Standardized Loads Acting in Hip Implants
Open accessAbstract
With the increasing success of hip joint replacements, the average age of patients has decreased, patients have become more active and their expectations of the implant durability have risen. Thus, pre-clinical endurance tests on hip implants require defining realistic in vivo loads from younger and more active patients. These loads require simplifications to be applicable for simulator tests and numerical analyses. Here, the contact forces in the joint were measured with instrumented hip implants in ten subjects during nine of the most physically demanding and frequent activities of daily living. Typical levels and directions of average and high joint loads were extracted from the intra- and inter-individually widely varying individual data. These data can also be used to analyse bone remodelling at the implant-bone interface, evaluate tissue straining in finite element studies or validate analytical loading predictions, among other uses. The current ISO standards for endurance tests of implant stems and necks are based on historic analytical data from the 1970s. Comparisons of these test forces with in vivo loads unveiled that their unidirectional orientations deviate from the time-dependent in vivo directions during walking and most other activities. The ISO force for testing the stem is substantially too low while the ISO force for the neck better matches typical in vivo magnitudes. Because the magnitudes and orientations of peak forces substantially vary among the activities, load scenarios that reflect a collection of time-dependent high forces should be applied rather than using unidirectional forces. Based on data from ten patients, proposals for the most demanding activities, the time courses of the contact forces and the required cycle numbers for testing are given here. Friction moments in the joint were measured in addition to the contact forces. The moment data were also standardized and can be applied to wear tests of the implant. It was shown that friction only very slightly influences the stresses in the implant neck and shaft.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Wu, G., Siegler, S., Allard, P., Kirtley, C., Leardini, A., Rosenbaum, D., Whittle, M., D’Lima, D. D., Cristofolini, L., Witte, H., et al. (2002) · Journal of Biomechanics · doi:10.1016/s0021-9290(01)00222-6 · Cited by 3,242
Hip contact forces and gait patterns from routine activities
Bergmann, G., Deuretzbacher, G., Heller, M., Graichen, F., Rohlmann, A., Strauss, J., & Duda, G. N. (2001) · Journal of Biomechanics · doi:10.1016/s0021-9290(01)00040-9 · Cited by 2,215
Self-citationAnthropometric Reference Data for Children and Adults: United States, 2003-2006: (623932009-001)
McDowell, M. A., Fryar, C. D., Ogden, C. L., & Flegal, K. M. (2008) · PsycEXTRA Dataset · doi:10.1037/e623932009-001 · Cited by 1,378
The mechanics of the knee joint in relation to normal walking
Morrison, J. B. (1970) · Journal of Biomechanics · doi:10.1016/0021-9290(70)90050-3 · Cited by 1,055
Loading of the Knee Joint during Activities of Daily Living Measured In Vivo in Five Subjects
Kutzner, I., Heinlein, B., Graichen, F., Bender, A., Rohlmann, A., Halder, A., Beier, A., & Bergmann, G. (2010) · Journal of Biomechanics · doi:10.1016/j.jbiomech.2010.03.046 · Cited by 798
Grand challenge competition to predict in vivo knee loads
Fregly, B. J., Besier, T. F., Lloyd, D. G., Delp, S. L., Banks, S. A., Pandy, M. G., & D'Lima, D. D. (2012) · Journal of Orthopaedic Research · doi:10.1002/jor.22023 · Cited by 596
The prediction of muscular load sharing and joint forces in the lower extremities during walking
Seireg, A., & Arvikar, R. J. (1975) · Journal of Biomechanics · doi:10.1016/0021-9290(75)90089-5 · Cited by 499
Musculo-skeletal loading conditions at the hip during walking and stair climbing
Heller, M. O., Bergmann, G., Deuretzbacher, G., Dürselen, L., Pohl, M., Claes, L., Haas, N. P., & Duda, G. N. (2001) · Journal of Biomechanics · doi:10.1016/s0021-9290(01)00039-2 · Cited by 446
Self-citationAnthropometric Reference Data for Children and Adults: United States, 2011-2014.
