Seven Direct Methods for Measuring HDL and LDL Cholesterol Compared with Ultracentrifugation Reference Measurement Procedures
Open accessAbstract
Background: Methods from 7 manufacturers and 1 distributor for directly measuring HDL cholesterol (C) and LDL-C were evaluated for imprecision, trueness, total error, and specificity in nonfrozen serum samples. Methods: We performed each direct method according to the manufacturer’s instructions, using a Roche/Hitachi 917 analyzer, and compared the results with those obtained with reference measurement procedures for HDL-C and LDL-C. Imprecision was estimated for 35 runs performed with frozen pooled serum specimens and triplicate measurements on each individual sample. Sera from 37 individuals without disease and 138 with disease (primarily dyslipidemic and cardiovascular) were measured by each method. Trueness and total error were evaluated from the difference between the direct methods and reference measurement procedures. Specificity was evaluated from the dispersion in differences observed. Results: Imprecision data based on 4 frozen serum pools showed total CVs <3.7% for HDL-C and <4.4% for LDL-C. Bias for the nondiseased group ranged from −5.4% to 4.8% for HDL-C and from −6.8% to 1.1% for LDL-C, and for the diseased group from −8.6% to 8.8% for HDL-C and from −11.8% to 4.1% for LDL-C. Total error for the nondiseased group ranged from −13.4% to 13.6% for HDL-C and from −13.3% to 13.5% for LDL-C, and for the diseased group from −19.8% to 36.3% for HDL-C and from −26.6% to 31.9% for LDL-C. Conclusions: Six of 8 HDL-C and 5 of 8 LDL-C direct methods met the National Cholesterol Education Program total error goals for nondiseased individuals. All the methods failed to meet these goals for diseased individuals, however, because of lack of specificity toward abnormal lipoproteins.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Most discordances in dyslipidemic samples are observed at lower concentration ranges of LDLC (< 1.8 mmol/L) and HDLC (< 1.0 mmol/L) [42].”
From Quantifying atherogenic lipoproteins for lipid-lowering strategies: Consensus-based recommendations from EAS and EFLM (2020) · cites this paper as [42]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Direct measurements of dLDLC and dHDLC in normolipidemic samples usually meet the National Cholesterol Education Program (NCEP) total error goals of ≤12% and ≤13%, respectively, but total error ranged from -26% to +32% for dLDLC and -20% to +36% for dHDLC in a comprehensive study of different assays in dyslipidemic samples [42].”
From Quantifying atherogenic lipoproteins for lipid-lowering strategies: Consensus-based recommendations from EAS and EFLM (2020) · cites this paper as [42]
Experimental — accuracy not yet measuredIn the citing paper
“It is important to note that direct LDL-C measurements also have limitations, including systematic bias and inaccuracy in patients with dyslipidaemia, especially for high TG levels [88-90].”
From 2019 ESC/EAS guidelines for the management of dyslipidaemias: lipid modification to reduce cardiovascular risk (2019) · cites this paper as [89]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Grundy, S. M., Bilheimer, D. W., Chait, A., Clark, L. T., Denke, M. A., Havel, R. J., Hazzard, W. R., Hulley, S. B., Hunninghake, D. B., Kreisberg, R. A., et al. (1993) · JAMA · doi:10.1001/jama.269.23.3015 · Cited by 27,944
Expert Panel on Detection, E. (2001) · JAMA: The Journal of the American Medical Association · doi:10.1001/jama.285.19.2486 · Cited by 26,011
Abell, L. L., Levy, B. B., Brodie, B. B., & Kendall, F. E. (1952) · Journal of Biological Chemistry · doi:10.1016/s0021-9258(19)50907-3 · Cited by 3,085
Nissen, S. E., Nicholls, S. J., Sipahi, I., Libby, P., Raichlen, J. S., Ballantyne, C. M., Davignon, J., Erbel, R., Fruchart, J. C., Tardif, J.-C., et al. (2006) · JAMA · doi:10.1001/jama.295.13.jpc60002 · Cited by 2,047
National Cholesterol Education Program
Lenfant, C. (1991) · Archives of Internal Medicine · doi:10.1001/archinte.1991.00400060019005 · Cited by 1,188
Lipid screening and cardiovascular health in childhood
Daniels, S. R., Greer, F. R., & and the Committee on Nutrition (2008) · Pediatrics · doi:10.1542/peds.2008-1349 · Cited by 1,178
Lowe, C. U., Coursin, D. B., Heald, F. P., Holliday, M. A., OʼBrien, D., Owen, G. M., Pearson, H. A., Scriver, C. R., Filer, L. J., & Kline, O. L. (1967) · PEDIATRICS · doi:10.1542/peds.40.1.130 · Cited by 478
Nauck, M., Warnick, G. R., & Rifai, N. (2002) · Clinical Chemistry · doi:10.1093/clinchem/48.2.236 · Cited by 442
Self-citationHandbook of Lipoprotein Testing
Rifai, N., Warnick, G. R., & Dominiczak, M. H. (2001) · Medical Entomology and Zoology · Cited by 384
Self-citationWarnick, G. R., Nauck, M., & Rifai, N. (2001) · Clinical Chemistry · doi:10.1093/clinchem/47.9.1579 · Cited by 228
Self-citation
Research works printed but not linked: 4
- 2009, Clin Chem, 55, 88
- 1999, Preparation and validation of commutable frozen human serum pools as secondary reference materials for cholesterol measurement procedures; approved guideline. CLSI document C37-A
- 1990, Scand J Clin Lab Invest Suppl, 198, 9
- 1999, Am J Cardiol, 83, 3F
Cited by
Sorted by citations
Items 201–300 of 383
Martin’s, Friedewald’s and Cordova’s formulas compared to LDL-C directly measured in Southern Brazil
de Córdova, C. M. M., Portal, A. S., & de Cordova, M. M. (2020) · Jornal Brasileiro de Patologia e Medicina Laboratorial · doi:10.5935/1676-2444.20200003 · Cited by 8
“All 10,664 135 ± 41 (24-423) 146 ± 40 (-133-377) 131 ± 34 (32-370) 144 ± 40 (2-376) LDL-C: low-density lipoprotein cholesterol; TG: triglycerides; TC: total cholesterol; (1)applied to samples with triglycerides ≤ 400 mg/dl.”
