Hemoperfusion: technical aspects and state of the art
Open accessThe DOI matches its Crossref record; 45 of the 45 references it cites were found in a registry or an index.
Abstract
Background: Blood purification through the removal of plasma solutes by adsorption to beads of charcoal or resins contained in a cartridge (hemoperfusion) has a long and imperfect history. Developments in production and coating technology, however, have recently increased the biocompatibility of sorbents and have spurred renewed interest in hemoperfusion. Methods: We performed a narrative assessment of the literature with focus on the technology, characteristics, and principles of hemoperfusion. We assessed publications in ex vivo, animal, and human studies. We synthesized such literature in a technical and state-of-the-art summary. Results: Early hemoperfusion studies were hampered by bioincompatibility. Recent technology, however, has improved its safety. Hemoperfusion has been used with positive effects in chronic dialysis and chronic liver disease. It has also demonstrated extraction of a variety of toxins and drugs during episodes of overdose. Trials with endotoxin binding polymyxin B have shown mixed results in septic shock and are under active investigation. The role of non-selective hemoperfusion in sepsis or inflammation remains. Although new technologies have made sorbents more biocompatible, the research agenda in the field remains vast. Conclusion: New sorbents markedly differ from those used in the past because of greater biocompatibility and safety. Initial studies of novel sorbent-based hemoperfusion show some promise in specific chronic conditions and some acute states. Systematic studies of novel sorbent-based hemoperfusion are now both necessary and justified.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Cruz, D. N., Antonelli, M., Fumagalli, R., Foltran, F., Brienza, N., Donati, A., Malcangi, V., Petrini, F., Volta, G., Bobbio Pallavicini, F. M., et al. (2009) · JAMA · doi:10.1001/jama.2009.856 · Cited by 832 · cited 1×
“It randomized 34 patients to polymyxin (PMX) B hemoperfusion and 30 to conventional care [53].”
BackgroundSelf-citationDellinger, R. P., Bagshaw, S. M., Antonelli, M., Foster, D. M., Klein, D. J., Marshall, J. C., Palevsky, P. M., Weisberg, L. S., Schorr, C. A., Trzeciak, S., et al. (2018) · JAMA · doi:10.1001/jama.2018.14618 · Cited by 502
The ABDOMIX Group, Payen, D. M., Guilhot, J., Launey, Y., Lukaszewicz, A. C., Kaaki, M., Veber, B., Pottecher, J., Joannes-Boyau, O., Martin-Lefevre, L., et al. (2015) · Intensive Care Medicine · doi:10.1007/s00134-015-3751-z · Cited by 407
Schädler, D., Pausch, C., Heise, D., Meier-Hellmann, A., Brederlau, J., Weiler, N., Marx, G., Putensen, C., Spies, C., Jörres, A., et al. (2017) · PLOS ONE · doi:10.1371/journal.pone.0187015 · Cited by 367 · cited 1×
“A multicenter randomized trial compared Cytosorb® therapy with conventional care [62] in 100 ventilated patients with sepsis or septic shock and either acute lung injury or ARDS.”
MethodRonco, C., Tetta, C., Mariano, F., Wratten, M. L., Bonello, M., Bordoni, V., Cardona, X., Inguaggiato, P., Pilotto, L., D’Intini, V., et al. (2003) · Artificial Organs · doi:10.1046/j.1525-1594.2003.07289.x · Cited by 356 · cited 1×
“Thus, any broad removal technique will remove more toxic than protective molecules [28].”
BackgroundSelf-citationEffectiveness of polymyxin B-immobilized fiber column in sepsis: a systematic review
Cruz, D. N., Perazella, M. A., Bellomo, R., de Cal, M., Polanco, N., Corradi, V., Lentini, P., Nalesso, F., Ueno, T., Ranieri, V. M., et al. (2007) · Critical Care · doi:10.1186/cc5780 · Cited by 347
Self-citationIn vitro comparison of the adsorption of inflammatory mediators by blood purification devices
Malard, B., Lambert, C., & Kellum, J. A. (2018) · Intensive Care Medicine Experimental · doi:10.1186/s40635-018-0177-2 · Cited by 273 · cited 1×
“Nonetheless, the concept of blood purification therapy is supported by multiple ex vivo studies [19–21] and experimental animal studies [22, 23].”
BackgroundKlein, D. J., Foster, D., Walker, P. M., Bagshaw, S. M., Mekonnen, H., & Antonelli, M. (2018) · Intensive Care Medicine · doi:10.1007/s00134-018-5463-7 · Cited by 249 · cited 1×
“However, a post hoc assessment of patients without extreme endotoxemia [56] found a survival advantage on time to event analysis.”