Fryar, C. D., Gu, Q., Ogden, C. L., & Flegal, K. M. (2016) · PubMed · Cited by 392
Tibio‐femoral loading during human gait and stair climbing
Taylor, W. R., Heller, M. O., Bergmann, G., & Duda, G. N. (2004) · Journal of Orthopaedic Research · doi:10.1016/j.orthres.2003.09.003 · Cited by 366
Self-citationDuration and frequency of every day activities in total hip patients
Morlock, M., Schneider, E., Bluhm, A., Vollmer, M., Bergmann, G., Müller, V, & Honl, M. (2001) · Journal of Biomechanics · doi:10.1016/s0021-9290(01)00035-5 · Cited by 360
Self-citationRealistic loads for testing hip implants
Bergmann, G., Graichen, F., Rohlmann, A., Bender, A., Heinlein, B., Duda, G. N., Heller, M. O., & Morlock, M. M. (2010) · Bio-Medical Materials and Engineering · doi:10.3233/bme-2010-0616 · Cited by 285
Self-citationMuscle, Ligament, and Joint-Contact Forces at the Knee during Walking
Shelburne, K. B., Torry, M. R., & Pandy, M. G. (2005) · Medicine & Science in Sports & Exercise · doi:10.1249/01.mss.0000180404.86078.ff · Cited by 281
Evaluation of the TEOM® Method for Measurement of Ambient Particulate Mass in Urban Areas
Allen, G., Sioutas, C., Koutrakis, P., Reiss, R., Lurmann, F. W., & Roberts, P. T. (1997) · Journal of the Air & Waste Management Association · doi:10.1080/10473289.1997.10463923 · Cited by 273
Mündermann, A., Dyrby, C. O., D'Lima, D. D., Colwell, C. W., & Andriacchi, T. P. (2008) · Journal of Orthopaedic Research · doi:10.1002/jor.20655 · Cited by 249
Taylor, S. J. G., Walker, P. S., Perry, J. S., Cannon, S. R., & Woledge, R. (1998) · The Journal of Arthroplasty · doi:10.1016/s0883-5403(98)90009-2 · Cited by 246
Tibial Forces Measured In Vivo After Total Knee Arthroplasty
D'Lima, D. D., Patil, S., Steklov, N., Slamin, J. E., & Colwell, C. W. (2006) · The Journal of Arthroplasty · doi:10.1016/j.arth.2005.07.011 · Cited by 231
Individual muscle contributions to the axial knee joint contact force during normal walking
Sasaki, K., & Neptune, R. R. (2010) · Journal of Biomechanics · doi:10.1016/j.jbiomech.2010.06.011 · Cited by 192
In vivo knee moments and shear after total knee arthroplasty
D’Lima, D. D., Patil, S., Steklov, N., Chien, S., & Colwell, C. W. (2007) · Journal of Biomechanics · doi:10.1016/j.jbiomech.2007.03.004 · Cited by 188
JOINT LOAD CONSIDERATIONS IN TOTAL KNEE REPLACEMENT
Kuster, M. S., Wood, G. A., Stachowiak, G., & Gächter, A. (1997) · Journal of Bone and Joint Surgery - British Volume · doi:10.1302/0301-620x.79b1.6978 · Cited by 179
Implantable 9-Channel Telemetry System for In Vivo Load Measurements With Orthopedic Implants
Graichen, F., Arnold, R., Rohlmann, A., & Bergmann, G. (2007) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2006.886857 · Cited by 173
Self-citationInfluence of Limb Alignment on Mediolateral Loading in Total Knee Replacement
Halder, A. M., Kutzner, I., Graichen, F., Heinlein, B., Beier, A., & Bergmann, G. (2012) · Journal of Bone and Joint Surgery · doi:10.2106/jbjs.k.00927 · Cited by 156
Self-citationDesign, calibration and pre-clinical testing of an instrumented tibial tray
Heinlein, B., Graichen, F., Bender, A., Rohlmann, A., & Bergmann, G. (2007) · Journal of Biomechanics · doi:10.1016/j.jbiomech.2007.02.014 · Cited by 150
Self-citationBioengineering analysis of force actions transmitted by the knee joint
Morrison, J. D. (1968) · Cited by 145
THE CHITRANJAN RANAWAT AWARD: In Vivo Knee Forces after Total Knee Arthroplasty
DʼLima, D. D., Patil, S. S., Steklov, N., Slamin, J. E., & Colwell, C. W. (2005) · Clinical Orthopaedics and Related Research · doi:10.1097/01.blo.0000186559.62942.8c · Cited by 128
Multichannel strain gauge telemetry for orthopaedic implants
