THE RADIOIMMUNOASSAY OF OXYTOCIN AND VASOPRESSIN
Abstract
The first radioimmunoassay was described by Yalow & Berson in 1960, since when the technique has been widely used in physiological and patho-physiological studies of the protein hormones. The success in this field has led to a number of attempts to develop similar assays for the small peptide hormones (molecular weight less than 3000). However, these compounds present certain difficulties which have considerably retarded progress. The purpose of the present review is to discuss the problems of the radioimmunoassay of the neurohypophysial peptides, with particular reference to oxytocin and vasopressin. The problems of the radioimmunoassay of small peptide hormones: There are two major problems. (1) Being of low molecular weight (about 1000) they are poor immunogens. As a result, the preparation of high-affinity antisera is considerably more exacting than in the case of larger molecules. (2) Their levels in the circulation are, in molar terms, considerably lower than those of
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
An antibody might recognise elements unrelated to the antigen against which it was raised, and unrelated molecules might cross-react.
“In some cases, an antibody might recognise elements unrelated to the antigen against which it was raised. Antibodies are sensitive to differences in amino acid sequence, although they recognise the structure of a molecule, not its amino acid sequence per se, and so unrelated molecules might cross‐react 76, 79.”
From Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers (2016) · cites this paper as [76]
Bioassays are less specific than immunoassays because of the cross-reactivity of vasopressin and oxytocin with their respective receptors.
“However, bioassays are less specific because of the cross‐reactivity of vasopressin and oxytocin with their respective receptors 76.”
From Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers (2016) · cites this paper as [76]
In 1973, Chard noted that the best immunoassays for vasopressin were no more sensitive than the best bioassays; the same is still true today for most radioimmunoassays and most enzyme-linked immunoassays (ELISAs).
“In 1973, Chard 76 noted that best immunoassays for vasopressin were no more sensitive than the best bioassays; the same is still true today for most radioimmunoassays and most enzyme‐linked immunoassays (ELISAs).”
From Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers (2016) · cites this paper as [76]
Problems in immunoassays can come from factors such as heparin or low pH or high osmolality that interfere with the antigen-antibody reaction.
“Problems can come from factors such as heparin or low pH or high osmolality that interfere with the antigen–antibody reaction.”
From Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers (2016) · cites this paper as [76]
There were considerable difficulties in developing immunoassays for vasopressin and oxytocin because they have a low molecular weight, are poor immunogens, making high-affinity antisera difficult, and circulating levels were clearly very low.
“There were considerable difficulties in developing immunoassays for vasopressin and oxytocin. Because these have a low molecular weight, they are poor immunogens, making the preparation of high‐affinity antisera difficult, and circulating levels were clearly very low.”
From Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers (2016) · cites this paper as [76]
Antibodies do not recognise the whole molecule, so an immunoassay may overestimate hormone concentrations by measuring fragments of no physiological significance.
“Antibodies do not recognise the whole molecule, and so an immunoassay may overestimate hormone concentrations by measuring fragments of no physiological significance 76.”
From Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers (2016) · cites this paper as [76]
Close attention was given to extraction and concentration procedures in developing immunoassays for vasopressin and oxytocin.
“Thus, close attention was given to extraction and concentration procedures 76.”
From Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers (2016) · cites this paper as [76]
When oxytocin is incubated in late-pregnancy plasma, there is a rapid loss of biological activity but after 2 h oxytocin immunoreactivity exceeds bioactivity by three-fold; the explanation is that oxytocinase degrades labelled oxytocin, reducing label binding to antibody and leading to overestimation.
“When oxytocin is incubated in this plasma, there is a rapid loss of biological activity but, after 2 h, oxytocin ‘immunoreactivity’ exceeds bioactivity by three‐fold. The obvious explanation is that oxytocinase in the sample also degrades the labelled oxytocin. The consequence will be that less label will bind to antibody, leading to overestimation of the amount of oxytocin 76.”
From Measuring Oxytocin and Vasopressin: Bioassays, Immunoassays and Random Numbers (2016) · cites this paper as [76]
The citation lists haven’t been fetched yet
Cited by 51
- Authors
- CHARD, T.
- Corresponding author
- CHARD, T.
- Venue
- Journal of Endocrinology
- Publisher
- Bioscientifica
- Published
- 1973-07-01
- Volume
- 58
- Issue
- 1
- Pages
- 143-160
- ISSN
- 0022-0795 (ISSN-L)
- 1479-6805
- OpenAlex
- W2003210454
- Cited by
- 51
- References
- 0
- Citations by year
- 2024: 2 · 2016: 1
- FWCI
- 9.48
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Journal of Endocrinology (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Social Sciences › Psychology › Social Psychology › Neuroendocrine regulation and behavior
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuropeptides and Animal Physiology
- Health Sciences › Medicine › Pulmonary and Respiratory Medicine › Electrolyte and hormonal disorders
- Keywords
- radioimmunoassay, oxytocin, vasopressin, neurohypophysial peptides, peptide hormones, antisera preparation, immunogenicity, hormone quantification
- MeSH terms
- Aminopeptidases, Animals, Antibody Formation, Binding, Competitive, Biological Assay, Chromatography, Gel, Humans, Immune Sera, Iodine Isotopes, Ion Exchange, Methods, Molecular Weight, Oxytocin, Radioimmunoassay, Vasopressins, Rats
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review