Biopsy-based calibration of T2* magnetic resonance for estimation of liver iron concentration and comparison with R2 Ferriscan
Open accessThe DOI matches its Crossref record; 31 of the 31 references it cites were found in a registry or an index.
Abstract
BACKGROUND: There is a need to standardise non-invasive measurements of liver iron concentrations (LIC) so clear inferences can be drawn about body iron levels that are associated with hepatic and extra-hepatic complications of iron overload. Since the first demonstration of an inverse relationship between biopsy LIC and liver magnetic resonance (MR) using a proof-of-concept T2* sequence, MR technology has advanced dramatically with a shorter minimum echo-time, closer inter-echo spacing and constant repetition time. These important advances allow more accurate calculation of liver T2* especially in patients with high LIC. METHODS: Here, we used an optimised liver T2* sequence calibrated against 50 liver biopsy samples on 25 patients with transfusional haemosiderosis using ordinary least squares linear regression, and assessed the method reproducibility in 96 scans over an LIC range up to 42 mg/g dry weight (dw) using Bland-Altman plots. Using mixed model linear regression we compared the new T2*-LIC with R2-LIC (Ferriscan) on 92 scans in 54 patients with transfusional haemosiderosis and examined method agreement using Bland-Altman approach. RESULTS: Strong linear correlation between ln(T2*) and ln(LIC) led to the calibration equation LIC = 31.94(T2*)-1.014. This yielded LIC values approximately 2.2 times higher than the proof-of-concept T2* method. Comparing this new T2*-LIC with the R2-LIC (Ferriscan) technique in 92 scans, we observed a close relationship between the two methods for values up to 10 mg/g dw, however the method agreement was poor. CONCLUSIONS: New calibration of T2* against liver biopsy estimates LIC in a reproducible way, correcting the proof-of-concept calibration by 2.2 times. Due to poor agreement, both methods should be used separately to diagnose or rule out liver iron overload in patients with increased ferritin.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Measuring agreement in method comparison studies
Bland, J. M., & Altman, D. G. (1999) · Statistical Methods in Medical Research · doi:10.1177/096228029900800204 · Cited by 9,619 · cited 2×
“For inter-observer liver R2* agreement, 96 liver R2* scans from 38 patients were independently evaluated in triplicate by two observers as described above (Figure 1) and analysed using BlandAltman plots [25].”
MethodMeasuring agreement in method comparison studies
Bland, J. M., & Altman, D. G. (1999) · Statistical Methods in Medical Research · doi:10.1191/096228099673819272 · Cited by 9,619
Cardiovascular T2-star (T2*) magnetic resonance for the early diagnosis of myocardial iron overload
Anderson, L. (2001) · European Heart Journal · doi:10.1053/euhj.2001.2822 · Cited by 1,682 · cited 7×
“A variety of non-invasive methods have been used to measure LIC including liver susceptometry (SQUID), [11,12], T2-weighted spin-echo MR with [13] or without signal intensity ratios (SIR) to adjacent tissues, [14] and T2* gradient-echo MR with [15] or without SIR, [16-18].”
BackgroundMethodInfluentialWood, J. C., Enriquez, C., Ghugre, N., Tyzka, J. M., Carson, S., Nelson, M. D., & Coates, T. D. (2005) · Blood · doi:10.1182/blood-2004-10-3982 · Cited by 1,045 · cited 7×
“A variety of non-invasive methods have been used to measure LIC including liver susceptometry (SQUID), [11,12], T2-weighted spin-echo MR with [13] or without signal intensity ratios (SIR) to adjacent tissues, [14] and T2* gradient-echo MR with [15] or without SIR, [16-18].”
BackgroundMethodInfluentialNoninvasive measurement and imaging of liver iron concentrations using proton magnetic resonance
St. Pierre, T. G., Clark, P. R., Chua-anusorn, W., Fleming, A. J., Jeffrey, G. P., Olynyk, J. K., Pootrakul, P., Robins, E., & Lindeman, R. (2005) · Blood · doi:10.1182/blood-2004-01-0177 · Cited by 909 · cited 6×
“A variety of non-invasive methods have been used to measure LIC including liver susceptometry (SQUID), [11,12], T2-weighted spin-echo MR with [13] or without signal intensity ratios (SIR) to adjacent tissues, [14] and T2* gradient-echo MR with [15] or without SIR, [16-18].”
