, pp. 1561– 1574, Dec. 2003. 14649746
- [26]Markelj P, “ A review of 3D/2D registration methods for image-guided interventions,” Med. Image Anal, vol. 16, no. 3, pp. 642– 661, 2012. 20452269
- [27]Otake Y, “ Intraoperative image-based multi-view 2D/3D registration for image-guided orthopaedic surgery: Incorporation of fiducial-based C-arm tracking and GPU-acceleration,” IEEE Trans. Med. Imag, vol. 31, no. 4, pp. 948– 962, Apr. 2012. PMC4451116 22113773
- [28]Otake Y, “ Robust patella motion tracking using intensity-based 2D-3D registration on dynamic bi-plane fluoroscopy: Toward quantitative assessment in MPFL reconstruction surgery,” Proc. SPIE, vol. 9786, 2016, Art. no. 97 860B.
- [29]Penney GP, “ A comparison of similarity measures for use in 2-D-3-D medical image registration,” IEEE Trans. Med. Imag, vol. 17, no. 4, pp. 586– 595, Aug. 1998. 9845314
- [30]Sheng-fu LL, “ Automatic localization of target vertebrae in spine surgery: Clinical evaluation of the LevelCheck registration algorithm,” Spine, vol. 40, no. 8, 2015, Art. no. E476. PMC4433144 25646750
- [31]Yao J, “ A C-arm fluoroscopy-guided progressive cut refinement strategy using a surgical robot,” Comput. Aided Surg, vol. 5, no. 6, pp. 373– 390, 2000. 11295851
- [32]Yi T, “ Robotic drill guide positioning using known-component 3D–2D image registration,” J. Med. Imag, vol. 5, no. 2, 2018, Art. no. 021212. PMC5800483 29430481
- [33]Langlotz F, “ The first twelve cases of computer assisted periacetabular osteotomy,” Comput. Aided Surg, vol. 2, no. 6, pp. 317– 326, 1997. 9587693
- [34]Naito M, “ Curved periacetabular osteotomy for treatment of dysplastic hip,” Clin. Orthopaedics Related Res, vol. 433, pp. 129– 135, 2005. 15805948
- [35]Radermacher K, “ Computer assisted orthopaedic surgery with image based individual templates,” Clin. Orthopaedics Relat. Res, vol. 354, pp. 28– 38, 1998. 9755761
- [36]Tönnis D, “ Triple pelvic osteotomy,” J. Pediatric Orthopaedics B, vol. 3, no. 1, pp. 54– 67, 1994.
- [37]Otsuki B, “ Developing a novel custom cutting guide for curved periacetabular osteotomy,” Int. Orthopaedics, vol. 37, no. 6, pp. 1033– 1038, 2013. PMC3664171 23568142
- [38]Murphy RJ, “ Clinical evaluation of a biomechanical guidance system for periacetabular osteotomy,” J. Orthopaedics Surg. Res, vol. 11, no. 1, p. 36, 2016. PMC4812624 27029935
- [39]Liu L, “ Periacetabular osteotomy through the pararectus approach: technical feasibility and control of fragment mobility by a validated surgical navigation system in a cadaver experiment,” Int. Orthopaedics, vol. 40, no. 7, pp. 1389– 1396, 2016. 26162984
- [40]Ninomiya S and Tagawa H, “ Rotational acetabular osteotomy for the dysplastic hip,” J. Bone Joint Surg. Amer, vol. 66, no. 3, pp. 430– 436, 1984. 6699061
- [41]Knaan D and Joskowicz L, “ Effective intensity-based 2D/3D rigid registration between fluoroscopic X-ray and CT,” in Medical Image Computing and Computer-Assisted Intervention. Berlin Germany: Springer, 2003, pp. 351– 358.
- [42]Grupp RB, Armand M, and Taylor RH, “ Patch-based image similarity for intraoperative 2D/3D pelvis registration during periacetabular osteotomy,” in Proc. Int. Workshop Clin. Image Based Procedures, 2018, pp. 153– 163.
- [44]Krčah M, Székely G, and Blanc R, “ Fully automatic and fast segmentation of the femur bone from 3D-CT images with no shape prior,” in Proc. IEEE Int. Symp. Biomed. Imag., 2011, pp. 2087– 2090.
- [45]Nikou C, “ Description of anatomic coordinate systems and rationale for use in an image-guided total hip replacement system,” in Proc. Med. Image Comput. Comput. Assisted Intervention, 2000, pp. 1188– 1194.
