Gene electrotransfer with flow-through microchannel and lower alternating voltage generated induced pluripotent cells from human lymphoblastoid cell lines
Open accessThe DOI matches its Crossref record; 53 of the 53 references it cites were found in a registry or an index; none of the 133 sources its claims cite is checked for support yet.
Abstract
Induced pluripotent stem cells (iPSCs) are useful for studying genetic and rare diseases and can be generated by reprogramming immortalized lymphoblastoid cell lines (LCLs) stored in global repositories with detailed genotype and phenotype data. Traditional bulk-type electroporators are commonly used for gene electrotransfer in reprogramming, but they have major drawbacks, including high costs associated with electric pulse generators and the requirement for fixed volumes for costly reprogramming factors. These limitations hinder cost-effective and scalable iPSC generation, particularly when working with large numbers of LCLs with diverse genotypes. We aimed to develop a flow-through-type electroporator utilizing microchannels for the generation of iPSCs from LCLs, to reduce the costs associated with traditional bulk-type electroporators and enable parallel processing for LCLs with various genotypes. We applied a continuous wave of biphasic alternating voltage (~10 V one-sided amplitude) to micro-scaled electrodes within the microchannel to develop a flow-through electroporator. Numerical simulations were conducted to assess the electric field distribution and its applicability to pore formation in the plasma membrane. To optimize electroporation and flow conditions, we used plasmid pCXLE-EGFP (encoding Green Fluorescent Protein, GFP) for gene electrotransfer to LCLs. Reprogramming factors (pCXLE-hSK, pCXLE-hOCT3/4-shp53-F, pCXLE-hUL) were also delivered to the cells via the same system. The flow-through electroporator achieved 31% transfection efficiency with 78% cell viability, 2 d post-electroporation. In each condition, only 3 µL of cell suspension was used with 10 7 cells/mL of cells and 500 ng/µL plasmid vector. A reprogramming efficiency of 0.048% was obtained, which is comparable to that achieved using bulk-type electroporators. This developed flow-through electroporator with microchannel technology offers significant advantages over traditional methods, including the potential to reduce costs and the ability to process small volumes of cell suspension, making it suitable for parallel processing of LCLs with diverse genotypes. The system provides a promising approach for scalable and potentially cost-effective iPSC generation.
Overview
CoChat Summary
TL;DR
A flow-through PDMS microchannel electroporator using ~10 V biphasic alternating voltage delivered episomal plasmids into human lymphoblastoid cell lines and generated iPSC colonies, with 31% GFP transfection/78% viability under a triangle waveform and a maximum reprogramming efficiency of 0.048%. The system uses small sample volumes and low voltage, but its transfection efficiency remains below commercial bulk electroporators.
Key findings
- Numerical simulation showed an electric field >1 kV/cm above the electrode pair, with a pore-formation threshold around 6 V; at 5 V, pore density dropped by more than two orders of magnitude, and transmembrane potential reached approximately 0.8 V.
- GFP electrotransfer had an estimated threshold between 5.0 and 7.5 V; at 7.5 and 10.0 V, transfection efficiency was >10% with >50% viability. Maximum transfection was 39% with 58% viability using a sinusoidal waveform, and 31% with 78% viability using a triangle waveform.
- iPSC reprogramming reached a maximum efficiency of 0.048% (12.5 V triangle, 10 kPa, 57 ± 6 colonies, n = 3), comparable to reported LCL reprogramming ranges of 0.001–0.06%; other conditions yielded 0.009–0.043%.
- High transfection occurred at cell velocities of approximately 50–100 mm/s; average velocity was 61 ± 12 mm/s, giving about 0.66 ms exposure and ~6.6 cycles per cell pass. Increasing pressure to 20 kPa improved sinusoidal 12.5 V reprogramming efficiency to 0.043%.
Methods
- The device used a PDMS microchannel (designed width 50 µm, height 30 µm, length 5.8 mm) with Pt/Ti electrodes (50 µm width, 40 µm gap), pipette-tip sample loading, pneumatic pressure-driven flow, and a 10 kHz biphasic sinusoidal or triangle voltage. GFP experiments used 3 µL of 10^7 cells/mL LCLs with 500 ng/µL pCXLE-EGFP; reprogramming used 12 µL with pCXLE-hSK, pCXLE-hOCT3/4-shp53-F, and pCXLE-hUL.
