Estimation of cardiac output and systemic vascular resistance using a multivariate regression model with features selected from the finger photoplethysmogram and routine cardiovascular measurements
Open accessAbstract
BACKGROUND: Cardiac output (CO) and systemic vascular resistance (SVR) are two important parameters of the cardiovascular system. The ability to measure these parameters continuously and noninvasively may assist in diagnosing and monitoring patients with suspected cardiovascular diseases, or other critical illnesses. In this study, a method is proposed to estimate both the CO and SVR of a heterogeneous cohort of intensive care unit patients (N=48). METHODS: Spectral and morphological features were extracted from the finger photoplethysmogram, and added to heart rate and mean arterial pressure as input features to a multivariate regression model to estimate CO and SVR. A stepwise feature search algorithm was employed to select statistically significant features. Leave-one-out cross validation was used to assess the generalized model performance. The degree of agreement between the estimation method and the gold standard was assessed using Bland-Altman analysis. RESULTS: The Bland-Altman bias ±precision (1.96 times standard deviation) for CO was -0.01 ±2.70 L min-1 when only photoplethysmogram (PPG) features were used, and for SVR was -0.87 ±412 dyn.s.cm-5 when only one PPG variability feature was used. CONCLUSIONS: These promising results indicate the feasibility of using the method described as a non-invasive preliminary diagnostic tool in supervised or unsupervised clinical settings.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Cardiac output has been estimated from pulse wave shape and low-frequency PPG variations.
“Cardiac output, monitored in peri-operative and critical care settings, has been estimated from pulse wave shape and low-frequency PPG variations”
From Wearable Photoplethysmography for Cardiovascular Monitoring (2022) · cites this paper as [178]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965 · see citing works
Measuring agreement in method comparison studies
Bland, J. M., & Altman, D. G. (1999) · Statistical Methods in Medical Research · doi:10.1177/096228029900800204 · Cited by 9,619 · see citing works
Measuring agreement in method comparison studies
Bland, J. M., & Altman, D. G. (1999) · Statistical Methods in Medical Research · doi:10.1191/096228099673819272 · Cited by 9,619 · see citing works
Measurement in Medicine: The Analysis of Method Comparison Studies
Altman, D. G., & Bland, J. M. (1983) · The Statistician · doi:10.2307/2987937 · Cited by 4,300 · see citing works
Comparing methods of measurement: why plotting difference against standard method is misleading
Bland, J. M., & Altman, D. G. (1995) · The Lancet · doi:10.1016/s0140-6736(95)91748-9 · Cited by 2,343 · see citing works
Critchley, L. A. H., & Critchley, J. (1999) · Journal of Clinical Monitoring and Computing · doi:10.1023/a:1009982611386 · Cited by 1,328 · see citing works
Critchley, L. A. H., Lee, A. S. Y., & Ho, A. M. ‐. (2010) · Anesthesia & Analgesia · doi:10.1213/ane.0b013e3181f08a5b · Cited by 513 · see citing works
Minimally Invasive Measurement of Cardiac Output during Surgery and Critical Care
Peyton, P. J., & Chong, S. W. (2010) · Anesthesiology · doi:10.1097/aln.0b013e3181ee3130 · Cited by 453 · see citing works
