Trabecular Bone Tissue Strains in the Healthy and Osteoporotic Human Femur
The DOI matches its Crossref record; 18 of the 18 references it cites were found in a registry or an index.
Abstract
Quantitative information about bone tissue‐level loading is essential for understanding bone mechanical behavior. We made microfinite element models of a healthy and osteoporotic human femur and found that tissue‐level strains in the osteoporotic femoral head were 70% higher on average and less uniformly distributed than those in the healthy one. Introduction: Bone tissue stresses and strains in healthy load‐adapted trabecular architectures should be distributed rather evenly, because no bone tissue is expected to be overloaded or unused. In this study, we evaluate this paradigm with the use of microfinite element (μFE) analyses to calculate tissue‐level stresses and strains for the human femur. Our objectives were to quantify the strain distribution in the healthy femur, to investigate to what extent this distribution is affected by osteoporosis, to determine if osteoporotic bone is simply bone adapted to lower load levels, and to determine the “safety factor” for trabecular bone. Materials and Methods: μFE models of a healthy and osteoporotic proximal femur were made from microcomputed tomography images. The models consisted of over 96 and 71 million elements for the healthy and osteoporotic femur, respectively, and represented their internal and external morphology in detail. Stresses and strains were calculated for each element and their distributions were calculated for a volume of interest (VOI) of trabecular bone in the femoral head. Results: The average tissue‐level principal strain magnitude in the healthy VOI was 304 ± 185 microstrains and that in the osteoporotic VOI was 520 ± 355 microstrains. Calculated safety factors were 8.6 for the healthy and 4.9 for the osteoporotic femurs. After reducing the force applied to the osteoporotic model to 59%, the average strain compared with that of the healthy femur, but the SD was larger (208 microstrains). Conclusions: Strain magnitudes in the osteoporotic bone were much higher and less uniformly distributed than those in the healthy one. After simulated joint‐load reduction, strain magnitudes in the osteoporotic femur were very similar to those in the healthy one, but their distribution is still wider and thus less favorable.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Hip contact forces and gait patterns from routine activities
Bergmann, G., Deuretzbacher, G., Heller, M., Graichen, F., Rohlmann, A., Strauss, J., & Duda, G. N. (2001) · Journal of Biomechanics · doi:10.1016/s0021-9290(01)00040-9 · Cited by 2,215
Das Gesetz der Transformation der Knochen
Wolff, J. (1893) · DMW - Deutsche Medizinische Wochenschrift · doi:10.1055/s-0028-1144106 · Cited by 1,999
Young's modulus of trabecular and cortical bone material: Ultrasonic and microtensile measurements
Rho, J. Y., Ashman, R. B., & Turner, C. H. (1993) · Journal of Biomechanics · doi:10.1016/0021-9290(93)90042-d · Cited by 1,283 · cited 1×
“5 GPa for cortical and 15 GPa for trabecular bone tissue.((9,10)) A physiological distributed load was applied normal to the femoral head surfaces.”
Backgroundvan Rietbergen, B., Weinans, H., Huiskes, R., & Odgaard, A. (1995) · Journal of Biomechanics · doi:10.1016/0021-9290(95)80008-5 · Cited by 865
Self-citationPistoia, W., van Rietbergen, B., Lochmüller, E.-M., Lill, C. A., Eckstein, F., & Rüegsegger, P. (2002) · Bone · doi:10.1016/s8756-3282(02)00736-6 · Cited by 643
Self-citationTurner, C. H., Rho, J., Takano, Y., Tsui, T. Y., & Pharr, G. M. (1999) · Journal of Biomechanics · doi:10.1016/s0021-9290(98)00177-8 · Cited by 547
Niebur, G. L., Feldstein, M. J., Yuen, J. C., Chen, T. J., & Keaveny, T. M. (2000) · Journal of Biomechanics · doi:10.1016/s0021-9290(00)00149-4 · Cited by 431
Hollister, S. J., Brennan, J. M., & Kikuchi, N. (1994) · Journal of Biomechanics · doi:10.1016/0021-9290(94)90019-1 · Cited by 322
If bone is the answer, then what is the question?
HUISKES, R. (2000) · Journal of Anatomy · doi:10.1046/j.1469-7580.2000.19720145.x · Cited by 296
Self-citationFinite element analysis of trabecular bone structure: a comparison of image-based meshing techniques
Ulrich, D., van Rietbergen, B., Weinans, H., & Rüegsegger, P. (1998) · Journal of Biomechanics · doi:10.1016/s0021-9290(98)00118-3 · Cited by 279
Self-citation
Cited by
Recent advances in X-ray microtomography applied to materials
Stock, S. R. (2008) · International Materials Reviews · doi:10.1179/174328008x277803 · Cited by 517
MacNeil, J. A., & Boyd, S. K. (2008) · Bone · doi:10.1016/j.bone.2008.01.017 · Cited by 440
Relation between age, femoral neck cortical stability, and hip fracture risk
Mayhew, P. M., Thomas, C. D., Clement, J. G., Loveridge, N., Beck, T. J., Bonfield, W., Burgoyne, C. J., & Reeve, J. (2005) · The Lancet · doi:10.1016/s0140-6736(05)66870-5 · Cited by 397
A theoretical framework for strain-related trabecular bone maintenance and adaptation
Ruimerman, R., Hilbers, P., van Rietbergen, B., & Huiskes, R. (2005) · Journal of Biomechanics · doi:10.1016/j.jbiomech.2004.03.037 · Cited by 300
Self-citationIf bone is the answer, then what is the question?