- Human EBV-transformed B lymphocyte LCL HEV0019 (Riken BRC, 25 passages) was cultured in RPMI1640 with 10% FBS and 1% penicillin-streptomycin, then suspended in Opti-MEM I. GFP transfection was assessed 2 days after processing by Hoechst 33342/PI/GFP imaging; iPSC colonies were maintained in Stemfit medium and evaluated at 21 days by alkaline phosphatase staining.
- Electric field and pore formation were simulated in COMSOL Multiphysics 6.2. Transfection efficiency was calculated as nGFP/(nHoechst−nPI), viability was calculated with negative-control subtraction, and reprogramming efficiency was colonies divided by initial processed cells (12 µL × 10^7 cells/mL = 1.2 × 10^5 cells). Statistical comparisons used two-tailed Student’s t-tests.
Limitations
- The paper states that transfection efficiency was lower than commercial bulk electroporators (>90%) and other microchannel systems (~75% for plasmid, 95% for mRNA), likely due to LCLs being hard-to-transfect and to the heterogeneous electric field from planar printed electrodes; the tapered channel may also let some cells pass with insufficient electrical exposure.
- The paper reports that channel clogging prevented completion of one 12.5 V/20 kPa condition, limiting it to n = 2; no transplantation of generated iPSCs into living organisms was performed.
- Evident from the text: experiments used a single LCL line (HEV0019) with small replicate numbers (n = 2–3 in reprogramming conditions), and pluripotency was assessed only by alkaline phosphatase staining, with no in vivo differentiation or teratoma assay reported.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Figures
![Fig 1 Experimental setup of gene electrotransfer with microchannel. (a) A schematic drawing of a microchannel chip for gene electrotransfer. Cells and plasmid vector suspension were introduced into single-lined microchannel via pipette tip and pneumatic pressure. An alternating biphasic wave, either sinusoidal or triangular, was applied across the electrodes, forming an electric field around them that facilitated plasmid vector transfer into cells. (b) Fabricated microchannel chip with inlet and outlet biopsy punches of 1- and 3-mm diameter, respectively. An unused branch channel for droplet formation was also visible [36]. The scale bar is 200 µm. (c) Experimental setup for flow generation and voltage application. A pneumatic pump (Flow EZ 7000 mbar, Fluigent) was attached to the pipette tip at the inlet. The electric power source consisted of a waveform generator (33500B, Agilent) connected to a bipolar amplifier (BA4825, NF Co.). A 1 kΩ resistor was connected to the electrodes, and an oscilloscope (PicoScope 6404E, P2056, Pico Technology) monitored the electrode voltage. The voltage amplitude decreased as the solution contacts the electrodes, and amplitude fluctuations indicated cell passage, which helped to monitor channel activity.](https://figures.cochat.ai/paper-figures/4682853b7b854538a78bbe6b457af495/gene-electrotransfer-with-flow-through-microchannel-and-lower-alternating-figure-1.jpg)


![Fig 4 Experimental results of gene electrotransfer with pCXLE-EGFP plasmid vector as a parameter of cell velocity. (a) Measured cell velocity in each applied pneumatic pressure. The theoretical value was calculated based on the flow speed of the rectangular channel [49]. Transfection efficiency and cell viability as a parameter of flow velocity for (b) sinusoidal wave and (c) triangle wave. Vin = 10 V was applied in each experiment.](https://figures.cochat.ai/paper-figures/5186c03e6b934d1e9187542a73050bb5/gene-electrotransfer-with-flow-through-microchannel-and-lower-alternating-figure-4.jpg)

Verification
Claims the paper makes
0 Valid · 0 Not valid · 44 Not checked · 93 Own results and interpretations
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“with GFP expressions observed at voltages exceeding 5V.”
§ Gene electrotransfer with flow-through microchannel and lower alternating voltage generated induced pluripotent cells from human lymphoblastoid cell lines > Results > Gene electrotransfer of pCXLE-EGFP plasmid vector
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“with Vp = 10 V producing the highest GFP expression.”
§ Gene electrotransfer with flow-through microchannel and lower alternating voltage generated induced pluripotent cells from human lymphoblastoid cell lines > Results > Gene electrotransfer of pCXLE-EGFP plasmid vector
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“At lower voltages (Vp = 2.5 and 5.0 V), transfection efficiency was low.”
