Biopsy-based calibration of T2* magnetic resonance for estimation of liver iron concentration and comparison with R2 Ferriscan
Open accessThe DOI matches its Crossref record; 31 of the 31 references it cites were found in a registry or an index.
Abstract
BACKGROUND: There is a need to standardise non-invasive measurements of liver iron concentrations (LIC) so clear inferences can be drawn about body iron levels that are associated with hepatic and extra-hepatic complications of iron overload. Since the first demonstration of an inverse relationship between biopsy LIC and liver magnetic resonance (MR) using a proof-of-concept T2* sequence, MR technology has advanced dramatically with a shorter minimum echo-time, closer inter-echo spacing and constant repetition time. These important advances allow more accurate calculation of liver T2* especially in patients with high LIC. METHODS: Here, we used an optimised liver T2* sequence calibrated against 50 liver biopsy samples on 25 patients with transfusional haemosiderosis using ordinary least squares linear regression, and assessed the method reproducibility in 96 scans over an LIC range up to 42 mg/g dry weight (dw) using Bland-Altman plots. Using mixed model linear regression we compared the new T2*-LIC with R2-LIC (Ferriscan) on 92 scans in 54 patients with transfusional haemosiderosis and examined method agreement using Bland-Altman approach. RESULTS: Strong linear correlation between ln(T2*) and ln(LIC) led to the calibration equation LIC = 31.94(T2*)-1.014. This yielded LIC values approximately 2.2 times higher than the proof-of-concept T2* method. Comparing this new T2*-LIC with the R2-LIC (Ferriscan) technique in 92 scans, we observed a close relationship between the two methods for values up to 10 mg/g dw, however the method agreement was poor. CONCLUSIONS: New calibration of T2* against liver biopsy estimates LIC in a reproducible way, correcting the proof-of-concept calibration by 2.2 times. Due to poor agreement, both methods should be used separately to diagnose or rule out liver iron overload in patients with increased ferritin.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Measuring agreement in method comparison studies
Bland, J. M., & Altman, D. G. (1999) · Statistical Methods in Medical Research · doi:10.1177/096228029900800204 · Cited by 9,619 · cited 2×
“For inter-observer liver R2* agreement, 96 liver R2* scans from 38 patients were independently evaluated in triplicate by two observers as described above (Figure 1) and analysed using BlandAltman plots [25].”
MethodMeasuring agreement in method comparison studies
Bland, J. M., & Altman, D. G. (1999) · Statistical Methods in Medical Research · doi:10.1191/096228099673819272 · Cited by 9,619
Cardiovascular T2-star (T2*) magnetic resonance for the early diagnosis of myocardial iron overload
Anderson, L. (2001) · European Heart Journal · doi:10.1053/euhj.2001.2822 · Cited by 1,682 · cited 7×
“A variety of non-invasive methods have been used to measure LIC including liver susceptometry (SQUID), [11,12], T2-weighted spin-echo MR with [13] or without signal intensity ratios (SIR) to adjacent tissues, [14] and T2* gradient-echo MR with [15] or without SIR, [16-18].”
BackgroundMethodInfluentialWood, J. C., Enriquez, C., Ghugre, N., Tyzka, J. M., Carson, S., Nelson, M. D., & Coates, T. D. (2005) · Blood · doi:10.1182/blood-2004-10-3982 · Cited by 1,045 · cited 7×
“A variety of non-invasive methods have been used to measure LIC including liver susceptometry (SQUID), [11,12], T2-weighted spin-echo MR with [13] or without signal intensity ratios (SIR) to adjacent tissues, [14] and T2* gradient-echo MR with [15] or without SIR, [16-18].”
BackgroundMethodInfluentialNoninvasive measurement and imaging of liver iron concentrations using proton magnetic resonance
St. Pierre, T. G., Clark, P. R., Chua-anusorn, W., Fleming, A. J., Jeffrey, G. P., Olynyk, J. K., Pootrakul, P., Robins, E., & Lindeman, R. (2005) · Blood · doi:10.1182/blood-2004-01-0177 · Cited by 909 · cited 6×
“A variety of non-invasive methods have been used to measure LIC including liver susceptometry (SQUID), [11,12], T2-weighted spin-echo MR with [13] or without signal intensity ratios (SIR) to adjacent tissues, [14] and T2* gradient-echo MR with [15] or without SIR, [16-18].”