Bergmann, G., Graichen, F., Siraky, J., Jendrzynski, H., & Rohlmann, A. (1988) · Journal of Biomechanics · doi:10.1016/0021-9290(88)90009-7 · Cited by 127
Self-citationTaylor, W. R., Kornaropoulos, E., Duda, G. N., Kratzenstein, S., Ehrig, R. M., Arampatzis, A., & Heller, M. O. W. (2010) · Gait & Posture · doi:10.1016/j.gaitpost.2010.05.005 · Cited by 120
Self-citationInstrumented implant for measuring tibiofemoral forces
Kaufman, K. R., Kovacevic, N., Irby, S. E., & Colwell, C. W. (1996) · Journal of Biomechanics · doi:10.1016/0021-9290(95)00124-7 · Cited by 117
Total hip joint prosthesis for in vivo measurement of forces and moments
Damm, P., Graichen, F., Rohlmann, A., Bender, A., & Bergmann, G. (2009) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2009.10.003 · Cited by 114
Self-citationDie Entwicklung der Adipositas in Deutschland im Zeitraum von 1985 bis 2002
Helmert, U., & Strube, H. (2004) · Das Gesundheitswesen · doi:10.1055/s-2004-813324 · Cited by 106
Patellofemoral joint contact forces during activities with high knee flexion
Trepczynski, A., Kutzner, I., Kornaropoulos, E., Taylor, W. R., Duda, G. N., Bergmann, G., & Heller, M. O. W. (2011) · Journal of Orthopaedic Research® · doi:10.1002/jor.21540 · Cited by 104
Self-citationBody Weight Distributions for Risk Assessment
Portier, K. M., Tolson, J. K., & Roberts, S. M. (2007) · Risk Analysis · doi:10.1111/j.1539-6924.2006.00856.x · Cited by 103
Di Monaco, M., & Castiglioni, C. (2013) · PubMed · Cited by 83
The effect of displacement control input parameters on tibiofemoral prosthetic knee wear
Johnson, T. S., Laurent, M. P., Yao, J. Q., & Gilbertson, L. N. (2001) · Wear · doi:10.1016/s0043-1648(01)00650-0 · Cited by 75
Model analysis of factors influencing the prediction of muscle forces at the knee
Mikosz, R. P., Andriacchi, T. P., & Andersson, G. B. J. (1988) · Journal of Orthopaedic Research® · doi:10.1002/jor.1100060207 · Cited by 57
Comparison of Fatigue Properties and Fatigue Crack Growth Rates of Various Implantable Metals
Okazaki, Y. (2012) · Materials · doi:10.3390/ma5122981 · Cited by 54
Popoola, O. O., Yao, J. Q., Johnson, T. S., & Blanchard, C. R. (2010) · Journal of Orthopaedic Research® · doi:10.1002/jor.21104 · Cited by 53
Schwiesau, J., Schilling, C., Kaddick, C., Utzschneider, S., Jansson, V., Fritz, B., Blömer, W., & Grupp, T. M. (2012) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2012.07.003 · Cited by 52
Determination of typical patterns from strongly varying signals
Bender, A., & Bergmann, G. (2011) · Computer Methods in Biomechanics & Biomedical Engineering · doi:10.1080/10255842.2011.560841 · Cited by 51
Self-citationEstimation of bone-on-bone contact forces in the tibiofemoral joint during walking
Thambyah, A., Pereira, B. P., & WYSS, U. P. (2005) · The Knee · doi:10.1016/j.knee.2004.12.005 · Cited by 48
Fabry, C., Herrmann, S., Kaehler, M., Klinkenberg, E., Woernle, C., & Bader, R. (2012) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2012.07.014 · Cited by 42
Comparison of ISO Standard and TKR patient axial force profiles during the stance phase of gait
Lundberg, H. J., Ngai, V., & Wimmer, M. A. (2011) · Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine · doi:10.1177/0954411911431650 · Cited by 18
TKA wear testing input after kinematic and dynamic meta-analysis: Technique and proof of concept
Franta, L., Kronek, J., & Suchánek, J. (2011) · Wear · doi:10.1016/j.wear.2011.02.026 · Cited by 15
Metadata
- Authors
- Bergmann, G., Bender, A., Dymke, J., Duda, G., & Damm, P.
- Venue
- PLOS ONE
- Publisher
- Public Library of Science (PLoS)
- Published
- 2016-05-19
- Volume
- 11
- Issue
- 5
- Pages
- e0155612
- ISSN
- 1932-6203 (ISSN-L)
- OpenAlex
- W2161748213
- Cited by
- 562
- References
- 43
- FWCI
- 50.12
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.