A Side by Side Comparison of Bruker Biotyper and VITEK MS: Utility of MALDI-TOF MS Technology for Microorganism Identification in a Public Health Reference Laboratory

Open access

The DOI matches its Crossref record; 34 of the 34 references it cites were found in a registry or an index.

Abstract

Matrix-assisted laser desorption ionization-time of flight mass spectrometry (MALDI-TOF MS) has emerged as a rapid, highly accurate, and cost-effective method for routine identification of a wide range of microorganisms. We carried out a side by side comparative evaluation of the performance of Bruker Biotyper versus VITEK MS for identification of a large and diverse collection of microorganisms. Most difficult and/or unusual microorganisms, as well as commonly encountered microorganisms were selected, including Gram-positive and negative bacteria, mycobacteria, actinomycetes, yeasts and filamentous fungi. Six hundred forty two strains representing 159 genera and 441 species from clinical specimens previously identified at the Laboratoire de santé publique du Québec (LSPQ) by reference methods were retrospectively chosen for the study. They included 254 Gram-positive bacteria, 167 Gram-negative bacteria, 109 mycobacteria and aerobic actinomycetes and 112 yeasts and moulds. MALDI-TOF MS analyses were performed on both systems according to the manufacturer's instructions. Of the 642 strains tested, the name of the genus and / or species of 572 strains were referenced in the Bruker database while 406 were present in the VITEK MS IVD database. The Biotyper correctly identified 494 (86.4%) of the strains, while the VITEK MS correctly identified 362 (92.3%) of the strains (excluding 14 mycobacteria that were not tested). Of the 70 strains not present in the Bruker database at the species level, the Biotyper correctly identified 10 (14.3%) to the genus level and 2 (2.9%) to the complex/group level. For 52 (74.2%) strains, we obtained no identification, and an incorrect identification was given for 6 (8.6%) strains. Of the 178 strains not present in the VITEK MS IVD database at the species level (excluding 71 untested mycobacteria and actinomycetes), the VITEK MS correctly identified 12 (6.8%) of the strains each to the genus and to the complex/group level. For 97 (54.5%) strains, no identification was given and for 69 (38.7%) strains, an incorrect identification was obtained. Our study demonstrates that both systems gave a high level (above 85%) of correct identification for a wide range of microorganisms. However, VITEK MS gave more misidentification when the microorganism analysed was not present in the database, compared to Bruker Biotyper. This should be taken into account when this technology is used alone for microorganism identification in a public health laboratory, where isolates received are often difficult to identify and/or unusual microorganisms.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1371/journal.pone.0144878

Cites

34 of 34 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 34 references

Research works printed but not linked: 4

  1. TC Dingle, 2013, Maldi-tof mass spectrometry for microorganism identification, Clin Lab Med, 33, 589, 10.1016/j.cll.2013.03.001 · DOI · Paper page
  2. TE McElvania, 2014, Evaluation of the Bruker Biotyper and VITEK MS MALDI-TOF MS systems for the identification of unusual and/or difficult-to-identify microorganisms isolated from clinical specimens, Eur J Clin Microbiol Infect Dis, 33, 2163, 10.1007/s10096-014-2183-y · DOI
  3. E Deak, 2015, Comparison of the Vitek MS and Bruker Microflex LT MALDI-TOF MS platforms for routine identification of commonly isolated bacteria and yeast in the clinical microbiology laboratory, Diagn Microbiol Infect Dis, 81, 27, 10.1016/j.diagmicrobio.2014.09.018 · DOI
  4. E Deak, 2015, Comparison of the Vitek MS and Bruker Microflex LT MALDI-TOF MS platforms for routine identification of commonly isolated bacteria and yeast in the clinical microbiology laboratory, Diagn Microbiol Infect Dis, 81, 27, 10.1016/j.diagmicrobio.2014.09.018 · DOI

Cited by

See all 172

Metadata

Authors
Lévesque, S., Dufresne, P. J., Soualhine, H., Domingo, M.-C., Bekal, S., Lefebvre, B., & Tremblay, C.
Affiliations
Institut National de Santé Publique du Québec (CA)
Corresponding author
Lévesque, Simon
Venue
PLOS ONE
Published
2015-12-10
Volume
10
Issue
12
Pages
e0144878
ISSN
1932-6203 (ISSN-L)
OpenAlex
W2209387386
Cited by
172
References
34
Citations by year
2026: 7 · 2025: 12 · 2024: 15 · 2023: 20 · 2022: 17 · 2021: 20 · 2020: 16 · 2019: 16 · 2018: 22 · 2017: 19 · 2016: 6
FWCI
12.87
Type
journal article
Language
en
Open access
Locations
  • PLoS ONE (publishedVersion, cc-by)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
  • Figshare (submittedVersion, cc-by)
Topics
  • Health Sciences › Medicine › Microbiology (medical) › Bacterial Identification and Susceptibility Testing
  • Physical Sciences › Chemistry › Spectroscopy › Mass Spectrometry Techniques and Applications
  • Health Sciences › Medicine › Microbiology (medical) › Infections and bacterial resistance
Keywords
MALDI-TOF mass spectrometry, microbial identification, Bruker Biotyper, VITEK MS, method comparison, bacterial identification, mycobacteria, actinomycetes, molds and yeasts, Gram-positive bacteria, Gram-negative bacteria, reference laboratory, clinical isolates, identification accuracy, public health microbiology
MeSH terms
Fungi, Gram-Negative Bacteria, Humans, Mycobacterium, Public Health, Quebec, Yeasts, Reproducibility of Results, Bacterial Typing Techniques, Mycological Typing Techniques, Spectrometry, Mass, Matrix-Assisted Laser Desorption-Ionization, Clinical Laboratory Techniques, Actinobacteria, Clinical Laboratory Services

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: A Side by Side Comparison of Bruker Biotyper and VITEK MS: Utility of MALDI-TOF MS Technology for Microorganism Identification in a Public Health Reference Laboratory. https://app.cochat.ai/papers/10.1371/journal.pone.0144878
The paper
Lévesque, S., Dufresne, P. J., Soualhine, H., Domingo, M.-C., Bekal, S., Lefebvre, B., & Tremblay, C. (2015). A Side by Side Comparison of Bruker Biotyper and VITEK MS: Utility of MALDI-TOF MS Technology for Microorganism Identification in a Public Health Reference Laboratory. PLOS ONE. https://doi.org/10.1371/journal.pone.0144878

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.