Body composition analysis by bioelectrical impedance: effect of skin temperature
The DOI matches its Crossref record.
Abstract
Bioelectrical impedance analysis (BIA) was used to estimate body water and composition under both cool (14.4°C, dry bulb) and warm (35.0°C) ambient conditions in eight healthy adult men. The prediction equation provided with the commercially available instrument (RJL Systems) was used with the BIA measurements to estimate body composition. Skin temperature increased from 24.1 ± 1.81°C in the cool condition to 33.4 ± 1.36°C in the warm condition. (Mean increase was 9.3 ± 1.75°C, t = 15.05, P < 0.01.) The corresponding BIA resistances were 461 ± 48 Ω and 426 ± 47Ω, respectively. (Mean reduction was 35.0 ± 9.8 Ω, t = 10.13, P < 0.01.) This resulted in a significant increase in predicted total body water (cool 47.4 ± 5.51 vs. warm 49.9 ± 5.61, t = 3.88, P < 0.01). Consequently, predicted fat mass was significantly lower in the warm than in the cool condition (8.8 ± 3.2 kg vs. 11.0 ± 3.7 kg; mean difference 2.23 ± 0.69 kg, t = 9.22, P < 0.01). These findings indicate that varying skin temperature by altering ambient temperature significantly changes resistance measurements and the estimation of total body water and percent fat by BIA. The observed changes in resistance are consistent with an apparent expansion of conductor volume in the warm environment and a reduction in the cooler condition. In this regard, the temperature-induced change in resistance could be due to alterations in cutaneous blood flow and/or compartmental distribution of body water. Thus, BIA measurements should be taken only under well-standardized ambient conditions.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Bioelectrical Impedance Analysis: A Review of Principles and Applications
Kushner, R. F. (1992) · Journal of the American College of Nutrition · doi:10.1080/07315724.1992.12098245 · Cited by 625
Is bioelectrical impedance accurate for use in large epidemiological studies?
Dehghan, M., & Merchant, A. T. (2008) · Nutrition Journal · doi:10.1186/1475-2891-7-26 · Cited by 505
Why bioelectrical impedance analysis should be used for estimating adiposity.
Houtkooper, L., Lohman, T., Going, S., & Howell, W. (1996) · The American Journal of Clinical Nutrition · doi:10.1093/ajcn/64.3.436s · Cited by 361
Clinical characteristics influencing bioelectrical impedance analysis measurements
Kushner, R., Gudivaka, R., & Schoeller, D. (1996) · The American Journal of Clinical Nutrition · doi:10.1093/ajcn/64.3.423s · Cited by 338
Sergi, G., De Rui, M., Stubbs, B., Veronese, N., & Manzato, E. (2017) · Aging Clinical and Experimental Research · doi:10.1007/s40520-016-0622-6 · Cited by 291
“[39] observed a mean increase in resistance value of 8 % after a temperature reduction from 35 to 14 C, and Garby et al.”
Metadata
- Authors
- CATON, J. R., MOL??, P. A., ADAMS, W. C., & HEUSTIS, D. S.
- Venue
- Medicine & Science in Sports & Exercise
- Published
- 1988-10
- Volume
- 20
- Issue
- 5
- Pages
- 489???491
- ISSN
- 0195-9131 (ISSN-L)
- OpenAlex
- W4322703285
- Cited by
- 65
- References
- 0
- FWCI
- 0.56
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Body composition analysis by bioelectrical impedance: effect of skin temperature. https://app.cochat.ai/papers/10.1249/00005768-198810000-00010
- The paper
- CATON, J. R., MOL??, P. A., ADAMS, W. C., & HEUSTIS, D. S. (1988). Body composition analysis by bioelectrical impedance: effect of skin temperature. Medicine & Science in Sports & Exercise. https://doi.org/10.1249/00005768-198810000-00010
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.