Gene electrotransfer with flow-through microchannel and lower alternating voltage generated induced pluripotent cells from human lymphoblastoid cell lines

Open access

The DOI matches its Crossref record; 53 of the 53 references it cites were found in a registry or an index; none of the 133 sources its claims cite is checked for support yet.

Abstract

Induced pluripotent stem cells (iPSCs) are useful for studying genetic and rare diseases and can be generated by reprogramming immortalized lymphoblastoid cell lines (LCLs) stored in global repositories with detailed genotype and phenotype data. Traditional bulk-type electroporators are commonly used for gene electrotransfer in reprogramming, but they have major drawbacks, including high costs associated with electric pulse generators and the requirement for fixed volumes for costly reprogramming factors. These limitations hinder cost-effective and scalable iPSC generation, particularly when working with large numbers of LCLs with diverse genotypes. We aimed to develop a flow-through-type electroporator utilizing microchannels for the generation of iPSCs from LCLs, to reduce the costs associated with traditional bulk-type electroporators and enable parallel processing for LCLs with various genotypes. We applied a continuous wave of biphasic alternating voltage (~10 V one-sided amplitude) to micro-scaled electrodes within the microchannel to develop a flow-through electroporator. Numerical simulations were conducted to assess the electric field distribution and its applicability to pore formation in the plasma membrane. To optimize electroporation and flow conditions, we used plasmid pCXLE-EGFP (encoding Green Fluorescent Protein, GFP) for gene electrotransfer to LCLs. Reprogramming factors (pCXLE-hSK, pCXLE-hOCT3/4-shp53-F, pCXLE-hUL) were also delivered to the cells via the same system. The flow-through electroporator achieved 31% transfection efficiency with 78% cell viability, 2 d post-electroporation. In each condition, only 3 µL of cell suspension was used with 10 7 cells/mL of cells and 500 ng/µL plasmid vector. A reprogramming efficiency of 0.048% was obtained, which is comparable to that achieved using bulk-type electroporators. This developed flow-through electroporator with microchannel technology offers significant advantages over traditional methods, including the potential to reduce costs and the ability to process small volumes of cell suspension, making it suitable for parallel processing of LCLs with diverse genotypes. The system provides a promising approach for scalable and potentially cost-effective iPSC generation.

Overview

CoChat Summary

TL;DR

A flow-through PDMS microchannel electroporator using ~10 V biphasic alternating voltage delivered episomal plasmids into human lymphoblastoid cell lines and generated iPSC colonies, with 31% GFP transfection/78% viability under a triangle waveform and a maximum reprogramming efficiency of 0.048%. The system uses small sample volumes and low voltage, but its transfection efficiency remains below commercial bulk electroporators.

Key findings

  • Numerical simulation showed an electric field >1 kV/cm above the electrode pair, with a pore-formation threshold around 6 V; at 5 V, pore density dropped by more than two orders of magnitude, and transmembrane potential reached approximately 0.8 V.
  • GFP electrotransfer had an estimated threshold between 5.0 and 7.5 V; at 7.5 and 10.0 V, transfection efficiency was >10% with >50% viability. Maximum transfection was 39% with 58% viability using a sinusoidal waveform, and 31% with 78% viability using a triangle waveform.
  • iPSC reprogramming reached a maximum efficiency of 0.048% (12.5 V triangle, 10 kPa, 57 ± 6 colonies, n = 3), comparable to reported LCL reprogramming ranges of 0.001–0.06%; other conditions yielded 0.009–0.043%.
  • High transfection occurred at cell velocities of approximately 50–100 mm/s; average velocity was 61 ± 12 mm/s, giving about 0.66 ms exposure and ~6.6 cycles per cell pass. Increasing pressure to 20 kPa improved sinusoidal 12.5 V reprogramming efficiency to 0.043%.