BackgroundMethodResultInfluentialCappellini, M. D. (2006) · Blood · doi:10.1182/blood-2005-08-3430 · Cited by 717 · cited 6×
“Such differences have also been an issue with SQUID; in a recent large-scale study an approximately two-fold correction factor was required between biopsy and SQUID methodology [2,4].”
BackgroundMethodResultInfluentialNon-invasive assessment of hepatic iron stores by MRI
Gandon, Y., Olivié, D., Guyader, D., Aubé, C., Oberti, F., Sebille, V, & Deugnier, Y. (2004) · The Lancet · doi:10.1016/s0140-6736(04)15436-6 · Cited by 671
Hepatic Iron Concentration and Total Body Iron Stores in Thalassemia Major
Angelucci, E., Brittenham, G. M., McLaren, C. E., Ripalti, M., Baronciani, D., Giardini, C., Galimberti, M., Polchi, P., & Lucarelli, G. (2000) · New England Journal of Medicine · doi:10.1056/nejm200008033430503 · Cited by 607
R2* magnetic resonance imaging of the liver in patients with iron overload
Hankins, J. S., McCarville, M. B., Loeffler, R. B., Smeltzer, M. P., Onciu, M., Hoffer, F. A., Li, C.-S., Wang, W. C., Ware, R. E., & Hillenbrand, C. M. (2009) · Blood · doi:10.1182/blood-2008-12-191643 · Cited by 372 · cited 3×
“A variety of non-invasive methods have been used to measure LIC including liver susceptometry (SQUID), [11,12], T2-weighted spin-echo MR with [13] or without signal intensity ratios (SIR) to adjacent tissues, [14] and T2* gradient-echo MR with [15] or without SIR, [16-18].”
BackgroundMethodLiver iron concentrations and urinary hepcidin in -thalassemia
Origa, R., Galanello, R., Ganz, T., Giagu, N., Maccioni, L., Faa, G., & Nemeth, E. (2007) · Haematologica · doi:10.3324/haematol.10842 · Cited by 365 · cited 1×
“Conversely, serum ferritin is disproportionately decreased relative to LIC in ascorbate deficiency [7] and in conditions where iron preferentially loads hepatocytes rather than macrophages, such as thalassemia intermedia [8].”
BackgroundCulpin, M. (1926) · The Lancet · Cited by 312
Cappellini, M. D., Porter, J., El-Beshlawy, A., Li, C. K., Seymour, J. F., Elalfy, M., Gattermann, N., Giraudier, S., Lee, J. W., Chan, L. L., et al. (2010) · Haematologica · doi:10.3324/haematol.2009.014696 · Cited by 288
Self-citationMEASUREMENT OF LIVER-IRON CONCENTRATION IN NEEDLE-BIOPSY SPECIMENS
BARRY, M. (1971) · The Lancet · doi:10.1016/s0140-6736(71)90838-5 · Cited by 288 · cited 2×
“In the original T2* publication by Anderson, [16] biochemical determination of LIC was performed on formalin fixed cores but not on paraffin embedded tissue, [29] which could explain in part why there is an approximately 2-fold difference with the calibration in this paper where paraffin embedded biopsies were used.”
BackgroundMethodBrittenham, G. M., Cohen, A. R., McLaren, C. E., Martin, M. B., Griffith, P. M., Nienhuis, A. W., Young, N. S., Allen, C. J., Farrell, D. E., & Harris, J. W. (1993) · American Journal of Hematology · doi:10.1002/ajh.2830420116 · Cited by 270 · cited 1×
“ferritin as a single measure of iron overload and response to chelation treatment, and also because variability in LIC accounts for only 57% of variability in serum ferritin [5].”
BackgroundPorter, J., Galanello, R., Saglio, G., Neufeld, E. J., Vichinsky, E., Cappellini, M. D., Olivieri, N., Piga, A., Cunningham, M. J., Soulières, D., et al. (2008) · European Journal of Haematology · doi:10.1111/j.1600-0609.2007.00985.x · Cited by 237 · cited 5×
“Such differences have also been an issue with SQUID; in a recent large-scale study an approximately two-fold correction factor was required between biopsy and SQUID methodology [2,4].”