Non‐invasive pulsatile arterial pressure and stroke volume changes from the human finger
Bogert, L. W. J., & LIESHOUT, J. J. V. (2005) · Experimental Physiology · doi:10.1113/expphysiol.2005.030262 · Cited by 413 · see citing works
Kubicek, W. G., PATTERSON, R. P., & Witsoe, D. A. (1970) · Annals of the New York Academy of Sciences · doi:10.1111/j.1749-6632.1970.tb17735.x · Cited by 396 · see citing works
Fortin, J., Habenbacher, W., Heller, A., Hacker, A., Grüllenberger, R., Innerhofer, J., Passath, H., Wagner, C., Haitchi, G., Flotzinger, D., et al. (2005) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2005.06.001 · Cited by 227 · see citing works
The relationship between the photoplethysmographic waveform and systemic vascular resistance
Alian, A. A., Haddadin, A. S., Tantawy, H., Badr, T., Stout, R. G., Silverman, D. G., & Shelley, K. H. (2007) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-007-9097-5 · Cited by 136 · see citing works
Mathews, L., & Singh, K. A. (2008) · Annals of Cardiac Anaesthesia · doi:10.4103/0971-9784.38455 · Cited by 124 · see citing works
Remmen, J. J., Aengevaeren, W. R. M., Verheugt, F. W. A., van der Werf, T., Luijten, H. E., Bos, A., & Jansen, R. W. M. M. (2002) · Clinical Science · doi:10.1042/cs1030143 · Cited by 77 · see citing works
Foo, J. Y. A., & Wilson, S. J. (2006) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-005-0008-y · Cited by 74 · see citing works
Cardiac output measurements. A review of current techniques and research
Ehlers, K. C., Mylrea, K. C., Waterson, C. K., & Calkins, J. M. (1986) · Annals of Biomedical Engineering · doi:10.1007/bf02584272 · Cited by 61 · see citing works
Tang, C. H. H., Middleton, P. M., Savkin, A. V., Chan, G. S. H., Bishop, S. M., & Lovell, N. H. (2010) · Physiological Measurement · doi:10.1088/0967-3334/31/6/004 · Cited by 47 · see citing works
Self-citationPeripheral photoplethysmography variability analysis of sepsis patients
Middleton, P. M., Tang, C. H. H., Chan, G. S. H., Bishop, S. M., Savkin, A. V., & Lovell, N. H. (2010) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-010-0713-z · Cited by 45 · see citing works
Self-citationAnalysis of the Ear Pulse Oximeter Waveform
Alian, A. A., Stout, R. G., Ghobashy, M. A. M., Rezkanna, H., Silverman, D. G., & Shelley, K. H. (2006) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-006-9018-z · Cited by 45 · see citing works
Critchley, L. A. H. (2011) · Anesthesiology · doi:10.1097/aln.0b013e318215e1e5 · Cited by 42 · see citing works
Middleton, P. M., Chan, G. S. H., Steel, E., Malouf, P. J., Critoph, C., Flynn, G. M., O’Lone, E., Celler, B. G., & Lovell, N. H. (2011) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-011-0749-8 · Cited by 41 · see citing works
Self-citationWang, L., Poon, C. C. Y., & Zhang, Y. (2010) · Physiological Measurement · doi:10.1088/0967-3334/31/5/008 · Cited by 33 · see citing works
Remmen, J. J., Aengevaeren, W. R. M., Verheugt, F. W. A., et al. (2002) · Clinical Science · doi:10.1042/cs20010357 · Cited by 27 · see citing works
The History and Evolution of Circulatory Shock
Manji, R. A., Wood, K. E., & Kumar, A. (2009) · Critical Care Clinics · doi:10.1016/j.ccc.2008.12.013 · Cited by 24 · see citing works
Multivariate classification of systemic vascular resistance using photoplethysmography
