Seven Direct Methods for Measuring HDL and LDL Cholesterol Compared with Ultracentrifugation Reference Measurement Procedures
Open accessAbstract
Background: Methods from 7 manufacturers and 1 distributor for directly measuring HDL cholesterol (C) and LDL-C were evaluated for imprecision, trueness, total error, and specificity in nonfrozen serum samples. Methods: We performed each direct method according to the manufacturer’s instructions, using a Roche/Hitachi 917 analyzer, and compared the results with those obtained with reference measurement procedures for HDL-C and LDL-C. Imprecision was estimated for 35 runs performed with frozen pooled serum specimens and triplicate measurements on each individual sample. Sera from 37 individuals without disease and 138 with disease (primarily dyslipidemic and cardiovascular) were measured by each method. Trueness and total error were evaluated from the difference between the direct methods and reference measurement procedures. Specificity was evaluated from the dispersion in differences observed. Results: Imprecision data based on 4 frozen serum pools showed total CVs <3.7% for HDL-C and <4.4% for LDL-C. Bias for the nondiseased group ranged from −5.4% to 4.8% for HDL-C and from −6.8% to 1.1% for LDL-C, and for the diseased group from −8.6% to 8.8% for HDL-C and from −11.8% to 4.1% for LDL-C. Total error for the nondiseased group ranged from −13.4% to 13.6% for HDL-C and from −13.3% to 13.5% for LDL-C, and for the diseased group from −19.8% to 36.3% for HDL-C and from −26.6% to 31.9% for LDL-C. Conclusions: Six of 8 HDL-C and 5 of 8 LDL-C direct methods met the National Cholesterol Education Program total error goals for nondiseased individuals. All the methods failed to meet these goals for diseased individuals, however, because of lack of specificity toward abnormal lipoproteins.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“It is important to note that direct LDL-C measurements also have limitations, including systematic bias and inaccuracy in patients with dyslipidaemia, especially for high TG levels [88-90].”
From 2019 ESC/EAS guidelines for the management of dyslipidaemias: lipid modification to reduce cardiovascular risk (2019) · cites this paper as [89]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Most discordances in dyslipidemic samples are observed at lower concentration ranges of LDLC (< 1.8 mmol/L) and HDLC (< 1.0 mmol/L) [42].”
From Quantifying atherogenic lipoproteins for lipid-lowering strategies: Consensus-based recommendations from EAS and EFLM (2020) · cites this paper as [42]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Direct measurements of dLDLC and dHDLC in normolipidemic samples usually meet the National Cholesterol Education Program (NCEP) total error goals of ≤12% and ≤13%, respectively, but total error ranged from -26% to +32% for dLDLC and -20% to +36% for dHDLC in a comprehensive study of different assays in dyslipidemic samples [42].”
From Quantifying atherogenic lipoproteins for lipid-lowering strategies: Consensus-based recommendations from EAS and EFLM (2020) · cites this paper as [42]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Grundy, S. M., Bilheimer, D. W., Chait, A., Clark, L. T., Denke, M. A., Havel, R. J., Hazzard, W. R., Hulley, S. B., Hunninghake, D. B., Kreisberg, R. A., et al. (1993) · JAMA · doi:10.1001/jama.269.23.3015 · Cited by 27,944
Expert Panel on Detection, E. (2001) · JAMA: The Journal of the American Medical Association · doi:10.1001/jama.285.19.2486 · Cited by 26,011