MethodShort-Term Effects of Screening for Cardiovascular Risk in the Deaf Community: A Pilot Study
Patel, J. V., Gill, P. S., Chackathayil, J., Ojukwu, H., Stemman, P., Sheldon, L., Meelu, S., Lane, D. A., Tracey, I., Lip, G. Y. H., et al. (2011) · Cardiology Research and Practice · doi:10.4061/2011/493546 · Cited by 8
“There are increasing concerns raised with respect to serum HDL cholesterol determination using direct measurement methods [16].”
BackgroundIdoko, A. (2018) · Universal Journal of Pharmaceutical Research · doi:10.22270/ujpr.v3i1.r7 · Cited by 8
“Twenty (20) rats used for the study were grouped into four groups of five (5) rats each.”
MethodNakajima, K., Higuchi, R., Iwane, T., Shibata, M., Takada, K., Sugiyama, M., Matsuda, M., & Nakamura, T. (2019) · Journal of Clinical Medicine · doi:10.3390/jcm8030381 · Cited by 8
“precipitation method) [29,30], and the remainder were measured using other methods, following”
MethodSelf-citationKenkre, J. S., Mazaheri, T., Neely, R. D. G., Soran, H., Datta, D. B. N., Penson, P. E., Downie, P., Yates, A. M., Hayden, K., Patel, M., et al. (2025) · Annals of Clinical Biochemistry International Journal of Laboratory Medicine · doi:10.1177/00045632251315303 · Cited by 8
“70 Although it has a possible role to quantify LDL-c with signi fi cant hypertriglyceridaemia (>9.0 mmol/L), when direct LDL-c measurement was compared to the CDC reference method, marked and clinically relevant bias was seen.”
InfluentialCardiovascular disease lipids and lipoproteins biomarker standardization
Lyle, A. N., Danilenko, U., Sugahara, O., & Vesper, H. W. (2025) · Critical Reviews in Clinical Laboratory Sciences · doi:10.1080/10408363.2025.2462817 · Cited by 7
“Freeze-thaw can affect LDL-C measurements and can lead to incorrect assessments of measurement bias [74].”
Stress, Cortisol, and Lipid Profiles Among Rwandan Undergraduate Students: A Cross-Sectional Study
Mushumba, P., Uwineza, D. N., Nsanzimana, V., Mapira, H. T., Gori, E., & Musarurwa, C. (2025) · Risk Management and Healthcare Policy · doi:10.2147/rmhp.s518801 · Cited by 9
Evaluation of current indirect methods for measuring LDL-cholesterol
Drobnik, S., Scharnagl, H., Samani, N. J., Braund, P. S., Nelson, C. P., Hollstein, T., Kassner, U., Dressel, A., Drobnik, W., & März, W. (2025) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2025-0024 · Cited by 7
Naser, A., Isgandarov, K., Güvenç, T. S., Güvenç, R. Ç., & Şahin, M. (2022) · Arquivos Brasileiros de Cardiologia · doi:10.36660/abc.20210641 · Cited by 7
Heidemann, B. E., Koopal, C., van Lennep, J. E. R., Stroes, E. S. G., Riksen, N. P., Mulder, M. T., van der Zee, L. C. V. V. –., Blackhurst, D., Visseren, F. L. J., & Marais, A. D. (2022) · Clinica Chimica Acta · doi:10.1016/j.cca.2022.11.035 · Cited by 7
Ephraim, R. K. D., Ativi, E., Ashie, S. A., Abaka‐Yawson, A., & Darkwah, K. O. (2023) · Bulletin of the National Research Centre/Bulletin of the National Research Center · doi:10.1186/s42269-023-01046-5 · Cited by 7
“Direct measurements of LDL-c are still costly, time-consuming, not fully standardized and not universally available (Nauck et al. 2002; Miller et al. 2010; Langlois et al. 2018; Piani et al. 2021).”