BackgroundMethodResultInfluentialCappellini, M. D. (2006) · Blood · doi:10.1182/blood-2005-08-3430 · Cited by 717 · cited 6×
“Such differences have also been an issue with SQUID; in a recent large-scale study an approximately two-fold correction factor was required between biopsy and SQUID methodology [2,4].”
BackgroundMethodResultInfluentialNon-invasive assessment of hepatic iron stores by MRI
Gandon, Y., Olivié, D., Guyader, D., Aubé, C., Oberti, F., Sebille, V, & Deugnier, Y. (2004) · The Lancet · doi:10.1016/s0140-6736(04)15436-6 · Cited by 671
Hepatic Iron Concentration and Total Body Iron Stores in Thalassemia Major
Angelucci, E., Brittenham, G. M., McLaren, C. E., Ripalti, M., Baronciani, D., Giardini, C., Galimberti, M., Polchi, P., & Lucarelli, G. (2000) · New England Journal of Medicine · doi:10.1056/nejm200008033430503 · Cited by 607
R2* magnetic resonance imaging of the liver in patients with iron overload
Hankins, J. S., McCarville, M. B., Loeffler, R. B., Smeltzer, M. P., Onciu, M., Hoffer, F. A., Li, C.-S., Wang, W. C., Ware, R. E., & Hillenbrand, C. M. (2009) · Blood · doi:10.1182/blood-2008-12-191643 · Cited by 372 · cited 3×
“A variety of non-invasive methods have been used to measure LIC including liver susceptometry (SQUID), [11,12], T2-weighted spin-echo MR with [13] or without signal intensity ratios (SIR) to adjacent tissues, [14] and T2* gradient-echo MR with [15] or without SIR, [16-18].”
BackgroundMethodLiver iron concentrations and urinary hepcidin in -thalassemia
Origa, R., Galanello, R., Ganz, T., Giagu, N., Maccioni, L., Faa, G., & Nemeth, E. (2007) · Haematologica · doi:10.3324/haematol.10842 · Cited by 365 · cited 1×
“Conversely, serum ferritin is disproportionately decreased relative to LIC in ascorbate deficiency [7] and in conditions where iron preferentially loads hepatocytes rather than macrophages, such as thalassemia intermedia [8].”
Background
Cited by
Farmakis, D., Porter, J., Taher, A., Domenica Cappellini, M., Angastiniotis, M., Eleftheriou, A., & Taskforce, F. T. 2. T. G. (2022) · HemaSphere · doi:10.1097/hs9.0000000000000732 · Cited by 307
Self-citationLiver Iron Quantification with MR Imaging: A Primer for Radiologists
Labranche, R., Gilbert, G., Cerny, M., Vu, K.-N., Soulières, D., Olivié, D., Billiard, J.-S., Yokoo, T., & Tang, A. (2018) · RadioGraphics · doi:10.1148/rg.2018170079 · Cited by 171
“However, there has not been consensus on the recommended acquisition parameters for assessment of R2* (30,67,68).”
BackgroundQuantification of Liver Iron Overload with MRI: Review and Guidelines from the ESGAR and SAR
Reeder, S. B., Yokoo, T., França, M., Hernando, D., Alberich-Bayarri, Á., Alústiza, J. M., Gandon, Y., Henninger, B., Hillenbrand, C., Jhaveri, K., et al. (2023) · Radiology · doi:10.1148/radiol.221856 · Cited by 167
Iron imaging reveals tumor and metastasis macrophage hemosiderin deposits in breast cancer
Leftin, A., Ben-Chetrit, N., Klemm, F., Joyce, J. A., & Koutcher, J. A. (2017) · PLOS ONE · doi:10.1371/journal.pone.0184765 · Cited by 36
Leftin, A., Zhao, H., Turkekul, M., de Stanchina, E., Manova, K., & Koutcher, J. A. (2017) · Scientific Reports · doi:10.1038/s41598-017-11899-2 · Cited by 33
Metadata
- Authors
- Garbowski, M. W., Carpenter, J.-P., Smith, G., Roughton, M., Alam, M. H., He, T., Pennell, D. J., & Porter, J. B.
- Venue
- Journal of Cardiovascular Magnetic Resonance
- Publisher
- Elsevier BV
- Published
- 2014-12
- Volume
- 16
- Issue
- 1
- Pages
- 40
- ISSN
- 1097-6647 (ISSN-L)
- OpenAlex
- W2133662247
- Cited by
- 203
- References
- 31
- FWCI
- 10.20
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.