Measurement and Correction of Microscopic Head Motion during Magnetic Resonance Imaging of the Brain
Open accessThe DOI matches its Crossref record; 17 of the 17 references it cites were found in a registry or an index.
Abstract
Magnetic resonance imaging (MRI) is a widely used method for non-invasive study of the structure and function of the human brain. Increasing magnetic field strengths enable higher resolution imaging; however, long scan times and high motion sensitivity mean that image quality is often limited by the involuntary motion of the subject. Prospective motion correction is a technique that addresses this problem by tracking head motion and continuously updating the imaging pulse sequence, locking the imaging volume position and orientation relative to the moving brain. The accuracy and precision of current MR-compatible tracking systems and navigator methods allows the quantification and correction of large-scale motion, but not the correction of very small involuntary movements in six degrees of freedom. In this work, we present an MR-compatible tracking system comprising a single camera and a single 15 mm marker that provides tracking precision in the order of 10 m and 0.01 degrees. We show preliminary results, which indicate that when used for prospective motion correction, the system enables improvement in image quality at both 3 T and 7 T, even in experienced and cooperative subjects trained to remain motionless during imaging. We also report direct observation and quantification of the mechanical ballistocardiogram (BCG) during simultaneous MR imaging. This is particularly apparent in the head-feet direction, with a peak-to-peak displacement of 140 m.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
RARE imaging: A fast imaging method for clinical MR
Hennig, J., Nauerth, A., & Friedburg, H. (1986) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910030602 · Cited by 2,081 · cited 1×
“The 3 T study was performed using a 2D turbo spin echo (TSE, [25]) sequence (TE: 87 ms, TR: 6500 ms, echo train length: 21, resolution: 0.”
MethodThree‐dimensional magnetization‐prepared rapid gradient‐echo imaging (3D MP RAGE)
Mugler, J. P., & Brookeman, J. R. (1990) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910150117 · Cited by 1,265 · cited 3×
“The 7 T experiment was performed using a 3D MP-RAGE [26] sequence (TE: 2.69 ms, TR: 2500 ms, TI: 1050 ms, flip angle: 5 deg., resolution: 0.6 mm isotropic, acquisition time: 14 min.”
MethodInfluentialMotion correction with PROPELLER MRI: Application to head motion and free-breathing cardiac imaging
Pipe, J. G. (1999) · Magnetic Resonance in Medicine · doi:10.1002/(sici)1522-2594(199911)42:5<963::aid-mrm17>3.0.co;2-l · Cited by 928 · cited 1×
“Self-navigating methods, such as PROPELLER (Periodically Rotated Overlapping ParallEL Lines with Enhanced Reconstruction) [2], are popular, but are specific to particular imaging sequences and often increase scan times.”
MethodZaitsev, M., Dold, C., Sakas, G., Hennig, J., & Speck, O. (2006) · NeuroImage · doi:10.1016/j.neuroimage.2006.01.039 · Cited by 391 · cited 5×
“The precision of tracking data during simultaneous MR imaging represents a one to two order of magnitude improvement over previously reported commercial or custom-built systems [3,7,12].”
BackgroundMethodInfluentialSelf-citationPROMO: Real‐time prospective motion correction in MRI using image‐based tracking
White, N., Roddey, C., Shankaranarayanan, A., Han, E., Rettmann, D., Santos, J., Kuperman, J., & Dale, A. (2010) · Magnetic Resonance in Medicine · doi:10.1002/mrm.22176 · Cited by 376 · cited 1×
“Prospective motion correction has obvious potential for routine clinical imaging, where patients may have difficulty remaining still, and this is the likely application of promising new methods, such as PROMO (PROspective MOtion correction) [6].”
MethodThe future of ultra-high field MRI and fMRI for study of the human brain
Duyn, J. H. (2012) · NeuroImage · doi:10.1016/j.neuroimage.2011.10.065 · Cited by 261 · cited 1×
“There is a general trend to higher magnetic field strengths, which improves the image signal-to-noise ratio and therefore allows higher nominal resolution [1].”
BackgroundProspective real‐time correction for arbitrary head motion using active markers
Ooi, M. B., Krueger, S., Thomas, W. J., Swaminathan, S. V., & Brown, T. R. (2009) · Magnetic Resonance in Medicine · doi:10.1002/mrm.22082 · Cited by 165
Qin, L., van Gelderen, P., Derbyshire, J. A., Jin, F., Lee, J., de Zwart, J. A., Tao, Y., & Duyn, J. H. (2009) · Magnetic Resonance in Medicine · doi:10.1002/mrm.22076 · Cited by 152 · cited 2×
“A promising alternative is prospective motion correction [3–7].”
BackgroundProspective Real-Time Slice-by-Slice Motion Correction for fMRI in Freely Moving Subjects
Speck, O., Hennig, J., & Zaitsev, M. (2006) · Magnetic Resonance Materials in Physics Biology and Medicine · doi:10.1007/s10334-006-0027-1 · Cited by 125 · cited 2×
“corruption of Nyquist ghost correction in EPI [19].”