BackgroundNon‐high‐density lipoprotein cholesterol as a cardiovascular risk screening tool in children
Miyazaki, A., Oguri, A., & Ichida, F. (2015) · Pediatrics International · doi:10.1111/ped.12891 · Cited by 7
Teramoto, T. (2013) · The Japanese Journal of Nutrition and Dietetics · doi:10.5264/eiyogakuzashi.71.3 · Cited by 7
Chang, Y., Lin, K.-C., Chang, D., Hsieh, C., & Yau-Jiunn, L. (2013) · The Review of Diabetic Studies · doi:10.1900/rds.2013.10.213 · Cited by 7
“Serum HDL-C was measured by direct enzymatic methods using commercial kits (Cholestest N-HDL, Sekisui Medical, Tokyo, Japan) with less than 3% intra-assay and inter-assay coefficients of variance [22-24].”
MethodKim, J., Cho, Y., Lee, S., & Yun, Y. (2019) · Imsang byeong'ri wa jeongdo gwanri/Journal of laboratory medicine and quality assurance · doi:10.15263/jlmqa.2019.41.3.121 · Cited by 7
Schneider, E. E., Sarkar, S., Margolick, J. B., Martin, S. S., Post, W. S., & Brown, T. T. (2019) · AIDS Research and Human Retroviruses · doi:10.1089/aid.2019.0220 · Cited by 7
“Other methods of direct LDL-C measurement have been developed, but these have several disadvantages including lack of specificity toward abnormal lipoproteins, variability in comparison against ultracentrifugation, and are discordant with other measures of atherogenic burden, including apoB and non-high density lipopr…”
Reijnders, E., Bossuyt, P. M., Jukema, J. W., Ruhaak, L. R., Romijn, F. P. H. T. M., Szarek, M. J., Trompet, S., Bhatt, D. L., Bittner, V., Díaz, R., et al. (2025) · Arteriosclerosis Thrombosis and Vascular Biology · doi:10.1161/atvbaha.124.322336 · Cited by 6
Second-line treatments for dyslipidemia in patients at risk of cardiovascular disease
Kondo, Y., Hamai, J., Nezu, U., Shigematsu, E., Kamiko, K., Yamazaki, S., Yoshii, T., Takahashi, M. M., Takano, T., Kawasaki, S., et al. (2014) · Endocrine Journal · doi:10.1507/endocrj.ej13-0404 · Cited by 6
Shi, B., Wang, H., Yin, D., Zhu, C., Feng, L., Wang, H., Jia, L., Fu, R., Song, C., Zhou, Z., et al. (2022) · Journal of Cardiovascular Development and Disease · doi:10.3390/jcdd9100342 · Cited by 7
Sun, C., McCudden, C. R., Brisson, D., Shaw, J. L. V., Gaudet, D., & Ooi, T. C. (2022) · Canadian Journal of Cardiology · doi:10.1016/j.cjca.2022.12.025 · Cited by 6
Rasaei, N., Fallah, M., Nemati, M., Gholami, F., Ghaffarian-Ensaf, R., & Mirzaei, K. (2023) · Scientific Reports · doi:10.1038/s41598-023-38988-9 · Cited by 7
Chung, S. (2023) · Journal of Clinical Laboratory Analysis · doi:10.1002/jcla.24963 · Cited by 5
Zhang, K., Wei, C., Shao, Y., Li, W., Zhao, Z., Song, Y., Tang, X., Li, Y., & Gou, Z. (2024) · Heliyon · doi:10.1016/j.heliyon.2024.e28155 · Cited by 7
“Examples of such conditions include diabetes and chronic renal disease [44].”
BackgroundTransitional Medicine of Intractable Primary Dyslipidemias in Japan
Ogura, M., Okazaki, S., Okazaki, H., Tada, H., Dobashi, K., Nakamura, K., Matsunaga, K., Miida, T., Minamino, T., Yokoyama, S., et al. (2024) · Journal of Atherosclerosis and Thrombosis · doi:10.5551/jat.rv22016 · Cited by 6
“The frequency It should be noted that some patients with LCAT deficiency have serum HDL-C levels > 20 mg/dL 43 , 46) .”
