AutoSet nasal CPAP titration: constancy of pressure, compliance and effectiveness at 8 month follow-up
Open accessThe DOI matches its Crossref record; 4 of the 4 references it cites were found in a registry or an index.
Abstract
We have previously shown that AutoSet satisfactorily improves sleep-disordered breathing and sleep architecture in subjects with obstructive sleep apnoea (OSA) syndrome. The aim of this study was to determine, in subjects treated with long-term conventional fixed pressure continuous positive airway pressure (CPAP) at the AutoSet recommended pressure, whether: the long-term compliance is satisfactory; the improvement persists once initial rebound is over; the titration pressure is stable with time; and the titration pressure is comparable with manual titration pressure using a similar end-point. Twenty males with OSA, previously studied with full polysomnography on their diagnostic night, at manual and AutoSet titration, and at the AutoSet recommended fixed pressure, were re-studied after a mean of 3 and 8 months of treatment at the recommended fixed pressure. Re-study included home respiratory monitoring (Nellcor EdenTrace), and repeated manual and AutoSet titration with polysomnography. Compliance was assessed with hour-meter readings. Mean (+/-SEM) usage was 5.7 +/- 0.1 h.night-1 at 3 and 8 months. The arousal index remained normalized. Diagnostic respiratory disturbance index (RDI) was 60.3 +/- 5.7 events.h-1. On AutoSet at fixed CPAP, RDI was initially 2.6 +/- 0.7 events.h-1, then rose slightly (p < 0.001) to 4.3 +/- 0.6 events.h-1 at 3 months, and was 3.6 +/- 0.5 events.h-1 at 8 months. AutoSet titration pressure was: 9.9 +/- 0.4 cmH2O initially, 10.6 +/- 0.4 cmH2O at 3 months, and 9.7 +/- 0.5 cmH2O at 8 months (NS). Manual titration pressure at 8 months was 10.4 +/- 0.4 cmH2O. The standard deviation of the discrepancy with AutoSet was 0.84 cmH2O. In conclusion, the AutoSet recommended pressure varies little with time, and closely predicts the final manual titration pressure; the improvement in respiratory disturbance index was largely maintained, and compliance was good, although probably enhanced by close supervision.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Statistical methods for assessing agreement between two methods of clinical measurement
Bland, J. M., & Altman, D. G. (2010) · International Journal of Nursing Studies · doi:10.1016/j.ijnurstu.2009.10.001 · Cited by 5,220
Sullivan, C., Berthon-Jones, M., Issa, F., & Eves, L. (1981) · The Lancet · doi:10.1016/s0140-6736(81)92140-1 · Cited by 2,731
Self-citationSleep apnea and body position during sleep
George, C. F., Millar, T. W., & Kryger, M. H. (1988) · Sleep · doi:10.1093/sleep/11.1.90 · Cited by 136 · cited 1×
“2 cmH2O variability will have been due to genuine random transient changes in requirement, as seen in the subject with a viral infection at 3 months, and random changes in sleep position [15] and sleep stage distribution probably account for further variability.”
Background
Cited by
Autoadjusting CPAP Therapy Based on Impedance Efficacy, Compliance and Acceptance
RANDERATH, W. J., SCHRAEDER, O., GALETKE, W., FELDMEYER, F., & RÜHLE, K.-H. (2001) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/ajrccm.163.3.2006168 · Cited by 143
Hui, D. S. C., Chan, J. K. W., Choy, D. K. L., Ko, F. W. S., Li, T. S. T., Leung, R. C. C., & Lai, C. K. W. (2000) · Chest · doi:10.1378/chest.117.5.1410 · Cited by 141
Two months auto-adjusting versus conventional nCPAP for obstructive sleep apnoea syndrome
Teschler, H., Wessendorf, T. E., Farhat, A. A., Konietzko, N., & Berthon-Jones, M. (2000) · European Respiratory Journal · doi:10.1034/j.1399-3003.2000.01503.x · Cited by 97
Self-citationAuto-Adjusting Versus Fixed Positive Pressure Therapy in Mild to Moderate Obstructive Sleep Apnoea
Nolan, G. M., Doherty, L. S., & Mc Nicholas, W. T. (2007) · Sleep · doi:10.1093/sleep/30.2.189 · Cited by 66
Evaluation of autoCPAP devices in home treatment of sleep apnea/hypopnea syndrome
Meurice, J. C., Cornette, A., Philip-Joet, F., Pepin, J. L., Escourrou, P., Ingrand, P., & Veale, D. (2007) · Sleep Medicine · doi:10.1016/j.sleep.2007.03.019 · Cited by 52
“Those studies usually found results comparable to conventional CPAP devices, but with a reduction in the mean pressure level required during the night, except with the use of Autoset (ResMed), with which pressure spent during 95% of the time (P 95) was generally higher than effective pressure on manual titration [12].”
MethodInfluential
Metadata
- Authors
- Teschler, H., Farhat, A., Exner, V, Konietzko, N., & Berthon-Jones, M.
- Venue
- European Respiratory Journal
- Publisher
- European Respiratory Society (ERS)
- Published
- 1997-09-01
- Volume
- 10
- Issue
- 9
- Pages
- 2073-2078
- ISSN
- 0903-1936 (ISSN-L)
- 1399-3003
- OpenAlex
- W2166611413
- Cited by
- 77
- References
- 4
- FWCI
- 4.02
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: AutoSet nasal CPAP titration: constancy of pressure, compliance and effectiveness at 8 month follow-up. https://app.cochat.ai/papers/10.1183/09031936.97.10092073
- The paper
- Teschler, H., Farhat, A., Exner, V, Konietzko, N., & Berthon-Jones, M. (1997). AutoSet nasal CPAP titration: constancy of pressure, compliance and effectiveness at 8 month follow-up. European Respiratory Journal. https://doi.org/10.1183/09031936.97.10092073
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.