Numerical simulation of ISFET structures for biosensing devices with TCAD tools

Open access
Passeri, Daniele, Morozzi, Arianna, Kanxheri, Keida, Scorzoni, Andrea · 2015 · BioMedical Engineering OnLine

The DOI matches its Crossref record; 16 of the 16 references it cites were found in a registry or an index; none of the 14 sources its claims cite is checked for support yet.

Abstract

Numerical simulation approach to ISFET structures for BioFET design optimization using Synopsys Sentaurus TCAD.

Overview

CoChat Summary

## TL;DR

The paper presents a TCAD-based numerical simulation framework for ISFET/BioFET biosensors by modeling the electrolyte as an intrinsic semiconductor in Synopsys Sentaurus. The simulations show that current modulation is strongly non-linear, depends on target-charge position, and can be greatly affected by reference-electrode bias.

## Key findings

  • In an ISFET pH-sensing simulation, the authors report a sensitivity of about 50 mV/pH in threshold-voltage shift, “in agreement with literature data.”
  • For a general BioFET, the paper reports no linear relationship between the number of charges and current modulation; target position matters strongly.
  • Charges localized near the source are more effective than those near the drain, and randomly distributed targets are more efficient than locally grouped targets for modulating current.
  • Increasing the reference electrode voltage can increase the drain current by around four orders of magnitude; with VREF = 1 V the transistor turns on, and VREF = 2 V enhances the effect further.

## Methods

  • The study uses Synopsys Sentaurus TCAD to simulate ISFET/BioFET structures at the device level.
  • The electrolyte is modeled by replacing the solution with a custom intrinsic semiconductor, with parameters (including permittivity, refractive index, mobility, and density of states) tuned to mimic water and ionic solutions.
  • The authors simulate ISFET transfer/output characteristics and a general BioFET with a p-type low-doped Si substrate, thin SiO2, electrolyte, reference electrode, and localized fixed charges represented by dielectric blocks; they vary pH / ion concentration, target number (from 1 to 10), charge position, and reference voltage.

## Limitations

  • The paper states that the results for a general BioFET require simulating many receptor-target combinations, which is described as an “overwhelming computational effort”; they therefore used “a still huge number of several randomly selected sample cases.”
  • The text makes evident that the electrolyte representation is an approximation: a real electrolyte is replaced by an intrinsic semiconductor, so the model is not a direct physical simulation of the solution itself.
  • The text also makes evident that the paper does not report quantitative validation against experimental data for the general BioFET case beyond stating that some current magnitudes are in agreement with literature.

Model-written summary. Check it against the paper before relying on it.

Integrity signals

Registry

Crossref · 10.1186/1475-925x-14-s2-s3

Cites

16 of 16 references verified0 citation errors · 0 not checkable

Statistics

Nothing to recompute.