BackgroundLanglois, M. R., Delanghe, J., Buyzere, M. D., Rietzschel, E. R., & Bacquer, D. D. (2012) · Atherosclerosis · doi:10.1016/j.atherosclerosis.2012.10.036 · Cited by 6
Cholesterol levels of Japanese dyslipidaemic patients with various comorbidities: BioBank Japan
Yokomichi, H., Noda, H., Nagai, A., Hirata, M., Tamakoshi, A., Kamatani, Y., Kiyohara, Y., Matsuda, K., Muto, K., Ninomiya, T., et al. (2017) · Journal of Epidemiology · doi:10.1016/j.je.2016.12.014 · Cited by 6
Osegbe, I., Ugonabo, M. C., Chukwuka, C., Meka, I. A., & Nwosu, N. I. (2017) · Journal of Laboratory Physicians · doi:10.4103/0974-2727.199632 · Cited by 6
“Homogeneous assays which do not require pretreatment have been developed by manufacturers for routine determination of dLDLC,[12] albeit costly as well.”
BackgroundTakashima, M., Tanaka, W., Matsuyama, H., Tajiri, H., & Sakakibara, H. (2021) · Nutrients · doi:10.3390/nu13041242 · Cited by 6
“reported that HDL-cholesterol levels resulted in more than 15% differences, when analyzing with eight kinds of commercial kits [39].”
BackgroundBiochemistry laboratories should routinely report non-HDL-cholesterol
Preiss, D., & Neely, D. (2015) · Annals of Clinical Biochemistry International Journal of Laboratory Medicine · doi:10.1177/0004563215594818 · Cited by 6
Influence of remnant cholesterol on direct LDL and HDL cholesterol measurement accuracy
Kwon, S. S., Jang, H., Rim, J. H., Cho, J., Seo, J. D., Yun, Y., Park, Y. M., Kwon, G. C., Cho, C.-I., Lee, S., et al. (2025) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2025-0593 · Cited by 5
“Factors such as dyslipidemia, high TG levels, low HDL-C levels, and paraproteins can contribute to measurement bias [21, 29, 30].”
Increased Cardiovascular Risk in Young Patients with CKD and the Role of Lipid-Lowering Therapy
Kratky, V., Valerianova, A., Hruskova, Z., Tesař, V., & Malík, J. (2024) · Current Atherosclerosis Reports · doi:10.1007/s11883-024-01191-w · Cited by 5
“The situation is also complicated by the changing volume status of CKD patients (hypo-or hypervolemia with concentration or dilution of lipoproteins) and by frequent analytical interferences [23].”
Concordance between LDL-C estimated by various formulas and directly measured LDL-C
David-Pardo, D. G., Ruiz, Á. J., Múñoz, Ó., Peña, Á. A. G., García, D. X. S., & Arcila, J. (2024) · Journal of clinical lipidology · doi:10.1016/j.jacl.2024.08.009 · Cited by 5
Steyn, N., Rossouw, H. M., Martins, J., & Pillay, T. S. (2023) · Clinica Chimica Acta · doi:10.1016/j.cca.2023.117373 · Cited by 5
Cemin, R., Casablanca, S., Ermacora, D., & Daves, M. (2023) · Journal of Clinical Medicine · doi:10.3390/jcm12134422 · Cited by 5
Solnica, B., Sygitowicz, G., Sitkiewicz, D., Jozwiak, J. J., Kasperczyk, S., Broncel, M., Wolska, A. T., Odrowąż‐Sypniewska, G., & Banach, M. (2024) · Diagnostyka Laboratoryjna · doi:10.5604/01.3001.0054.4954 · Cited by 7
Determination of cholesterol in blood. Part 3
Buzanovskii, V. A. (2016) · Review Journal of Chemistry · doi:10.1134/s2079978016030018 · Cited by 5
Khan, S. H., Niazi, N. K., Sobia, F., Fazal, N., Manzoor, S. M., & Nadeem, A. J. (2019) · Pakistan Journal of Medical Sciences · doi:10.12669/pjms.35.2.679 · Cited by 5
Manita, D., Yoshida, H., Koyama, I., Nakamura, M., & Hirowatari, Y. (2020) · The Journal of Applied Laboratory Medicine · doi:10.1093/jalm/jfaa144 · Cited by 5
Self-citationShi, B., Wang, H., Liu, J., Cai, Z., Song, C., Jia, L., Yin, D., Wang, H., Dou, K., & Song, W. (2022) · Frontiers in Cardiovascular Medicine · doi:10.3389/fcvm.2022.932878 · Cited by 5
“The accuracy of these dLDLC methods, especially when employed in individuals with severe dyslipidemia, remains controversial (15, 16).”
BackgroundMethodChallenges in Routine Clinical Chemistry Testing
Sepulveda, J. L. (2013) · Elsevier eBooks · doi:10.1016/b978-0-12-415783-5.00008-6 · Cited by 5
Opinion paper: Is the Friedewald equation obsolete?