BackgroundSupady, A., Weber, E., Rieder, M., Lother, A., Niklaus, T., Zahn, T., Frech, F., Müller, S., Kuhl, M., Benk, C., et al. (2021) · The Lancet Respiratory Medicine · doi:10.1016/s2213-2600(21)00177-6 · Cited by 206 · cited 5×
“In this regard, we recommend that a hemoperfusion registry be set up to collect data as in ECMO registries.”
MethodInfluentialBrouwer, W. P., Duran, S., Kuijper, M., & Ince, C. (2019) · Critical Care · doi:10.1186/s13054-019-2588-1 · Cited by 191
Ankawi, G., Neri, M., Zhang, J., Breglia, A., Ricci, Z., & Ronco, C. (2018) · Critical Care · doi:10.1186/s13054-018-2181-z · Cited by 187 · cited 1×
“Diffusion, as in standard hemodialysis (HD), convection as in hemofiltration or their combination as in hemodiafiltration (HDF) [2].”
BackgroundSelf-citationMoriyama, K., & Nishida, O. (2021) · International Journal of Molecular Sciences · doi:10.3390/ijms22168882 · Cited by 174 · cited 1×
“, kidneys, liver, and immune system) malfunction, toxic solutes accumulate [18].”
BackgroundRoberts, J. A., Joynt, G. M., Lee, A., Choi, G., Bellomo, R., Kanji, S., Mudaliar, M. Y., Peake, S. L., Stephens, D., Taccone, F. S., et al. (2021) · Clinical Infectious Diseases · doi:10.1093/cid/ciaa224 · Cited by 151 · cited 1×
“However, while extensive clearance data exists for different forms of renal replacement therapy [29], the data for antibiotics and antifungal drugs removal during hemoperfusion is scant or absent.”
BackgroundSelf-citationAnkawi, G., Fan, W., Montin, D. P., Lorenzin, A., Neri, M., Caprara, C., de Cal, M., & Ronco, C. (2018) · Blood Purification · doi:10.1159/000493523 · Cited by 128 · cited 1×
“The hemoperfusion research agenda and recommendations New sorbent materials has now paved the way for a resurgence of research into the clinical application of hemoperfusion [74].”
BackgroundSelf-citationHuang, Z., Wang, S., Su, W., & Liu, J. (2010) · Therapeutic Apheresis and Dialysis · doi:10.1111/j.1744-9987.2010.00825.x · Cited by 97
Alharthy, A. M., Faqihi, F. A., Memish, Z. A., Balhamar, A., Nasim, N., Shahzad, A., Tamim, H., Alqahtani, S. A., Brindley, P. G., & Karakitsos, D. (2020) · Artificial Organs · doi:10.1111/aor.13864 · Cited by 85
Huang, Z., Wang, S., Yang, Z., & Liu, J. (2012) · Therapeutic Apheresis and Dialysis · doi:10.1111/j.1744-9987.2012.01083.x · Cited by 82
Stockmann, H., Thelen, P., Stroben, F., Pigorsch, M., Keller, T., Krannich, A., Spies, C., Treskatsch, S., Ocken, M., Kunz, J. V., et al. (2022) · Critical Care Medicine · doi:10.1097/ccm.0000000000005493 · Cited by 74 · cited 1×
“The second reported the effect of Cytosorb® therapy for 3 to 7 days in 50 COVID-19 patients with vasoplegic shock on time to resolution of shock [73].”
BackgroundHemoperfusion with CytoSorb as Adjuvant Therapy in Critically Ill Patients with SARS-CoV2 Pneumonia
Rampino, T., Gregorini, M., Perotti, L., Ferrari, F., Pattonieri, E. F., Grignano, M. A., Valente, M., Garrone, A., Islami, T., Libetta, C., et al. (2020) · Blood Purification · doi:10.1159/000511725 · Cited by 71
Scharf, C., Schroeder, I., Paal, M., Winkels, M., Irlbeck, M., Zöller, M., & Liebchen, U. (2021) · Annals of Intensive Care · doi:10.1186/s13613-021-00905-6 · Cited by 67
From Multiple Organ Support Therapy to Extracorporeal Organ Support in Critically Ill Patients
Ronco, C., Ricci, Z., & Husain‐Syed, F. (2019) · Blood Purification · doi:10.1159/000490694 · Cited by 66 · cited 1×
“Hemoperfusion: characteristics and principles Extracorporeal blood purification can be achieved by different mass separation processes [1].”
BackgroundSelf-citationKöhler, T., Schwier, E., Praxenthaler, J., Kirchner, C., Henzler, D., & Eickmeyer, C. (2021) · International Journal of Molecular Sciences · doi:10.3390/ijms222312786 · Cited by 60 · cited 1×
“Initial reports in the treatment of chronic dialysis patients and of chronic liver patients with pruritus have been encouraging [49, 75].”
