Quantitative Digital Radiography Versus Dual Photon Absorptiometry of the Lumbar Spine*
Abstract
Lumbar spine bone mineral density (BMD) was measured by quantitative digital radiography, a new dual energy x-ray technique, and by 153Gd dual photon absorptiometry (DPA) in 85 patients. Each patient was measured twice by the new method and once by DPA on the same day, with repositioning between measurements. Serial measurements were made on an hydroxyapatite spine phantom embedded in tissue-equivalent plastic to evaluate the long term reproducibility of each instrument. The spinal BMD measurements with the 2 techniques were linearly related and highly correlated (r = 0.98) over a range from severely osteopenic to high normal. This correlation was not affected by the age, weight, or BMD of the patient measured. Quantitative digital radiography's long-term reproducibility using the spine phantom was stable for 180 days (coefficient of variation, 0.23%); DPA values were 3 times as variable for 170 days (coefficient of variation 0.73%) and increased 1.0% (P less than 0.0001) after a software change. The short term reproducibility of quantitative digital radiography, estimated from paired patient measurements, was 2-fold better than reported values for DPA and was independent of the patient's age, weight, or BMD. Measurement time by quantitative digital radiography was 5-8 min, with a maximum radiation exposure of 3 mrem, significantly lower than the corresponding DPA values. Quantitative digital radiography's image resolution was superior to that of DPA, enabling it to measure more bones. These advantages along with the elimination of 153Gd source changes and Nuclear Regulatory Commission licensing requirements indicate that quantitative digital radiography is the superior method for spinal BMD measurements.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
New approaches for interpreting projected bone densitometry data
Professor Carter, D. R., Bouxsein, M. L., & Marcus, R. (1992) · Journal of Bone and Mineral Research · doi:10.1002/jbmr.5650070204 · Cited by 1,065
GLASTRE, C., BRAILLON, P., DAVID, L., COCHAT, P., MEUNIER, P. J., & DELMAS, P. D. (1990) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jcem-70-5-1330 · Cited by 590
Heymsfield, S. B., Gonzalez, M. C., Lu, J., Jia, G., & Zheng, J. (2015) · Proceedings of the Nutrition Society · doi:10.1017/s0029665115000129 · Cited by 483
Postmenopausal hormone replacement therapy prevents central distribution of body fat after menopause
Haarbo, J., Marslew, U., Gotfredsen, A., & Christiansen, C. (1991) · Metabolism · doi:10.1016/0026-0495(91)90037-w · Cited by 441
Validation of body composition by dual energy X‐ray absorptiometry (DEXA)
Haarbo, J., Gotfredsen, A., Hassager, C., & Christiansen, C. (1991) · Clinical Physiology · doi:10.1111/j.1475-097x.1991.tb00662.x · Cited by 419
Metadata
- Authors
- KELLY, T. L., SLOVIK, D. M., SCHOENFELD, D. A., & NEER, R. M.
- Venue
- The Journal of Clinical Endocrinology & Metabolism
- Publisher
- The Endocrine Society
- Published
- 1988-10
- Volume
- 67
- Issue
- 4
- Pages
- 839-844
- ISSN
- 0021-972X (ISSN-L)
- 1945-7197
- OpenAlex
- W2001374121
- Cited by
- 214
- References
- 0
- FWCI
- 22.68
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Quantitative Digital Radiography Versus Dual Photon Absorptiometry of the Lumbar Spine*. https://app.cochat.ai/papers/10.1210/jcem-67-4-839
- The paper
- KELLY, T. L., SLOVIK, D. M., SCHOENFELD, D. A., & NEER, R. M. (1988). Quantitative Digital Radiography Versus Dual Photon Absorptiometry of the Lumbar Spine*. The Journal of Clinical Endocrinology & Metabolism. https://doi.org/10.1210/jcem-67-4-839
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.