Simultaneous Quantification of Apolipoprotein A-I and Apolipoprotein B by Liquid-Chromatography–Multiple- Reaction–Monitoring Mass Spectrometry

Open access
Agger, Sean A, Marney, Luke C, Hoofnagle, Andrew N·2010·Clinical Chemistry

Abstract

BACKGROUND: If liquid-chromatography–multiple-reaction–monitoring mass spectrometry (LC-MRM/MS) could be used in the large-scale preclinical verification of putative biomarkers, it would obviate the need for the development of expensive immunoassays. In addition, the translation of novel biomarkers to clinical use would be accelerated if the assays used in preclinical studies were the same as those used in the clinical laboratory. To validate this approach, we developed a multiplexed assay for the quantification of 2 clinically well-known biomarkers in human plasma, apolipoprotein A-I and apolipoprotein B (apoA-I and apoB). METHODS: We used PeptideAtlas to identify candidate peptides. Human samples were denatured with urea or trifluoroethanol, reduced and alkylated, and digested with trypsin. We compared reversed-phase chromatographic separation of peptides with normal flow and microflow, and we normalized endogenous peptide peak areas to internal standard peptides. We evaluated different methods of calibration and compared the final method with a nephelometric immunoassay. RESULTS: We developed a final method using trifluoroethanol denaturation, 21-h digestion, normal flow chromatography-electrospray ionization, and calibration with a single normal human plasma sample. For samples injected in duplicate, the method had intraassay CVs <6% and interassay CVs <12% for both proteins, and compared well with immunoassay (n = 47; Deming regression, LC-MRM/MS = 1.17 × immunoassay − 36.6; Sx|y = 10.3 for apoA-I and LC-MRM/MS = 1.21 × immunoassay + 7.0; Sx|y = 7.9 for apoB). CONCLUSIONS: Multiplexed quantification of proteins in human plasma/serum by LC-MRM/MS is possible and compares well with clinically useful immunoassays. The potential application of single-point calibration to large clinical studies could simplify efforts to reduce day-to-day digestion variability.

