Quantification of Hepcidin-25 in Human Serum by Isotope Dilution Micro-HPLC–Tandem Mass Spectrometry
Open accessThe DOI matches its Crossref record.
Abstract
Hepcidin-25, a liver-produced peptide hormone, was initially isolated from human urine and blood ultrafiltrate. Hepcidin-25 is thought to be the central regulator of iron metabolism (1). Iron deficiency is associated with low hepcidin-25 concentrations and anemia of chronic disease with high concentrations, but the true diagnostic value of hepcidin-25 is still under investigation. A recently published review(2) stated that only semiquantitative methods for comparative studies based on mass spectrometry (MS) have been used for the determination of hepcidin in serum and urine. Isotope-dilution MS is generally accepted as yielding high analytical specificity and accuracy. We report an isotope-dilution–micro-HPLC–tandem MS (MS/MS) method that allows the quantification of hepcidin-25 present at less than nanomole-per-liter concentrations. During sample preparation, hepcidin-25 undergoes strong nonspecific binding to surfaces. A stable isotope-labeled internal standard helps to compensate for any matrix effects. Micro-HPLC-MS/MS with monolithic capillary columns ensures highest resolution and limits of quantification (3). Native hepcidin-25 (Mr 2789) was purchased from Bachem AG. The internal standard, hepcidin-25 [DTHFPI(13C615N1)CI(13C615N1)FCCGCCHRSKCGMCCKT-OH] was synthesized with isotopically labeled isoleucines by use of the FastMoc/tBu-strategy. Renaturation and purification were performed according to a previously reported method (4). Correct folding was verified by testing bioactivity with an in vitro ferroportin internalization assay(1). Blood samples were collected and anonymized in-house according to the Roche Diagnostics policy, and informed consent was obtained from all sample donors; results were not used for regulatory purposes. Calibrators with concentrations of 0.1–100.0 nmol/L were prepared from hepcidin-25. For sample preparation we added 5 μL concentrated formic acid and 50 μL internal standard solution (1450 nmol/L in water, 0.04% acetic acid) to 45 μL human serum or calibrator. Samples were ultrafiltered with a Microcon Ultracel YM-10 filter (Millipore). The flow-through was transferred into HPLC vials. A Thermo Electron Quantum-Ultra triple-quadrupole mass spectrometer, Dionex Ultimate 3000 micro-HPLC, and PS-DVB Monolithic 200-μm internal diameter × 5 cm column (P/N 161409) were used for online micro-HPLC-MS/MS. A 2-μL sample was injected. The mobile phases were (eleunt A) 1% formic acid/0.025% trifluoroacetic acid in water and (eleunt B) 1% formic acid/0.025% trifluoroacetic acid in acetonitrile, flow rate 3 μL/min, with a linear gradient from 0% to 80% eleunt B during 7 min and then held at 80% eleunt B until minute 11. We used microelectrospray ionization in the positive mode; recorded selected reaction-monitoring transitions were m/z 930.8→ 1145.5 and m/z 935.5→1152.6, collision energy 33 V, and argon collision gas 2.0 mTorr. Carryover was excluded by blank injections. Samples were kept in long-term storage at −80 °C and were stable for at least 2 days at 6 °C. Processed samples were stable for 48 h at 6 °C. We observed no ion suppression attributable to changing elution conditions and monitoring-signal intensities; postcolumn infusion experiments are not compatible with the used microflow ion-source and could not be performed. We verified linearity and recovery by using spiked human serum with 7 concentrations from 0.1 to 80 nmoL/L. A correlation coefficient (r2) of 0.9968 was observed between expected and observed hepcidin-25, and recoveries of 95%–100% were calculated. To determine the total imprecision of the method we used 3 native samples containing 3 different concentrations, measured in triplicate on 6 different days. The CVs were 21% (0.3 nmol/L), 17% (1.2 nmol/L), and 4.8% (5.9 nmol/L), with a limit of quantification of 0.3 nmol/L. At a concentration of 0.1 nmol/L the signal:noise ratio was 10:1. With samples from 5 of the donors, no differences in the measured values were observed for results from 6 sampling devices (BD Serum SST II, BD Serum Z, BD Serum SST, Sarstedt Serum, Sarstedt Serum Gel, and Sarstedt Serum EDTA). Hepcidin-25 concentrations from 10 apparently healthy fasting persons drawn in the morning on 4 consecutive days were measured (Table 1 ). Observed hepcidin-25 concentrations were in the lower nanomole-per-liter range. Concentrations were higher in males than females. Day-to-day variation of hepcidin-25 values (nmol/L) for serum samples colected in the morning from apparently healthy fasting individuals. F, female; M, male. Day-to-day variation of hepcidin-25 values (nmol/L) for serum samples colected in the morning from apparently healthy fasting individuals. F, female; M, male. The concentrations we observed were in agreement with results of the ferroportin internalization assay. Ferroportin internalization was detected at concentrations below 1 nmol/L hepcidin-25, and 50% internalization occurred at 4.6 nmol/L. These results indicate that the measured hep
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Hepcidin in Human Iron Disorders: Diagnostic Implications
Kroot, J. J., Tjalsma, H., Fleming, R. E., & Swinkels, D. W. (2011) · Clinical Chemistry · doi:10.1373/clinchem.2009.140053 · Cited by 246
“HPLC/UPLC MS/MS (87;88) c-ELISA (13;29;98;99) (Bachem) Q-TOF LC-MS (92)”
A fully human anti-hepcidin antibody modulates iron metabolism in both mice and nonhuman primates
Cooke, K. S., Hinkle, B., Salimi-Moosavi, H., Foltz, I., King, C., Rathanaswami, P., Winters, A., Steavenson, S., Begley, C. G., Molineux, G., et al. (2013) · Blood · doi:10.1182/blood-2013-06-505792 · Cited by 113
Isaacs, J. D., Harari, O., Kobold, U., Lee, J. S., & Bernasconi, C. (2013) · Arthritis Research & Therapy · doi:10.1186/ar4397 · Cited by 105
“Serum hepcidin was measured by high-performance liquid chromatography–tandem mass spectroscopy assay [18].”
MethodSelf-citationSofroniadou, S., & Goldsmith, D. (2011) · Drug Safety · doi:10.2165/11585040-000000000-00000 · Cited by 60
Sonnweber, T., Theurl, I., Seifert, M., Schroll, A., Eder, S., Mayer, G., & Weiss, G. (2011) · Nephrology Dialysis Transplantation · doi:10.1093/ndt/gfq483 · Cited by 55
Metadata
- Authors
- Kobold, U., Dülffer, T., Dangl, M., Escherich, A., Kubbies, M., Röddiger, R., & Wright, J. A.
- Venue
- Clinical Chemistry
- Publisher
- Oxford University Press (OUP)
- Published
- 2008-09-01
- Volume
- 54
- Issue
- 9
- Pages
- 1584-1586
- ISSN
- 0009-9147 (ISSN-L)
- 1530-8561
- OpenAlex
- W2123576922
- Cited by
- 58
- References
- 0
- FWCI
- 6.87
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Quantification of Hepcidin-25 in Human Serum by Isotope Dilution Micro-HPLC–Tandem Mass Spectrometry. https://app.cochat.ai/papers/10.1373/clinchem.2008.107029
- The paper
- Kobold, U., Dülffer, T., Dangl, M., Escherich, A., Kubbies, M., Röddiger, R., & Wright, J. A. (2008). Quantification of Hepcidin-25 in Human Serum by Isotope Dilution Micro-HPLC–Tandem Mass Spectrometry. Clinical Chemistry. https://doi.org/10.1373/clinchem.2008.107029
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.