- [46]Stone JE, Gohara D, and Shi G, “ OpenCL: A parallel programming standard for heterogeneous computing systems,” Comput. Sci. Eng, vol. 12, no. 3, pp. 66– 73, 2010. 21037981 PMC2964860
- [47]Powell MJ, “ The BOBYQA algorithm for bound constrained optimization without derivatives,” University of Cambridge, Cambridge, U.K., Rep. NA2009/06, 2009.
- [48]Hartley R and Zisserman A, Multiple View Geometry in Computer Vision. Cambridge, U.K.: Cambridge Univ. Press, 2003.
- [49]Armiger RS, “ Evaluation of a computerized measurement technique for joint alignment before and during periacetabular osteotomy,” Comput. Aided Surg, vol. 12, no. 4, pp. 215– 224, 2007. 17786597 PMC2716292
- [50]Murray RM, A Mathematical Introduction to Robotic Manipulation. Boca Raton, FL, USA: CRC Press, 1994.
- [51]Markelj P, Likar B, and Pernuš F, “ Standardized evaluation methodology for 3D/2D registration based on the visible human data set,” Med. Phys, vol. 37, no. 9, pp. 4643– 4647, 2010. 20964183
- [52]Horn BK, “ Closed-form solution of absolute orientation using unit quaternions,” J. Opt. Soc. Amer. A, vol. 4, no. 4, pp. 629– 642, 1987.
- [53]Armand M, “ Biomechanical guidance system for periacetabular osteotomy,” in Intelligent Orthopaedics. Berlin, Germany: Springer, 2018, pp. 169– 179. 30306481
- [54]Murphy R, “ Computer-assisted X-ray image-based navigation of periacetabular osteotomy with fiducial based 3D acetabular fragment tracking,” Orthopaedics Proc, vol. 95, no. SUPP_28, 2013, Art. no. 84.
- [55]Bier B, “ X-ray-transform invariant anatomical landmark detection for pelvic trauma surgery,” in Medical Image Computing and Computer Assisted Intervention. Berlin, Germany: Springer, 2018, pp. 55– 63.
- [56]Kendall A, Grimes M, and Cipolla R, “ PoseNet: A convolutional network for real-time 6-DOF camera relocalization,” in Proc. IEEE Int. Conf. Comput. Vis, 2015, pp. 2938– 2946.
- [57]Callanan MC, “ The john charnley award: Risk factors for cup malpositioning: Quality improvement through a joint registry at a tertiary hospital,” Clin. Orthopaedics Relat. Res, vol. 469, no. 2, pp. 319– 329, 2011. PMC3018230 20717858
- [58]Otake Y, “ Comparison of optimization strategy and similarity metric in Atlas-to-subject registration using statistical deformation model,” Proc. SPIE, vol. 94 15, 2015, Art. no. 94 150Q.
- [59]Sadowsky O, Chintalapani G, and Taylor R, “ Deformable 2D-3D registration of the pelvis with a limited field of view, using shape statistics,” in Proc. Med. Image Comput. Comput. Assisted Interv, 2007, pp. 519– 526. 18044608
- [60]Grupp R, “ Pelvis surface estimation from partial CT for computer-aided pelvic osteotomies,” Orthopaedics Proc, vol. 98, no. SUPP_5, 2016, Art. no. 55–55.
- [61]Chintalapani G, “ Statistical atlas based extrapolation of CT data,” Proc. SPIE, vol. 7625, 2010, Art. no. 762 539.
Cited by
Sorted by citations
Unberath, M., Gao, C., Hu, Y., Judish, M., Taylor, R. H., Armand, M., & Grupp, R. (2021) · Frontiers in Robotics and AI · doi:10.3389/frobt.2021.716007 · Cited by 102 · see citing works
“…as: kinematic measurements of bone [99, 100, 101], rib motion and respiratory analysis [102], intra-operative assessment of adjusted bone fragments [20, 18, 103], confirmation of screw implant placement during spine surgery [104] and the positioning of a surgical robot with respect to target…”
BackgroundSelf-citationAutomatic Annotation of Hip Anatomy in Fluoroscopy for Robust and Efficient 2D/3D Registration
Grupp, R. B., Unberath, M., Gao, C., Hegeman, R. A., Murphy, R. J., Alexander, C. P., Otake, Y., McArthur, B. A., Armand, M., & Taylor, R. H. (2020) · International Journal of Computer Assisted Radiology and Surgery · doi:10.1007/s11548-020-02162-7 · Cited by 81 · see citing works
“These landmarks were previously identified as being useful for obtaining initial registration estimates of the pelvis [7].”