Abell, L. L., Levy, B. B., Brodie, B. B., & Kendall, F. E. (1952) · Journal of Biological Chemistry · doi:10.1016/s0021-9258(19)50907-3 · Cited by 3,085
Nissen, S. E., Nicholls, S. J., Sipahi, I., Libby, P., Raichlen, J. S., Ballantyne, C. M., Davignon, J., Erbel, R., Fruchart, J. C., Tardif, J.-C., et al. (2006) · JAMA · doi:10.1001/jama.295.13.jpc60002 · Cited by 2,047
National Cholesterol Education Program
Lenfant, C. (1991) · Archives of Internal Medicine · doi:10.1001/archinte.1991.00400060019005 · Cited by 1,188
Lipid screening and cardiovascular health in childhood
Daniels, S. R., Greer, F. R., & and the Committee on Nutrition (2008) · Pediatrics · doi:10.1542/peds.2008-1349 · Cited by 1,178
Lowe, C. U., Coursin, D. B., Heald, F. P., Holliday, M. A., OʼBrien, D., Owen, G. M., Pearson, H. A., Scriver, C. R., Filer, L. J., & Kline, O. L. (1967) · PEDIATRICS · doi:10.1542/peds.40.1.130 · Cited by 478
Nauck, M., Warnick, G. R., & Rifai, N. (2002) · Clinical Chemistry · doi:10.1093/clinchem/48.2.236 · Cited by 442
Self-citationHandbook of Lipoprotein Testing
Rifai, N., Warnick, G. R., & Dominiczak, M. H. (2001) · Medical Entomology and Zoology · Cited by 384
Self-citationWarnick, G. R., Nauck, M., & Rifai, N. (2001) · Clinical Chemistry · doi:10.1093/clinchem/47.9.1579 · Cited by 228
Self-citationMora, S., Rifai, N., Buring, J. E., & Ridker, P. M. (2009) · Clinical Chemistry · doi:10.1373/clinchem.2008.117929 · Cited by 176
Brewer, H. B. (1999) · The American Journal of Cardiology · doi:10.1016/s0002-9149(99)00308-2 · Cited by 158
Performance of Four Homogeneous Direct Methods for LDL-Cholesterol
Miller, W. G., Waymack, P. P., Anderson, F. P., Ethridge, S. F., & Jayne, E. C. (2002) · Clinical Chemistry · doi:10.1093/clinchem/48.3.489 · Cited by 104
Self-citationEsteban-Salán, M., Guimon-Bardesi, A., de la Viuda-Unzueta, J. M., Azcarate-Ania, M. N., Pascual-Usandizaga, P., & Amoroto-Del-Río, E. (2000) · Clinical Chemistry · doi:10.1093/clinchem/46.8.1121 · Cited by 64
Sakaue, T., Hirano, T., Yoshino, G., Sakai, K., Takeuchi, H. T., & Adachi, M. (2000) · Clinica Chimica Acta · doi:10.1016/s0009-8981(00)00200-x · Cited by 50
Usui, S., Kakuuchi, H., Okamoto, M., Mizukami, Y., & Okazaki, M. (2002) · Clinical Chemistry · doi:10.1093/clinchem/48.11.1946 · Cited by 42
Benlian, P., Cansier, C., Hennache, G., Khallouf, O., Bayer, P., Duron, F., Carrat, F., Couderc, R., Chazouillères, O., Bardet, J., et al. (2000) · Clinical Chemistry · doi:10.1093/clinchem/46.4.493 · Cited by 35
Halloran, P., Roetering, H. A., Pisani, T., van den Berg, B., & Cobbaert, C. M. (1999) · Archives of Pathology & Laboratory Medicine · doi:10.5858/1999-123-0317-rsaapo · Cited by 21
Wägner, A. M., Zapico, E., Bonet, R., Pérez, A., & Ordóñez‐Llanos, J. (2003) · Clinical Biochemistry · doi:10.1016/s0009-9120(03)00006-7 · Cited by 17
Laboratory Measurement of Lipid and Lipoprotein Risk Factors
Warnick, G. R. (1990) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.3109/00365519009091079 · Cited by 13
Self-citationCobbaert, C. M., Mulder, P. G. H., Baadenhuijsen, H., Zwang, L., Weykamp, C. W., & Demacker, P. N. M. (1999) · Clinical Chemistry · doi:10.1093/clinchem/45.3.360 · Cited by 8
Charlton-Menys, V., Liu, Y., Moorhouse, A., & Durrington, P. N. (2007) · Clinica Chimica Acta · doi:10.1016/j.cca.2007.04.009 · Cited by 6
Hypertriglyceridemia and low HDL: therapeutic considerations
Larsen, M. L. (1994) · Current Opinion in Lipidology · doi:10.1097/00041433-199402000-00008 · Cited by 2
Research works printed but not linked: 4
- 2009, Clin Chem, 55, 88
