Three of 7 Hemoglobin A1c Point-of-Care Instruments Do Not Meet Generally Accepted Analytical Performance Criteria

Lenters-Westra, Erna, Slingerland, Robbert J · 2014 · Clinical Chemistry
Cited by 130

The DOI matches its Crossref record; 3 of the 3 references it cites were found in a registry or an index.

Abstract

BACKGROUND: In 2009, we investigated the conformance of 8 hemoglobin A1c (Hb A1c) point-of-care (POC) instruments. Since then, instruments have improved and new devices are available on the market. In this second study, we evaluated the performance of DCA Vantage, Afinion, InnovaStar, Quo-Lab, Quo-Test, Cobas B101, and B-analyst Hb A1c POC instruments. METHODS: Clinical and Laboratory Standards Institute protocols EP-5 and EP-9 were applied to investigate imprecision, accuracy, and bias. We assessed bias using the mean of 3 certified secondary reference measurement procedures (SRMPs). Assay conformance with the National Glycohemoglobin Standardization Program (NGSP) certification criteria was also evaluated. Interference of common Hb variants was investigated for methods that could work with hemolysed material. RESULTS: The total CVs for all instruments, except for the DCA Vantage at a high Hb A1c value, were ≤3.1% in SI units and ≤2.1% in Diabetes Control and Complications Trial (DCCT) units. Afinion, DCA Vantage, B-analyst, and Cobas B101 instruments passed the NGSP criteria with 2 different reagent lot numbers. Quo-Test, Quo-Lab, and InnovaStar instruments had a negative bias compared to the mean of the 3 SRMPs and failed NGSP criteria. Most of the common Hb variants did not interfere with the investigated instruments, except Hb AE for the Cobas B101. CONCLUSIONS: Afinion, DCA Vantage, Cobas B101, and B-analyst instruments met the generally accepted performance criteria for Hb A1c. Quo-Test, Quo-Lab, and InnovaStar met the criteria for precision but not for bias. Proficiency testing should be mandated for users of Hb A1c POC assays to ensure quality.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1373/clinchem.2014.224311

Cites

3 of 3 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 3 references

Cited by

See all 130

Metadata

Authors
Lenters-Westra, E., & Slingerland, R. J.
Venue
Clinical Chemistry
Published
2014-08-01
Volume
60
Issue
8
Pages
1062-1072
ISSN
  • 0009-9147 (ISSN-L)
  • 1530-8561
OpenAlex
W2104992851
Cited by
130
References
3
FWCI
9.91
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Three of 7 Hemoglobin A1c Point-of-Care Instruments Do Not Meet Generally Accepted Analytical Performance Criteria. https://app.cochat.ai/papers/10.1373/clinchem.2014.224311
The paper
Lenters-Westra, E., & Slingerland, R. J. (2014). Three of 7 Hemoglobin A1c Point-of-Care Instruments Do Not Meet Generally Accepted Analytical Performance Criteria. Clinical Chemistry. https://doi.org/10.1373/clinchem.2014.224311

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.