BackgroundMethodInfluentialSelf-citationIntraoperative 2D/3D Image Registration via Differentiable X-Ray Rendering
Gopalakrishnan, V., Dey, N., & Golland, P. (2024) · 2024 IEEE/CVF Conference on Computer Vision and Pattern Recognition (CVPR) · doi:10.1109/cvpr52733.2024.01108 · Cited by 48 · see citing works
“Traditional gradient-free optimization (xReg [24]) is able to successfully recover the poses of some X-rays but frequently converges to suboptimal minima.”
BackgroundMethodInfluentialA Fully Differentiable Framework for 2D/3D Registration and the Projective Spatial Transformers
Gao, C., Feng, A., Liu, X., Taylor, R. H., Armand, M., & Unberath, M. (2024) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2023.3299588 · Cited by 36 · see citing works
“…method for pose estimation of bone anatomy and surgical devices has been tested to meet clinical requirements in various orthopedic applications, including osteotomy [2], [13], femoroplasty [4], [41], core decompression of the hip [42], and transforaminal lumbar epidural injections [43], etc.”
BackgroundMethodSelf-citationKu, P.-C., Martin-Gomez, A., Gao, C., Grupp, R., Mears, S. C., & Armand, M. (2023) · Computer Methods in Biomechanics and Biomedical Engineering: Imaging & Visualization · doi:10.1080/21681163.2022.2152375 · Cited by 8 · see citing works
“Registration pipelines between CT and 2D radiographs have been proposed for surgical treatments around the hip region such as periacetabular osteotomy (Grupp et al. 2020) and femoroplasty (Gao et al. 2020).”
BackgroundMethodSelf-citationRapid patient-specific neural networks for intraoperative X-ray to volume registration
Gopalakrishnan, V., Chlorogiannis, D.-D., Abumoussa, A., Larson, A. M., Haouchine, N., Orbach, D. B., Frisken, S., Dey, N., & Golland, P. (2025) · arXiv · doi:10.48550/arxiv.2503.16309 · Cited by 2 · see citing works
- Authors
- Grupp, R. B., Hegeman, R. A., Murphy, R. J., Alexander, C. P., Otake, Y., McArthur, B. A., Armand, M., & Taylor, R. H.
- Affiliations
- Johns Hopkins University (US)
- Nara Institute of Science and Technology (JP)
- ORCID
- Venue
- IEEE Transactions on Biomedical Engineering
- Published
- 2020-02
- Volume
- 67
- Issue
- 2
- Pages
- 441-452
- ISSN
- 0018-9294 (ISSN-L)
- 1558-2531
- PMC
- PMC7297497
- OpenAlex
- W2925003514
- Cited by
- 38
- References
- 2
- Citations by year
- 2026: 4 · 2025: 8 · 2024: 4 · 2023: 5 · 2022: 5 · 2021: 3 · 2020: 9
- FWCI
- 1.90
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Licence
- Locations
- IEEE Transactions on Biomedical Engineering (publishedVersion)
- PubMed (publishedVersion)
- PubMed Central (acceptedVersion)
- arXiv (Cornell University) (submittedVersion)
- arXiv (Cornell University)
- Funders
- NIH/NIBIB (R01EB006839, R21EB020113)
- MEXT/JSPS KAKENHI (26108004)
- Johns Hopkins University Internal Funds
- Johns Hopkins University Applied Physics Laboratory
- Topics
- Physical Sciences › Computer Science › Computer Vision and Pattern Recognition › Medical Image Segmentation Techniques
- Health Sciences › Medicine › Surgery › Hip disorders and treatments
- Physical Sciences › Engineering › Biomedical Engineering › Medical Imaging and Analysis
- Keywords
- periacetabular osteotomy, fluoroscopy, pelvic osteotomy, femoral head coverage, lateral center-edge angle, preoperative surgical planning, cadaveric surgery
- MeSH terms
- Acetabulum, Aged, Aged, 80 and over, Female, Fluoroscopy, Hip Joint, Humans, Male, Middle Aged, Osteotomy, Phantoms, Imaging, Imaging, Three-Dimensional, Surgery, Computer-Assisted
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review