BackgroundPaul, R. K., Sathe, P., Kumar, S., Prasad, S. M., Aleem, M. A., & Sakhalvalkar, P. (2021) · World Journal of Critical Care Medicine · doi:10.5492/wjccm.v10.i1.22 · Cited by 60
Therapeutic plasma exchange in liver failure
Chris-Olaiya, A., Kapoor, A., Ricci, K. S., & Lindenmeyer, C. C. (2021) · World Journal of Hepatology · doi:10.4254/wjh.v13.i8.904 · Cited by 59 · cited 1×
“These techniques remove a broad array of molecules that are believed to be toxic as in sepsis [16] or severe liver failure [17].”
BackgroundTreatment for Severe Coronavirus Disease 2019 With the Seraph-100 Microbind Affinity Blood Filter
Olson, S. W., Oliver, J. D., Collen, J. F., Bunin, J. L., Gleeson, T. D., Foster, B. E., Simmons, M. P., Chen, H. W., Ficke, J. B., Brown, T. E., et al. (2020) · Critical Care Explorations · doi:10.1097/cce.0000000000000180 · Cited by 51 · cited 1×
“This device contains adsorptive beads of ultrahigh molecular weight which, in vitro, remove the COVID-19 virus [70].”
BackgroundLi, W., Yin, Y.-M., Chen, H., Wang, X., Yun, H., Li, H., Luo, J., & Wang, J. (2017) · Medicine · doi:10.1097/md.0000000000006160 · Cited by 50
Li, X., Liu, C., Mao, Z., Qi, S., Song, R., & Zhou, F. (2020) · Journal of Critical Care · doi:10.1016/j.jcrc.2020.09.007 · Cited by 48 · cited 1×
“The concept of endotoxin adsorption has been based on several trials of the endotoxin-binding ability of polymyxin B as discussed in detail below [31].”
MethodCoupled Plasma Filtration Adsorption: A Multipurpose Extracorporeal Detoxification Therapy
Manna, G. L., & Donati, G. (2018) · Blood Purification · doi:10.1159/000490234 · Cited by 46
Schmidt, J. J., Borchina, D. N., Klooster, M. V. ́., Soki, K., Okioma, R., Herbst, L., Rodríguez, D., Premužić, V., Büttner, S., Bader, B. D., et al. (2021) · Nephrology Dialysis Transplantation · doi:10.1093/ndt/gfab347 · Cited by 43
Blood Purification in Non-Renal Critical Illness
Kellum, J. A., Bellomo, R., Mehta, R. L., et al. (2003) · Blood Purification · doi:10.1159/000067862 · Cited by 41 · cited 1×
“Hemoperfusion (direct hemoadsorption) (HP): Blood is circulated by a pump through a sorbent unit (cartridge) and enters in direct contact with the sorbent particles [38] (Fig.”
BackgroundSelf-citationSazonov, V., Abylkassov, R., Tobylbayeva, Z., Saparov, A., Mironova, O., & Poddighe, D. (2021) · Frontiers in Pediatrics · doi:10.3389/fped.2021.672260 · Cited by 40 · cited 1×
“Hemoperfusion with the JAFRON HA cartridge series (Jafron Biomedical, Guangdong, China) has been used in sepsis and reported in case series [64, 65].”
MethodThe effect of coupled haemofiltration and adsorption on inflammatory cytokines in an ex vivo model
Cole, L., Bellomo, R., Davenport, P., Tipping, P., Uchino, S., Tetta, C., & Ronco, C. (2002) · Nephrology Dialysis Transplantation · doi:10.1093/ndt/17.11.1950 · Cited by 38
Self-citationBlood purification with CytoSorb in critically ill COVID‐19 patients: A case series of 26 patients
Nassiri, A. A., Hakemi, M. S., Miri, M. M., Shahrami, R., Koomleh, A. A., & Sabaghian, T. (2021) · Artificial Organs · doi:10.1111/aor.14024 · Cited by 35
Boss, K., Jahn, M., Wendt, D., Haidari, Z., Demircioğlu, E., Thielmann, M., Ruhparwar, A., Kribben, A., & Tyczynski, B. (2021) · PLoS ONE · doi:10.1371/journal.pone.0246299 · Cited by 35 · cited 1×
“At the same time, clinical applications of sorbents in sepsis, acute kidney injury, and other inflammatory states have provided useful data on feasibility and safety, forming the basis for future technical, procedural, and manufacturing optimization [58, 76, 77].”