HUISKES, R. (2000) · Journal of Anatomy · doi:10.1046/j.1469-7580.2000.19720145.x · Cited by 296
Self-citationHomminga, J., Van-Rietbergen, B., Lochmüller, E. M., Weinans, H., Eckstein, F., & Huiskes, R. (2004) · Bone · doi:10.1016/j.bone.2003.12.001 · Cited by 227
Self-citationExtending DXA beyond bone mineral density: Understanding hip structure analysis
Beck, T. J. (2007) · Current Osteoporosis Reports · doi:10.1007/s11914-007-0002-4 · Cited by 215
“van Rietbergen et al. [ 6 ] have demonstrated that it is possible to create realistic FE models of the proximal end of the femur, but the complexity requires tens of millions of elements and hundreds of hours on a supercomputer to process a single loading condition [7].”
BackgroundAn Improved Segmentation Method for In Vivo μCT Imaging
Waarsing, J. H., Day, J. S., & Weinans, H. (2004) · Journal of Bone and Mineral Research · doi:10.1359/jbmr.040705 · Cited by 212
“In addition, CT provides the possibility to set up finite element (FE) models to determine the strength and stiffness of the bone sample based solely on the trabecular architecture.((3,4)) Furthermore, the 3D representations of bone can be used as the input for computer simulations of bone remodeling.”
BackgroundBlack, D. M., Bouxsein, M. L., Marshall, L. M., Cummings, S. R., Lang, T. F., Cauley, J. A., Ensrud, K. E., Nielson, C. M., & Orwoll, E. S. (2008) · Journal of Bone and Mineral Research · doi:10.1359/jbmr.080316 · Cited by 209
“A variety of structural parameters have been identified as potential predictors of hip fracture, including bone width,((11,20)) cortical thickness, hip axis length,((10,24)) neck shaft angle,((21)) and trabecular density and structure.((25)) In addition, some recent studies have suggested that it may be important to a…”
BackgroundQuantitative MRI for the assessment of bone structure and function
Wehrli, F. W., Song, H. K., Saha, P. K., & Wright, A. C. (2006) · NMR in Biomedicine · doi:10.1002/nbm.1066 · Cited by 207
“Most work reported is based on high-resolution m-CT (162,163) or m-MRI of TB specimens (144), including intact bone (158).”
BackgroundTrabecular bone ontogeny in the human proximal femur
Ryan, T. M., & Krovitz, G. E. (2006) · Journal of Human Evolution · doi:10.1016/j.jhevol.2006.06.004 · Cited by 193
Finite Element Method (FEM), Mechanobiology and Biomimetic Scaffolds in Bone Tissue Engineering
Boccaccio, A., Ballini, A., Pappalettere, C., Tullo, D., Cantore, S., & Desiate, A. (2011) · International Journal of Biological Sciences · doi:10.7150/ijbs.7.112 · Cited by 178
Load distribution in the healthy and osteoporotic human proximal femur during a fall to the side
Verhulp, E., van Rietbergen, B., & Huiskes, R. (2008) · Bone · doi:10.1016/j.bone.2007.08.039 · Cited by 174
“Based on the results of these and earlier [3] analyses, we conclude that the osteoporotic femur analyzed is not ‘overadapted’.”
InfluentialSelf-citationBone Strength: The Whole Is Greater Than the Sum of Its Parts
Davison, K. S., Siminoski, K., Adachi, J. D., Hanley, D. A., Goltzman, D., Hodsman, A. B., Josse, R., Kaiser, S., Olszynski, W. P., Papaioannou, A., et al. (2006) · Seminars in Arthritis and Rheumatism · doi:10.1016/j.semarthrit.2006.04.002 · Cited by 171
Whole Bone Mechanics and Bone Quality
Cole, J. H., & van der Meulen, M. C. H. (2011) · Clinical Orthopaedics & Related Research · doi:10.1007/s11999-011-1784-3 · Cited by 169
“effect of architecture on stiffness of whole bones such as the vertebral body, distal radius, or proximal femur [42, 57, 98, 126].”
Ural, A., & Vashishth, D. (2007) · Journal of Biomechanical Engineering · doi:10.1115/1.2768377 · Cited by 62
Metadata
- Authors
- Van Rietbergen, B., Huiskes, R., Eckstein, F., & Rüegsegger, P.
- Venue
- Journal of Bone and Mineral Research
- Publisher
- Oxford University Press (OUP)
- Published
- 2003-10
- Volume
- 18
- Issue
- 10
- Pages
- 1781-1788
- ISSN
- 0884-0431 (ISSN-L)
- 1523-4681
- OpenAlex
- W2144424358
- Cited by
- 256
- References
- 18
- FWCI
- 12.73
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.