Management of radiation oncology patients with a pacemaker or ICD: A new comprehensive practical guideline in The Netherlands
Open accessThe DOI matches its Crossref record; 18 of the 18 references it cites were found in a registry or an index.
Abstract
Current clinical guidelines for the management of radiotherapy patients having either a pacemaker or implantable cardioverter defibrillator (both CIEDs: Cardiac Implantable Electronic Devices) do not cover modern radiotherapy techniques and do not take the patient's perspective into account. Available data on the frequency and cause of CIED failure during radiation therapy are limited and do not converge. The Dutch Society of Radiotherapy and Oncology (NVRO) initiated a multidisciplinary task group consisting of clinical physicists, cardiologists, radiation oncologists, pacemaker and ICD technologists to develop evidence based consensus guidelines for the management of CIED patients. CIED patients receiving radiotherapy should be categorised based on the chance of device failure and the clinical consequences in case of failure. Although there is no clear cut-off point nor a clear linear relationship, in general, chances of device failure increase with increasing doses. Clinical consequences of device failures like loss of pacing, carry the most risks in pacing dependent patients. Cumulative dose and pacing dependency have been combined to categorise patients into low, medium and high risk groups. Patients receiving a dose of less than 2 Gy to their CIED are categorised as low risk, unless pacing dependent since then they are medium risk. Between 2 and 10 Gy, all patients are categorised as medium risk, while above 10 Gy every patient is categorised as high risk. Measures to secure patient safety are described for each category. This guideline for the management of CIED patients receiving radiotherapy takes into account modern radiotherapy techniques, CIED technology, the patients' perspective and the practical aspects necessary for the safe management of these patients. The guideline is implemented in The Netherlands in 2012 and is expected to find clinical acceptance outside The Netherlands as well.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Wilkoff, B. L., Love, C. J., Byrd, C. L., Bongiorni, M. G., Carrillo, R. G., Crossley, G. H., Epstein, L. M., Friedman, R. A., Kennergren, C. E. H., Mitkowski, P., et al. (2009) · Heart Rhythm · doi:10.1016/j.hrthm.2009.05.020 · Cited by 1,024
Authors/Task Force Members, Vardas, P. E., Auricchio, A., Blanc, J.-J., Daubert, J.-C., Drexler, H., Ector, H., Gasparini, M., Linde, C., Morgado, F. B., et al. (2007) · European Heart Journal · doi:10.1093/eurheartj/ehm305 · Cited by 902
[Guidelines for cardiac pacing and cardiac resynchronization therapy].
Vardas, P. E., Auricchio, A., Blanc, J., Daubert, J., Drexler, H., Ector, H., Gasparini, M., Linde, C. M., Morgado, F. B., Oto, A., et al. (2008) · PubMed · Cited by 550
Crossley, G. H., Poole, J. E., Rozner, M. A., Asirvatham, S. J., Cheng, A., Chung, M. K., Ferguson, T. B., Gallagher, J. D., Gold, M. R., Hoyt, R. H., et al. (2011) · Heart Rhythm · doi:10.1016/j.hrthm.2010.12.023 · Cited by 340 · cited 5×
“11 1 [1] @ 0.”
BackgroundInfluentialCrossley, G. H., Poole, J. E., Rozner, M. A., Asirvatham, S. J., Cheng, A., Chung, M. K., Ferguson, T. B., Gallagher, J. D., Gold, M. R., Hoyt, R. H., et al. (2011) · Heart Rhythm · doi:10.1016/j.hrthm.2011.05.010 · Cited by 300
Gould, P. A. (2006) · JAMA · doi:10.1001/jama.295.16.1907 · Cited by 288 · cited 1×
“8% experienced a serious complication [50].”
Sears, S. F., Hauf, J. D., Kirian, K., Hazelton, G., & CONTI, J. B. (2011) · Circulation Arrhythmia and Electrophysiology · doi:10.1161/circep.110.957670 · Cited by 155 · cited 1×
“Such ICD shocks in patients are uncomfortable [24,45,46], although not lethal.”
BackgroundElectromagnetic interference in patients with implanted pacemakers or cardioverter-defibrillators
NIEHAUS, M., & Tebbenjohanns, J. (2001) · Heart · doi:10.1136/heart.86.3.246 · Cited by 115 · cited 1×
“; licensee BioMed Cent Commons Attribution License (http://creativec reproduction in any medium, provided the or patients receiving radiotherapy with a CIED [1-10] are not widely implemented [2,11].”
BackgroundHudson, F. R., Coulshed, D. S., Dsouza, E. I., & Baker, C. (2010) · Journal of Medical Imaging and Radiation Oncology · doi:10.1111/j.1754-9485.2010.02138.x · Cited by 88 · cited 2×
“Electromagnetic fields around modern linear accelerators have decreased, reducing the concern to patients with a CIED [4,9,24-27].”
BackgroundBorleffs, C. J. W., Thijssen, J., de Bie, M. K., van Rees, J. B., van Welsenes, G. H., van Erven, L., Bax, J. J., Cannegieter, S. C., & Schalij, M. J. (2010) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2010.02780.x · Cited by 82 · cited 1×
“Other studies have shown that the chance of infection increases with subsequent CIED replacement [51].”
BackgroundCardiac device infections are associated with a significant mortality risk
de Bie, M. K., van Rees, J. B., Thijssen, J., Borleffs, C. J. W., Trines, S. A. I. P., Cannegieter, S. C., Schalij, M. J., & van Erven, L. (2011) · Heart Rhythm · doi:10.1016/j.hrthm.2011.10.034 · Cited by 82 · cited 1×
“7%) [49].”