§ Gene electrotransfer with flow-through microchannel and lower alternating voltage generated induced pluripotent cells from human lymphoblastoid cell lines > Results > Gene electrotransfer of pCXLE-EGFP plasmid vector
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper states — it cites no source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“As cell velocity increased, transfection efficiencies decreased for both waveforms, especially for the triangle waveform.”
§ Gene electrotransfer with flow-through microchannel and lower alternating voltage generated induced pluripotent cells from human lymphoblastoid cell lines > Results > Gene electrotransfer of pCXLE-EGFP plasmid vector
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The transfection efficiency of GFP expression and cell viability results are shown in Fig 3(c), showing similar responses to voltage across both sinusoidal and triangle waveforms.”
§ Gene electrotransfer with flow-through microchannel and lower alternating voltage generated induced pluripotent cells from human lymphoblastoid cell lines > Results > Gene electrotransfer of pCXLE-EGFP plasmid vector
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
51 more claims
Create a free account to see all 137 claims and to verify the 132 references still open.
- It may state two separate claims as one. 57
- Probably not a substantial claim. 36
- The quote only partly states this claim. 1
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Rated major: it would touch a result — a value a result rests on, reported and recomputed p-values on opposite sides of the significance level, two passages that conflict with no sign which is right, or a cited work retracted before this paper was published.
In the paper
§ Gene electrotransfer with flow-through microchannel and lower alternating voltage generated induced pluripotent cells from human lymphoblastoid cell lines > Results > Gene electrotransfer of pCXLE-EGFP plasmid vector
“For the sinusoidal waveform, maximum transfection efficiency reached 59% with a viability rate of 39%”
§ Gene electrotransfer with flow-through microchannel and lower alternating voltage generated induced pluripotent cells from human lymphoblastoid cell lines > Discussion
“The maximum transfection efficiency observed was 39% with viability of 58% using a sinusoidal waveform”
Rated major: it would touch a result — a value a result rests on, reported and recomputed p-values on opposite sides of the significance level, two passages that conflict with no sign which is right, or a cited work retracted before this paper was published.
In the paper
§ Gene electrotransfer with flow-through microchannel and lower alternating voltage generated induced pluripotent cells from human lymphoblastoid cell lines > Results > Gene electrotransfer of pCXLE-EGFP plasmid vector
“for the triangle waveform, the maximum efficiency was 52% with a viability rate of 85%.”
§ Gene electrotransfer with flow-through microchannel and lower alternating voltage generated induced pluripotent cells from human lymphoblastoid cell lines > Discussion
“The maximum transfection efficiency observed was 39% with viability of 58% using a sinusoidal waveform, and 31% with viability of 78% using a triangular wave.”
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ Gene electrotransfer with flow-through microchannel and lower alternating voltage generated induced pluripotent cells from human lymphoblastoid cell lines > Conclusion
“A voltage range of 7.5–12.5 V allowed for efficient gene transfer without clogging or bubbling”
§ Gene electrotransfer with flow-through microchannel and lower alternating voltage generated induced pluripotent cells from human lymphoblastoid cell lines > Discussion
“The observed result of “small amount of bubble is generated on the electrodes at the beginning of voltage application” in 12.5 V may be due to bubble nuclei forming on the electrodes.”
Citations
Cites
Takahashi, K., & Yamanaka, S. (2006) · Cell · doi:10.1016/j.cell.2006.07.024 · Cited by 26,673
Gene transfer into mouse lyoma cells by electroporation in high electric fields.
Neumann, E., Schaefer-Ridder, M., Wang, Y., & Hofschneider, P. H. (1982) · The EMBO Journal · doi:10.1002/j.1460-2075.1982.tb01257.x · Cited by 2,502
A more efficient method to generate integration-free human iPS cells
Okita, K., Matsumura, Y., Sato, Y., Okada, A., Morizane, A., Okamoto, S., Hong, H., Nakagawa, M., Tanabe, K., Tezuka, K.-I., et al. (2011) · Nature Methods · doi:10.1038/nmeth.1591 · Cited by 2,061
Self-citationModeling Electroporation in a Single Cell
Krassowska, W., & Filev, P. D. (2007) · Biophysical Journal · doi:10.1529/biophysj.106.094235 · Cited by 501 · cited 1×
“Additional parameter values were taken from the literature, as shown in Table A in S1 Appendix [39–42].”