Lee, Q. Y., Chan, G. S. H., Redmond, S. J., Middleton, P. M., Steel, E., Malouf, P. J., Critoph, C., Flynn, G. M., O’Lone, E., & Lovell, N. H. (2011) · Physiological Measurement · doi:10.1088/0967-3334/32/8/008 · Cited by 24 · see citing works
Self-citationSakka, S. G., & Meier‐Hellmann, A. (1999) · Journal of Cardiothoracic and Vascular Anesthesia · doi:10.1016/s1053-0770(99)90254-9 · Cited by 20 · see citing works
Prognostic value of hemodynamic findings from impedance cardiography in hypertensive stroke
RAMIREZ, M., Tibayan, R., Marinas, C. E., YAMAMOTO, M., & Caguioa, E. V. (2005) · American Journal of Hypertension · doi:10.1016/j.amjhyper.2004.11.027 · Cited by 20 · see citing works
Continuous measurement of systemic vascular resistance
O'Dwyer, J., KING, J. E., Wood, C. E., Taylor, B. L., & Smith, G. B. (1994) · Anaesthesia · doi:10.1111/j.1365-2044.1994.tb14225.x · Cited by 5 · see citing works
Gawlikowski, M., & Pustelny, T. P. (2012) · BioMedical Engineering OnLine · doi:10.1186/1475-925x-11-24 · Cited by 4 · see citing works
Flagged first, then by citations
Research works printed but not linked: 1
- Draper NR Smith H Pownell E Applied Regression Analysis, Volume 3 1966 New York: Wiley
Cited by
Sorted by citations
Photoplethysmogram Analysis and Applications: An Integrative Review
Park, J., Seok, H. S., Kim, S.-S., & Shin, H. (2022) · Frontiers in Physiology · doi:10.3389/fphys.2021.808451 · Cited by 450 · see citing works
Flexible graphene photodetectors for wearable fitness monitoring
Polat, E. O., Mercier, G., Nikitskiy, I., Puma, E., Galán, T., Gupta, S., Marques, M. M., Piqueras, J. J., Bouwens, M., Durduran, T., et al. (2019) · Science Advances · doi:10.1126/sciadv.aaw7846 · Cited by 299 · see citing works
Wearable Photoplethysmography for Cardiovascular Monitoring
Charlton, P. H., Kyriacou, P. A., Mant, J., Marozas, V., Chowienczyk, P., & Alastruey, J. (2022) · Proceedings of the IEEE · doi:10.1109/jproc.2022.3149785 · Cited by 250 · see citing works
Transducer Technologies for Biosensors and Their Wearable Applications
Polat, E. O., Çetin, M. M., Tabak, A. F., Güven, E. B., Uysal, B. Ö., Arsan, T., Kabbani, A., Hamed, H., & GÜL, S. B. (2022) · Biosensors · doi:10.3390/bios12060385 · Cited by 209 · see citing works
Efficacy of flavonoids in the management of high blood pressure
Clark, J. L., Zahradka, P. C., & Taylor, C. G. (2015) · Nutrition Reviews · doi:10.1093/nutrit/nuv048 · Cited by 135 · see citing works
Wearable Photoplethysmography for Cardiovascular Monitoring.
Charlton, P. H., Kyriacou, P. A., Mant, J., Marozas, V., Chowienczyk, P., & Alastruey, J. (2022) · KTUePubl (Repository of Kaunas University of Technology) · doi:10.17863/cam.81148 · Cited by 113 · see citing works
Cuffless blood pressure estimation using only a smartphone
Matsumura, K., Rolfe, P. J., Toda, S., & Yamakoshi, T. (2018) · Scientific Reports · doi:10.1038/s41598-018-25681-5 · Cited by 102 · see citing works
Systolic blood pressure estimation using PPG and ECG during physical exercise
Sun, S., Bezemer, R., Long, X., Muehlsteff, J., & Aarts, R. M. (2016) · Physiological Measurement · doi:10.1088/0967-3334/37/12/2154 · Cited by 78 · see citing works
Clinical Validation of Heart Rate Apps: Mixed-Methods Evaluation Study