Hauser, R. G., Hayes, D. L., Epstein, A. E., Cannom, D. S., Vlay, S. C., SONG, S. L., & Tyers, G. F. O. (2006) · Heart Rhythm · doi:10.1016/j.hrthm.2006.02.011 · Cited by 74 · cited 1×
“In a mixed population, around 10% (range: 2-63%) of patients are so-called pacing dependent, meaning they have no intrinsic or escape rhythm and may become symptomatic (syncope, arrhythmia, serious injury or even sudden death) when the CIED pacing function fails [41-43].”
BackgroundImplanted Cardiac Defibrillator Care in Radiation Oncology Patient Population
Gelblum, D. Y., & Amols, H. I. (2008) · International Journal of Radiation Oncology*Biology*Physics · doi:10.1016/j.ijrobp.2008.06.1903 · Cited by 74 · cited 1×
“reported on 33 patients with an ICD, with dose to ICD ranging from 1 cGy to 300 cGy [37].”
MethodRadiotherapy to patients with artificial cardiac pacemakers
Sundar, S., Symonds, R. P., & Deehan, C. (2005) · Cancer Treatment Reviews · doi:10.1016/j.ctrv.2005.05.002 · Cited by 72 · cited 3×
“, [8] suggested a similar categorisation, making the distinction between pacing dependent and independent patients.”
BackgroundHashii, H., Hashimoto, T., Okawa, A., Shida, K., Isobe, T., Hanmura, M., Nishimura, T., Aonuma, K., Sakae, T., & Sakurai, H. (2012) · International Journal of Radiation Oncology*Biology*Physics · doi:10.1016/j.ijrobp.2012.05.043 · Cited by 55 · cited 2×
“Most probably, effects observed at very low dose levels are caused by neutrons evoking changes in the memory or the logical circuits of the CIED [32-35].”
BackgroundAn Updated Review of Implantable Cardioverter/Defibrillators, Induced Anxiety, and Quality of Life
Bostwick, J. M., & Sola, C. L. (2010) · Heart Failure Clinics · doi:10.1016/j.hfc.2010.10.003 · Cited by 54
Habibović, M., van den Broek, K. C., Alings, M., Van der Voort, P. H., & Denollet, J. (2012) · Health Psychology · doi:10.1037/a0024701 · Cited by 49 · cited 1×
“Such ICD shocks in patients are uncomfortable [24,45,46], although not lethal.”
BackgroundWadasadawala, T., Pandey, A., Agarwal, J. P., Jalali, R., Laskar, S. G., Chowdhary, S., Budrukkar, A., Sarin, R., Deshpande, D., & Munshi, A. (2010) · Clinical Oncology · doi:10.1016/j.clon.2010.08.031 · Cited by 49 · cited 3×
“1994 Souliman in vitro research 18 0 70 11 [18] @ 7,0 Gy 2 [18] @ 1,7 Gy 5 [18] @ 2,5 Gy 1) temporary change to interference or safety mode pacing lasting for the duration of the irradiation only [2] change to interference mode pacing—from which recovery may occur after reprogramming the pacemaker [3] severe damage”
Background
Cited by
Lyon, A. R., López-Fernández, T., Couch, L. S., Asteggiano, R., Aznar, M. C., Bergler-Klein, J., Boriani, G., Cardinale, D., Cordoba, R., Cosyns, B., et al. (2022) · European Heart Journal · doi:10.1093/eurheartj/ehac244 · Cited by 3,076
2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy
Glikson, M., Nielsen, J. C., Kronborg, M. B., Michowitz, Y., Auricchio, A., Barbash, I. M., Barrabés, J. A., Boriani, G., Braunschweig, F., Brignole, M., et al. (2021) · European Heart Journal · doi:10.1093/eurheartj/ehab364 · Cited by 2,019
2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy
Glikson, M., Nielsen, J. C., Kronborg, M. B., Michowitz, Y., Auricchio, A., Barbash, I. M., Barrabés, J. A., Boriani, G., Braunschweig, F., Brignole, M., et al. (2022) · EP Europace · doi:10.1093/europace/euab232 · Cited by 914
ATUALIZAÇÃO DA DIRETRIZ BRASILEIRA DE DISLIPIDEMIAS E PREVENÇÃO DA ATEROSCLEROSE - 2017
Faludi, A., Izar, M., Saraiva, J., Chacra, A., Bianco, H., Afiune Neto, A., Bertolami, A., Pereira, A., Lottenberg, A., Sposito, A., et al. (2017) · Arquivos Brasileiros de Cardiologia · doi:10.5935/abc.20170121 · Cited by 633
Lyon, A. R., López-Fernández, T., Couch, L. S., Asteggiano, R., Aznar, M. C., Bergler-Klein, J., Boriani, G., Cardinale, D., Cordoba, R., Cosyns, B., et al. (2022) · European Heart Journal - Cardiovascular Imaging · doi:10.1093/ehjci/jeac106 · Cited by 507
Metadata
- Authors
- Hurkmans, C. W., Knegjens, J. L., Oei, B. S., Maas, A. J., Uiterwaal, G., van der Borden, A. J., Ploegmakers, M. M., & van Erven, L.
- Venue
- Radiation Oncology
- Published
- 2012-11-24
- Volume
- 7
- Issue
- 1
- Pages
- 198-198
- ISSN
- 1748-717X (ISSN-L)
- OpenAlex
- W2113616938
- Cited by
- 175
- References
- 18
- FWCI
- 5.18
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.