BackgroundMethodResultInfluentialSelf-citationJensen, P. D., Jensen, F. T., Christensen, T., Eiskjær, H., Baandrup, U., & Nielsen, J. L. (2003) · Blood · doi:10.1182/blood-2002-09-2754 · Cited by 174
Westwood, M. A., Anderson, L. J., Firmin, D. N., Gatehouse, P. D., Lorenz, C. H., Wonke, B., & Pennell, D. J. (2003) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.10396 · Cited by 166 · cited 1×
“Other sources of variability related to repeatability and reproducibility are unlikely to explain the poor agreement between both methods as reproducibility studies for T2* [26,31,32] and R2 LIC indicate (excellent single centre inter-scan [14], although not multi-centre with broad LoA −71 to 74% [33]).”
ResultSelf-citationTanner, M. A., He, T., Westwood, M. A., Firmin, D. N., & Pennell, D. J. (2006) · PubMed · Cited by 166
Self-citationKirk, P., He, T., Anderson, L. J., Roughton, M., Tanner, M. A., Lam, W. W. M., Au, W. Y., Chu, W. C. W., Chan, G., Galanello, R., et al. (2010) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.22245 · Cited by 159
Self-citationQuantitative Study of the Variability of Hepatic Iron Concentrations
Emond, M. J., Bronner, M. P., Carlson, T. H., Lin, M., Labbé, R. F., & Kowdley, K. V. (1999) · Clinical Chemistry · doi:10.1093/clinchem/45.3.340 · Cited by 137 · cited 1×
“The variability of needle biopsy LIC is dependent on the biopsy specimen size [27] (CV of approximately 19% when <4 mg/g dw [28] and 9% when >9 mg/g dw [29]), on fibrosis content, [27] and the presence of cirrhosis (CV >40%) [28,30].”
BackgroundAngelucci, E., BARONCIANI, D., Lucarelli, G., Baldassarri, M., GALIMBERTI, M., Giardini, C., Martinelli, F., Polchi, P., Polizzi, V., Ripalti, M., et al. (1995) · British Journal of Haematology · doi:10.1111/j.1365-2141.1995.tb08412.x · Cited by 133 · cited 1×
“The variability of needle biopsy LIC is dependent on the biopsy specimen size [27] (CV of approximately 19% when <4 mg/g dw [28] and 9% when >9 mg/g dw [29]), on fibrosis content, [27] and the presence of cirrhosis (CV >40%) [28,30].”
BackgroundPierre, T. G. S., El-Beshlawy, A. M., Elalfy, M. S., Al‐Jefri, A., Zir, K. A., Daar, S., Habr, D., Kriemler‐Krahn, U., & Taher, A. T. (2013) · Magnetic Resonance in Medicine · doi:10.1002/mrm.24854 · Cited by 129 · cited 1×
“Other sources of variability related to repeatability and reproducibility are unlikely to explain the poor agreement between both methods as reproducibility studies for T2* [26,31,32] and R2 LIC indicate (excellent single centre inter-scan [14], although not multi-centre with broad LoA −71 to 74% [33]).”
ResultGhugre, N. R., & Wood, J. C. (2010) · Magnetic Resonance in Medicine · doi:10.1002/mrm.22657 · Cited by 97 · cited 1×
“associated with different sensitivity to noise of both methods, to iron particle size, clumping, distribution and resultant proton mobility [35] within tissue is perhaps a more likely reason for lack of within-patient agreement.”
BackgroundChapman, R. W., Hussain, M. A. M., Gorman, A., Laulicht, M., Politis, D., Flynn, D. M., Sherlock, S., & Hoffbrand, A. V. (1982) · Journal of Clinical Pathology · doi:10.1136/jcp.35.5.487 · Cited by 87 · cited 1×
“Conversely, serum ferritin is disproportionately decreased relative to LIC in ascorbate deficiency [7] and in conditions where iron preferentially loads hepatocytes rather than macrophages, such as thalassemia intermedia [8].”
BackgroundBrittenham, G. M., Sheth, S., Allen, C. J., & Farrell, D. E. (2001) · Seminars in Hematology · doi:10.1016/s0037-1963(01)90059-9 · Cited by 77 · cited 1×
“A variety of non-invasive methods have been used to measure LIC including liver susceptometry (SQUID), [11,12], T2-weighted spin-echo MR with [13] or without signal intensity ratios (SIR) to adjacent tissues, [14] and T2* gradient-echo MR with [15] or without SIR, [16-18].”