Methods

  • The device used a PDMS microchannel (designed width 50 µm, height 30 µm, length 5.8 mm) with Pt/Ti electrodes (50 µm width, 40 µm gap), pipette-tip sample loading, pneumatic pressure-driven flow, and a 10 kHz biphasic sinusoidal or triangle voltage. GFP experiments used 3 µL of 10^7 cells/mL LCLs with 500 ng/µL pCXLE-EGFP; reprogramming used 12 µL with pCXLE-hSK, pCXLE-hOCT3/4-shp53-F, and pCXLE-hUL.
  • Human EBV-transformed B lymphocyte LCL HEV0019 (Riken BRC, 25 passages) was cultured in RPMI1640 with 10% FBS and 1% penicillin-streptomycin, then suspended in Opti-MEM I. GFP transfection was assessed 2 days after processing by Hoechst 33342/PI/GFP imaging; iPSC colonies were maintained in Stemfit medium and evaluated at 21 days by alkaline phosphatase staining.
  • Electric field and pore formation were simulated in COMSOL Multiphysics 6.2. Transfection efficiency was calculated as nGFP/(nHoechst−nPI), viability was calculated with negative-control subtraction, and reprogramming efficiency was colonies divided by initial processed cells (12 µL × 10^7 cells/mL = 1.2 × 10^5 cells). Statistical comparisons used two-tailed Student’s t-tests.

Limitations

  • The paper states that transfection efficiency was lower than commercial bulk electroporators (>90%) and other microchannel systems (~75% for plasmid, 95% for mRNA), likely due to LCLs being hard-to-transfect and to the heterogeneous electric field from planar printed electrodes; the tapered channel may also let some cells pass with insufficient electrical exposure.
  • The paper reports that channel clogging prevented completion of one 12.5 V/20 kPa condition, limiting it to n = 2; no transplantation of generated iPSCs into living organisms was performed.
  • Evident from the text: experiments used a single LCL line (HEV0019) with small replicate numbers (n = 2–3 in reprogramming conditions), and pluripotency was assessed only by alkaline phosphatase staining, with no in vivo differentiation or teratoma assay reported.

Model-written summary. Check it against the paper before relying on it.

Integrity signals

Registry

Crossref · 10.1371/journal.pone.0333491

Cites

53 of 53 references verified0 citation errors · 0 not checkable

Statistics

Nothing to recompute.