Jialal, I., & Winter, W. E. (2020) · Clinica Chimica Acta · doi:10.1016/j.cca.2020.12.032 · Cited by 5
Farnier, M., Dong, Q., Shah, A. K., Johnson‐Levonas, A. O., & Brudi, P. (2011) · Lipids in Health and Disease · doi:10.1186/1476-511x-10-212 · Cited by 5
Mendieta, E. J. L., & Timón-Zapata, J. (2013) · Annals of Clinical Biochemistry International Journal of Laboratory Medicine · doi:10.1177/0004563213489965 · Cited by 5
“Since homogeneous assays became available, variability in both LDL-c and HDL-c reagents has been reported.(3,4) This heterogeneity could be explained by the distinct determination principles and the different reactivities to lipoproteins, especially towards various forms of modified LDL particles.”
BackgroundReporting LDL-cholesterol levels in the era of intensive lipid management: a clarion call
Jialal, I., & Gounden, V. (2017) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2017-0639 · Cited by 5
Carelse, A., Rossouw, H. M., Steyn, N., Martins, J., & Pillay, T. S. (2023) · Journal of Clinical Pathology · doi:10.1136/jcp-2023-208916 · Cited by 4
Liang, S., Lang, X., Li, X., Zhang, H., Zhao, Y., Tao, C., & Gan, W. (2025) · Clinica Chimica Acta · doi:10.1016/j.cca.2025.120607 · Cited by 5
New joint consensus initiative on quantifying atherogenic lipoproteins
Stock, J. K. (2018) · Atherosclerosis · doi:10.1016/j.atherosclerosis.2018.08.037 · Cited by 4
Labordiagnostik von Fettstoffwechselstörungen
März, W., Scharnagl, H., Kleber, M. E., Silbernagel, G., Nauck, M., Müller‐Wieland, D., & von Eckardstein, A. (2023) · DMW - Deutsche Medizinische Wochenschrift · doi:10.1055/a-1516-2511 · Cited by 5
“Vorsicht ist auch bei der Interpretation von Lipidbefunden geboten, wenn atypische Lipoproteine vorliegen (HTG), Typ III-HLP, Leber-und Nierenerkrankungen [31, 32].”
BackgroundPleiotropic functions and clinical importance of circulating HDL-PON1 complex
Mahrooz, A. (2024) · Advances in clinical chemistry · doi:10.1016/bs.acc.2024.04.003 · Cited by 4
Aldana, S. I., Demateis, D., Valvi, D., Just, A. C., Gutiérrez-Ávila, I., Estrada‐Gutiérrez, G., Téllez‐Rojo, M. M., Wright, R. O., Baccarelli, A., Wu, H., et al. (2024) · Environmental Research · doi:10.1016/j.envres.2025.121198 · Cited by 4
Fatigued with Friedewald: why isn't everyone onboard yet with the new LDL-C equations?
Narasimhan, M., Cao, J., Meeusen, J. W., Remaley, A. T., Martin, S. S., & Muthukumar, A. (2025) · Frontiers in Cardiovascular Medicine · doi:10.3389/fcvm.2025.1534460 · Cited by 4
Self-citationTimón-Zapata, J., Mendieta, E. J. L., Sáenz-Mateos, L. F., Ruiz-Trujillo, L., Fernández, A. A., Palomino-Muñoz, T. J., Loeches-Jiménez, M. P., & Gómez-Serranillos, M. (2014) · Clinica Chimica Acta · doi:10.1016/j.cca.2013.12.043 · Cited by 4
Farmanbar, R., Zeidi, I. M., & Afkar, A. (2018) · Iranian Red Crescent Medical Journal · doi:10.5812/ircmj.64581 · Cited by 4
Performance of HDL-C measurements assessed by a 4-year trueness-based EQA/PT program in China
Zhou, W., Luo, W., Yu, S., Li, H., Wang, D., Zhang, J., Wang, S., Zeng, J., Zhang, C., Zhao, H., et al. (2022) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2020-0658 · Cited by 4
Triglyceride-rich lipoprotein cholesterol and cardiovascular risk
Bilgic, S., Remaley, A. T., & Sniderman, A. D. (2023) · Current Opinion in Lipidology · doi:10.1097/mol.0000000000000905 · Cited by 5
Self-citationGomes, D., Paiva, M. S., Freitas, P., Albuquerque, F., Lima, M. R., Santos, R. R., Presume, J., Trabulo, M., Aguíar, C., Ferreira, J., et al. (2024) · Arquivos Brasileiros de Cardiologia · doi:10.36660/abc.20230242 · Cited by 4
Yoshida, H. (2017) · Journal of Atherosclerosis and Thrombosis · doi:10.5551/jat.ed076 · Cited by 4
Jialal, I. (2014) · American Journal of Clinical Pathology · doi:10.1309/ajcped9x7rnmyaeq · Cited by 4
Integrated Measure for Atherogenic Lipoproteins in the Modern Era
Rosenson, R. S., Hegele, R. A., & Gotto, A. M. (2016) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2015.10.059 · Cited by 4
Gcingca, T. A., Sampson, M. L., Zubirán, R., Wolska, A. T., Meeusen, J. W., Donato, L. J., Jaffe, A. S., & Remaley, A. T. (2025) · The Journal of Applied Laboratory Medicine · doi:10.1093/jalm/jfaf093 · Cited by 3
Self-citationCocco, C., Siotto, M., Guerrini, A., Germanotta, M., Falchini, F., Cipollini, V., Cortellini, L., Pavan, A., Lattanzi, S., Insalaco, S., et al. (2025) · Neurology International · doi:10.3390/neurolint17120193 · Cited by 3
“Total cholesterol was measured by means of oxidation from a cholesteroxidase to cholest-4-en-3-one [43] while direct HDL cholesterol was assessed by a transformation of the HDL portion into a quinone derivative [44]; triglycerides were measured by a peroxidase-coupled method [45].”