- 1999, Preparation and validation of commutable frozen human serum pools as secondary reference materials for cholesterol measurement procedures; approved guideline. CLSI document C37-A
- 1990, Scand J Clin Lab Invest Suppl, 198, 9
- 1999, Am J Cardiol, 83, 3F
Cited by
Sorted by citations
Mach, F., Baigent, C., Catapano, A. L., Koskinas, K. C., Casula, M., Badimon, L., Chapman, M. J., De Backer, G. G., Delgado, V., Ference, B. A., et al. (2019) · Atherosclerosis · doi:10.1016/j.atherosclerosis.2019.08.014 · Cited by 4,854
Sampson, M. L., Ling, C., Sun, Q., Harb, R., Ashmaig, M., Warnick, R., Sethi, A. A., Fleming, J. K., Otvos, J. D., Meeusen, J. W., et al. (2020) · JAMA Cardiology · doi:10.1001/jamacardio.2020.0013 · Cited by 593
Self-citationRosenson, R. S., Brewer, H. B., Chapman, M. J., Fazio, S., Hussain, M. M., Kontush, A., Krauss, R. M., Otvos, J. D., Remaley, A. T., & Schaefer, E. J. (2011) · Clinical Chemistry · doi:10.1373/clinchem.2010.155333 · Cited by 523
Self-citationLDL cholesterol: controversies and future therapeutic directions
Ridker, P. M. (2014) · The Lancet · doi:10.1016/s0140-6736(14)61009-6 · Cited by 274
Crimarco, A., Springfield, S., Petlura, C., Streaty, T., Cunanan, K., Lee, J., Fielding-Singh, P., Carter, M. M., Topf, M. A., Wastyk, H. C., et al. (2020) · The American Journal of Clinical Nutrition · doi:10.1093/ajcn/nqaa203 · Cited by 286
Metadata
- Authors
- Miller, W. G., Myers, G. L., Sakurabayashi, I., Bachmann, L. M., Caudill, S. P., Dziekonski, A., Edwards, S., Kimberly, M. M., Korzun, W. J., Leary, E. T., Nakajima, K., Nakamura, M., Nilsson, G., Shamburek, R. D., Vetrovec, G. W., Warnick, G. R., & Remaley, A. T.
- Affiliations
- Virginia Commonwealth University (US)
- Centers for Disease Control and Prevention (US)
- Jichi Medical University (JP)
- Virginia Commonwealth University, Richmond, VA
- Pacific Biomarkers (United States) (US)
- Otsuka (Japan) (JP)
- Osaka Medical Center for Health Science and Promotion, Osaka, Japan
- Nilsson Measurement Quality, Uppsala, Sweden
- National Institutes of Health (US)
- Health Diagnostics Laboratory, Richmond, VA
- Corresponding author
- Miller, W Greg
- ORCID
- Venue
- Clinical Chemistry
- Publisher
- Oxford University Press (OUP)
- Published
- 2010-06-01
- Volume
- 56
- Issue
- 6
- Pages
- 977-986
- ISSN
- 0009-9147 (ISSN-L)
- 1530-8561
- PMC
- PMC4687457
- OpenAlex
- W2130132335
- Cited by
- 386
- References
- 23
- Citations by year
- 2026: 22 · 2025: 31 · 2024: 25 · 2023: 31 · 2022: 30 · 2021: 30 · 2020: 31 · 2019: 26 · 2018: 22 · 2017: 21 · 2016: 15 · 2015: 14 · 2014: 27 · 2013: 16 · 2012: 17
- FWCI
- 13.22
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Licence
- academic.oup.com/journals/pages/open_access/funder_policies/chorus/standard_publication_model (vor)
- Locations
- Clinical Chemistry (publishedVersion)
- PubMed (publishedVersion)
- PubMed Central (acceptedVersion)
- Funders
- Merck
- Pfizer
- Abbott
- NIH
- Topics
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Lipoproteins and Cardiovascular Health
- Physical Sciences › Physics and Astronomy › Instrumentation › Scientific Measurement and Uncertainty Evaluation
- Keywords
- HDL cholesterol, LDL cholesterol, direct measurement method, method comparison, total error, imprecision, trueness, specificity, dyslipidemia, cardiovascular disease
- MeSH terms
- Blood Chemical Analysis, Humans, Cholesterol, HDL, Cholesterol, LDL, Reference Standards, Sensitivity and Specificity, Ultracentrifugation
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review