Figures

Fig 1 Experimental setup of gene electrotransfer with microchannel. (a) A schematic drawing of a microchannel chip for gene electrotransfer. Cells and plasmid vector suspension were introduced into single-lined microchannel via pipette tip and pneumatic pressure. An alternating biphasic wave, either sinusoidal or triangular, was applied across the electrodes, forming an electric field around them that facilitated plasmid vector transfer into cells. (b) Fabricated microchannel chip with inlet and outlet biopsy punches of 1- and 3-mm diameter, respectively. An unused branch channel for droplet formation was also visible [36]. The scale bar is 200 µm. (c) Experimental setup for flow generation and voltage application. A pneumatic pump (Flow EZ 7000 mbar, Fluigent) was attached to the pipette tip at the inlet. The electric power source consisted of a waveform generator (33500B, Agilent) connected to a bipolar amplifier (BA4825, NF Co.). A 1 kΩ resistor was connected to the electrodes, and an oscilloscope (PicoScope 6404E, P2056, Pico Technology) monitored the electrode voltage. The voltage amplitude decreased as the solution contacts the electrodes, and amplitude fluctuations indicated cell passage, which helped to monitor channel activity.
Fig 1 Experimental setup of gene electrotransfer with microchannel. (a) A schematic drawing of a microchannel chip for gene electrotransfer. Cells and plasmid vector suspension were introduced into single-lined microchannel via pipette tip and pneumatic pressure. An alternating biphasic wave, either sinusoidal or triangular, was applied across the electrodes, forming an electric field around them that facilitated plasmid vector transfer into cells. (b) Fabricated microchannel chip with inlet and outlet biopsy punches of 1- and 3-mm diameter, respectively. An unused branch channel for droplet formation was also visible [36]. The scale bar is 200 µm. (c) Experimental setup for flow generation and voltage application. A pneumatic pump (Flow EZ 7000 mbar, Fluigent) was attached to the pipette tip at the inlet. The electric power source consisted of a waveform generator (33500B, Agilent) connected to a bipolar amplifier (BA4825, NF Co.). A 1 kΩ resistor was connected to the electrodes, and an oscilloscope (PicoScope 6404E, P2056, Pico Technology) monitored the electrode voltage. The voltage amplitude decreased as the solution contacts the electrodes, and amplitude fluctuations indicated cell passage, which helped to monitor channel activity.CC BY · PubMed Central
Fig 2 Results of numerical simulation of the electric field in the microchannel under a 10-V application. (a) Intensity of electric field in the microchannel without cells. (b) Intensity of electric field when a cell is centered in the channel between electrodes at 1 ms. (c) Transmembrane potential distribution on cell membrane at 1 ms based on asymptotic Smoluchowski equations. (d) Distribution of pore density on the plasma membrane. (e) Distribution of pore density on plasma membrane circumference at various application voltages at a time of 1 ms. (f) Pore density at q = 0, and 1 ms in various applications.
Fig 2 Results of numerical simulation of the electric field in the microchannel under a 10-V application. (a) Intensity of electric field in the microchannel without cells. (b) Intensity of electric field when a cell is centered in the channel between electrodes at 1 ms. (c) Transmembrane potential distribution on cell membrane at 1 ms based on asymptotic Smoluchowski equations. (d) Distribution of pore density on the plasma membrane. (e) Distribution of pore density on plasma membrane circumference at various application voltages at a time of 1 ms. (f) Pore density at q = 0, and 1 ms in various applications.CC BY · PubMed Central
Fig 3 Experimental results of gene electrotransfer with pCXLE-EGFP plasmid vector as a parameter of applied voltage. (a) Images of LCLs with bright field (BF), the fluorescence of Hoechst 33342 (blue), PI (red), and GFP expression (green). Bright-field images and fluorescent-filtered images with the filters of FITC, mCherry, and DAPI are taken, respectively. Positions of cells were slightly moved while taking with each fluorescent filter because LCLs are floating cells. Scale bar = 200 µm. (b) Microscopic images of microchannel around electrode pair. White arrows show the location of electrode elution. Bubbling and elution of electrodes are sometimes observed above 12.5 V application. Scale bar = 100 µm. (c) Transfection efficiency and viability of LCLs as a parameter of application voltages for sinusoidal wave and triangle wave. Significant differences between 0V application and other voltages are indicated by #. Significant differences between triangle and sinusoidal waves are depicted by *.
Fig 3 Experimental results of gene electrotransfer with pCXLE-EGFP plasmid vector as a parameter of applied voltage. (a) Images of LCLs with bright field (BF), the fluorescence of Hoechst 33342 (blue), PI (red), and GFP expression (green). Bright-field images and fluorescent-filtered images with the filters of FITC, mCherry, and DAPI are taken, respectively. Positions of cells were slightly moved while taking with each fluorescent filter because LCLs are floating cells. Scale bar = 200 µm. (b) Microscopic images of microchannel around electrode pair. White arrows show the location of electrode elution. Bubbling and elution of electrodes are sometimes observed above 12.5 V application. Scale bar = 100 µm. (c) Transfection efficiency and viability of LCLs as a parameter of application voltages for sinusoidal wave and triangle wave. Significant differences between 0V application and other voltages are indicated by #. Significant differences between triangle and sinusoidal waves are depicted by *.CC BY · PubMed Central
Fig 4 Experimental results of gene electrotransfer with pCXLE-EGFP plasmid vector as a parameter of cell velocity. (a) Measured cell velocity in each applied pneumatic pressure. The theoretical value was calculated based on the flow speed of the rectangular channel [49]. Transfection efficiency and cell viability as a parameter of flow velocity for (b) sinusoidal wave and (c) triangle wave. Vin = 10 V was applied in each experiment.
Fig 4 Experimental results of gene electrotransfer with pCXLE-EGFP plasmid vector as a parameter of cell velocity. (a) Measured cell velocity in each applied pneumatic pressure. The theoretical value was calculated based on the flow speed of the rectangular channel [49]. Transfection efficiency and cell viability as a parameter of flow velocity for (b) sinusoidal wave and (c) triangle wave. Vin = 10 V was applied in each experiment.CC BY · PubMed Central
Fig 5 Images of generated iPSCs reprogrammed by gene electrotransfer to LCLs. (a) Microscopy images of iPSCs 21 days after gene electrotransfer. Flat-shaped cells are reprogrammed iPSCs, while sphererical cells are LCLs. Electrotransfer was performed using a triangle waverform at Vp = 12.5 V with an air pressure of 10 kPa for acceleration. Scale bar = 100 µm. (b) Image of a 12-well plates containing cultured and fixed iPSCs. Alkaline-phosphatase-stained cell colonies are depicted as black points.
Fig 5 Images of generated iPSCs reprogrammed by gene electrotransfer to LCLs. (a) Microscopy images of iPSCs 21 days after gene electrotransfer. Flat-shaped cells are reprogrammed iPSCs, while sphererical cells are LCLs. Electrotransfer was performed using a triangle waverform at Vp = 12.5 V with an air pressure of 10 kPa for acceleration. Scale bar = 100 µm. (b) Image of a 12-well plates containing cultured and fixed iPSCs. Alkaline-phosphatase-stained cell colonies are depicted as black points.CC BY · PubMed Central