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1373/clinchem.2010.152264

Cites

4 of 4 references verified0 citation errors · 0 not checkable

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Cites

Flagged first, then by citations

Research works printed but not linked: 29

  1. [1]Hoofnagle AN Wener MH The fundamental flaws of immunoassays and potential solutions using tandem mass spectrometry. J Immunol Methods 2009 347 3 11 19538965 PMC2720067
  2. [2]Addona TA Abbatiello SE Schilling B Skates SJ Mani DR Bunk DM Multi-site assessment of the precision and reproducibility of multiple reaction monitoring-based measurements of proteins in plasma. Nat Biotechnol 2009 27 633 41 19561596 PMC2855883
  3. [3]Bondar OP Barnidge DR Klee EW Davis BJ Klee GG LC-MS/MS quantification of Zn-alpha2 glycoprotein: a potential serum biomarker for prostate cancer. Clin Chem 2007 53 673 8 17317883
  4. [5]Kuhn E Addona T Keshishian H Burgess M Mani DR Lee RT Developing multiplexed assays for troponin I and interleukin-33 in plasma by peptide immunoaffinity enrichment and targeted mass spectrometry. Clin Chem 2009 55 1108 17 19372185 PMC2865473
  5. [6]Kuhn E Wu J Karl J Liao H Zolg W Guild B Quantification of C-reactive protein in the serum of patients with rheumatoid arthritis using multiple reaction monitoring mass spectrometry and 13C-labeled peptide standards. Proteomics 2004 4 1175 86 15048997
  6. [7]Kuzyk MA Smith D Yang J Cross TJ Jackson AM Hardie DB Multiple reaction monitoring-based, multiplexed, absolute quantitation of 45 proteins in human plasma. Mol Cell Proteomics 2009 8 1860 77 19411661 PMC2722777
  7. [8]Neubert H Gale J Muirhead D Online high-flow peptide immunoaffinity enrichment and nanoflow liquid chromatography/tandem mass spectrometry: assay development for total salivary pepsin/pepsinogen. Clin Chem 2010 56 1413 23 20472821
  8. [9]Whiteaker JR Zhao L Anderson L Paulovich AG An automated and multiplexed method for high throughput peptide immunoaffinity enrichment and multiple reaction monitoring mass spectrometry-based quantification of protein bio-markers. Mol Cell Proteomics 2010 9 184 96 19843560 PMC2808264
  9. [11]Brun V Masselon C Garin J Dupuis A Isotope dilution strategies for absolute quantitative proteomics. J Proteomics 2009 72 740 9 19341828
  10. [12]Elliott MH Smith DS Parker CE Borchers C Current trends in quantitative proteomics. J Mass Spectrom 2009 44 1637 60 19957301
  11. [13]Anderson NL Anderson NG Pearson TW Borchers CH Paulovich AG Patterson SD A human proteome detection and quantitation project. Mol Cell Proteomics 2009 8 883 6 19131327 PMC2689772
  12. [14]Carr SA Anderson L Protein quantitation through targeted mass spectrometry: the way out of biomarker purgatory? Clin Chem 2008 54 1749 52 18957555 PMC2848668
  13. [15]Paulovich AG Whiteaker JR Hoofnagle AN Wang P The interface between biomarker discovery and clinical validation: the tar pit of the protein biomarker pipeline. Proteom Clin Appl 2008 2 1386 402 PMC2957839 20976028
  14. [17]Davidson MH Apolipoprotein measurements: is more widespread use clinically indicated? Clin Cardiol 2009 32 482 6 19743499 PMC6653425
  15. [18]Holme I Aastveit AH Hammar N Jungner I Walldius G Lipoprotein components and risk of congestive heart failure in 84,740 men and women in the Apolipoprotein MOrtality RISk study (AMORIS). Eur J Heart Fail 2009 11 1036 42 19801574
  16. [19]Lamarche B Moorjani S Lupien PJ Cantin B Bernard PM Dagenais GR Despres JP Apolipoprotein A-I and B levels and the risk of ischemic heart disease during a five-year follow-up of men in the Quebec cardiovascular study. Circulation 1996 94 273 8 8759066
  17. [21]Movva R Rader DJ Laboratory assessment of HDL heterogeneity and function. Clin Chem 2008 54 788 800 18375481
  18. [22]Walldius G Jungner I The apoB/apoA-I ratio: a strong, new risk factor for cardiovascular disease and a target for lipid-lowering therapy—a review of the evidence. J Intern Med 2006 259 493 519 16629855
  19. [23]Hoofnagle AN Heinecke JW Lipoproteomics: using mass spectrometry-based proteomics to explore the assembly, structure, and function of lipoproteins. J Lipid Res 2009 50 1967 75 19738003 PMC2739765
  20. [24]Kay RG Gregory B Grace PB Pleasance S The application of ultra-performance liquid chromatography/tandem mass spectrometry to the detection and quantitation of apolipoproteins in human serum. Rapid Commun Mass Spectrom 2007 21 2585 93 17639571
  21. [25]Deutsch EW Eng JK Zhang H King NL Nesvizhskii AI Lin B Human plasma PeptideAtlas. Proteomics 2005 5 3497 500 16052627
  22. [27]Abbatiello SE Mani DR Keshishian H Carr SA Automated detection of inaccurate and imprecise transitions in peptide quantification by multiple reaction monitoring mass spectrometry. Clin Chem 2010 56 291 305 20022980 PMC2851178
  23. [28]Rodbard D Statistical quality control and routine data processing for radioimmunoassays and immunoradiometric assays. Clin Chem 1974 20 1255 70 4370388
  24. [29]La Belle M McCall MR Krauss RM Forte TM Unique structural properties of apolipoprotein B in low-density lipoproteins produced by several human hepatoma-derived cell lines. Biochim Biophys Acta 1990 1046 288 93 2171671
  25. [30]Lopez MF Rezai T Sarracino DA Prakash A Krastins B Athanas M Selected reaction monitoring-mass spectrometric immunoassay responsive to parathyroid hormone and related variants. Clin Chem 2010 56 281 90 20022981
  26. [31]Anderson NL Anderson NG Haines LR Hardie DB Olafson RW Pearson TW Mass spectrometric quantitation of peptides and proteins using Stable Isotope Standards and Capture by Anti-Peptide Antibodies (SISCAPA). J Proteome Res 2004 3 235 44 15113099
  27. [32]Hoofnagle AN Quantitative clinical proteomics by liquid chromatography-tandem mass spectrometry: assessing the platform. Clin Chem 2010 56 161 4 19959622
  28. [33]Hoofnagle AN Peptide lost and found: internal standards and the mass spectrometric quantification of peptides. Clin Chem 2010 56 1515 17 20739635
  29. [34]Barr JR Maggio VL Patterson DG Jr Cooper GR Henderson LO Turner WE Isotope dilution–mass spectrometric quantification of specific proteins: model application with apolipoprotein A-I. Clin Chem 1996 42 1676 82 8855153

Cited by

Sorted by citations

Authors
Agger, S. A., Marney, L. C., & Hoofnagle, A. N.
Affiliations
University of Washington (US)
Corresponding author
Hoofnagle, Andrew N
Venue
Clinical Chemistry
Published
2010-12-01
Volume
56
Issue
12
Pages
1804-1813
ISSN
  • 0009-9147 (ISSN-L)
  • 1530-8561
OpenAlex
W2137885916
Cited by
143
References
4
Citations by year
2026: 1 · 2025: 1 · 2024: 2 · 2023: 2 · 2022: 2 · 2021: 2 · 2020: 8 · 2019: 3 · 2018: 9 · 2017: 13 · 2016: 13 · 2015: 15 · 2014: 16 · 2013: 21 · 2012: 24
FWCI
8.08
Type
journal article
Language
en
Open access
Locations
  • Clinical Chemistry (publishedVersion)
  • PubMed (publishedVersion)
  • PubMed Central (acceptedVersion)
Funders
University of Washington
Topics
  • Physical Sciences › Chemistry › Spectroscopy › Advanced Proteomics Techniques and Applications
  • Physical Sciences › Chemistry › Spectroscopy › Mass Spectrometry Techniques and Applications
  • Life Sciences › Biochemistry, Genetics and Molecular Biology › Molecular Biology › Metabolomics and Mass Spectrometry Studies
Keywords
apolipoprotein A-I, apolipoprotein B, multiplexed protein quantification, one-point calibration, immunoassay comparison, peptide identification, trypsin digestion, internal standard normalization
MeSH terms
Apolipoproteins B, Calibration, Chromatography, High Pressure Liquid, Humans, Immunoassay, Nephelometry and Turbidimetry, Protein Denaturation, Trifluoroethanol, Urea, Reproducibility of Results, Biomarkers, Apolipoprotein A-I, Spectrometry, Mass, Electrospray Ionization, Tandem Mass Spectrometry

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review