Direct visualization at the single-cell level of electrically mediated gene delivery
Golzio, M., Teissié, J., & Rols, M.-P. (2002) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.022646499 · Cited by 411
Teissié, J., & Rols, M. P. (1993) · Biophysical Journal · doi:10.1016/s0006-3495(93)81052-x · Cited by 374
Cellular membrane potentials induced by alternating fields
Grosse, C., & Schwan, H. P. (1992) · Biophysical Journal · doi:10.1016/s0006-3495(92)81740-x · Cited by 247
Gene Electrotransfer: A Mechanistic Perspective
Rosazza, C., Haberl Meglic, S., Zumbusch, A., Rols, M.-P., & Miklavcic, D. (2016) · Current Gene Therapy · doi:10.2174/1566523216666160331130040 · Cited by 209
Pucihar, G., Miklavcic, D., & Kotnik, T. (2009) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2009.2014244 · Cited by 175
Electroporation of mammalian cells in a microfluidic channel with geometric variation
Wang, H.-Y., & Lu, C. (2006) · Analytical Chemistry · doi:10.1021/ac060733n · Cited by 141
Research works printed but not linked: 1
- HS Rho, 2015, Direct delivery of reagents from a pipette tip to a PDMS microfluidic device, Chips and tips
Cited by
Microfluidic transfection systems
Mohammadi, R., Stavrakis, S., & deMello, A. J. (2026) · Advanced Drug Delivery Reviews · doi:10.1016/j.addr.2026.115818 · Cited by 6
Ishii-Teshima, M., Maeda, K., Hanauchi, K., Asechi, E., Setyawan, D., Niki, T., NAKASHIMA, K., Kurita, H., Numano, R., & Shibata, T. (2025) · PLoS ONE · doi:10.1371/journal.pone.0339487
Self-citationIshii-Teshima, M., Kawamoto, I., NAKASHIMA, K., Kurita, H., Numano, R., & Shibata, T. (2026) · SSRN Electronic Journal · doi:10.2139/ssrn.6920319
Self-citationReprogramming lymphoblastoid cell lines into induced pluripotent stem cells for disease modeling
Gao, F., Sun, H., & Sun, Y. (2026) · Discover Medicine · doi:10.1007/s44337-026-00745-y
Metadata
- Authors
- Ishii-Teshima, M., Maeda, K., Hanauchi, K., Asechi, E., Setyawan, D., Niki, T., Nakashima, K., Kurita, H., Numano, R., & Shibata, T.
- Affiliations
- Toyohashi University of Technology (JP)
- National Institute of Technology, Sasebo College (JP)
- Corresponding author
- Ishii-Teshima, Miho
- Shibata, Takayuki
- Venue
- PLOS One
- Publisher
- Public Library of Science (PLoS)
- Published
- 2025-09-26
- Volume
- 20
- Issue
- 9
- Pages
- e0333491
- ISSN
- 1932-6203 (ISSN-L)
- PMC
- PMC12468887
- OpenAlex
- W4414539127
- Cited by
- 4
- References
- 53
- Citations by year
- 2026: 3 · 2025: 1
- FWCI
- 1.24
- Type
- journal article
- Language
- en
- Update notices
- correction (2025-12-22)
- Open access
- gold
- doi.org
- Licence
- Locations
- PLoS ONE (publishedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Funders
- Japan Society for the Promotion of Science (22J40077)
- Adaptable and Seamless Technology Transfer Program through Target-Driven R and D (AS2915129U)
- Naito Science and Engineering Foundation
- The Hibi Science Foundation
- Topics
- Life Sciences › Biochemistry, Genetics and Molecular Biology › Biotechnology › Microbial Inactivation Methods
- Life Sciences › Biochemistry, Genetics and Molecular Biology › Molecular Biology › Pluripotent Stem Cells Research
- Life Sciences › Biochemistry, Genetics and Molecular Biology › Molecular Biology › RNA Interference and Gene Delivery
- Keywords
- induced pluripotent stem cells, gene electrotransfer, lymphoblastoid cell lines, iPSC reprogramming, electric field distribution, reprogramming factors
- MeSH terms
- Cell Line, Humans, Lymphocytes, Gene Transfer Techniques, Electroporation, Induced Pluripotent Stem Cells, Cellular Reprogramming
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
A language model drew the claims from the paper on 2026-11-21, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-11-21 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›