Figures

Figure 1 ISFET sketch. Cross-section of the simulated ISFET device
Figure 1 ISFET sketch. Cross-section of the simulated ISFET deviceCC BY · PubMed Central
Figure 2 ISFET simulation. Output characteristics (ID vs. VDS) of the ISFET device.
Figure 2 ISFET simulation. Output characteristics (ID vs. VDS) of the ISFET device.CC BY · PubMed Central
Figure 3 ISFET as a pH sensor - 1. Drain current as a function of the drain voltage at different pH solution concentrations (VREF = 0).
Figure 3 ISFET as a pH sensor - 1. Drain current as a function of the drain voltage at different pH solution concentrations (VREF = 0).CC BY · PubMed Central
Figure 4 ISFET as a pH sensor - 2. Drain current as a function of the reference voltage at different pH solution concentrations (VDS = 2 V).
Figure 4 ISFET as a pH sensor - 2. Drain current as a function of the reference voltage at different pH solution concentrations (VDS = 2 V).CC BY · PubMed Central
Figure 5 BioFET simulated structure. Sketch of the simulated structure: charge localization at the SiO2 surface is modelled with localized dielectric block with or without charge.
Figure 5 BioFET simulated structure. Sketch of the simulated structure: charge localization at the SiO2 surface is modelled with localized dielectric block with or without charge.CC BY · PubMed Central
Figure 6 BioFET channel modulation. Channel modulation as a function of the number of the occupied receptors: electron density along the channel.
Figure 6 BioFET channel modulation. Channel modulation as a function of the number of the occupied receptors: electron density along the channel.CC BY · PubMed Central
Figure 7 BioFET electrical potential. Electrostatic potential along a vertical structure cut (on the right) at different solution concentrations.
Figure 7 BioFET electrical potential. Electrostatic potential along a vertical structure cut (on the right) at different solution concentrations.CC BY · PubMed Central
Figure 8 BioFET drain current. Drain current as a function of the reference voltage, depending on the position of a single charged receptor.
Figure 8 BioFET drain current. Drain current as a function of the reference voltage, depending on the position of a single charged receptor.CC BY · PubMed Central
Figure 9 BioFET channel charge distribution. Channel charge density: localized vs. distributed charged receptors.
Figure 9 BioFET channel charge distribution. Channel charge density: localized vs. distributed charged receptors.CC BY · PubMed Central
Figure 10 BioFET current evaluation. Drain current as a function of the reference electrode voltage with three active receptors: localized vs. distributed charged receptor effect on the ID current.
Figure 10 BioFET current evaluation. Drain current as a function of the reference electrode voltage with three active receptors: localized vs. distributed charged receptor effect on the ID current.CC BY · PubMed Central
Figure 11 BioFET channel modulation: voltage reference effects. Electron current densities at different reference electrode voltages.
Figure 11 BioFET channel modulation: voltage reference effects. Electron current densities at different reference electrode voltages.CC BY · PubMed Central
Figure 12 BioFET current at different reference electrode voltages. ID as a function of VDS at different VREF.
Figure 12 BioFET current at different reference electrode voltages. ID as a function of VDS at different VREF.CC BY · PubMed Central
Figure 13 Drain current as a function of the number of charged receptors - 0 V. Drain current as a function of the number of charged receptors: coloured circles in each column (e.g. for the same number of charged receptors) correspond to different random position combinations (VDS = 100mV, VREF = 0V).
Figure 13 Drain current as a function of the number of charged receptors - 0 V. Drain current as a function of the number of charged receptors: coloured circles in each column (e.g. for the same number of charged receptors) correspond to different random position combinations (VDS = 100mV, VREF = 0V).CC BY · PubMed Central
Figure 14 Drain current as a function of the number of charged receptors - 1 V. Drain current as a function of the number of charged receptors: coloured circles in each column (e.g. for the same number of charged receptors) correspond to different random position combinations(VDS = 100mV, VREF = 1V).
Figure 14 Drain current as a function of the number of charged receptors - 1 V. Drain current as a function of the number of charged receptors: coloured circles in each column (e.g. for the same number of charged receptors) correspond to different random position combinations(VDS = 100mV, VREF = 1V).CC BY · PubMed Central

Verification

Claims the paper makes

0 Valid · 0 Not valid · 11 Not checked · 31 Own results and interpretations

24 more claims

Create a free account to see all 42 claims and every source verdict.

See all claims — free account
  • Probably not a substantial claim. 8
  • It may state two separate claims as one. 8

Create a free account to see these claims.

Citations

Cites

Show all 16 references

Cited by

See all 53

Metadata

Authors
Passeri, D., Morozzi, A., Kanxheri, K., & Scorzoni, A.
Affiliations
University of Perugia (IT)
Corresponding author
Passeri, Daniele
Venue
BioMedical Engineering OnLine
Published
2015
Volume
14
Issue
Suppl 2
Pages
S3
ISSN
1475-925X (ISSN-L)
OpenAlex
W1916200669
Cited by
53
References
16
Citations by year
2026: 3 · 2025: 5 · 2024: 3 · 2023: 5 · 2022: 7 · 2021: 10 · 2020: 3 · 2019: 6 · 2018: 3 · 2017: 5 · 2016: 1 · 2015: 2
Type
conference abstract
Language
en
Open access
Licence
cc-by
Locations
  • BioMedical Engineering OnLine (publishedVersion, cc-by)
  • Europe PMC (PubMed Central) (publishedVersion, cc-by)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
Topics
  • Physical Sciences › Chemistry › Analytical Chemistry › Analytical Chemistry and Sensors
  • Physical Sciences › Engineering › Electrical and Electronic Engineering › Advancements in Semiconductor Devices and Circuit Design
  • Physical Sciences › Engineering › Electrical and Electronic Engineering › Analog and Mixed-Signal Circuit Design
Keywords
ion-sensitive field-effect transistor, BioFET, biosensors, electrolyte modeling, electrolyte replacement, charge capture, current modulation, transfer characteristics, electrolyte solutions, device simulation
MeSH terms
Models, Theoretical, Time Factors, Transistors, Electronic, Biosensing Techniques, Computer-Aided Design

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).

A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Numerical simulation of ISFET structures for biosensing devices with TCAD tools. https://app.cochat.ai/papers/10.1186/1475-925x-14-s2-s3
The paper
Passeri, D., Morozzi, A., Kanxheri, K., & Scorzoni, A. (2015). Numerical simulation of ISFET structures for biosensing devices with TCAD tools. BioMedical Engineering OnLine. https://doi.org/10.1186/1475-925x-14-s2-s3

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.