Noise reduction using a Bayesian penalized-likelihood reconstruction algorithm on a time-of-flight PET-CT scanner

Open access
Caribé, Paulo R. R. V., Koole, M., D’Asseler, Yves, Van Den Broeck, B., Vandenberghe, S.·2019·EJNMMI Physics

Abstract

Purpose: Q.Clear is a block sequential regularized expectation maximization (BSREM) penalized-likelihood reconstruction algorithm for PET. It tries to improve image quality by controlling noise amplification during image reconstruction. In this study, the noise properties of this BSREM were compared to the ordered-subset expectation maximization (OSEM) algorithm for both phantom and patient data acquired on a state-of-the-art PET/CT. Methods: The NEMA IQ phantom and a whole-body patient study were acquired on a GE DMI 3-rings system in list mode and different datasets with varying noise levels were generated. Phantom data was evaluated using four different contrast ratios. These were reconstructed using BSREM with different β-factors of 300–3000 and with a clinical setting used for OSEM including point spread function (PSF) and time-of-flight (TOF) information. Contrast recovery (CR), background noise levels (coefficient of variation, COV), and contrast-to-noise ratio (CNR) were used to determine the performance in the phantom data. Findings based on the phantom data were compared with clinical data. For the patient study, the SUV ratio, metabolic active tumor volumes (MATVs), and the signal-to-noise ratio (SNR) were evaluated using the liver as the background region. Results: Based on the phantom data for the same count statistics, BSREM resulted in higher CR and CNR and lower COV than OSEM. The CR of OSEM matches to the CR of BSREM with β = 750 at high count statistics for 8:1. A similar trend was observed for the ratios 6:1 and 4:1. A dependence on sphere size, counting statistics, and contrast ratio was confirmed by the CNR of the ratio 2:1. BSREM with β = 750 for 2.5 and 1.0 min acquisition has comparable COV to the 10 and 5.0 min acquisitions using OSEM. This resulted in a noise reduction by a factor of 2–4 when using BSREM instead of OSEM. For the patient data, a similar trend was observed, and SNR was reduced by at least a factor of 2 while preserving contrast. Conclusion: The BSREM reconstruction algorithm allowed a noise reduction without a loss of contrast by a factor of 2–4 compared to OSEM reconstructions for all data evaluated. This reduction can be used to lower the injected dose or shorten the acquisition time.

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1186/s40658-019-0264-9

Cites

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • PL reconstruction has the potential to systematically improve SNR compared to OSEM-based algorithms.

    “Based on the potential of PL reconstruction to systematically improve SNR compared to OSEM-based algorithms [108,113,168]”

    From Influences on PET Quantification and Interpretation (2022) · cites this paper as [168]