Vandenberk, T., Stans, J., Mortelmans, C., Haelst, R. V., Schelvergem, G. V., Pelckmans, C., Smeets, C. J. P. P., Lanssens, D., Cannière, H. D., Storms, V., et al. (2017) · JMIR mhealth and uhealth · doi:10.2196/mhealth.7254 · Cited by 58 · see citing works
Andreozzi, E., Sabbadini, R., Centracchio, J., Bifulco, P., Irace, A., Breglio, G., & Riccio, M. (2022) · Sensors · doi:10.3390/s22197566 · Cited by 36 · see citing works
Michael, S., Jay, O., Graham, K. S., & Davis, G. M. (2017) · Applied Physiology, Nutrition, and Metabolism · doi:10.1139/apnm-2017-0049 · Cited by 28 · see citing works
Michael, S., Jay, O., Graham, K. S., & Davis, G. M. (2017) · European Journal of Applied Physiology · doi:10.1007/s00421-017-3673-2 · Cited by 25 · see citing works
Non-invasive vascular resistance monitoring with a piezoelectric sensor and photoplethysmogram
Wang, L., Ansari, S., Slavin, D., Ward, K., Najarian, K., & Oldham, K. R. (2017) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2017.06.007 · Cited by 16 · see citing works
Xu, X., Tang, Q., & Chen, Z. (2023) · Sensors · doi:10.3390/s23229057 · Cited by 16 · see citing works
Quality controlled compression technique for Photoplethysmogram monitoring applications
Alam, S., Gupta, R., & Bera, J. (2018) · Measurement · doi:10.1016/j.measurement.2018.07.091 · Cited by 15 · see citing works
Assessment of postoperative pain intensity by using photoplethysmography
Peng, L., Yu, S., Wei, W., Gong, Q., & Gao, B. (2014) · Journal of Anesthesia · doi:10.1007/s00540-014-1837-3 · Cited by 14 · see citing works
Altamirano‐Diaz, L., Kassay, A., Serajelahi, B., McIntyre, C. W., Filler, G., & Kharche, S. (2019) · Frontiers in Physiology · doi:10.3389/fphys.2019.01391 · Cited by 10 · see citing works
Rapp, E. S., Pawar, S. R., Gohean, J. R., Larson, E. R., Smalling, R. W., & Longoria, R. G. (2019) · Journal of Engineering and Science in Medical Diagnostics and Therapy · doi:10.1115/1.4045204 · Cited by 10 · see citing works
Classification of photoplethysmogram signal using self organizing map
Ghosal, P., & Gupta, R. (2015) · International Conference on Research in Computational Intelligence and Communication Networks (ICRCICN) · doi:10.1109/icrcicn.2015.7434220 · Cited by 10 · see citing works
Deep Learning Model for Blood Pressure Estimation from PPG Signal
Kim, M., Lee, H., Kim, K., & Kim, K. H. (2022) · IEEE International Conference on Metrology for eXtended Reality, Artificial Intelligence and Neural Engineering (MetroXRAINE) · doi:10.1109/metroxraine54828.2022.9967606 · Cited by 8 · see citing works
Piezoelectric Nanofibers for Noninvasive Wearable Blood Pressure Monitoring
Closson, A. B., Xu, Z. Y., Bargamian, J., Bryan, A. C., Abess, A., & Zhang, J. X. J. (2024) · ACS Applied Electronic Materials · doi:10.1021/acsaelm.4c00840 · Cited by 6 · see citing works
Zhang, Z., Ansari, S., Wang, L., Aaronson, K. D., Golbus, J. R., & Oldham, K. R. (2021) · IFAC-PapersOnLine · doi:10.1016/j.ifacol.2021.11.190 · Cited by 6 · see citing works
Vahdani-Manaf, N. (2015) · Journal of Applied Sciences · doi:10.3923/jas.2015.1312.1317 · Cited by 5 · see citing works
Momeni, M., Wuthe, S., Mølmer, M. B., Svendsen, E. L., Brabrand, M., Biesenbach, P., & Teichmann, D. (2025) · IEEE Sensors Journal · doi:10.1109/jsen.2025.3553368 · Cited by 3 · see citing works
