IFCC Guideline for sampling, measuring and reporting ionized magnesium in plasma

Ben Rayana, International Federation of Clinica, Burnett, Robert W., Covington, Arthur K., D'Orazio, Paul, Fogh-Andersen, Niels, Jacobs, Ellis, Külpmann, Wolf R., Kuwa, Katsuhiko, Larsson, Lasse, Lewenstam, Andrzej, Maas, Anton H.J., Mager, Gerhard, Naskalski, Jerzy W., Okorodudu, Anthony O., Ritter, Christoph, St John, Andrew · 2008 · Clinical Chemistry and Laboratory Medicine
Cited by 58

The DOI matches its Crossref record; 3 of the 3 references it cites were found in a registry or an index.

Abstract

Analyzers with ion-selective electrodes (ISEs) for ionized magnesium (iMg) should yield comparable and unbiased results for iMg. This IFCC guideline on sampling, measuring and reporting iMg in plasma provides a prerequisite to achieve this goal [in this document, "plasma" refers to circulating plasma and the forms in which it is sampled, namely the plasma phase of anticoagulated whole blood (or "blood"), plasma separated from blood cells, or serum]. The guideline recommends measuring and reporting ionized magnesium as a substance concentration relative to the substance concentration of magnesium in primary aqueous calibrants with magnesium, sodium, and calcium chloride of physiological ionic strength. The recommended name is "the concentration of ionized magnesium in plasma". Based on this guideline, results will be approximately 3% higher than the true substance concentration and 4% lower than the true molality in plasma. Calcium ions interfere with all current magnesium ion-selective electrodes (Mg-ISEs), and thus it is necessary to determine both ions simultaneously in each sample and correct the result for Ca2+ interference. Binding of Mg in plasma is pH-dependent. Therefore, pH should be measured simultaneously with iMg to allow adjustment of the result to pH 7.4. The concentration of iMg in plasma may be physiologically and clinically more relevant than the concentration of total magnesium. Furthermore, blood-gas analyzers or instruments for point-of-care testing are able to measure plasma iMg using whole blood (with intact blood cells) as the sample, minimizing turn-around time compared to serum and plasma, which require removal of blood cells.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1515/cclm.2008.001

Cites

3 of 3 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Cited by

See all 58

Metadata

Authors
Ben Rayana, I. F. O. C., Burnett, R. W., Covington, A. K., D'Orazio, P., Fogh-Andersen, N., Jacobs, E., Külpmann, W. R., Kuwa, K., Larsson, L., Lewenstam, A., Maas, A. H. J., Mager, G., Naskalski, J. W., Okorodudu, A. O., Ritter, C., & St John, A.
Venue
Clinical Chemistry and Laboratory Medicine
Published
2008-01-01
Volume
46
Issue
1
Pages
21-6
ISSN
  • 1434-6621 (ISSN-L)
  • 1437-4331
OpenAlex
W2021850993
Cited by
58
References
3
FWCI
3.36
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.