Measurement of late-night salivary cortisol with an automated immunoassay system

Vogeser, Michael, Durner, Jürgen, Seliger, Ewald, Auernhammer, Christoph · 2006 · Clinical Chemistry and Laboratory Medicine (CCLM)
Cited by 51

The DOI matches its Crossref record; 11 of the 11 references it cites were found in a registry or an index.

Abstract

BACKGROUND: Measurement of late-night salivary cortisol concentrations is increasingly used as a screening test in suspected Cushing's syndrome. Cortisol concentrations are typically extremely low in late-night samples and discordant assay-specific reference ranges have been reported. Therefore, the aim of our study was to assess the analytical performance of the first automated cortisol immunoassay specified for salivary measurements and to establish late-night sampling reference-range data for this test. METHODS: Salivary cortisol was measured using the Roche Cobas Cortisol assay (Roche Diagnostics). Five salivary pools in different concentration ranges were used to assess the inter-assay imprecision of this test in a two-centre evaluation protocol including two reagent lots. Linearity was tested by serial dilution. Salivary samples were obtained at 23:00 h from 100 apparently healthy volunteers using a commercially available salivary sampling device (Salivette, Sarstedt). A subset of 20 samples was used for method comparison with isotope dilution liquid chromatography-tandem mass spectrometry. RESULTS: Inter-assay coefficients of variation (n=20) between 11.6% and 40.4% were found for mean cortisol concentrations between 12.9 and 2.6 nmol/L, with an estimated functional sensitivity of approximately 5.0 nmol/L. The test also gave linear results in the lowest concentration range between 1.0 and 8.3 nmol/L. Mean late-night salivary cortisol of 5.0 nmol/L was found for healthy individuals; the absolute range was 1.4-16.7 nmol/L, and the 95th percentile was 8.9 nmol/L. Substantially lower concentrations were found with isotope dilution LC-MS/MS compared to immunoassay results (mean concentrations 1.8 and 4.4 nmol/L, respectively). CONCLUSIONS: The automated assay investigated was found to offer acceptable analytical performance in the very low concentration range required for late-night salivary cortisol, despite a very short turn-around time. Using this assay, late-night salivary cortisol concentrations below 8.9 nmol/L are typically found in healthy volunteers.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1515/cclm.2006.244

Cites

11 of 11 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Cited by

See all 51

Metadata

Authors
Vogeser, M., Durner, J., Seliger, E., & Auernhammer, C.
Venue
Clinical Chemistry and Laboratory Medicine (CCLM)
Published
2006-01-01
Volume
44
Issue
12
Pages
1441-5
ISSN
  • 1434-6621 (ISSN-L)
  • 1437-4331
OpenAlex
W2042501074
Cited by
51
References
11
FWCI
2.55
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.