Kamrani, F., Imannezhad, M., Rezaee, A. M., Darroudi, S., Kondori, A. Y., Sadatian, S., Shahri, B., Esmaily, H., Bakavoli, A. H., Ghayour‐Mobarhan, M., et al. (2025) · Scientific Reports · doi:10.1038/s41598-025-27681-8 · Cited by 3
“Even in healthy individuals, calculated LDL-C may vary by 13.3–13.5% 7 .”
InfluentialFarheen, F., Ambiger, S., Jaalam, K., & Javali, S. B. (2023) · Journal of Laboratory Physicians · doi:10.1055/s-0043-1768951 · Cited by 3
Global Standardization of Lipid/Lipoprotein Testing
Miida, T. (2024) · Juntendō Igaku/Juntendo igaku · doi:10.14789/jmj.jmj24-0019-p · Cited by 3
Györfi, I., Oprea, O. R., Mănescu, I. B., Curici, A., & Dobreanu, M. (2025) · Diagnostics · doi:10.3390/diagnostics15182298 · Cited by 3
“However, most dLDL-C methods are standardized and traceable to beta-quantification, which makes them a reliable alternative for routine use in clinical laboratories [1,20–22].”
Pillay, T. S., van Deventer, B. S., Gwiliza, S., van Niekerk, C., & Twomey, P. J. (2025) · Clinical Chemistry · doi:10.1093/clinchem/hvaf118 · Cited by 3
Chapter 8 – Challenges in Routine Clinical Chemistry Testing: Analysis of Small Molecules
Sepulveda, J. L. (2013) · Cited by 3
Is data mining approach a best fit formula for estimation of low-density lipoprotein cholesterol?
Sarangi, R., Bahinipati, J., Pathak, M., & Mahapatra, S. (2021) · Journal of Family Medicine and Primary Care · doi:10.4103/jfmpc.jfmpc_1734_20 · Cited by 3
“Journal of Family Medicine and Primary Care 328 Volume 10 : Issue 1 : January 2021 lipoproteins influences the ability of a direct method to specifically measure the cholesterol contents of one lipoprotein class in presence of other types of lipoprotein.[6] Sometimes direct LDL value is overestimated and even some dir…”
BackgroundCalculated values of serum LDL-cholesterol (LDL-C) – for better or worse?
Ramasamy, J., Job, V., Mani, T., & Jacob, M. (2021) · Nutrition Metabolism and Cardiovascular Diseases · doi:10.1016/j.numecd.2021.01.016 · Cited by 3
Tan, H. T., Yong, S., Liu, H., Liu, Q., Teo, T. L., & Sethi, S. K. (2023) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2022-1301 · Cited by 3
The necessity for improving lipid testing reagents: A real world study
Zheng, C., Zhou, W., Zhou, R., Zhang, H., Liang, S., Zhang, W., Li, G., Zhang, C., & Gan, W. (2023) · Clinica Chimica Acta · doi:10.1016/j.cca.2023.117529 · Cited by 3
Kilpatrick, E. S., Kallner, A., Atkin, S. L., & Sathyapalan, T. (2024) · Annals of Clinical Biochemistry International Journal of Laboratory Medicine · doi:10.1177/00045632241305936 · Cited by 3
Gupta, H., Schiros, C. G., Sharifov, O. F., Jain, A., & Denney, T. S. (2016) · BMC Cardiovascular Disorders · doi:10.1186/s12872-016-0352-x · Cited by 3
“These operating characteristics may not hold true for many commercial assays [23].”