Verification

Claims the paper makes

0 Valid · 0 Not valid · 44 Not checked · 93 Own results and interpretations

51 more claims

Create a free account to see all 137 claims and to verify the 132 references still open.

See all claims — free account
  • It may state two separate claims as one. 57
  • Probably not a substantial claim. 36
  • The quote only partly states this claim. 1

Create a free account to see these claims.

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

Each contradiction between two passages listed here was confirmed by an independent check.

3 of 3

Citations

Cites

Research works printed but not linked: 1

  1. HS Rho, 2015, Direct delivery of reagents from a pipette tip to a PDMS microfluidic device, Chips and tips

Cited by

See all 4

Metadata

Authors
Ishii-Teshima, M., Maeda, K., Hanauchi, K., Asechi, E., Setyawan, D., Niki, T., Nakashima, K., Kurita, H., Numano, R., & Shibata, T.
Affiliations
  • Toyohashi University of Technology (JP)
  • National Institute of Technology, Sasebo College (JP)
Corresponding author
  • Ishii-Teshima, Miho
  • Shibata, Takayuki
Venue
PLOS One
Published
2025-09-26
Volume
20
Issue
9
Pages
e0333491
ISSN
1932-6203 (ISSN-L)
OpenAlex
W4414539127
Cited by
4
References
53
Citations by year
2026: 3 · 2025: 1
FWCI
1.24
Type
journal article
Language
en
Open access
Locations
  • PLoS ONE (publishedVersion, cc-by)
  • Europe PMC (PubMed Central) (publishedVersion, cc-by)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
Funders
  • Japan Society for the Promotion of Science (22J40077)
  • Adaptable and Seamless Technology Transfer Program through Target-Driven R and D (AS2915129U)
  • Naito Science and Engineering Foundation
  • The Hibi Science Foundation
Topics
  • Life Sciences › Biochemistry, Genetics and Molecular Biology › Biotechnology › Microbial Inactivation Methods
  • Life Sciences › Biochemistry, Genetics and Molecular Biology › Molecular Biology › Pluripotent Stem Cells Research
  • Life Sciences › Biochemistry, Genetics and Molecular Biology › Molecular Biology › RNA Interference and Gene Delivery
Keywords
induced pluripotent stem cells, gene electrotransfer, lymphoblastoid cell lines, iPSC reprogramming, electric field distribution, reprogramming factors
MeSH terms
Cell Line, Humans, Lymphocytes, Gene Transfer Techniques, Electroporation, Induced Pluripotent Stem Cells, Cellular Reprogramming

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

A language model drew the claims from the paper on 2026-11-21, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).

A second model labelled each claim on 2026-11-21 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.