MethodAydınok, Y., Ülger, Z., Nart, D., Terzi, A., Cetiner, N., Ellis, G., Zimmermann, A., & Manz, C. (2007) · Haematologica · doi:10.3324/haematol.11414 · Cited by 77
Measurement of iron in liver biopsies — a comparison of three analytical methods
Kreeftenberg, H. G., Koopman, B. J., Huizenga, J. R., van Vilsteren, T., Wolthers, B. G., & Gips, C. H. (1984) · Clinica Chimica Acta · doi:10.1016/0009-8981(84)90061-5 · Cited by 61 · cited 1×
“The variability of needle biopsy LIC is dependent on the biopsy specimen size [27] (CV of approximately 19% when <4 mg/g dw [28] and 9% when >9 mg/g dw [29]), on fibrosis content, [27] and the presence of cirrhosis (CV >40%) [28,30].”
BackgroundMonitoring Long‐Term Efficacy of Iron Chelation Treatment with Biomagnetic Liver Susceptometry
FISCHER, R., Piga, A. G., Harmatz, P., & NIELSEN, P. (2005) · Annals of the New York Academy of Sciences · doi:10.1196/annals.1345.043 · Cited by 60 · cited 1×
“A variety of non-invasive methods have been used to measure LIC including liver susceptometry (SQUID), [11,12], T2-weighted spin-echo MR with [13] or without signal intensity ratios (SIR) to adjacent tissues, [14] and T2* gradient-echo MR with [15] or without SIR, [16-18].”
MethodPorter, J. B., & Huehns, E. R. (2009) · Acta Haematologica · doi:10.1159/000205875 · Cited by 57 · cited 1×
“Serum ferritin may be disproportionately increased relative to LIC by hepatitis and liver damage, generalised inflammation, and by vaso-occlusive syndromes in sickle cell disorders [6].”
BackgroundSelf-citationButensky, E., Fischer, R., Hudes, M. L., Schumacher, L. B., Williams, R., Moyer, T. P., Vichinsky, E. P., & Harmatz, P. (2004) · American Journal of Clinical Pathology · doi:10.1309/puux-egxd-lh26-nxa2 · Cited by 56
R2 and R2* are equally effective in evaluating chronic response to iron chelation
Wood, J. C., Zhang, P., Rienhoff, H. Y., Abi‐Saab, W., & Neufeld, E. J. (2014) · American Journal of Hematology · doi:10.1002/ajh.23673 · Cited by 39
Cited by
Farmakis, D., Porter, J., Taher, A., Domenica Cappellini, M., Angastiniotis, M., Eleftheriou, A., & Taskforce, F. T. 2. T. G. (2022) · HemaSphere · doi:10.1097/hs9.0000000000000732 · Cited by 307
Self-citationLiver Iron Quantification with MR Imaging: A Primer for Radiologists
Labranche, R., Gilbert, G., Cerny, M., Vu, K.-N., Soulières, D., Olivié, D., Billiard, J.-S., Yokoo, T., & Tang, A. (2018) · RadioGraphics · doi:10.1148/rg.2018170079 · Cited by 171
“However, there has not been consensus on the recommended acquisition parameters for assessment of R2* (30,67,68).”
BackgroundIron imaging reveals tumor and metastasis macrophage hemosiderin deposits in breast cancer
Leftin, A., Ben-Chetrit, N., Klemm, F., Joyce, J. A., & Koutcher, J. A. (2017) · PLOS ONE · doi:10.1371/journal.pone.0184765 · Cited by 36
Leftin, A., Zhao, H., Turkekul, M., de Stanchina, E., Manova, K., & Koutcher, J. A. (2017) · Scientific Reports · doi:10.1038/s41598-017-11899-2 · Cited by 33
Metadata
- Authors
- Garbowski, M. W., Carpenter, J.-P., Smith, G., Roughton, M., Alam, M. H., He, T., Pennell, D. J., & Porter, J. B.
- Venue
- Journal of Cardiovascular Magnetic Resonance
- Publisher
- Elsevier BV
- Published
- 2014-12
- Volume
- 16
- Issue
- 1
- Pages
- 40
- ISSN
- 1097-6647 (ISSN-L)
- OpenAlex
- W2133662247
- Cited by
- 203
- References
- 31
- FWCI
- 10.20
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.