BackgroundCalculated low density lipoprotein cholesterol: Its relevance in Indian perspective
Nigam, P. K. (2017) · Indian Heart Journal · doi:10.1016/j.ihj.2017.01.007 · Cited by 3
“Fourthly, for the equations incorporating HDLc the procedure for HDLc(direct/precipitation) should also be taken into consideration as bias have been reported even in the directly assayedHDLc values by different direct assays as compared to reference HDLc values.(4,15)”
ResultOhba, K., Oki, Y., Fujita, K., Kameko, F., Moriyama, T., Horiike, Y., Morita, H., Matsushita, A., Iino, K., Sasaki, S., et al. (2013) · Clinica Chimica Acta · doi:10.1016/j.cca.2013.09.003 · Cited by 3
Gerique, J. A. G., Ballesteros, B. M., Sardina, R. G., Salán, M. E., Romero, F. F., Puzo, J., Doreste, J. Á. A., Calmarza, P., Vázquez, M. D., & Romero, C. (2011) · Revista del Laboratorio Clínico · doi:10.1016/j.labcli.2011.09.005 · Cited by 3
Immunoprecipitation of apolipoprotein B-containing lipoproteins for isolation of HDL particles
Contois, J. H., Albert, A. L., & Nguyen, R.-A. (2014) · Clinica Chimica Acta · doi:10.1016/j.cca.2014.06.017 · Cited by 3
Seki, R., Inoue, K., Yamamoto, S., & Akimoto, K. (2017) · Biomedical Research and Clinical Practice · doi:10.15761/brcp.1000141 · Cited by 3
“Further, direct homogeneous methods have been shown to yield inconsistent data when IDL is elevated [28].”
BackgroundEngineering of a miniaturized, robotic clinical laboratory
Nourse, M. B., Engel, K., Anekal, S. G., Bailey, J. A., Bhatta, P., Bhave, D. P., Chandrasekaran, S., Chen, Y., Chow, S., Das, U., et al. (2017) · Bioengineering & Translational Medicine · doi:10.1002/btm2.10084 · Cited by 3
Pallazola, V. A., Quispe, R., Elshazly, M. B., Vakil, R. M., Sathiyakumar, V., Jones, S. R., & Martin, S. S. (2018) · Current Cardiovascular Risk Reports · doi:10.1007/s12170-018-0590-9 · Cited by 3
Agha, A. M., & Ballantyne, C. M. (2022) · Contemporary cardiology · doi:10.1007/978-3-030-98824-1_4 · Cited by 3
A failure of standardization or a failure of the process of standardization
Sniderman, A. D., & McQueen, M. (2018) · Journal of clinical lipidology · doi:10.1016/j.jacl.2018.05.009 · Cited by 3
Calculating LDL cholesterol in familial combined hyperlipidemia: Out with the old, in with the new?
Martin, S. S. (2018) · Atherosclerosis · doi:10.1016/j.atherosclerosis.2018.07.034 · Cited by 3
Lee, J. S., Lee, H., Cha, H., Seok, J., & Jeong, I. C. (2025) · PLoS ONE · doi:10.1371/journal.pone.0327169 · Cited by 2
“Although a previous study [39] reported that the Sekisui assay failed to meet the NCEP–ATP III total allowable error (TAE) criterion of ±12% in certain diseased populations, more recent evidence from a multicenter study by Miida et al. [40] demonstrated that the assay satisfied the NCEP–ATP III…”
Kathariya, G., Aggarwal, J., Nandal, R., & Reddy, S. (2025) · Heart India · doi:10.4103/heartindia.heartindia_77_25 · Cited by 2
Kuklenyik, Z., Ivanova, A. A., Drinkard, L. E., Schieltz, D. M., Jones, J. I., Bierbaum, K., Toth, C. A., Gardner, M. S., Parks, B. A., Andrews, M., et al. (2026) · Analytical and Bioanalytical Chemistry · doi:10.1007/s00216-026-06390-9 · Cited by 2
Meikle, T. G., Wu, J., Wang, T., Dakic, A., Yi, C., Mellett, N. A., van Buuren-Milne, M., Beyene, H. B., Cinel, M. A., Duong, T., et al. (2026) · Analytical Chemistry · doi:10.1021/acs.analchem.5c04984 · Cited by 2
Diederiks, N. M., Romijn, F. P. H. T. M., van Neer, N. J. M., Pieterse, M., Dittrich, J., Ceglarek, U., Cobbaert, C. M., Ruhaak, L. R., the IFCC Working Group for Standardization of Apolipoproteins by Mass Spectrometry, Ruhaak, L. R., et al. (2026) · Clinical Chemistry · doi:10.1093/clinchem/hvag041 · Cited by 2
Non-HDL Estimation Methods: Advancing Cardiovascular Disease Prediction
Akbar, A. M., Jha, D., Ali, H., & Ahmad, T. (2023) · Indian Journal of Pharmacy Practice · doi:10.5530/ijopp.16.4.52 · Cited by 2
Which LDL Value Should Clinicians Look at?
Malve, H., & Asalkar, A. (2021) · Journal of Laboratory Physicians · doi:10.1055/s-0041-1729131 · Cited by 2
Lipid and Lipoprotein Abnormalities in Chronic Renal Insufficiency: Review
Rácz, O., Gaško, R., & Klimov, E. (2011) · InTech eBooks · doi:10.5772/22122 · Cited by 2
“Another unsolved problem is that the assay methods from different providers are not yet harmonized at all (Fuentes-Arderiu et al, 2009, Miller et al, 2010).”