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

The citation lists haven’t been fetched yet

References 30 · Cited by 51

Research works printed but not linked: 24

  1. [1]Muehllehner G Karp JS Positron emission tomography Phys Med Biol 2006 51 R117 10.1088/0031-9155/51/13/R08 16790899 · DOI
  2. [2]Vandenberghe S Mikhaylova E D’Hoe E Mollet P Karp JS Recent developments in time-of-flight PET EJNMMI Physics. 2016 3 1 3 10.1186/s40658-016-0138-3 26879863 PMC4754240 · DOI
  3. [3]Alessio AM Stearns CW Tong S Application and evaluation of a measured spatially variant system model for PET image reconstruction IEEE Trans Med Imaging. 2010 29 3 938 949 10.1109/TMI.2010.2040188 20199927 PMC2903538 · DOI
  4. [4]Hudson HM Larkin RS Accelerated image reconstruction using ordered subsets of projection data IEEE Transactions on Medical Imaging 1994 13 601 609 10.1109/42.363108 18218538 · DOI
  5. [5]Vandenberghe S D’Asseler Y van de Walle R Iterative reconstruction algorithms in nuclear medicine Comput Med Imaging Graph 2001 25 105 111 10.1016/S0895-6111(00)00060-4 11137786 · DOI
  6. [6]Armstrong IS Kelly MD Williams HA Matthews JC Impact of point spread function modelling and time of flight on FDG uptake measurements in lung lesions using alternative filtering strategies EJNMMI Phys. 2014 1 99 10.1186/s40658-014-0099-3 26501457 PMC4545221 · DOI
  7. [7]Prieto E Domínguez-Prado I García-Velloso MJ Peñuelas I Richter JÁ Martí-Climent JM Impact of time-of-flight and point-spread-function in SUV quantification for oncological PET Clin Nucl Med. 2013 38 2 103 109 10.1097/RLU.0b013e318279b9df 23334123 · DOI
  8. [8]Gaitanis A Kontaxakis G Spyrou G Panayiotakis G Tzanakos G PET image reconstruction: a stopping rule for the MLEM algorithm based on properties of the updating coefficients Comput Med Imaging Graph 2010 34 131 141 10.1016/j.compmedimag.2009.07.006 19744826 · DOI
  9. [9]Wang G Qi J Penalized likelihood PET image reconstruction using patch-based edge-preserving regularization IEEE Trans Med Imaging. 2012 31 12 2194 2204 10.1109/TMI.2012.2211378 22875244 PMC4080915 · DOI
  10. [10]Tsoumpas C Polycarpou I Thielemans K Buerger C King AP Schaeffter T Marsden PK The effect of regularization in motion compensated PET image reconstruction: a realistic numerical 4D simulation study Phys Med Biol 2013 58 1759 10.1088/0031-9155/58/6/1759 23442264 · DOI
  11. [11]Ahn S Fessler JA Globally convergent image reconstruction for emission tomography using relaxed ordered subsets algorithms IEEE Trans Med Imaging 2003 22 613 626 10.1109/TMI.2003.812251 12846430 · DOI
  12. [12]Nuyts J Bequé D Dupont P Mortelmans L A concave prior penalizing relative differences for maximum-a-posteriori reconstruction in emission tomography IEEE Trans Nucl Sci. 2002 49 56 60 10.1109/TNS.2002.998681 · DOI
  13. [13]Teoh EJ McGowan DR Bradley KM Belcher E Black E Gleeson FV Novel penalised likelihood reconstruction of PET in the assessment of histologically verified small pulmonary nodules Eur Radiol 2016 26 576 10.1007/s00330-015-3832-y 25991490 PMC4551414 · DOI · Paper page
  14. [14]Sangtae A Steven GR Evren A Jun M Xiao J Lishui C Scott DW Ravindra MM Quantitative comparison of OSEM and penalized likelihood image reconstruction using relative difference penalties for clinical PET Phys Med Biol 2015 60 5733 10.1088/0031-9155/60/15/5733 26158503 · DOI
  15. [15]Parvizi N Franklin JM McGowan DR Teoh EJ Bradley KM Gleeson FV Does a novel penalized likelihood reconstruction of 18F-FDG PET-CT improve signal-to-background in colo-rectal liver metastases? Eur J Radiology 2015 84 1873 1878 10.1016/j.ejrad.2015.06.025 26163992 · DOI
  16. [19]Caribé P Koole M Bertin H D’Asseler Y Vandenberghe S NEMA NU 2012 performance evaluation of silicon-photomultiplier–based and conventional PMT-based time-of-flight systems EJNMMI Physics 2018 5 Suppl 1, A29 19 10.1186/s40658-018-0218-7 29956037 PMC6023806 · DOI
  17. [21]Boellaard R FDG PET/CT: EANM procedure guidelines for tumour imaging: version 2.0 Eur J Nucl Med Mol Imaging 2015 42 328 354 10.1007/s00259-014-2961-x 25452219 PMC4315529 · DOI · Paper page
  18. [22]van der Vos CS Koopman D Rijnsdorp S Quantification, improvement, and harmonization of small lesion detection with state-of-the-art PET Eur J Nucl Med Mol Imaging. 2017 44 4 16 10.1007/s00259-017-3727-z 28687866 PMC5541089 · DOI
  19. [23]Nguyen NC Vercher-Conejero JL Sattar A Image quality and diagnostic performance of a digital PET prototype in patients with oncologic diseases: initial experience and comparison with analog PET J Nucl Med. 2015 56 1378 1385 10.2967/jnumed.114.148338 26159588 · DOI
  20. [25]Rahmim A Qi J Sossi V Resolution modeling in PET imaging: theory, practice, benefits, and pitfalls Med Phys. 2013 40 6 064301 10.1118/1.4800806 23718620 PMC3663852 · DOI
  21. [26]Spasic E Jehanno N Bblondeel-Gomes S Huchet V Luporsi M Mounat TC Phantom and clinical evaluation for new PET/CT reconstruction algorithm: Bayesian penalized likelihood reconstruction algorithm Q.Clear J Nucl Med Radiat Ther 2018 9 4 371 10.4172/2155-9619.1000371 · DOI
  22. [27]Messerli M Stolzmann P Egger-Sigg Impact of a Bayesian penalized likelihood reconstruction algorithm on image quality in novel digital PET/CT: clinical implications for the assessment of lung tumors EJNMMI Physics 2018 5 27 10.1186/s40658-018-0223-x 30255439 PMC6156690 · DOI · Paper page
  23. [28]Messerli M Kotasidis F Burger IA Ferraro DA Muehlematter UJ Weyermann C Impact of different image reconstructions on PET quantification in non-small cell lung cancer: a comparison of adenocarcinoma and squamous cell carcinoma Br J Radiol 2019 92 20180792 10.1259/bjr.20180792 30673302 PMC6540860 · DOI
  24. [29]Caribé PRRV Koole M D’Asseler Y NEMA NU 2–2007 performance characteristics of GE Signa integrated PET/MR for different PET isotopes EJNMMI Phys 2019 6 11 10.1186/s40658-019-0247-x 31273558 PMC6609673 · DOI

Authors
Caribé, P. R. R. V., Koole, M., D’Asseler, Y., Van Den Broeck, B., & Vandenberghe, S.
Affiliations
  • Universitair Ziekenhuis Leuven (BE)
  • Ghent University Hospital (BE)
  • Ghent University (BE)
Venue
EJNMMI Physics
Published
2019-12
Volume
6
Issue
1
Pages
22-22
ISSN
2197-7364 (ISSN-L)
OpenAlex
W2995364088
Cited by
51
References
30
Citations by year
2026: 3 · 2025: 8 · 2024: 11 · 2023: 9 · 2022: 5 · 2021: 9 · 2020: 5 · 2018: 1
FWCI
3.14
Type
journal article
Language
en
Open access
Locations
  • EJNMMI Physics (publishedVersion, cc-by)
  • Europe PMC (PubMed Central) (publishedVersion, cc-by)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • Ghent University Academic Bibliography (Ghent University) (submittedVersion, cc-by)
  • Lirias (KU Leuven) (submittedVersion, other-oa)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
Funders
Conselho Nacional de Desenvolvimento Científico e Tecnológico (process number 235040/2014-2)
Topics
  • Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Medical Imaging Techniques and Applications
  • Physical Sciences › Computer Science › Computer Vision and Pattern Recognition › Image and Signal Denoising Methods
  • Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Radiation Dose and Imaging
Keywords
PET image reconstruction, noise reduction, Bayesian penalized likelihood reconstruction, Q.Clear, BSREM, OSEM, time-of-flight PET, PET/CT, contrast recovery, coefficient of variation, contrast-to-noise ratio, point spread function, acquisition time reduction

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review