Evidence for Endotoxin Contamination in Plastic Na+-Heparin Blood Collection Tube Lots

Newhall, Kathryn J, Diemer, Geoffrey S, Leshinsky, Natalia, Kerkof, Keith, Chute, Hilary T, Russell, Chris B, Rees, William, Welcher, Andrew A, Patterson, Scott D, Means, Gary D · 2010 · Clinical Chemistry
Cited by 28

The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.

Abstract

BACKGROUND: Biomarker assays are often conducted on whole blood samples in the course of drug development studies. Because bacterial lipopolysaccharide (LPS) (endotoxin) contamination is known to cause spontaneous cytokine production by monocytes, contamination of blood collection tubes may interfere with biomarker assay results. METHODS: Whole blood from healthy donors was collected into plastic or glass sodium (Na+)-heparin Vacutainer™ blood collection tubes and heparinized syringes. Samples were analyzed for phosphoprotein response, cytokine production, and RNA expression. Tubes were tested for endotoxin contamination by use of the limulus amoebocyte lysate assay. RESULTS: Results of phospho-flow cytometry, branched DNA (bDNA), and ELISA assays indicated that a specific lot (#5339582) of plastic Na+-heparin Vacutainer tubes was highly contaminated with an endotoxinlike substance, and contamination was confirmed by the limulus amoebocyte lysate assay. Analysis of multiple-analyte panels revealed that analytes whose changed expression was predictive of LPS stimulation were increased when whole blood was incubated in contaminated tubes for 6 or 18 h. Two additional lots of plastic tubes tested had detectable amounts of endotoxin sufficient to strongly alter phospho-flow cytometry analyses, as determined by the fold change in phosphorylation of p38 mitogen-activated protein kinase in response to tumor necrosis factor α and LPS. In contrast, 3 lots of glass tubes had substantially lower levels of spontaneous blood activation. CONCLUSIONS: Endotoxin contamination associated with tubes from 3 lots of a particular type of plastic Na+-heparin Vacutainer tube dramatically affected biomarker assay measurements. Prescreening these tubes is suggested before their use in clinical sample analysis.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1373/clinchem.2006.144618

Cites

1 of 1 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Cited by

See all 28

Metadata

Authors
Newhall, K. J., Diemer, G. S., Leshinsky, N., Kerkof, K., Chute, H. T., Russell, C. B., Rees, W., Welcher, A. A., Patterson, S. D., & Means, G. D.
Venue
Clinical Chemistry
Published
2010-09-01
Volume
56
Issue
9
Pages
1483-1491
ISSN
  • 0009-9147 (ISSN-L)
  • 1530-8561
OpenAlex
W2021224403
Cited by
28
References
1
FWCI
0.95
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.