BackgroundClinical effects of hemoperfusion combined with pulse high-volume hemofiltration on septic shock
Chu, L., Li, G., Yu, Y., Bao, X., Wei, H., & Hu, M. (2020) · Medicine · doi:10.1097/md.0000000000019058 · Cited by 30
Therapeutic Plasma Exchange in the Critically Ill Patient: Technology and Indications
Sanchez, A. P., & Balogun, R. A. (2021) · Advances in Chronic Kidney Disease · doi:10.1053/j.ackd.2021.03.005 · Cited by 29
Sarti, E., Chenet, T., Stevanin, C., Costa, V., Cavazzini, A., Catani, M., Martucci, A., Precisvalle, N., Beltrami, G., & Pasti, L. (2020) · Molecules · doi:10.3390/molecules25153331 · Cited by 29 · cited 1×
“These factors may result in earlier saturation or prolonged efficiency of the hemoperfusion unit [11].”
BackgroundSurviving Sepsis Campaign: Research Opportunities for Infection and Blood Purification Therapies
Martín‐Loeches, I., Nunnally, M., Hellman, J., Lat, I., Martin, G. S., Jog, S., Kesecioğlu, J., Backer, D. D., & Coopersmith, C. M. (2021) · Critical Care Explorations · doi:10.1097/cce.0000000000000511 · Cited by 23 · cited 1×
“Selective vs non‐selective hemoperfusion in sepsis Blood purification in sepsis has been a key area of investigation because of the view that soluble mediators of injury are a major contributor to morbidity and mortality in septic patients [30].”
BackgroundTomescu, D. R., Popescu, M., David, C., & Dima, S. (2019) · The International Journal of Artificial Organs · doi:10.1177/0391398818823762 · Cited by 22 · cited 1×
“At the same time, clinical applications of sorbents in sepsis, acute kidney injury, and other inflammatory states have provided useful data on feasibility and safety, forming the basis for future technical, procedural, and manufacturing optimization [58, 76, 77].”
BackgroundKittanamongkolchai, W., Zoghby, Z., Hay, J., Wiesner, R. H., Kamath, P. S., LaRusso, N. F., Watt, K. D. S., Cramer, C. H., & Leung, N. (2016) · Hepatology International · doi:10.1007/s12072-016-9775-9 · Cited by 20 · cited 3×
“Liver disease There is very limited information or research in the use of hemoperfusion for severe liver failure (either acute or acute on chronic) even though there is a robust rationale for targeting ammonia or bilirubin in this setting [49].”
BackgroundEfficiency of Double Plasma Molecular Absorption System on the Acute Severe Cholestatic Hepatitis
Wu, M., Zhang, H., Huang, Y., Wu, W., Huang, J., & Yan, D. (2021) · Blood Purification · doi:10.1159/000513161 · Cited by 19 · cited 1×
“The nature of the sorbents and the characteristics of the cartridges depend on the indications and the degree of severity [41] Hemoperfusion in combination with ECMO: In patients undergoing veno-venous or veno-arterial extracorporeal membrane oxygenation (VV-ECMO or VA-ECMO) hemoperfusion may be connected to the ECMO…”
BackgroundAdsorbents: From Basic Structure to Clinical Application
Ronco, C., Bordoni, V., & Levin, N. (2002) · KARGER eBooks · doi:10.1159/000060242 · Cited by 17 · cited 1×
“In the last 50 years, however, organic polymer ion exchange resins and finally synthetic porous polymers (styrene or acrylic acid based) have been applied to blood purification (Table 1) [4].”
BackgroundSelf-citationRosa, S. D., Samoni, S., & Ronco, C. (2019) · Blood Purification · doi:10.1159/000505146 · Cited by 17
Self-citationRodeia, S., Martins, F., Fortuna, P., & Bento, L. F. N. (2021) · Blood Purification · doi:10.1159/000518712 · Cited by 13 · cited 1×
“However, the sorbent and pressure gradients should be adjusted to achieve adequate flows while avoiding perturbations of the main circuit [42] (Fig.”
BackgroundA case of hemoperfusion and l-carnitine management in valproic acid overdose
Jung, J.-H., Eo, E. K., & Ahn, K.-O. (2008) · The American Journal of Emergency Medicine · doi:10.1016/j.ajem.2007.07.032
Cited by
Zarbock, A., Nadim, M. K., Pickkers, P., Gomez, H., Bell, S., Joannidis, M., Kashani, K., Koyner, J. L., Pannu, N., Meersch, M., et al. (2023) · Nature Reviews Nephrology · doi:10.1038/s41581-023-00683-3 · Cited by 767
Self-citation
Metadata
- Authors
- Ronco, C., & Bellomo, R.
- Venue
- Critical Care
- Published
- 2022-05-12
- Volume
- 26
- Issue
- 1
- Pages
- 135-135
- ISSN
- 1364-8535 (ISSN-L)
- OpenAlex
- W4280555470
- Cited by
- 188
- References
- 45
- FWCI
- 19.30
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.