Ling, P., Gong, Q., Yu, S., Bo, G., & Wei, W. (2017) · Journal of Anesthesia · doi:10.1007/s00540-017-2338-y · Cited by 3 · see citing works
Haemodynamic determinants of supine hypertension in patients with classical orthostatic hypotension
van der Stam, A. H., Gagaouzova, B. S., Kerkhof, F. I., van Rossum, I. A., Shmuely, S., Reijntjes, R. H. A. M., van Houwelingen, M. J., Thijs, R. D., & van Dijk, J. G. (2025) · Journal of Hypertension · doi:10.1097/hjh.0000000000004194 · Cited by 2 · see citing works
Towards Cardiac Output Estimation Using Earbud Photoplethysmography Sensor
Zhang, L., Rahman, M. M., Morshed, M. B., Zhu, L., Zhou, H., O’Bryan, J. F., Mendes, W. B., & Kuang, J. (2025) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc58623.2025.11254388 · Cited by 2 · see citing works
Castro, A., Carvalho, P., Muehlsteff, J., Mattos, S. D. S., & Coimbra, M. T. (2016) · Advances in bioinformatics and biomedical engineering book series · doi:10.4018/978-1-5225-0660-7.ch002 · Cited by 1 · see citing works
Yan, T., Yang, L., Zhang, S., Yang, Y., & Li, X. (2015) · Bio-Medical Materials and Engineering · doi:10.3233/bme-151327
Sumaratih, L., Soedono, R., & Aditianingsih, D. (2019) · Bioscientia Medicina Journal of Biomedicine and Translational Research · doi:10.32539/bsm.v3i4.99
Estimation of Non-invasive Cuff-less Blood Pressure Using the Photoplethysmogram SignalPreprint
Rajagopal, S., Manimegalai, P., & Ramachandaran, D. (2020) · doi:10.21203/rs.3.rs-22297/v1
Castro, A., Carvalho, P., Muehlsteff, J., Mattos, S. D. S., & Coimbra, M. T. (2017) · IGI Global eBooks · doi:10.4018/978-1-5225-3158-6.ch045
Momeni, M., Wuthe, S., Mølmer, M. B., Svendsen, E. L., Brabrand, M., Peimankar, A., Biesenbach, P., & Teichmann, D. (2025) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc58623.2025.11253932
Khodaei, S., Daeian, M., Maftoon, N., & Keshavarz‐Motamed, Z. (2026) · Elsevier eBooks · doi:10.1016/b978-0-443-27526-5.00002-8
- Authors
- Lee, Q. Y., Redmond, S. J., Chan, G. S., Middleton, P. M., Steel, E., Malouf, P., Critoph, C., Flynn, G., O’Lone, E., & Lovell, N. H.
- Affiliations
- UNSW Sydney (AU)
- The University of Sydney (AU)
- Prince of Wales Hospital (AU)
- Corresponding author
- Lee, Qim Y
- Venue
- BioMedical Engineering OnLine
- Published
- 2013
- Volume
- 12
- Issue
- 1
- Pages
- 19
- ISSN
- 1475-925X (ISSN-L)
- PMC
- PMC3649882
- OpenAlex
- W2136022512
- Cited by
- 34
- References
- 27
- Citations by year
- 2026: 1 · 2025: 4 · 2024: 1 · 2023: 1 · 2022: 6 · 2021: 1 · 2020: 1 · 2019: 4 · 2018: 2 · 2017: 6 · 2016: 2 · 2015: 4 · 2014: 1
- FWCI
- 1.29
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- cc-by
- Locations
- BioMedical Engineering OnLine (publishedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Rate Variability and Autonomic Control
- Keywords
- cardiac output estimation, systemic vascular resistance, noninvasive hemodynamic monitoring, multivariate regression, feature selection, intensive care unit patients, mean arterial pressure, heart rate
- MeSH terms
- Aged, Blood Pressure, Cardiac Output, Cardiovascular Diseases, Cardiovascular System, Female, Fingers, Heart Rate, Humans, Male, Monitoring, Physiologic, Regression Analysis, Signal Processing, Computer-Assisted, Vascular Resistance, Reproducibility of Results, Multivariate Analysis, Photoplethysmography, Arterial Pressure
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review