BackgroundThe New ACC/AHA Cardiovascular Risk Guidelines: Impact and Controversies
Remaley, A. T., Shamburek, R. D., & Amar, M. J. (2014) · Clinical Chemistry · doi:10.1373/clinchem.2014.224964 · Cited by 2
Self-citationA Critical Review of LDL Cholesterol and HDL Cholesterol Measurement
Contois, J. H. (2012) · Cited by 2
Prevalence of Elevated Serum Homocysteine and Serum Lipoprotein ‘a’ in Women
Dhat, V. (2014) · JOURNAL OF CLINICAL AND DIAGNOSTIC RESEARCH · doi:10.7860/jcdr/2014/9607.4995 · Cited by 2
Le point sur les différentes méthodes pour déterminer le cholestérol-LDL
Duvillard, L. (2011) · Médecine des Maladies Métaboliques · doi:10.1016/s1957-2557(11)70277-2 · Cited by 2
Solnica, B., Sygitowicz, G., Sitkiewicz, D., Cybulska, B., Jozwiak, J. J., Odrowąż‐Sypniewska, G., & Banach, M. (2020) · Diagnostyka Laboratoryjna · doi:10.5604/01.3001.0014.1296 · Cited by 2
Introduction of HDL Molecules, Past and Brief Future
Tabuchi, M., & Komoda, T. (2013) · Elsevier eBooks · doi:10.1016/b978-0-12-407867-3.00001-9 · Cited by 1
Michenaud, L., Marrié, N., Rimbert, A., Marmontel, O., Charrière, S., Gibert, C., Bouveyron, C., Mammi, J., Cariou, B., Moulin, P., et al. (2024) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2024-0587 · Cited by 2
Ercan, Ş., & Manav, A. (2025) · Biochemia Medica · doi:10.11613/bm.2025.020704 · Cited by 3
Kuti, M. A., Adeleye, J. O., Akinyemi, J. O., Ogundeji, O. A., Omoyele, O. O., Obe, O. A., Adewoyin, A. S., & Soriyan, O. O. (2025) · African Journal of Laboratory Medicine · doi:10.4102/ajlm.v14i1.2729 · Cited by 1
“4,5 There are, however, challenges with the use of the FF. Firstly, its inability to accurately estimate LDL cholesterol when the triglycerides are ≥ 4.52 mmol/L means that the laboratory needs to have an alternative means of determining and reporting LDL cholesterol in such scenarios.”
Influential
Metadata
- Authors
- Miller, W. G., Myers, G. L., Sakurabayashi, I., Bachmann, L. M., Caudill, S. P., Dziekonski, A., Edwards, S., Kimberly, M. M., Korzun, W. J., Leary, E. T., Nakajima, K., Nakamura, M., Nilsson, G., Shamburek, R. D., Vetrovec, G. W., Warnick, G. R., & Remaley, A. T.
- Affiliations
- Virginia Commonwealth University (US)
- Centers for Disease Control and Prevention (US)
- Jichi Medical University (JP)
- Virginia Commonwealth University, Richmond, VA
- Pacific Biomarkers (United States) (US)
- Otsuka (Japan) (JP)
- Osaka Medical Center for Health Science and Promotion, Osaka, Japan
- Nilsson Measurement Quality, Uppsala, Sweden
- National Institutes of Health (US)
- Health Diagnostics Laboratory, Richmond, VA
- Corresponding author
- Miller, W Greg
- ORCID
- Venue
- Clinical Chemistry
- Publisher
- Oxford University Press (OUP)
- Published
- 2010-06-01
- Volume
- 56
- Issue
- 6
- Pages
- 977-986
- ISSN
- 0009-9147 (ISSN-L)
- 1530-8561
- PMC
- PMC4687457
- OpenAlex
- W2130132335
- Cited by
- 386
- References
- 23
- Citations by year
- 2026: 22 · 2025: 31 · 2024: 25 · 2023: 31 · 2022: 30 · 2021: 30 · 2020: 31 · 2019: 26 · 2018: 22 · 2017: 21 · 2016: 15 · 2015: 14 · 2014: 27 · 2013: 16 · 2012: 17
- FWCI
- 13.22
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Licence
- academic.oup.com/journals/pages/open_access/funder_policies/chorus/standard_publication_model (vor)
- Locations
- Clinical Chemistry (publishedVersion)
- PubMed (publishedVersion)
- PubMed Central (acceptedVersion)
- Funders
- Merck
- Pfizer
- Abbott
- NIH
- Topics
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Lipoproteins and Cardiovascular Health
- Physical Sciences › Physics and Astronomy › Instrumentation › Scientific Measurement and Uncertainty Evaluation
- Keywords
- HDL cholesterol, LDL cholesterol, direct measurement method, method comparison, total error, imprecision, trueness, specificity, dyslipidemia, cardiovascular disease
- MeSH terms
- Blood Chemical Analysis, Humans, Cholesterol, HDL, Cholesterol, LDL, Reference Standards, Sensitivity and Specificity, Ultracentrifugation
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review