BackgroundSelf-citationFrauenrath, T., Hezel, F., Renz, W., d'Orth, T. D. G., Dieringer, M. A., von Knobelsdorff‐Brenkenhoff, F., Prothmann, M., Schulz‐Menger, J., & Niendorf, T. (2010) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-12-67 · Cited by 117 · cited 1×
“Signal quality after application of the bandpass filter is very high, and MPT data might provide a more robust means of gating than ECG at high fields [22].”
MethodReal‐time optical motion correction for diffusion tensor imaging
Aksoy, M., Forman, C., Straka, M., Skare, S., Holdsworth, S., Hornegger, J., & Bammer, R. (2011) · Magnetic Resonance in Medicine · doi:10.1002/mrm.22787 · Cited by 116 · cited 1×
“The precision of tracking data during simultaneous MR imaging represents a one to two order of magnitude improvement over previously reported commercial or custom-built systems [3,7,12].”
3D cine displacement‐encoded MRI of pulsatile brain motion
Soellinger, M., Rutz, A., Kozerke, S., & Boesiger, P. (2008) · Magnetic Resonance in Medicine · doi:10.1002/mrm.21802 · Cited by 91
Spin‐history artifact during functional MRI: Potential for adaptive correction
Yancey, S. E., Rotenberg, D. J., Tam, F., Chiew, M., Ranieri, S., Biswas, L., Anderson, K. J. T., Baker, S. N., Wright, G. A., & Graham, S. J. (2011) · Medical Physics · doi:10.1118/1.3583814 · Cited by 52
Echo‐planar imaging with prospective slice‐by‐slice motion correction using active markers
Ooi, M. B., Krueger, S., Muraskin, J., Thomas, W. J., & Brown, T. R. (2011) · Magnetic Resonance in Medicine · doi:10.1002/mrm.22780 · Cited by 50 · cited 1×
“Examples include B0 field distortions [17,18] and Figure 4.”
BackgroundBoegle, R., Maclaren, J., & Zaitsev, M. (2010) · Magnetic Resonance Materials in Physics Biology and Medicine · doi:10.1007/s10334-010-0225-8 · Cited by 46 · cited 1×
“Examples include B0 field distortions [17,18] and Figure 4.”
BackgroundSelf-citationCombined prospective and retrospective motion correction to relax navigator requirements
Maclaren, J., Lee, K. J., Luengviriya, C., Speck, O., & Zaitsev, M. (2011) · Magnetic Resonance in Medicine · doi:10.1002/mrm.22754 · Cited by 32 · cited 1×
“If the latency cannot be reduced, a further option could be predictive filtering or correction of the residual motion in post-processing [15].”
BackgroundSelf-citationArmstrong, B., Verron, T., Heppe, L., Reynolds, J., & Schmidt, K. (2003) · Proceedings - IEEE International Conference on Robotics and Automation/Proceedings · doi:10.1109/robot.2002.1013678 · Cited by 21 · cited 2×
“The system, which we call moiré phase tracking (MPT), uses a passive marker that generates moiré patterns to quantify out-of-plane rotations from a single image [10,11].”
BackgroundSelf-citation
Cited by
Parkes, L., Fulcher, B., Yücel, M., & Fornito, A. (2018) · NeuroImage · doi:10.1016/j.neuroimage.2017.12.073 · Cited by 801
Motion artifacts in MRI: A complex problem with many partial solutions
Zaitsev, M., Maclaren, J., & Herbst, M. (2015) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.24850 · Cited by 774
Self-citationPros and cons of ultra-high-field MRI/MRS for human application
Ladd, M. E., Bachert, P., Meyerspeer, M., Moser, E., Nagel, A. M., Norris, D. G., Schmitter, S., Speck, O., Straub, S., & Zaiss, M. (2018) · Progress in Nuclear Magnetic Resonance Spectroscopy · doi:10.1016/j.pnmrs.2018.06.001 · Cited by 544
Self-citationQuantitative multi-parameter mapping of R1, PD*, MT, and R2* at 3T: a multi-center validation
Weiskopf, N., Suckling, J., Williams, G., Correia, M. M., Inkster, B., Tait, R., Ooi, C., Bullmore, E. T., & Lutti, A. (2013) · Frontiers in Neuroscience · doi:10.3389/fnins.2013.00095 · Cited by 538
Low‐field MRI: An MR physics perspective
Marques, J. P., Simonis, F. F. J., & Webb, A. G. (2019) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.26637 · Cited by 386
Metadata
- Authors
- Maclaren, J., Armstrong, B. S. R., Barrows, R. T., Danishad, K. A., Ernst, T., Foster, C. L., Gumus, K., Herbst, M., Kadashevich, I. Y., Kusik, T. P., Li, Q., Lovell-Smith, C., Prieto, T., Schulze, P., Speck, O., Stucht, D., & Zaitsev, M.
- Venue
- PLoS ONE
- Publisher
- Public Library of Science (PLoS)
- Published
- 2012-11-07
- Volume
- 7
- Issue
- 11
- Pages
- e48088
- ISSN
- 1932-6203 (ISSN-L)
- OpenAlex
- W2015093290
- Cited by
- 215
- References
- 17
- FWCI
- 8.68
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.